PASCAL MEDICAL AB MENIETT 20

K991562 · Pascal Medical AB · ETY · Dec 28, 1999 · Ear, Nose, Throat

Device Facts

Record IDK991562
Device NamePASCAL MEDICAL AB MENIETT 20
ApplicantPascal Medical AB
Product CodeETY · Ear, Nose, Throat
Decision DateDec 28, 1999
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 874.1090
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Meniett 20 is indicated for symptomatic treatment of Menière's disease. The therapeutic effect of Meniett 20 is achieved by applying low frequency, low amplitude pressure pulses to the middle ear whereby inner ear endolymphatic fluids are assumed to be evacuated from the cochlea and thus relieve the patient of the symptoms associated with endolymphatic hydrops.

Device Story

Meniett 20 is an electronically controlled membrane pump used for home-based symptomatic treatment of Menière's disease. Following surgical myringotomy and placement of a ventilation tube, patients self-administer therapy by placing an ear-cuff in the external ear canal. The device generates low-frequency, low-amplitude pressure pulses (10-16 mbar) delivered to the middle ear to evacuate endolymphatic fluids from the cochlea. Treatment typically involves three daily sessions of three minutes each. The device features a pump house, electronic hardware, software, and an ear-cuff interface. A pulse shape selector allows switching between three pulse patterns. Indicators on the front panel facilitate patient communication. The device is intended to relieve vertigo attacks and improve balance/functionality. Clinical benefit is demonstrated by symptom relief during active disease periods.

Clinical Evidence

Clinical study of 20 patients with definite Meniere's disease and active vestibular symptoms. Patients treated with daily middle ear pressure pulses for 4 weeks, with many continuing up to 4 months. Primary endpoints: relief of vertigo attacks and improvement in balance/functionality. Results: 18 of 20 patients achieved intended clinical effect; no adverse effects observed. Temporary symptom relapses occurred in patients who discontinued use, with relief returning upon treatment resumption.

Technological Characteristics

Electronically controlled membrane pump; plastic encapsulation; battery-powered. Delivers low-frequency, low-amplitude pressure pulses (10-16 mbar) via ear-cuff. Includes pulse shape selector (3 settings), on/off button, and status indicators. Operates as a standalone unit for patient self-use.

Indications for Use

Indicated for symptomatic treatment of Menière's disease in patients with active vestibular symptoms. Requires prior surgical placement of a ventilation tube in the eardrum.

Regulatory Classification

Identification

An auditory impedance tester is a device that is intended to change the air pressure in the external auditory canal and measure and graph the mobility characteristics of the tympanic membrane to evaluate the functional condition of the middle ear. The device is used to determine abnormalities in the mobility of the tympanic membrane due to stiffness, flaccidity, or the presence of fluid in the middle ear cavity. The device is also used to measure the acoustic reflex threshold from contractions of the stapedial muscle, to monitor healing of tympanic membrane grafts or stapedectomies, or to monitor followup treatment for inflammation of the middle ear.

