← Product Code [EBZ](/submissions/DE/subpart-g%E2%80%94miscellaneous-devices/EBZ) · K223414

# Curing Light, Model: MaxCure 9 (K223414)

_Guilin Refine Medical Instrument Co., Ltd. · EBZ · Nov 15, 2023 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/DE/subpart-g%E2%80%94miscellaneous-devices/EBZ/K223414

## Device Facts

- **Applicant:** Guilin Refine Medical Instrument Co., Ltd.
- **Product Code:** [EBZ](/submissions/DE/subpart-g%E2%80%94miscellaneous-devices/EBZ.md)
- **Decision Date:** Nov 15, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.6070
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Therapeutic

## Indications for Use

The MaxCure9 is intended to polymerize resinous dental materials, restorative composite materials, and orthodontic brackets, bonding and sealing materials that are photo-polymerized in the 385~515mm waveband of visible light.

## Device Story

MaxCure 9 is a dental curing light used by trained professionals to polymerize light-cured resins, composites, and orthodontic materials. The device consists of a handpiece with an integrated LED light source and a light guide. The operator selects a mode (Turbo or Normal) and irradiation time; the device emits high-intensity visible light (385-515nm) to initiate polymerization. The device is powered by a rechargeable lithium-ion battery, charged via contact. It is used in clinical settings to restore teeth or bond orthodontic brackets. The output is controlled by the operator to ensure proper curing of dental materials, directly affecting the structural integrity of the dental restoration or orthodontic appliance. Patient benefit is achieved through efficient, reliable hardening of dental materials.

## Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing, including electrical safety (ANSI AAMI ES60601-1), electromagnetic compatibility (IEC 60601-1-2), dental equipment safety (IEC 80601-2-60), and performance testing per ANSI ADA Standard No. 48-2004. Biocompatibility testing (cytotoxicity, skin sensitization, oral mucosa irritation) was performed.

## Technological Characteristics

Handheld LED curing light; wavelength 385-515nm; dual peak 400nm/452nm. Powered by rechargeable lithium-ion battery. Contact-based charging. Patient-contacting material: PE (polyethylene). Standards: IEC 60601-1, IEC 60601-1-2, IEC 80601-2-60, ANSI ADA No. 48, ISO 10993-5/10. Non-sterile.

## Regulatory Identification

An ultraviolet activator for polymerization is a device that produces ultraviolet radiation intended to polymerize (set) resinous dental pit and fissure sealants or restorative materials by transmission of light through a rod.

## Predicate Devices

- Diadent Group International, D-Lux+ ([K200809](/device/K200809.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 15, 2023

Guilin Refine Medical Instrument Co., Ltd. % Kevin Wang Consultant Chonconn Medical Device Consulting Co., Ltd. Room 504, Block C, No. 1029 Nanhai Avenue, Nanshan District Shenzhen, Guangdong 518067 China

Re: K223414

Trade/Device Name: Curing Light, Model: MaxCure 9 Regulation Number: 21 CFR 872.6070 Regulation Name: Ultraviolet Activator For Polymerization Regulatory Class: Class II Product Code: EBZ Dated: August 14, 2023 Received: August 15, 2023

Dear Kevin Wang:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Bobak Shirmoham madi -S

For Michael E. Adjodha, M. ChE., CQIA Assistant Director DHT1B: Division of Dental and

{2}------------------------------------------------

ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

## Indications for Use

510(k) Number (if known) K223414

Device Name Curing Light, Model: MaxCure 9

### Indications for Use (Describe)

The MaxCure9 is intended to polymerize resinous dental materials, restorative composite materials, and orthodontic brackets, bonding and sealing materials that are photo-polymerized in the 385~515mm waveband of visible light.

| Type of Use (Select one or both, as applicable)                                                                                                                                                                                                                                                                                                                          |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span> <span style="padding-right: 10px;"> <span style="border: 1px solid black; padding: 0px; margin-right: 2px;"><b>X</b></span>Prescription Use (Part 21 CFR 801 Subpart D)                 </span> <span> <span style="border: 1px solid black; padding: 0px; margin-right: 2px;"></span>Over-The-Counter Use (21 CFR 801 Subpart C)                 </span> </span> |

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services  

Food and Drug Administration  

Office of Chief Information Officer  

Paperwork Reduction Act (PRA) Staff  

[PRAStaff@fda.hhs.gov](mailto:PRAStaff@fda.hhs.gov)

*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."*

EF

PSC Publishing Services (301) 443-6740

{4}------------------------------------------------

## 510(K) Summary

Prepared in accordance with the requirements of 21 CFR Part 807.92

### Prepared Date: 2023/11/9

#### 1. Submission sponsor

Name: Guilin Refine Medical Instrument Co., Ltd.

