INVENTIS MIDDLE EAR ANALYZER, FLUTE BASIC, INVENTIS MIDDLE EAR ANALYZER, FLUTE PLUS, INVENTIS MIDDLE EAR ANALYZER, FLUTE

K133013 · Inventis S.R.L. · ETY · Jan 24, 2014 · Ear, Nose, Throat

Device Facts

Record IDK133013
Device NameINVENTIS MIDDLE EAR ANALYZER, FLUTE BASIC, INVENTIS MIDDLE EAR ANALYZER, FLUTE PLUS, INVENTIS MIDDLE EAR ANALYZER, FLUTE
ApplicantInventis S.R.L.
Product CodeETY · Ear, Nose, Throat
Decision DateJan 24, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 874.1090
Device ClassClass 2
AttributesPediatric

Indications for Use

The Inventis Middle Ear Analyzers, Flute and Viola, are intended for use by trained operators in hospitals, nurseries, ENT clinics and audiology offices in conducting diagnostic hearing evaluations and assisting in diagnosis of possible otologic disorders. The Inventis Middle Ear Analyzers are intended for use on individuals six months of age or older. Flute is a tympanometer. Viola is a combination of audiometer and tympanometer, which reduces the amount of necessary equipment.

Device Story

Inventis Flute and Viola are middle ear analyzers; Flute functions as a standalone tympanometer; Viola integrates tympanometry with diagnostic audiometry (air/bone conduction, speech). Devices input acoustic impedance/admittance data via probe tones (226 Hz, 1 kHz) and audiometric stimuli (pure tone, warble, speech). Operated by trained clinicians in hospitals, ENT clinics, and audiology offices. Devices process signals to measure middle ear function and hearing thresholds; output results via graphical color display or optional integrated thermal printer. Clinical utility includes assisting in diagnosis of otologic disorders. Benefits include streamlined diagnostic workflow through combined functionality and improved portability compared to traditional equipment.

Clinical Evidence

Bench testing only. Performance verified against ANSI S3.39 (impedance), ANSI S3.6 (audiometry), and IEC 60645 series standards. Electrical safety and electromagnetic compatibility testing confirmed compliance with IEC 60601-1 and IEC 60601-1-2.

Technological Characteristics

Impedance/admittance measurement via probe tones (226 Hz, 1 kHz). Audiometry (Viola) via air/bone conduction. Graphical color display. Internal flash memory for speech storage. Powered by AC. Safety standards: IEC 60601-1 (Class II, Type BF), IEC 60601-1-2. Performance standards: ANSI S3.39/IEC 60645-5 (Impedance), ANSI S3.6/IEC 60645-1/2 (Audiometry).

Indications for Use

Indicated for diagnostic hearing evaluations and diagnosis of otologic disorders in individuals 6 months of age or older. Used by trained operators in clinical settings.

Regulatory Classification

Identification

An auditory impedance tester is a device that is intended to change the air pressure in the external auditory canal and measure and graph the mobility characteristics of the tympanic membrane to evaluate the functional condition of the middle ear. The device is used to determine abnormalities in the mobility of the tympanic membrane due to stiffness, flaccidity, or the presence of fluid in the middle ear cavity. The device is also used to measure the acoustic reflex threshold from contractions of the stapedial muscle, to monitor healing of tympanic membrane grafts or stapedectomies, or to monitor followup treatment for inflammation of the middle ear.

