← Product Code [PLV](/submissions/DE/subpart-e%E2%80%94surgical-devices/PLV) · K211466

# Yomi Robotic System with YomiPlan Go (K211466)

_Neocis, Inc. · PLV · Jun 14, 2022 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/DE/subpart-e%E2%80%94surgical-devices/PLV/K211466

## Device Facts

- **Applicant:** Neocis, Inc.
- **Product Code:** [PLV](/submissions/DE/subpart-e%E2%80%94surgical-devices/PLV.md)
- **Decision Date:** Jun 14, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.4120
- **Device Class:** Class 2
- **Review Panel:** Dental

## Indications for Use

Yomi is a computerized robotic navigational system intended to provide assistance in both the planning (pre-operative) and the surgical (intra-operative) phases of dental implantation surgery. The system provides software to preoperatively plan dental implantation procedures and provides robotic navigational guidance of the surgical instruments. Yomi is intended for use in partially edentulous and fully edentulous adult patients who qualify for dental implants. When YomiPlan software is used for preplanning on third party PCs, it is intended to perform the planning (pre-operative) phase of dental implantation surgery. Yomi Plan provides pre-operative planning for dental implantation procedures using the Yomi Robotic System. The output of Yomi Plan is to be used with the Yomi Robotic System.

## Device Story

Yomi Robotic System with YomiPlan Go is a computerized robotic navigational system for dental implantation. Unlike the predicate, YomiPlan Go enables surgery without uploading a preoperative CT scan. The surgeon uses the robotic arm drill tip as a pointer to define the implant trajectory and osteotomy site based on direct visualization of patient anatomy; the system maintains this trajectory during the procedure. Used in dental clinics by dentists; provides robotic guidance, depth stop, and steadying of the drill to prevent skiving. Output is real-time navigational guidance for surgical instruments. Benefits include reduced radiation exposure, faster workflow, and improved precision compared to freehand techniques. Clinical decision-making relies on surgeon's visual assessment and medical training, optionally supplemented by mental navigation using pre-operative X-rays or CTs.

## Clinical Evidence

Prospective, multi-center, single-arm study (G210363) with 15 patients (44 implants). Evaluated human factors and accuracy (pre-op vs. post-op CBCT). Results: Depth error 1.38 ± 0.71 mm, lateral error 2.04 ± 1.59 mm, angular error 12.91 ± 7.56°. No serious adverse events. One minor event (implant into sinus) attributed to user error. Investigators confirmed design/performance requirements met; 3rd party review confirmed no critical structure damage.

## Technological Characteristics

Robotic navigational system; Windows 10-based GUI; connectivity via Ethernet/Wi-Fi/USB. Precision specification: Upper 95% probability with 95% model fit < 1.00 mm. Wireless coexistence tested per AAMI TIR69:2017 and IEEE ANSI C63.27-2017. Software level of concern: Moderate.

## Regulatory Identification

A bone cutting instrument and accessories is a metal device intended for use in reconstructive oral surgery to drill or cut into the upper or lower jaw and may be used to prepare bone to insert a wire, pin, or screw. The device includes the manual bone drill and wire driver, powered bone drill, rotary bone cutting handpiece, and AC-powered bone saw.

## Predicate Devices

- Neocis Guidance System (NGS) with YomiPlan v2.0 ([K202264](/device/K202264.md))

## Reference Devices

- Orthosoft Navitrack System-Optical TKR CT-Less System ([K021760](/device/K021760.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Neocis Inc. William Tapia VP RA/QA 2800 Biscayne Blvd Suite 600 Miami, Florida 33137

Re: K211466

Trade/Device Name: Yomi Robotic System with YomiPlan Go Regulation Number: 21 CFR 872.4120 Regulation Name: Bone Cutting Instrument And Accessories Regulatory Class: Class II Product Code: PLV Dated: May 20, 2022 Received: May 20, 2022

Dear William Tapia:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Andrew I. Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K211466

Device Name Yomi Robotic System with YomiPlan Go

### Indications for Use (Describe)

Yomi is a computerized robotic navigational system intended to provide assistance in both the planning (pre-operative) and the surgical (intra-operative) phases of dental implantation surgery. The system provides software to preoperatively plan dental implantation procedures and provides robotic navigational guidance of the surgical instruments. Yomi is intended for use in partially edentulous adult patients who qualify for dental implants.

When YomiPlan software is used for preplanning on third party PCs, it is intended to perform the planning (pre-operative) phase of dental implantation surgery. Yomi Plan provides pre-operative planning for dental implantation procedures using the Yomi Robotic System. The output of Yomi Plan is to be used with the Yomi Robotic System.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

# CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Neocis. The word "neocis" is written in a sans-serif font, with the letters in gold. Above the word is a gray horizontal line that extends to the left and right of the word. To the right of the word is a gray vertical line that intersects the horizontal line, forming a cross shape.

# 510(k) Summary K211466

#### l. Submitter

Neocis Inc.

2800 Biscayne Blvd.

Suite 600

Miami, FL 33137

Tel: 1-855-9NEOCIS

Contact Person: William Tapia, VP RA/QA

Date Prepared: June 14, 2022

### II. Device

Trade Name: Yomi Robotic System with YomiPlan Go

Common Name: Dental Stereotaxic Instrument

Classification Name: Bone cutting instrument and accessories (21 CFR 872.4120)

Classification: Class II

Product Code: PLV

#### lll. Predicate Device

Neocis Guidance System (NGS) with YomiPlan v2.0 (K202264)

#### IV. Reference Device

Orthosoft Navitrack System-Optical TKR CT-Less System (K021760)

#### V. Indications for Use

Yomi is a computerized robotic navigational system intended to provide assistance in both the planning (pre-operative) and the surgical (intra-operative) phases of dental implantation surgery. The system provides software to preoperatively plan dental implantation procedures and provides robotic navigational guidance of the surgical instruments. Yomi is intended for use in partially edentulous and fully edentulous adult patients who qualify for dental implants.