Address: No.8-3, Information Industrial Park, High-Tech Zone, Qixing District, Guilin, GuangXi, China 541004

Contact person: Chen Judong

Title: Management Representative

E-mail: chenjudong(@refine-med.com

#### 2. Submission correspondent

Name: Chonconn Medical Device Consulting Co., Ltd.

Address: Room 504, Block C, No. 1029 Nanhai Avenue, Nanshan District, Shenzhen, Guangdong, P. R. China 518067

Contact person: Kevin Wang E-mail: kevin(@chonconn.com

Tel: +86-755 33941160

| Trade/Device Name | Curing Light, Model: MaxCure 9                                     |
|-------------------|--------------------------------------------------------------------|
| Model             | MaxCure9                                                           |
| Common Name       | Curing Light                                                       |
| Regulatory Class  | Class II                                                           |
| Classification    | 21 CFR 872.6070 / Activator, Ultraviolet, For Polymerization / EBZ |
| Submission type   | Traditional 510(K)                                                 |

#### 3 Subiect Device Information

#### 4. Predicate Device

- 1. Diadent Group International, D-Lux+ under K200809 (Primary predicate)
#### Device Description ട്.

This Curing Light is a device which generates high intensity light for polymerization of light curing materials used for dental curing purpose. This curing light is intended for use by trained dental professionals for the purpose of light curing dental resins and composites and used to restore teeth.

#### Intended use & Indication for use 6.

The MaxCure9 is intended to polymerize resinous dental materials, restorative composite materials, and orthodontic brackets, bonding and sealing materials that are photo-polymerized in the 385~515nm waveband of visible light.