Special Controls

*Classification.* Class II (special controls). The device, when it is a tympanometer or auditory impedance tester that complies with FDA-recognized consensus standard ANSI S3.39, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 874.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## INVENTIS MIDDLE EAR ANALYZER TRADITIONAL 510(K) PREMARKET NOTIFICATION # 510(k) Summary ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ # 1 - Information 510(k) Notification: K133013 # GENERAL INFORMATION | Applicant: | Inventis srl | |------------|-------------------------------------------| | | Corso Stati Uniti 1/3, 35127 Padova Italy | | | Tel: +39 049 8962844 | | | Fax: +39 049 8966 343 | | | Registration Number: 3008699634 | #### Massimo Martinelli Contact Person: General Manager E-mail: massimo.martinelli@inventis.it t Date: December 18, 2013 ## DEVICE INFORMATION The Inventis Middle Ear Analyzers (Flute and Viola) are conventional Auditory Impedance Testers. Below you can find the information related to the subject device. Katalogia (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) : And Controller Comments | Classification: | 21 CFR§874.1090<br>Class II | |-------------------------|---------------------------------------| | Product Code: | ETY | | Trade Name: | Middle Ear Analyzers, Flute and Viola | | Common name: | Middle Ear Analyzer | | Classification<br>Name: | Auditory Impedance Tester | JAN 2 4 2014 ・ {1}------------------------------------------------ ### PREDICATE DEVICES Inventis Middle Ear Analyzers are substantially equivalent to the Interacoustics Audio Traveller, Model AA222 for the constructive technological features and the performance. #### INDICATIONS FOR USE The Inventis Middle Ear Analyzers, Flute and Viola, are intended for use by trained operators in hospitals, nurseries, ENT clinics and audiology offices in conducting diagnostic hearing evaluations and assisting in diagnosis of possible otologic disorders. The Inventis Middle Ear Analyzers are intended for use on individuals six months of age or older. Flute is a tympanometer. Viola is a combination of audiometer and tympanometer. #### PRODUCT DESCRIPTION Flute is stand-alone tympanometry / reflex diagnostic device. Viola is a stand-alone combined unit, integrating a screening middle ear analyzer and a diagnostic audiometer capable of performing air and bone conduction threshold and speech exams. Flute and Viola are available with or without the integrated thermal printer. # 2 - Substantial Equivalence Inventis Middle Ear Analyzers (Flute and Viola) represent a modification of the cleared device, Interacoustics Audiotraveller, Model AA222 (510(k) number: K022728). The changes are of minor nature and do not concern the main functions of the predicate device. In particular, the differences regard: - the pressure range, which is -600 to +400 daPa instead of -600 to +300 daPa and . - the introduction of an Internal Flash Memory where the speech material is stored. . In further support of a substantial equivalence determination, Section 12-Substancial Equivalent Discussion provides a comparison table between the submitted device and the predicate device. The main comparison features are reported in the following table: {2}------------------------------------------------ · · . . | | Predicate Device | New Device | | |---------------------------------------------|-------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------|-----------------------------------------------------------------| | Features | Interacoustics Audiotraveller,<br>Model AA222<br>(510(k) n.: K022728) | Flute | Viola | | Type | 2 - Impedance<br>(ANSI S3.39)<br>2- Tone Audiometer<br>(ANSI S3.6)<br>B-E Speech Audiometer<br>(ANSI S3.6) | 2-Impedance<br>(ANSI S3.39) | 2-Impedance<br>(ANSI S3.39)<br>3-Tone Audiometer<br>(ANSI S3.6) | | Probe Tone Frequency | 226 Hz, 678 Hz, 800 Hz, 1 kHz | 226 Hz,<br>1kHz (model HF) | 226 Hz | | Compliance Range | up to 8.0 ml | Same | | | Pressure Range | -600 to +300 daPa | -600 to +400 daPa | | | Safety Limitations for<br>pressuring system | -800 daPa and +600 daPa | Same | | | Acoustic Reflex<br>outputs | Ipsi (I), Contra (C) | Same | | | Acoustic Reflex Stimuli | 250Hz (C), 500Hz, 1kHz, 2kHz,<br>3kHz, 4kHz, 6kHz (C), 8kHz (C),<br>BBN, LPN, HPN | 250Hz (C), 500Hz, 1kHz, 2kHz, 4kHz | | | Audiometry:<br>Available Frequencies | 125Hz, 250Hz, 500Hz, 750Hz,<br>1kHz, 1.5kHz, 2kHz, 3kHz, 4kHz,<br>6kHz, 8kHz | | Same | | Audiometry Outputs | AC, BC, FF, insert mask | | AC, BC, insert mask | | Audiometry Inputs | PTA: pure tone, warble | | PTA: pure tone,<br>warble | | | Speech: EXT1, EXT2, MIC<br>Masking: NBN, WN, SN | | Speech: EXT1, EXT2,<br>INT, MIC<br>Masking: NBN, WN,<br>SN | | Max Power<br>Consumption | 45VA | 19.2 