SAFESTEP SAFETY DENTAL CARTRIDGE INJECTOR, MODEL 33-00000-1
K022959 · MDC Research , Ltd. · EJI · Feb 5, 2003 · Dental
Device Facts
Record ID
K022959
Device Name
SAFESTEP SAFETY DENTAL CARTRIDGE INJECTOR, MODEL 33-00000-1
Applicant
MDC Research , Ltd.
Product Code
EJI · Dental
Decision Date
Feb 5, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.6770
Device Class
Class 2
Indications for Use
The SafeStep™ Safety Dental Cartridge Injector is a sterile, single use disposable injector barrel assembly and a reusable plunger assembly that is indicated for use with pre-filled, 1.8mL, anesthetic cartridges and existing screw-on dental needles for injection of anesthetic solution in oral tissues. The SafeStep™ Safety Dental Cartridge Injector aids in the prevention of needle stick injuries. The SafeStepTM Safety Dental Cartridge Injector incorporates a needle retraction and re-advancing mechanism which may be operating with one hand.
Device Story
SafeStep™ Safety Dental Cartridge Injector is a manual, single-use disposable injector barrel with a reusable plunger assembly; used by dental professionals for delivering anesthetic solution into oral tissues. Device accepts standard 1.8mL pre-filled anesthetic cartridges and screw-on dental needles. Key feature is a one-handed needle retraction and re-advancing mechanism designed to prevent needle stick injuries. Device is purely mechanical; no electronic or software components. Clinical benefit is reduced risk of sharps injuries to the clinician during dental procedures.
Clinical Evidence
Simulated use clinical study compared SafeStep™ Safety Dental Cartridge Injector to the Septodont Safety Plus predicate. Primary endpoint: prevalence of sharps injuries or safety feature failure. Results showed zero sharps injuries and zero failures of the protection mechanism. Evaluator preference and performance were equivalent to the predicate.
Technological Characteristics
Manual, mechanical dental cartridge syringe. Components: single-use disposable injector barrel, reusable plunger assembly. Materials tested for biocompatibility per ISO standards. Sterilization: gamma irradiation (25 kGy) per ANSI/AAMI/ISO 11137-1999. No electronic, software, or energy-based components.
Indications for Use
Indicated for use in patients requiring dental anesthetic injections in oral tissues. Compatible with 1.8mL pre-filled anesthetic cartridges and standard screw-on dental needles. Intended for use by dental professionals to aid in the prevention of needle stick injuries.
Regulatory Classification
Identification
A cartridge syringe is a device intended to inject anesthetic agents subcutaneously or intramuscularly. The device consists of a metal syringe body into which a disposable, previously filled, glass carpule (a cylindrical cartridge) containing anesthetic is placed. After attaching a needle to the syringe body and activating the carpule by partially inserting the plunger on the syringe, the device is used to administer an injection to the patient.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 872.9.
K013801 — ULTRASAFE DENTAL INJECTION SYSTEM · Safety Syringes, Inc. · Dec 13, 2001
K171254 — LeEject 2 Dental Syringe and Needle System · Advanced Technology and Capital, Inc. · Nov 16, 2017
K991039 — EASYSAFE RETRACTING NEEDLE SYRINGE · Saf-T-Med, Inc. · May 21, 1999
Submission Summary (Full Text)
{0}------------------------------------------------
FEB 0 5 2003
# SUMMARY OF SAFETY AND EFFECTIVENESS SECTION 16:
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and CFR 807.92.
## SUBMITTER INFORMATION 16.1
Original Premarket 510(k) Notification
| a. | Company Name: | MDC Research Ltd. |
|----|--------------------------------------|--------------------------------------------|
| b. | Company Address: | 2810 Bunsen Avenue<br>Ventura, CA 93003 |
| c. | Company Phone:<br>Company Facsimile: | (805) 339-0375<br>(805) 339-9751 |
| d. | Contact Person: | David Dowsett<br>Executive V.P. and C.O.O. |
| e. | Date Summary Prepared: | December 11, 2002 |
#### DEVICE IDENTIFICATION 16.2.
| a. Trade/Proprietary Name: | SafeStepTM Safety Dental Cartridge Injector |
|----------------------------|---------------------------------------------|
| b. Classification Name: | Cartridge Syringe<br>21 CFR 872.6770 |
# IDENTIFICATION OF PREDICATE DEVICE 16.3
| Company | Device | 510(k) No. | Date Cleared |
|-----------------------|-------------------------------------|------------|--------------|
| Specialties Septodont | Safety Plus Dental<br>Needle System | K913573 | 11/07/91 |
## 16.4 DEVICE DESCRIPTION
The SafeStep™ Safety Dental Cartridge Syringe consists of a single-use, disposable aspirating syringe-type injector with a separate re-usable plunger assembly indicated for the use in delivering dental anesthetics. The SafeStep™ Safety Dental Cartridge Syringe incorporates a needle safety mechanism that can be activated with one hand.
{1}------------------------------------------------
#### SUBSTANTIAL EQUIVALENCE 16.5
The SafeStep™ Safety Dental Cartridge Injector is substantially equivalent to the Septodont Safety Plus Dental Needle Injection System in terms of function, materials, and intended use. In a simulated use clinical study, the SafeStep™ Safety Dental Cartridge Injector was found to perform equivalently to the predicate device.