Section 5 510(k) Summary

{5}------------------------------------------------

| Features                                                                         | Predicate device<br>D-Lux+                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Subject device<br>Curing Light, Model:<br>MaxCure 9 (K223414)                                                                                                                                                                                                                                                                                                                                                                                  | Note                                                                                                                  |
|----------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|
| 510(k) number                                                                    | K200809                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | \                                                                                                                                                                                                                                                                                                                                                                                                                                              | Same                                                                                                                  |
| Product code                                                                     | EBZ                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | EBZ                                                                                                                                                                                                                                                                                                                                                                                                                                            | Same                                                                                                                  |
| Intended Use &<br>Indications for<br>use                                         | The D-Lux+ is intended to<br>polymerize resinous dental<br>materials, restorative composite<br>materials, and orthodontic<br>brackets, bonding and sealing<br>materials that are photo-<br>polymerized in the 385~515nm<br>waveband of visible light.                                                                                                                                                                                                                              | The MaxCure9 is intended<br>to polymerize resinous<br>dental materials, restorative<br>composite materials, and<br>orthodontic brackets,<br>bonding and sealing<br>materials that are photo-<br>polymerized in the<br>385~515nm waveband of<br>visible light.                                                                                                                                                                                  | Same                                                                                                                  |
| Principles of<br>operation                                                       | 1. Insert the Light Probe into the<br>handpiece. Attach the Light<br>Protector to the handpiece.<br>2. Cover the Light Probe with the<br>Disposable Sheath.<br>3. Turn on the power and select<br>the mode and set-up time.<br>4. Press the Operation button to<br>perform light irradiation for the<br>set time and press the button<br>once again to stop the light<br>irradiation. The light irradiation<br>progress time is displayed<br>numerically in the display<br>window. | a) Insert the light guide into<br>the mounting hole of the<br>curing light. Clamp and<br>install the light hood at the<br>root of the light guide.<br>b) Cover the Light Guide<br>with the disposable isolation<br>sleeve.<br>c)Turn on the power and<br>select the mode and set-up<br>time.<br>d) Press the mode key to<br>perform light irradiation for<br>the set time and press the<br>button once again to stop<br>the light irradiation. | Same                                                                                                                  |
| Delivery form                                                                    | -D-Lux+ Handpiece<br>-D-Lux+ Charger<br>-Light Probe<br>-Light Protector<br>-C-Battery (Included in the<br>handpiece)<br>-Adapter<br>-Power Cord<br>-Disposable Sheaths(200ea/Box)                                                                                                                                                                                                                                                                                                 | Main unit,<br>Charger,<br>Light hood,<br>Power adapter,<br>disposable PE isolation<br>sleeve (100ea/Box),<br>Instruction Manual.                                                                                                                                                                                                                                                                                                               | Same                                                                                                                  |
| Features                                                                         | Predicate device<br>D-Lux+                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Subject device<br>Curing Light, Model:<br>MaxCure 9                                                                                                                                                                                                                                                                                                                                                                                            | Note                                                                                                                  |
|                                                                                  | -User Manual                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                       |
| Technical<br>Specifications –<br>Wavelength<br>range                             | 385 – 515 nm                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | 385 – 515 nm                                                                                                                                                                                                                                                                                                                                                                                                                                   | Same                                                                                                                  |
| Peak<br>wavelength                                                               | Dual peak:<br>405nm, 460nm                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Dual peak:<br>400nm, 452 nm                                                                                                                                                                                                                                                                                                                                                                                                                    | SE, The<br>values are<br>similar. This<br>difference<br>does not<br>raise issues<br>of<br>substantial<br>equivalence. |
| Depth of cure                                                                    | ≥2mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | ≥2mm                                                                                                                                                                                                                                                                                                                                                                                                                                           | Same                                                                                                                  |
| Light<br>source/Probe                                                            | a light probe with an embedded<br>LED                                                                                                                                                                                                                                                                                                                                                                                                                                              | a light probe with an<br>embedded LED                                                                                                                                                                                                                                                                                                                                                                                                          | Same                                                                                                                  |
| Operation<br>Modes and<br>Light Intensity                                        | 5 modes:<br>STD (Standard) mode :900 mW/cm²,<br>5, 10, 15 and 20 seconds<br>SFT (Soft Start) mode: 1,300<br>mW/cm²,<br>10 and 15 seconds<br>HIG (High Power) mode: 1,300<br>mW/cm²,<br>5, 10 and 15 seconds<br>ORT (Orthodontic) mode: 1,800<br>mW/cm²,<br>3, 4, 5 seconds and P<br>MAX (Max Power) mode:2,400<br>mW/cm²,<br>1, 2 and 3 seconds                                                                                                                                    | 2 modes:<br>Turbo mode: 2000 mW/cm2,<br>1S, 3S<br>Normal mode: 1000<br>mW/cm2,<br>5S, 10S, 15S, 20S                                                                                                                                                                                                                                                                                                                                            | Different 1                                                                                                           |
| Cross Infection<br>Prevent                                                       | disposable sheath                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | disposable PE isolation<br>sleeve                                                                                                                                                                                                                                                                                                                                                                                                              | Same                                                                                                                  |
| Use                                                                              | Prescription / Hospital                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Prescription / Hospital                                                                                                                                                                                                                                                                                                                                                                                                                        | Same                                                                                                                  |
| Sterility                                                                        | Non-sterile                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Non-sterile                                                                                                                                                                                                                                                                                                                                                                                                                                    | Same                                                                                                                  |
| Power Source                                                                     | rechargeable lithium ion battery                                                                                                                                                                                                                                                                                                                                                                                                                                                   | rechargeable lithium ion                                                                                                                                                                                                                                                                                                                                                                                                                       | Same                                                                                                                  |
| Features                                                                         | Predicate device<br>D-Lux+                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Subject device<br>Curing Light, Model:<br>MaxCure 9                                                                                                                                                                                                                                                                                                                                                                                            | Note                                                                                                                  |
| Battery<br>Charging                                                              | wirelessly charged                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | charging method by the<br>contact                                                                                                                                                                                                                                                                                                                                                                                                              | Different 2                                                                                                           |
| Power supply                                                                     | 100-240VAC, 50-60Hz                                                                                                                                                                                                                                                                                                                                                                                                                                                                | 100V-240V AC 50Hz/60Hz                                                                                                                                                                                                                                                                                                                                                                                                                         | Same                                                                                                                  |
| outputs                                                                          | 6VDC/2A                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | DC 5V/1A                                                                                                                                                                                                                                                                                                                                                                                                                                       | Different 2                                                                                                           |
| Electrical Safety                                                                | IEC60601-1 and IEC 60601-1-1-<br>2                                                                                                                                                                                                                                                                                                                                                                                                                                                 | IEC60601-1 and IEC<br>60601-1-1-2                                                                                                                                                                                                                                                                                                                                                                                                              | Same                                                                                                                  |
| Chemical<br>composition of<br>patient<br>contacting<br>portions of the<br>device | N/A                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | PE                                                                                                                                                                                                                                                                                                                                                                                                                                             | Different 3                                                                                                           |
| Biocompatibility                                                                 | Direct contact with issue in not<br>intended. Therefore ISO10993-1<br>is not applicable                                                                                                                                                                                                                                                                                                                                                                                            | ISO 10993-5<br>ISO 10993-10                                                                                                                                                                                                                                                                                                                                                                                                                    | Different 3                                                                                                           |