W | | | Display Description | Graphic monochrome display | Graphical color display | | | Size | 16.5 x 12.5 x 4 in<br>(42 x 32 x 10.2 cm) | 12.6 x 12.6 x 3.5 in (32 x 32 x 9 cm);<br>w/printer: 12.6 x 15.4 x 3.5 in (32 x 39 x 9 cm) | | | Weight | 9 lbs (4,1 kg) | 4.19 lbs (1,9 kg);<br>w/printer: 5.29 lbs (2,4kg) | | | Safety and<br>Effectiveness | Impedance<br>ANSI S3.39/IEC 60645-5: Type 2 | Impedance<br>ANSI S3.39/IEC 60645-5: Type 2 | | | | Audiometer<br>ANSI S3.6/IEC 60645-1:<br>Type 2 Tone Audiometer<br>ANSI S3.6/IEC 60645-2:<br>Speech Type B-E | Audiometer (only Viola models)<br>ANSI S3.6/IEC 60645-1:<br>Type 3 Pure-tone Audiometer | | | | Safety<br>IEC 60601-1, Class I, Type B<br>IEC 60601-1-2 | Safety<br>IEC 60601-1, Class II, Type BF<br>IEC 60601-1-2 | | - {3}------------------------------------------------ Inventis Middle Ear Analyzers (Flute and Viola) are substantially equivalent to the predicate device with regard to design, function, safety and technological and performance characteristic, intended use. Any differences in the technological characteristics between the devices do not raise any new issues of safety or effectiveness. Thus, the proposed Inventis Middle Ear Analyzers are substantially equivalent to the predicate devices. ## TESTING IN SUPPORT OF SUBSTANTIAL EQUIVALENCE DETERMINATION All necessary bench testing was conducted on the Inventis Middle Ear Analyzers (Flute and Viola) to support a determination of substantial equivalence to the predicate device. The safety and performance testing included the following tests: - . Electrical and mechanical safety testing - . Electromagnetic compatibility safety testing - . Specifications for Instruments to Measure Aural Acoustic Impedance and Admittance (ANSI S3.39) - . Instruments for the measurement of aural acoustic impedance (IEC 60645-5) #### CONCLUSION STATEMENT All the necessary electromagnetic compatibility and electrical safety tests were performed and documented in the Section 17-Electromagnetic compatibility and electrical safety. The results demonstrate that the Inventis Middle Ear Analyzer is in compliance with both the standards IEC 60601-1-2:2007 and IEC 60601-1:2005 and is safe as the predicate device. All the necessary performance tests in support of substantial equivalence determination were conducted and documented in the Section 18-Performance testing - Bench. The tests demonstrate that the Inventis Middle Ear Analyzer is effective and performs as well as the predicate device. The minor differences in the technological characteristics between the devices do not raise any new issues of safety or efficacy. Based on these results, we conclude that the Inventis Middle Ear Analyzer is substantially equivalent to the existing legally marketed device Interacoustics Audiotraveller Model AA222 under Federal Food, Drug and Cosmetic Act. {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three wing-like shapes. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird-like figure. ### DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002 #### January 24, 2014 Inventis Srl c/o Mr. Massimo Martinelli General Manager Corso Stati Uniti 1/3 35127, Padova Italy Re: K133013 Trade/Device Name: Inventis Middle Ear Analyzers, Flute and Viola Regulation Number: 21 CFR 874.1090 Regulation Name: Auditory Impedance Tester Regulatory Class: Class II Product Code: ETY Dated: December 18, 2013 Received: December 23, 2013 Dear Mr. Martinelli: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {5}------------------------------------------------ Page 2 - Mr. Massimo Martinelli Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.goy/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. # Eric A. Mann -S for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ .. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . # SECTION 04 INDICATIONS FOR USE STATEMENTS INDICATIONS FOR USE 510(k) Number (if known): K133013 Device Name: Inventis Middle Ear Analyzers, Flute and Viola Indications for Use: The Inventis Middle Ear Analyzers, Flute and Viola, are intended for use by trained operators in hospitals, nurseries, ENT clinics and audiology offices in conducting diagnostic hearing evaluations and assisting in diagnosis of possible otologic disorders. The Inventis Middle Ear Analyzers are intended for use on individuals six months of age or older. Flute is a tympanometer. Viola is a combination of audiometer and tympanometer, which reduces the amount of necessary equipment. Prescription Use _ Y (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (Please DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Shuchen Peng -S Shuchen Peng -S 2014.01.28 13:37:04 -05:00, SECTION 04 Page 04 - 1 of 1 Pages
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...