K121980 · Kuraray Noritake Dental, Inc. · EIH · Oct 19, 2012 · Dental
Device Facts
Record ID
K121980
Device Name
ZIRPRIME
Applicant
Kuraray Noritake Dental, Inc.
Product Code
EIH · Dental
Decision Date
Oct 19, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.6660
Device Class
Class 2
Indications for Use
Zirprime is indicated in the anterior and posterior segments for: crowns telescopic primary crowns multi-unit bridges (with no more than two pontics between abutment crowns) Zirprime can be used as a substructure (framework) which is then veneered with a dental veneering ceramic or can be used for full-contour application (without veneering) as well. In the case of telescopic primary crowns the substructure is not veneered.
Device Story
Zirprime is a dental ceramic material used for fabricating crowns, telescopic primary crowns, and multi-unit bridges. It serves as a substructure framework for veneering with dental ceramic or as a full-contour restoration. Used by dental professionals in clinical settings to restore tooth structure in anterior and posterior segments. The device provides a durable framework for prosthetic dental applications, benefiting patients by restoring function and aesthetics.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Porcelain powder for clinical use (21 CFR 872.6660). Class II device. Product code EIH.
Indications for Use
Indicated for patients requiring dental restorations in anterior and posterior segments, including crowns, telescopic primary crowns, and multi-unit bridges (max 2 pontics). Used as a substructure for veneering or as full-contour restoration.
Regulatory Classification
Identification
Porcelain powder for clinical use is a device consisting of a mixture of kaolin, felspar, quartz, or other substances intended for use in the production of artificial teeth in fixed or removable dentures, of jacket crowns, facings, and veneers. The device is used in prosthetic dentistry by heating the powder mixture to a high temperature in an oven to produce a hard prosthesis with a glass-like finish.
Related Devices
K111383 — ZIRPRIME · Noritake Co., Inc. · Aug 12, 2011
K133211 — ZIRTOUGH · Kuraray Noritake Dental, Inc. · Mar 20, 2014
K071674 — GC-ZIRCONIA · GC America, Inc. · Aug 9, 2007
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three tail feathers, representing the three levels of government: federal, state, and local. The eagle is positioned to the right of a circular seal that contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Kuraray Noritake Dental, Incorporated C/O Mr. Ivan Wasserman Partner Manatt, Phelps & Phillips, Limited Liability Company 700 12TH Street North West, Suite 1100 Washington, District of Columbia 20010
Re: K121980
Trade/Device Name: Zirprime Regulation Number: 21 CFR 872.6660 Regulation Name: Porcelain Powder for Clinical Use Regulatory Class: II Product Code: EIH Dated: October 5, 2012 Received: October 9, 2012
Dear Mr. Wasserman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
OCT
19
2012
{1}------------------------------------------------
## Page 2- Mr. Wasserman
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Susan Russer.
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known):
Device Name: Zirprime
Indications For Use:
Zirprime is indicated in the anterior and posterior segments for:
ُ crowns
telescopic primary crowns
multi-unit bridges (with no more than two pontics between abutment crowns)
Zirprime can be used as a substructure (framework) which is then veneered with a dental veneering ceramic or can be used for full-contour application (without veneering) as well. In the case of telescopic primary crowns the substructure is not veneered.
121980
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Suser Russo
(Division Sign-Off) Division of Anesthestology, General Hospital Infection Control, Dental Devices
510(k) Number:
Page 1 of
7
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.