GRACE MEDICAL PC COATED TYMPANOSTOMY TUBES

K062672 · Grace Medical, Inc. · ETD · Oct 24, 2006 · Ear, Nose, Throat

Device Facts

Record IDK062672
Device NameGRACE MEDICAL PC COATED TYMPANOSTOMY TUBES
ApplicantGrace Medical, Inc.
Product CodeETD · Ear, Nose, Throat
Decision DateOct 24, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 874.3880
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Grace Medical PC Coated Tympanostomy Tubes have primary intended use as the predicate devices - A device that is intended to be implanted for ventilation or drainage of the middle ear.

Device Story

Tympanostomy tube implanted in tympanic membrane for middle ear ventilation and drainage. Device consists of medical-grade silicone body coated with PC1036 phosphorylcholine. Coating provides resistance to staphylococcal and pseudomonal biofilm formation. Used by physicians in clinical settings. Output is physical ventilation of middle ear space; aids in resolving effusion, reducing infection recurrence, and alleviating symptoms like hearing loss, otalgia, and vertigo.

Clinical Evidence

Bench testing only. Evidence includes literature citation (Berry JA, et al., 2000) demonstrating in vitro resistance of PC1036 coating to staphylococcal and pseudomonal biofilm formation.

Technological Characteristics

Medical-grade silicone substrate with PC1036 phosphorylcholine coating. Designed for middle ear ventilation/drainage. Supplied sterile. No software or electronic components.

Indications for Use

Indicated for patients with chronic otitis media with effusion (serous, mucoid, or purulent); recurrent acute otitis media despite medical treatment; persistent high negative middle ear pressure with symptomatic conductive hearing loss, persistent/recurrent otalgia, or persistent/recurrent vertigo/tinnitus; and tympanic membrane retraction pockets.

Regulatory Classification

Identification

A tympanostomy tube is a device that is intended to be implanted for ventilation or drainage of the middle ear. The device is inserted through the tympanic membrane to permit a free exchange of air between the outer ear and middle ear. A type of tympanostomy tube known as the malleous clip tube attaches to the malleous to provide middle ear ventilation. The device is made of materials such as polytetrafluoroethylene, polyethylene, silicon elastomer, or porous polyethylene.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Ko62672 ## 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS GRACE MEDICAL PC COATED TYMPANOSTOMY TUBES | Trade Name: | | |----------------------|--| | Common Name: | | | Classification Name: | | Tympanostomy Tubes Tympanostomy Tubes Tympanostomy Tubes (CFR 21 § 874.3880) Official Contact: Jeff Cobb OCT 2 4 2006 Telephone: Telefax: Date Prepared: Grace Medical, Inc. 8500 Wolf Lake Drive, Suite 110 Memphis, TN 38133 (901) 380-7000 (901) 380-7001 September 8, 2006 Predicate Devices - The Grace Medical PC Coated Tympanostomy equivalent to the current silicone tympanostomy tubes marketed by Grace Medical, Inc. (K943325 & K981575) and PC Coated Tympanostomy Tubes (Pacific™) (K000801) marketed by Gyrus ENT. Intended Use - The Grace Medical PC Coated Tympanostomy Tubes have primary intended use as the predicate devices - A device that is intended to be implanted for ventilation or drainage of the middle ear. - Indications for Use 1) - (a) Chronic otitis media with effusion (serous, mucoid, or purulent). - (b) Recurrent episodes of acute otitis media despite conventional medical treatment. - (c) A record of persistent high negative middle ear pressure associated with one or more of the following svmptoms: - (i) Conductive hearing loss that is symptomatic. - (ii) Persistent or recurrent otalgia. - (iii) Persistent or recurrent vertigo, tinnitus, or both. - (d) Retraction pocket of the tympanic membrane. Material of Tubes - The Grace Medical PC Coated Tympanostomy Tubes are manufactured from medical grade silicone. Each tube is coated with a layer of PC1036 phosphorylcholine. | Design Features - Various designs of tubes are available to meet physician preference. | | |----------------------------------------------------------------------------------------|--| | | | | | | | | PC Coated<br>Tympanostomy Tubes<br>(Grace Medical) | Uncoated<br>Tympanostomy Tubes<br>(Grace Medical) | PC Coated Tympanostomy<br>Tubes<br>(Gyrus ENT) | Uncoated Tympanostomy<br>Tubes (Gyrus ENT) | |----------------------------|----------------------------------------------------|---------------------------------------------------|------------------------------------------------|--------------------------------------------| | Intended Use | Ventilation &<br>Drainage of Middle<br>Ear | Ventilation &<br>Drainage of Middle<br>Ear | Ventilation & Drainage of<br>Middle Ear | Ventilation & Drainage<br>of Middle Ear | | Material | Silicone w/ PC1036<br>Coating | Fluoroplastic,<br>Titanium, & Silicone | Fluoroplastic/PC1036<br>Coating | Fluoroplastic | | Coating<br>Characteristics | Resists Biofilm<br>Formation1 | N/A | Resists Biofilm Formation1 | N/A | | How Shipped | Supplied Sterile | Supplied Sterile | Supplied Sterile | Supplied Sterile | PC1036 Coating Information - The PC1036 coating has been shown to be resistant to staphylococcal biofilm formation and pseudomonal biofilm formation. Differences between the Grace Medical PC Coated Tympanostomy Tubes and the predicate devices should not affect the safety or effectiveness. 4 Berry JA, et.al.: In vitro resistance to bacterial biofilm formation on coated fluoroplastic tympanostomy tubes. Otolaryngology Head and Neck Surgery. 2000 Sep;123(3):246-51. {1}------------------------------------------------ Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its body and wings. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Grace Medical, Inc. c/o Jeff Cobb Vice President, Regulatory Affairs & Quality 8500 Wolf Lake Drive Suite 110 Memphis, TN 38133 OCT 2 4 2006 Re: K062672 Trade/Device Name: Grace Medical PC Coated tympanostomy Tubes Regulation Number: 21 CFR 874.3880 Regulation Name: Tympanostomy tube Regulatory Class: Class II Product Code: ETD Dated: September 7, 2006 Received: September 26, 2006 Dear Mr. Cobb: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {2}------------------------------------------------ ## Page 2 - Mr. Jeff Cobb This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, M.B. Eychler, M.D. Malvina B Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ## 510(k) Number: Device Name: PC Coated Tympanostomy Tubes - 1) Indications for Use - (a) Chronic otitis media with effusion (serous, mucoid, or purulent). - (b) Recurrent episodes of acute otitis media despite conventional medical treatment. - (c) A record of persistent high negative middle ear pressure associated with one or more of the following symptoms: - (i) Conductive hearing loss that is symptomatic. - (ii) Persistent or recurrent otalgia. - (iii)Persistent or recurrent vertigo, tinnitus, or both. - (d) Retraction pocket of the tympanic membrane. ## (PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) | Prescription Use<br>(Per 21 CFR 801.109) | X | OR | Over-The-Counter Use<br>(Optional Format 1-2/96) | |------------------------------------------|---|----|--------------------------------------------------| |------------------------------------------|---|----|--------------------------------------------------| Seinus Norden (Division Sign-Off) Division of Ophthalmi Nose and Throat Devi Prescription Use (Per 21 CFR 801.109) X 510(k) Number K062672 Page 3
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...