Hummingbird Tympanostomy Tube

K151830 · Preceptis Medical · ETD · Sep 21, 2015 · Ear, Nose, Throat

Device Facts

Record IDK151830
Device NameHummingbird Tympanostomy Tube
ApplicantPreceptis Medical
Product CodeETD · Ear, Nose, Throat
Decision DateSep 21, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 874.3880
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Indicated where chronic Eustachian tube dysfunction does not respond to conventional therapy.

Device Story

Hummingbird Tympanostomy Tube is a medical-grade silicone tube designed for insertion through the tympanic membrane. It provides ventilation to the middle ear and facilitates drainage of fluids. The device is used in clinical settings to treat chronic Eustachian tube dysfunction. It is intended to be used in conjunction with the Hummingbird Tympanostomy Tube System. By maintaining an open pathway through the eardrum, the tube helps resolve middle ear fluid accumulation and pressure issues, potentially benefiting patients by restoring middle ear function and preventing complications associated with chronic dysfunction.

Clinical Evidence

Bench testing only. The device met all performance acceptance criteria.

Technological Characteristics

Medical-grade silicone ventilation tube; 1 mm inside lumen diameter; sterile (EO).

Indications for Use

Indicated for patients with chronic Eustachian tube dysfunction unresponsive to conventional therapy.

