KURZ precious metal tympanostomy tubes are intended to provide a means of aerating and draining the middle ear in patients with otitis media and to reduce the likelihood of recurrent postoperative middle ear infections for the duration in which the tube remains in situ. The MINIMAL Ventilation Tube (Benz) has been designed for 1. temporary ventilation of the middle ear in secretory otitis media in adult patients for whom paracentesis alone is not sufficient, yet retention time of conventional ventilation tubes would be too long. 2. It can also effectively be applied in cases of narrow meatus, atrophic drum and post-tympanoplasty tube dysfunctions. Due to its tiny size (dia. 0.6 mm), insertion after tympanoplasty considerably reduces risk of permanent tympanum perforation. Following surface anaesthesia, the tube is pushed through the tympanic membrane with alligator forceps without prior paracentesis.
Device Story
Tympanostomy tubes; bobbin-shaped or narrow-tube designs; materials include gold-platinum alloy, gold-plated silver, titanium, and gold-plated stainless steel. Implants placed in tympanic membrane via myringotomy to aerate/drain middle ear; reduce recurrent infection risk. MINIMAL tube features sharpened end for insertion without prior paracentesis; held by tissue tension. Operated by physicians using alligator forceps under surface anesthesia. Output is physical conduit for pressure equalization and fluid drainage. Benefits include reduced risk of permanent tympanum perforation and improved retention in specific anatomical cases.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Materials: gold-platinum alloy, gold-plated silver, titanium, gold-plated stainless steel (1.4301). Form factors: bobbin-shaped, narrow tube with sharpened end, cone-shaped external flange. Features: optional wire for placement/removal, unilateral/bilateral eye extensions. Single patient use; sterile packaging.
Indications for Use
Indicated for patients with otitis media requiring middle ear aeration and drainage. MINIMAL Ventilation Tube (Benz) indicated for adult patients with secretory otitis media, narrow meatus, atrophic drum, or post-tympanoplasty tube dysfunction.
Regulatory Classification
Identification
A tympanostomy tube is a device that is intended to be implanted for ventilation or drainage of the middle ear. The device is inserted through the tympanic membrane to permit a free exchange of air between the outer ear and middle ear. A type of tympanostomy tube known as the malleous clip tube attaches to the malleous to provide middle ear ventilation. The device is made of materials such as polytetrafluoroethylene, polyethylene, silicon elastomer, or porous polyethylene.
K071150 — TROCAR VENTILATION TUBE/WITH TROCAR TIP, TROCAR HANDLE, MODEL 1015 074, 1015 076, 1015 075, 1015 077, W/ RETENTION WIRE · Heinz Kurz GmbH Medizintechnik · Dec 12, 2007
K992222 — FLUOROPLASTIC VENTILATION TUBES · Gyrus Ent LLC · Aug 26, 1999
K062672 — GRACE MEDICAL PC COATED TYMPANOSTOMY TUBES · Grace Medical, Inc. · Oct 24, 2006
K103595 — ACCLARENT TYMPANOSTOMY TUBE AND TYMPANOSTOMY TUBE DELIVERY SYSTEM · Acclarent, Inc. · Apr 1, 2011
Submission Summary (Full Text)
{0}------------------------------------------------
KURZ
Heinz Kurz GmbH Medizintechnik
Hausanschrift: Tübinger Straße 3 D-72144 Dußlingen
Telefon (0 70 72) 91 79-0 Telefax (07072) 91 79-7
Heinz Kurz GmbH · Medizintechnik · Postfach 39 · D-72142 Dußlingen
K973226
NOV - 4 1997
# 6. 510(k) SUMMARY
As required by Section 807.92(c)
- Submitter 1807.92(a){ 1)} (1) Dagmar S. Mäser Business Support International Amstel 320-1 Amsterdam, 1017AP The Netherlands
Tel. 011 - 31 - 20 - 428 9591 -428 9429 Fax
bsi@xs4all.nl e-mail
Identification of Devices 1807.92(a)(2)} (2) Tübingen Ventilation Tube Trade Names a. Tympanic Ventilation Tube b. DIABOLO Tympanic Ventilation Tubes C. MINIMAL Typanic Ventilation Tube (Benz) d. Tympanostomy Tube; Ventilation Tube Common Name Tube, Tympanostomy Classification Name Product Code 77 ETD Class II Regulatory Number 874.3880 Classification Panel ENT Branch (HFZ-470)
#### Identification of Predicate Devices 1807.92(a)(3)] (3)
The KURZ ventilation tubes are substantially equivalent to a wide variety of ventilation tube designs produced by XOMED and XOMED-TREACE, Jacksonville, FL, USA and by Smith & Nephew Richards, Bartlett, TN, USA. In terms of indication, the short-term MINIMAL Ventilation Tube (Benz) is substantially equivalent to the Hubbard Airplane and the Tiny Tytan Vent Tubes, manufactured by Xomed-Treace.
These devices have been found substantially equivalent through the 510(k) premarket notification process.
6-1
Update: 10/06/97
Geschäftsführer: Heinz Kurz Trauce Kurz-Buczki USt - Id. Nr. DE 81 1570328
Volksbank Steinlach-Wiesaz eG 181 7 640 618 541 Konto-Nr. 70 150 001 Baden-Württembergische Bank Tübingen (BLZ 641 200 30) Konto-Nr. 120 140 2300 Amtsgericht Tübingen HRB 1424
{1}------------------------------------------------
# Heinz Kurz GmbH Medizintechnik
510(k) Summary (Cont'd)
.