Special Controls

*Classification.* Class II (special controls). The device, when it is a tympanometer or auditory impedance tester that complies with FDA-recognized consensus standard ANSI S3.39, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 874.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ PASCAL MEDICAL DEC 28 1999 Kg91562 Confidential Filename Meniett 20 510(k) Supplement FDA No. K991562 September 15th, 1999 #### SECTION XVII 510(K) SUMMARY 510(K) Summary | Submitter: | Pascal Medical AB<br>TeknoCenter<br>S-302 50 Halmstad, Sweden | | |----------------------|----------------------------------------------------------------------------------------------------|----------------------------------| | | Telephone:<br>FAX: | 46-35-17 47 40<br>46-35-12 69 21 | | Date prepared: | March 31, 1999 | | | Device name: | Pascal Medical AB Meniett 20 | | | Classification name: | Tympanometer | | | Predicate Devices: | American Electromedics Corporation Screening<br>Tympanometer (K822475) & Hyper/Hypo baric Chambers | | This medical device is being submitted in this 510(K) Notification by virtue of the fact that it is substantially equivalent to hypo / hyperbaric pressure chambers which have been used previously for the treatment and temporary relief of symptoms associated with Meniere's Disease. The Meniett 20 is a device which functions in a range of performance specifications which are similar to a conventional tympanometer and is used in a manner similar to a hypo / hyper baric pressure chamber to create localized overpressure in the middle ear to relieve the symptoms associated with Meniere's Disease. # Device Description: Meniett 20 is indicated for symptomatic treatment of Menière's Disease. The therapeutic effect of Meniett 20 is achieved by applying low frequency, low amplitude pressure pulses to the middle ear whereby inner ear endolymphatic fluids are assumed to be evacuated from the cochlea and thus relieve the patient of the symptoms associated with endolymphatic hydrops. {1}------------------------------------------------ Image /page/1/Picture/0 description: The image shows the logo for Pascal Medical. The logo consists of a nautilus shell on the left, followed by the words "PASCAL MEDICAL" in all capital letters. The text is in a simple, sans-serif font. Confidential Filename Meniett 20 510(k) Supplement FDA No. Date September 15th, 1999 K991562 Patient treatment The treatment consists of two phases: - 1. Myrongotomiy A surgical intervention whereby a conventional ventilation tube is surgically placed in the eardrum of the ear to be treated. This intervention is performed in exactly the same way and using the same materials as for other indications, e.g. otitis media. - 2. Application of pressure pulses to the middle ear The therapy starts after the ventilation tube has been placed in the eardrum. The patient is instructed in how to place the ear-cuff in the external ear canal to minimize leakage to the outer environment. The patient is instructed to treat him/herself three times daily, morning, midday and evening and for three minutes each time. The total daily exposure of pressure pulses in the range of 10-16 mbar with this dosage is 64.8 seconds. Alteration of the standard dosage may be done in after consultation with the treating doctor. The treatment is continued for as long as the patient finds him/herself in a period of attacks of vertigo. During periods of remission there is no need for treatment Meniett 20 is an electronically controlled membrane pump. Meniett 20 generates dynamic pressure and consists of a pump house, electronic hardware, software and encapsulation. In connection with the pump house there is a tube to which an ear cuff is connected and through which air pressure pulses from Meniett 20 are transferred to the patient's ear canal. The construction is encapsulated in a plastic cover designed specially for Meniett 20. There is an opening for replacing the battery. With the help of a pulse shape selector, concealed in the battery compartment, there is a possibility to switch between three different pulse shapes. To start and stop the treatment there is an on/off button on the front panel. On the front panel are four indicators for communication with the patient. # Intended Use: Meniett 20 is indicated for symptomatic treatment of Menière's disease. The therapeutic effect of Meniett 20 is achieved by applying low frequency, low amplitude pressure pulses to the middle ear whereby inner ear endolymphatic fluids are assumed to be evacuated from the cochlea and thus relieve the patient of the symptoms associated with endolymphatic hydrops. # Technological Characteristics: {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the logo for Pascal Medical. The logo consists of a nautilus shell on the left and the words "PASCAL MEDICAL" on the right. The text is in a sans-serif font and is all caps. Confidential Filename Meniett 20 510(k) Supplement Date FDA No. September 15th, 1999 K991562 Comparing the proposed device to the predicate devices, both devices utilize the same range of chemical compositions, packaging and formulations. There are no significant differences. ## Summary of Non-Clinical Tests: In vitro testing was performed to evaluate / assess the product life, presentation of data, environmental parameters, power supply and battery life, sound levels, and environmental operating temperatures. The results of these tests confirmed that the proposed device meets the specifications for these parameters. # Clinical Test Results: 20 patients with a definite history and diagnosis of Meniere's disease and active vestibular symptoms have been treated with middle ear pressure pulses using Meniett daily for a period of four weeks. Thereafter, the majority of patients have continued treatment up to 4 months. The dosage was standardized in all patients. The aim of treatment was to relieve patients from the attacks of vertigo and to improve balance and functionality. The results showed that the use of Meniett 20 gave the intended clinical effect in all the patients except two. None of the patients became worse while using Meniett 20. No other adverse effects of treatment have been observed in connection with the use of Meniett 20. Temporary relapses of symptoms have been observed in patients who were not able to continue the use Meniett 20. The symptoms were relieved again in these patients by a renewal of treatment with the Meniett 20. It was demonstrated that the Meniett 20 performed well under the normal conditions of daily clinical use. ### Conclusions: Testing performed on the Pascal Medical AB Meniett 20 indicates that it is safe, effective, and performs as intended, when used in accordance with the instructions for use. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines representing its body and wings. The eagle is facing right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 DEC 2 8 1999 Jeffrey R. Sideman Pascal Medical AB 7307 Glouchester Drive Minneapolis, Minnesota 55435 Re: K991562 Trade Name: Meniett 20 Regulatory Class: Class II Product Code: 77ETY/874.1090 Dated: September 24, 1999 Received: September 29, 1999 Dear Mr. Sideman: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic OS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {4}------------------------------------------------ Page 2 - Mr. Jeffrey R. Sideman This letter will allow you to begin marketing your devices as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your devices to a legally marketed predicate device results in a classification for your devices and thus, permits your devices to proceed to the market. If you desire specific advice for your devices on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your devices, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, A. Ralph Rosenthal A. Ralph Rosenthal, M.D. Director Division of Ophthalmic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ Page / of / 510(k) Number (if known): x 991562 Device Name: Pas call Midical AB Meniett 20 Indications For Use: Image /page/5/Picture/4 description: The image shows the logo for Pascal Medical. The logo consists of a nautilus shell on the left and the words "PASCAL MEDICAL" on the right. The text is in a serif font and is all uppercase. Pre-Market Notification #### Pascal Medical AB Meniett 20 INDICATIONS FOR USE STATEMENT SECTION XII Meniett 20 is indicated for symptomatic treatment of Menière's disease. The therapeutic effect of Meniett 20 is achieved by applying low frequency, low amplitude pressure pulses to the middle ear whereby inner ear endolymphatic fluids are assumed to be evacuated from the cochlea and thus relieve the patient of the symptoms associated with endolymphatic hydrops. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Ellsworth L. Sharpe (Division Sign-Off) Division of Ophthalmic Devices 510(k) Number K991562 Prescription Use (Per 21 CFR 801.109) OR Over-The-Counter Use (Optional Format 1-2-96)
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...