Regulatory Classification

Identification

A tympanostomy tube is a device that is intended to be implanted for ventilation or drainage of the middle ear. The device is inserted through the tympanic membrane to permit a free exchange of air between the outer ear and middle ear. A type of tympanostomy tube known as the malleous clip tube attaches to the malleous to provide middle ear ventilation. The device is made of materials such as polytetrafluoroethylene, polyethylene, silicon elastomer, or porous polyethylene.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle with three lines representing its wings and body. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 September 21, 2015 Preceptis Medical Mr. Keith Leland Vice President of Research and Developement 505 Highwav 169 North, #365 Plymouth, MN 55441 Re: K151830 Trade/Device Name: Hummingbird Tympanostomy Tube Regulation Number: 21 CFR 874.3880 Regulation Name: Tympanostomy Tube Regulatory Class: Class II Product Code: ETD Dated: July 22, 2015 Received: July 23, 2015 Dear Mr. Leland: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. Eric A. Mann -S for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Section 4 Indications For Use Statement # Indications for Use 510(k) Number (if known): Device Name: Hummingbird™ Tympanostomy Tube Indications For Use: Indicated where chronic Eustachian tube dysfunction does not respond to conventional therapy. X Prescription Use ______________________________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) {3}------------------------------------------------ # Section 5 510k Summary # K151830 | Submitter<br>Information: | Preceptis Medical, Inc.<br>505 Highway 169 North, #365<br>Plymouth, MN 55441<br>763.568.7819 | |----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------| | Contact: | Keith Leland, VP of R&D | | Date Prepared: | 1 July 2015 | | Trade Name | Hummingbird™ Tympanostomy Tube | | Product Code | ETD (21 CFR Part 874.3880) | | Common Name | Tympanostomy Tube | | Predicate Device | Summit Medical Ventilation Tubes, 510(k) #K830228 | | Device<br>Description | The tympanostomy tube is a silicone tube that is intended to provide<br>ventilation and drainage to the middle ear. | | Indications For<br>Use | Indicated where chronic Eustachian tube dysfunction does not respond<br>to conventional therapy. | | Technological<br>Characteristics | The tympanostomy tube is inserted through the tympanic membrane to<br>allow for ventilation of the middle ear and the drainage of fluid. | | Performance<br>Data | The tympanostomy tube met all performance acceptance criteria. | {4}------------------------------------------------ ## Section 5 510k Summary #### Overview of Device The Hummingbird™ Tympanostomy Tube is a silicone tympanostomy tube. It is intended to provide ventilation of the middle ear and drainage of fluids in instances where chronic Eustachian tube dysfunction does not respond to conventional therapy The Hummingbird Tympanostomy Tube is preceded by the predicate: Summit Medical ventilation tubes (cleared under K830228). Similar to the predicate device, the Hummingbird Tympanostomy Tube is designed to be inserted using the Hummingbird Tympanostomy Tube System. ### Substantial Equivalence Analysis The key performance, safety and design characteristics of the devices that are used to establish equivalence are identified and are listed in the substantial equivalence table. The device comparison table shows the similarities and differences between the Hummingbira™ Tympanostomy Tube and the predicate device. The indications for use for the devices are the same. The Hummingbird Tympanostomy tube is identical with respect to function and materials as the Summit Medical ventilation tubes. The Hummingbird Tympanostomy Tube is comparable dimensionally to Summit Medical ventilation tubes. An analysis of the differences between the devices showed minor dimensional differences between the Hummingbird Tympanostomy Tube and the Summit Medical ventilation tubes. The devices are substantially equivalent. ## Device Description The Hummingbird™ Tympanostomy Tube consists of the following model: | Component/ Model | Description | |------------------------------|---------------------------| | Ventilation Tube/05-1026-009 | Silicone Ventilation Tube | The ventilation tube is a standard, medical grade silicone otologic ventilation tube with a 1 mm inside lumen diameter as illustrated in diagram 1 (all dimensions are in millimeters.) Image /page/4/Picture/13 description: The image shows a 3D rendering of a light blue plastic object. The object has a cylindrical body with a semi-circular opening at one end. A rectangular prism extends vertically from the top of the cylindrical body. The object appears to be a component or part of a larger mechanism. Image /page/4/Figure/14 description: The image shows a technical drawing with measurements of an object. On the left, a side view shows a horizontal cylinder with a length of 5.45. A diagonal support with a length of 0.31 is angled at 55 degrees. A vertical rectangular prism with a width of 1.10 and height of 2.06 is placed on top of the cylinder. On the right, an end view shows a semi-circular base with a radius of 1.30 and a vertical cylinder with a diameter of 1.60 and a smaller cylinder with a diameter of 1.0. Diagram 1 - 05-1026-009 {5}------------------------------------------------ # Section 5 510k Summary #### Substantial Equivalence The Hummingbird "11 Tympanostomy Tube is substantially equivalent with respect to the Summit Medical predicate as confirmed through dimensional attributes and material equivalence. The device comparison table shows the similarities and differences between the Hummingbird™ Tympanostomy Tube and the predicate. The indications for use, materials, and method of sterilization are all the same. | Parameter | HummingbirdTM<br>Tympanostomy Tube<br>(PN 05-1026-009) | Summit Medical Otologic<br>Ventilation Tubes - Predicate | Comparison | |----------------------------------|----------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------|------------| | General<br>Description | Ventilation Tube | Ventilation Tube | | | Indications for<br>Use | Indicated where chronic<br>Eustachian tube dysfunction<br>does not respond to<br>conventional therapy. | Indicated where chronic<br>Eustachian tube dysfunction<br>does not respond to<br>conventional therapy. | Same | | Intended use | Indicated<br>where<br>chronic<br>Eustachian tube dysfunction<br>not<br>does<br>to<br>conventional therapy. | Indicated where chronic<br>Eustachian tube dysfunction<br>does not respond to<br>conventional therapy. | Same | | Design | Silicone ventilation tube | Silicone ventilation tube | Same | | Technological<br>Characteristics | Silicone vent tube to be<br>placed across the tympanic<br>membrane to allow for<br>middle ear ventilation and<br>fluid drainage. | Vent tube to be placed across<br>the tympanic membrane to<br>allow for middle ear ventilation<br>and fluid drainage. | Same | | Dimensions<br>(vent tube) | 1 mm lumen diameter | 0.89-1.14 mm lumen diameter | Similar | | Materials | Medical Grade Silicone | Medical Grade Silicone | Same | | Sterile | Sterile (EO) | Sterile (EO) | Same | ### Device Comparison Table #### Summary of Similarities - Indications for use are the same. ■ - Design and Technological Characteristics are the same. I - Materials are the same. - I Method of sterilization is the same. #### Summary of Differences - There are minor dimensional differences between the Hummingbird Tympanostomy Tube and the Summit Medical predicate tubes, but the dimensions of the Hummingbird tube fall within the range covered by the predicates. An analysis of the differences between the devices does not raise a question of safety or efficacy. The devices are substantially equivalent.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...