:
P
#### Description of Devices 1807.92(a)(4)] (4)
- Typical bobbin-shaped design made of gold-platinum, Tübingen a. gold-plated silver, and titanium; with or without a wire for secure placement and removal. Internal diameter of Size 2 is min. 0.2 mm wider than that of most predicate devices ( SE: Reuter Bobbin, Collar Button).
#### Tympanic Ventilation Tube b.
- Bobbin shape with unilateral eye extension of 1 Eye . inner flange; made of gold-platinum alloy (SE: Spoon Bobbin, Shah Type Tube)
- Bobbin shape with bi-lateral eye extensions of inner 2 Eyes flange; made of gold-platinum alloy (SE: Most T-Tubes)
#### DIABOLO Ventilation Tube C.
External flange is cone-shaped to reduce the length of the internal hollow cylinder; made of gold-platinum alloy (SE: Beveled Bobbin; Shepard Grommets)
#### MINIMAL Ventilation Tube (Benz) d.
Narrow tube with sharpened end for insertion without prior paracentesis. Material: gold-plated stainless steel (1.4301).
The devices are implants for single patient use. They are sold in an indivi dual sterile package in a sealed carton.
#### Intended Use of Devices [807.92(a)(5)] (5)
KURZ precious metal tympanostomy tubes are intended to provide a means of aerating and draining the middle ear in patients with otitis media and to reduce the likelihood of recurrent postoperative middle ear infections for the duration in which the tube remains in situ.
The MINIMAL Ventilation Tube (Benz) has been designed for
- temporary ventilation of the middle ear in secretory otitis media in adult ﺎ . patients for whom paracentesis alone is not sufficient, yet retention time of conventional ventilation tubes would be too long.
- 2. It can also effectively be applied in cases of narrow meatus, atrophic drum and post-tympanoplasty tube dysfunctions. Due to its tiny size (dia. 0.6 mm), insertion after tympanoplasty considerably reduces risk of permanent tympanum perforation.
Following surface anaesthesia, the tube is pushed through the tympanic membrane with alligator forceps without prior paracentesis.
{2}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image contains the text "Public Health Service". The text is in a bold, sans-serif font. The words are stacked on top of each other. The text is black against a white background.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Dagmar S. Maser Business Support International Heinz Kurz GmbH Medisintechnik Amstel 320-1 1017 AP Amsterdam The Netherlands
Re:
NOV - 4 1997
K973226 Tympanostomy Tubes; Tüububgen Ventilation Tubes (Standard) Tympanic Ventilation Tube (with one eye); Tympanic Ventilation Tube ( with 2 eyes); DIABLO Ventilation Tube and MINIMAL Ventilation Tube (Benz) Dated: July 20, 1997 Received: August 27, 1997 Regulatory class: II 21 CFR 874.3880/Procode: 77 ETD
Dear Mr. Maser:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and probibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marked predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in yitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for question and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
William Yu
Lillian Yin, Ph.D. Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Heinz Kurz GmbH Medizintechnik
| 510(k) Number | K973226 |
|---------------|-------------------------------------------------------------|
| Device Names | Tübingen, Tympanic (1 or 2 eyes), DIABOLO Ventilation Tubes |
# INDICATIONS FOR USE:
Tympanic tubes are inserted into the eardrum (myringotomy) for drainage and temporary ventilation of the middle ear subsequent to acute otitis media.
Tympanic tubes with one or two eyes are designed to increase the retention time in the eardrum.
The substantially shortened passage of the DIABOLO reduces the length of the internal hollow cylinder and the smooth surface of the gold seems to diminish the risk of luminal occlusion.
## Placement:
The tubes are placed in the tympanic membrane to provide the means for drainage of any fluid buildup in the middle ear while creating an avenue for the passage of air to equalize pressure on either side of the drum.
(Please do not write below this line - continue on another page if needed)
Concurrence of CDRH, Office of Device Evaluation (ODE)
David A. Saylor
(Division Sign-Off Division of Reproductive, Abdominal, ENT, and Radiological Devices 510(k) Number
Prescription Use (Per CFR 801 109) Over-The-Counter Use _
OR
{4}------------------------------------------------
Heinz Kurz GmbH Medizintechnik
973226 (10/06/97) 510(k) Number ________________________________________________________________________________________________________________________________________________________________
Device Names MINIMAL Tympanic Ventilation Tube (Benz)
### INDICATIONS FOR USE:
- Temporary ventilation of the middle ear in serous otitis media in adult patients. 1.
- In cases of narrow meatus, atrophic drum or post tympanoplasty tube dysfunction. 2. Due to its tiny size (Diameter 0.6 mm), insertion after tympanoplasty considerably reduces the risk of permanent perforation of the tympanum.
This tympanic ventilation tube, constructed of gold-plated stainless steel (1.4301), has been designed for temporary ventilation in special cases in which paracentesis alone is not sufficient, yet retention time of conventional tubes would be too long. The tube features a sharpened end that can be pushed through the tympanic membrane with alligator forceps without prior paracentesis.
If the effusion is somewhat thicker than expected, an additional small paracentesis beside the tube is recommended, via which the mucous tympanum can easily be aspirated due to the pressure balance which is achieved through the ventilation tube.
Should secretion encrustation occur, the MINIMAL Ventilation Tube can be replaced without the need for surface anaesthesia and without pain to the patient as it has no internal flange but is kept in position through tissue tension.
### Placement:
Following surface anaesthesia, the tube is pushed through the tympanic membrane with alligator forceps without prior paracentesis.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED);
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT,
and Radiological Devices
| 510(k) Number | K973226 |
|---------------|---------|
|---------------|---------|
| Prescription Use | OR | Over-The-Counter Use |
|-------------------|----|----------------------|
| (Per CFR 801 109) | | |
KRZ-INDM.ini
(Optional Format 1-2-96)
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.