# 7. Comparison to the Predicate Device

{6}------------------------------------------------

{7}------------------------------------------------

Different (1): Operation Modes and Light Intensity

The subject device has several modes corresponding to the light output intensity and available times. The light output safety and performance test was conducted according to IEC 60601-1, IEC60601-1-2 and FDA guidance performance testing requirements with the difference. The testing results show that these differences do not affect safety and effectiveness .

Different (2): Battery Charging and Power Supply

The subject device is charged by the charging method of contact. However, the predicate device is wirelessly charged without a contact using a wireless charger. As a power source for the charger, the AC/DC adapter provided by the manufacturer is used. The input specifications of the adapter(100-240VAC, 50-60Hz) are the same, but the outputs differ from each other at 5VDC/1A and 6VDC/2A respectively. These differences do not affect safety and effectiveness . Different (3): Biocompatibility

Biocompatibility testing of the subject device was conducted because although the device does not come in contact with oral tissue on the chance that contact does occur. However, the predicate device is not evaluated. The testing results show that these differences do not affect safety and effectiveness.

#### 8. Performance Data

The following performance data were provided in support of the substantial equivalence determination.

## Biocompatibility testing

Biocompatibility testing was conducted because although the device does not come in contact with oral tissue on the chance that contact does occur.

The biocompatibility evaluation for the Curing light was conducted in accordance with the FDA Guidance Section 5 510(k) Summary 5- 5 / 6

{8}------------------------------------------------

for Industry and Food and Drug Administration Staff: Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process". The battery of testing included the following tests:

- Cytoxicity
- Skin Sensitization
- Oral Mucosa Irritation

## Software Validation

The Curing light software has been fully validated and the device software meets its Design Specifications. FDA Guidance Document Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices has been followed in connection with the design of the device.

## Non-clinical data

The Curing light has been tested according to the following standards:

- ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012 (Consolidated Text) Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD)
- IEC 60601-1-2:2014 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests
- IEC 80601-2-60 Edition 2.0 2019-06 Medical electrical equipment - Part 2-60: Particular requirements for the basic safety and essential performance of dental equipment
- ANSI ADA Standard No. 48-2004 (R2015) Visible Light Curing Units

## Clinical data

The subject of this premarket submission, Curing light, did not require clinical studies to support substantial equivalence.

#### 9. Conclusion

Based on the similarities in intended use, principles of operation, design rationale, test results, and performance, the subject Curing light is considered to be substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/DE/subpart-g%E2%80%94miscellaneous-devices/EBZ/K223414](https://fda.innolitics.com/submissions/DE/subpart-g%E2%80%94miscellaneous-devices/EBZ/K223414)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
