21 CFR 874.3305 — Hearing Aid, Air-Conduction With Wireless Technology, Prescription

Ear, Nose, Throat (EN) · Part 874 Subpart D—Prosthetic Devices · § 874.3305

Identification

A wireless air-conduction hearing aid is a wearable sound-amplifying device, intended to compensate for impaired hearing that incorporates wireless technology in its programming or use. A wireless air-conduction hearing aid is subject to the requirements in § 800.30 or § 801.422 of this chapter, as applicable.

Classification Rationale

Class II (special controls). The special controls for this device are:

Product Codes

Product CodeDevice NameClassDevicesAttributes
OSMHearing Aid, Air-Conduction With Wireless Technology, Prescription21
QUGHearing Aid, Air-Conduction With Wireless Technology, Over The Counter20

Special Controls

OSM — Hearing Aid, Air-Conduction With Wireless Technology, Prescription

The C4-PA hearing aid, as a wireless air-conduction hearing aid, is subject to the following special controls:

De Novo Order DEN100025

OSM — Hearing Aid, Air-Conduction With Wireless Technology, Prescription

*Classification.* Class II (special controls). The special controls for this device are:(1) Performance data must demonstrate the electromagnetic compatibility (EMC), electrical safety, and thermal safety of the device; (2) Performance testing must validate safety of exposure to non-ionizing radiation; and (3) Performance data must validate wireless technology functions. (c) *Premarket notification.* The wireless air-conduction hearing aid is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 874.9.

eCFR

OSM — Hearing Aid, Air-Conduction With Wireless Technology, Prescription

(1) Performance data must demonstrate the electromagnetic compatibility (EMC), electrical safety, and thermal safety of the device; (2) Performance testing must validate safety of exposure to non-ionizing radiation; and (3) Performance data must validate wireless technology functions.

Ecfr Llm

QUG — Hearing Aid, Air-Conduction With Wireless Technology, Over The Counter

*Classification.* Class II (special controls). The special controls for this device are:(1) Performance data must demonstrate the electromagnetic compatibility (EMC), electrical safety, and thermal safety of the device; (2) Performance testing must validate safety of exposure to non-ionizing radiation; and (3) Performance data must validate wireless technology functions. (c) *Premarket notification.* The wireless air-conduction hearing aid is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 874.9.

eCFR

Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...