eXciteOSA without remote control (3000); eXciteOSA with remote control (6000)

K223446 · Signifier Medical Technologies, Ltd. · QNO · Jan 18, 2023 · Dental

Device Facts

Record IDK223446
Device NameeXciteOSA without remote control (3000); eXciteOSA with remote control (6000)
ApplicantSignifier Medical Technologies, Ltd.
Product CodeQNO · Dental
Decision DateJan 18, 2023
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 872.5575
Device ClassClass 2
AttributesTherapeutic

Indications for Use

eXciteOSA® is a removable tonque muscle that delivers neuromuscular stimulation to the tongue in order to reduce mild obstructive sleep apnea (AHI <15) and snoring for patients that are 18 years or older.

Device Story

Removable tongue muscle stimulation device; delivers neuromuscular stimulation via small electric currents to tongue muscles; improves tongue function to maintain upper airway patency during sleep and reduce throat vibration. Mouthpiece features two tabs positioned above and below tongue; held in place by gentle mouth closure. Operated by patient; used at any time during day. Controlled via smartphone application (iOS/Android) using Bluetooth Smart (BLE 4.0). Output consists of electrical stimulation pulses; healthcare provider uses device to treat mild OSA and snoring; patient benefits from reduced apnea events and snoring.

Clinical Evidence

No clinical or non-clinical tests were performed for this submission as the device is identical to the predicate with no physical changes.

Technological Characteristics

Removable mouthpiece with two tabs for tongue positioning; neuromuscular electrical stimulation; Bluetooth Smart (BLE 4.0) connectivity for smartphone app control; software-driven stimulation parameters.

Indications for Use

Indicated for patients 18 years or older with mild obstructive sleep apnea (AHI <15) and snoring.

Regulatory Classification

Identification

A neuromuscular tongue muscle stimulator for the reduction of snoring and obstructive sleep apnea consists of a removable intraoral mouthpiece that uses electrodes to deliver neuromuscular stimulation to the tongue to strengthen tongue musculature to reduce snoring and obstructive sleep apnea.

Special Controls

In combination with the general controls of the FD&C Act, the neuromuscular tongue stimulator for reduction of snoring and obstructive sleep apnea is subject to the following special controls:

*Classification.* Class II (special controls). The special controls for this device are:(1) Performance testing must demonstrate the wireless compatibility, electrical safety, battery safety, and electromagnetic compatibility of the device in its intended use environment. (2) Software verification, validation, and hazard analysis must be performed. (3) The patient-contacting components of the device must be demonstrated to be biocompatible. (4) Patient labeling must include: (i) Information on device components, setup, and use of the device including placement of sensors and mouthpieces, and images or illustrations; (ii) A summary of technical specifications; (iii) Instructions on how to clean and maintain the device; (iv) A statement that the patient should maintain regular follow up visits with dentist and sleep specialist; and (v) A statement that patients should have a comprehensive dental examination prior to using this device. (5) A human factors assessment must evaluate simulated use of the device to demonstrate that the user can correctly use device based on the labeling and instructions for use.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized emblem. To the right is the FDA logo, with the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text. January 18, 2023 Signifier Medical Technologies Ltd % Darren Scheer Principal RegChoice LLC 13014 N. Dale Mabry Hwy STE 803 Tampa, Florida 33618 Re: K223446 Trade/Device Name: eXciteOSA without remote control (3000); eXciteOSA with remote control (6000) Regulation Number: 21 CFR 872.5575 Regulation Name: Neuromuscular tongue muscle stimulator for the reduction of snoring and obstructive sleep apnea Regulatory Class: Class II Product Code: QNO Dated: November 14, 2022 Received: December 19, 2022 Dear Darren Scheer: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of {1}------------------------------------------------ Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Michael E. Adjodha -S Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. Submission Number (if known) K223446 Device Name eXciteOSA without remote control (3000); eXciteOSA with remote control (6000) Indications for Use (Describe) eXciteOSA® is a removable tongue muscle stimulation device that delivers neuromuscular stimulation to the tongue in order to reduce mild obstructive sleep apnea (AHI <15) and snoring for patients that are 18 years or older. Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary Prepared on: 2022-11-14 | | | | 21 CFR 807.92(a)(1) | |---------------------------------|--------------------------------------------------------------|--|---------------------| | Applicant Name | Signifier Medical Technologies Ltd | | | | Applicant Address | 5-17 Hammersmith Grove London W6 0LG United Kingdom | | | | Applicant Contact Telephone | 442070960586 | | | | Applicant Contact | Mr. Yasser Zayni | | | | Applicant Contact Email | yasser.z@signifiermedical.com | | | | Correspondent Name | RegChoice LLC | | | | Correspondent Address | 13014 N. Dale Mabry Hwy STE 803 Tampa FL 33618 United States | | | | Correspondent Contact Telephone | 8133633004 | | | | Correspondent Contact | Dr. Darren Scheer | | | | Correspondent Contact Email | dscheer@regchoice.com | | | Device Name | | | | 21 CFR 807.92(a)(2) | |---------------------|-------------------------------------------------------------------------------------------------|--|---------------------| | Device Trade Name | eXciteOSA without remote control (3000); eXciteOSA with remote control (6000) | | | | Common Name | Neuromuscular tongue muscle stimulator for the reduction of snoring and obstructive sleep apnea | | | | Classification Name | Neuromuscular Tongue Muscle Stimulator For The Reduction Of Snoring And Obstructive Sleep Apnea | | | | Regulation Number | 872.5575 | | | | Product Code | QNO | | | Legally Marketed Predicate Devices | | | | 21 CFR 807.92(a)(3) | |-------------|----------------------------------------------------------------|--------------|---------------------| | Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code | | | DEN200018 | eXciteOSA without remote control, eXciteOSA with remote contre | QNO | | Device Description Summary | | | | 21 CFR 807.92(a)(4) | |--|--|--|---------------------| |--|--|--|---------------------| eXciteOSA® is a removable tongue muscle stimulation device that delivers neuromuscular stimulation to the to reduce mild obstructive sleep apnea (AHI <15) and snoring for patients that are 18 years or older. The small electric currents delivered through the mouthpiece stimulate the tongue and improved function. The improved function of the tongue muscle will help in keeping the upper airway open during sleep and reduce the vibration of the throat region. Two tabs of the mouthpiece sit comfortably above and below the tongue. The mouthpiece is designed such that when the mouth is gently closed, it will naturally sit around the tongue and won't move during the therapy session. The device can be used at any point during the day. The device can be controlled by a smartphone application. This app can be downloaded from the App store (Apple iOS) or Play Store {4}------------------------------------------------ (Google Android). The mobile app software can be used on iPhone 6/6 Plus, iPhone 6/6s Plus, iPhone 7/7 Plus, iPhone 8/8 Plus, iPhone X, with iOS 11.0 and higher. The mobile app software can also be used with Android devices with BLE support (Bluetooth 4.0) and Android 7.0 and above. eXciteOSA® uses Bluetooth Smart; mobile devices used must be compatible with Bluetooth Smart. ## Intended Use/Indications for Use eXciteOSA® is a removable tonque muscle that delivers neuromuscular stimulation to the tongue in order to reduce mild obstructive sleep apnea (AHI <15) and snoring for patients that are 18 years or older. ## Indications for Use Comparison The indications for use are the same as the predicate device. ## Technological Comparison Other than for the deletion of some contraindications (difference #1) and for minor software/firmware updates (difference #2), this subject device has the same technological characteristics of the predicate device. The first difference, in contraindications, is not critical to the intended therapeutic use of the three supporting sources of data and the associated risk analysis. The second difference, in software/firmware minor updates, does not rise to the need of submitting a 510(k) in accordance with FDA Guidance Deciding When to Submit a 510(k) for a Software Change to an Existing Device. These changes are justified via risk analysis. #### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b) Non-clinical and/or clinical tests were not performed, as the subject device are the same and no physical changes were made. # 21 CFR 807.92(a)(5) ## 21 CFR 807.92(a)(5) ### 21 CFR 807.92(a)(6)
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...