#### 16.6 INTENDED USE
The SafeStep™ Safety Dental Cartridge Injector is a sterile, single use disposable injector barrel assembly and a reusable plunger assembly that is indicated for use with pre-filled, 1.8mL, anesthetic cartridges and existing screw-on dental needles for injection of anesthetic solution in oral tissues. The SafeStep™ Safety Dental Cartridge Injector aids in the prevention of needle stick injuries. The SafeStepTM Safety Dental Cartridge Injector incorporates a needle retraction and re-advancing mechanism which may be operating with one hand.
# TECHNOLOGICAL CHARACTERISTICS 16.7
The SafeStep™ Safety Dental Cartridge Injector is equivalent to the Septodont Safety Plus in terms of materials, function and intended use. Both devices are constructed of a single use, disposable injector barrel component and a reusable plunger assembly. Both devices utilize pre-filled 1.8 mL dental anesthetic cartridges and standard screw-on dental needles.
#### 16.8 NON-CLINICAL STUDIES
Performance testing of the SafeStep™ Safety Dental Cartridge Injector included thermal stress, functional attribute, injection force, compression/tension loading before and after retraction, release force, leakage/splatter and aspiration testing. The results of the performance evaluations of the SafeStep™ Safety Dental Cartridge Injector were found to be acceptable in all test cases.
{2}------------------------------------------------
The materials of the SafeSter™ Safety Dental Cartridge Injector were tested for biocompatibility in accordance with ISO standards. The results of the biocompatibility tests, including cytotoxicity, sensitization and acute intracutaneous reactivity, were found to be acceptable.
The SafeStep™ Safety Dental Cartridge Injector is sterilized by gamma irradiation using a dosage of 25 kGy. Sterilization validation is to be performed on the final form-fill-seal production package in accordance with ANSI/AAMI/ISO 1137-1999 using Method 1 validation. MDC Research certifies that the SafeStep™ Safety Dental Cartridge Injector will pass ANSV/AAMI/ISO 11137-1994, Method 1 validation prior to marketing of the device.
#### 16.9 CLINICAL STUDIES
MDC Research Ltd. conducted a simulated use clinical study to evaluate the safety and effectiveness of the MDC SafeStep™ Safety Dental Cartridge Injector (SIDCI), for use in the injection of dental anesthetic medication into a patient. The primary research hypothesis being tested is that the use of the MDC SDCI for injection of dental anesthetic medication causes zero prevalence of sharp injuries, or failures of the safety protection feature of the device that could reasonably lead to a sharps injury. The simulated use clinical study evaluated the MDC SafeStep™ Safety Dental Cartridge Injector in comparison to the Septodont Safety Plus predicate device.
The results of the simulated use study demonstrated that the MDC SDCI is equivalent to the Septodont Safety Plus predicate device in terms of performance and evaluator preference. Further, there were no adverse events reported (i.e., sharps injuries) and no failures of the protection feature of the device that could reasonably lead to a sharps injury.
{3}------------------------------------------------
# 16.10 CONCLUSIONS
This notification contains all information required by 21 CFR 807.87. A completed copy of the Premarket Notification 510(k) Reviewer's Checklist is provided in this submission. Performance testing, non-clinical and clinical evaluations of the SafeStep™ Safety Dental Cartridge Injector show that the device performs as intended. Comparison the SafeStep™ Safety Dental Cartridge Injector to the predicate devices show that the device is substantially equivalent.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's emblem, which is a stylized representation of a human figure in profile, with three curved lines above it, resembling a bird in flight. The emblem is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB 0 5 2003
MDC Research Limited C/O Ms. Carol Patterson Patterson Consulting Group, Incorporated 21911 Erie Lane Lake Forest, California 92630
Re: K022959
Trade/Device Name: SafeStep™ Safety Dental Cartridge Injector Regulation Number: 872.6770 Regulation Name: Cartridge Syringe Regulatory Class: II Product Code: EJI Dated: December 11, 2002 Received: December 13, 2002
Dear Ms. Patterson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Ms. Patterson
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4613. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely vours.
Susan Runne
Susan Runner, DDS, MA Interim Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{6}------------------------------------------------
K022959
# INDICATION FOR USE
| 510(k) Number: | K022959 |
|----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name: | SafeStep™ Safety Dental Cartridge Injector |
| Indications for Use: | The SafeStep™ Safety Dental Cartridge Injector is a sterile, single use disposable injector barrel assembly and a reusable plunger assembly that is indicated for use with pre-filled, 1.8mL, anesthetic cartridges and existing screw-on dental needles for injection of anesthetic solution in oral tissues. The SafeStep™ Safety Dental Cartridge Injector aids in the prevention of needle stick injuries. The SafeStep™ Safety Dental Cartridge Injector incorporates a needle retraction and re-advancing mechanism which may be operated with one hand. |
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use | X |
|------------------|---|
|------------------|---|
(Per 21 CFR 801.109)
OR
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________
Ken Mnhuy for M.Sir
(Division Sign-Off)
Division of Anesthesiology, General Hospital,
Infection Control, Dental Devices
510(k) Number: K022959
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.