When YomiPlan software is used for preplanning on third party PCs, it is intended to perform the planning (pre-operative) phase of dental implantation surgery. Yomi Plan provides pre-operative planning for dental implantation procedures using the Yomi Robotic System. The output of Yomi Plan is to be used with the Yomi Robotic System.

#### VI. Device Description

The subject of this submission is YomiPlan Go, a feature of the Yomi Robotic System which enables the use of the system without the uploading of a preoperative CT scan. The dynamic planning feature in K202264 requires a pre-operative CT scan for use. This submission includes a new workflow called YomiPlan Go and provides instructions on how to use this feature without the need of uploading a CT scan to the Yomi system. This planning involves placing the robotic arm drill tip to the point where an osteotomy is to be performed i.e., the surgeon performs

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Neocis. The word "neocis" is written in a sans-serif font, with the "neo" portion in a yellow-orange color and the "cis" portion in gray. A gray horizontal line extends above the word, and a gray vertical line intersects it at the "o" in "neocis", forming a crosshair-like symbol.

planning with their direct visualization of the anatomy and with the use of the robotic arm to select where the osteotomy is to be performed. YomiPlan Go gives surgeons the ability to perform an osteotomy under robotic guidance at the point that the surgeon selects on the patient's anatomy. The selected point, axis, and trajectory are maintained by the robotic arm while the surgeon performs the osteotomy.

# Principal of Operation

YomiPlan Go gives surgeons the ability to place implants on the trajectory defined by the axis and location of the robotic arm drill tip for any selected implant, without the use of a preoperative CT scan uploaded to the Yomi System. This feature allows surgeons to place implants with precision without a preloaded CT scan. With YomiPlan Go, the surgeon selects the implant and then creates the desired path for the planned osteotomy by positioning the robotic arm drill tip at the site of where the osteotomy is to be performed.

### Intended Patient Population

Based on a clinical study performed with YomiPlan Go (G210363), the intended population for YomiPlan Go was determined to be:

- . Healthy adults 18 years of age or older with sufficient bone height and width appropriate for dental implant surgery
- Partially or fully edentulous patients ●

# Preplanning with YomiPlan Go

Preplanning with YomiPlan Go is similar to that of traditional freehand technique as no CT image is actively referenced within YomiPlan. It is important to note that this does not preclude the surgeon from taking a pre-operative image and using it mentally as consideration for the freehand planning. The surgeon uses the system to indicate the plan on the patient by observing the patient's specific anatomy that is plainly visible.

The clinical study performed via G210363 demonstrated that YomiPlan Go could be used to successfully capture the trajectory of the osteotomy based mainly on the drill bit tip established by the surgeon and was adequate to provide placement of the implant at the position determined by the surgeon. In the study, YomiPlan Go was able to provide guidance which used a single point with a trajectory and resulting in placement of the implant at the position established by the surgeon.

### Labeling for YomiPlan Go

Labeling (i.e., instructions for use, training) is an important part of YomiPlan Go. The following are key points from these materials:

- . We recommend preoperative CT as standard of care for consultation and diagnosis. No imaging is required by Yomi Go software for preoperative planning or intraoperative guidance.
- . Accuracy is not a specification or an advantage of YomiPlan Go, however precision is an important specification which has been validated through performance testing.
- . The surgeon must use clinical judgment in determining the best plan (position, angle, and depth) based on any anatomy that is plainly visible as well as any pre-operative X-rays or CT they may have taken of the patient.
- Use of YomiPlan Go vields less position accuracy relative to a CBCT than the guided technique for osteotomy placement in which CBCT scanning and digital intraoral impressions are used

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Neocis. The word "neocis" is written in a sans-serif font, with the letters in yellow. Above the word is a gray horizontal line that extends to the left and right of the word. At the end of the horizontal line is a gray vertical line with a circle in the middle.

to generate a virtual representation of the patient's jaw and oral anatomy, and the exact position of the implant is determined in advance of treatment.

### Use of Anatomical Cues to Define Trajectory

The implant position and angle are defined by the surgeon by using the drill tip as a pointer. This enables the surgeon to indicate to the system the desired position and angle based on any anatomical cues they believe are relevant, e.g., neighboring teeth, bone ridge, extraction socket, etc. It is up to the surgeon to plan the implant appropriately based on the patient's clinical condition and rely on their medical training and ability to visualize the relevant anatomy of the patient.

### VII. Substantial Equivalence Discussion

The table below summarizes a comparison of YomiPlan Go with the predicate NGS with Yomi Plan 2.0 (K202264).

|                                                  | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                               |
|--------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Feature                                          | Yomi Robotic System with YomiPlan Go                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | NGS with Yomi Plan v2.0                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | SE Analysis                                                                                                                                                                                                   |
|                                                  | K211466                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | K202264                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                                                                                                                                                                                                               |
| Indications for use                              | Yomi is a computerized robotic navigational system<br>intended to provide assistance in both the planning<br>(pre-operative) and the surgical (intra- operative)<br>phases of dental implantation surgery. The system<br>provides software to preoperatively plan dental<br>implantation procedures and provides robotic<br>navigational guidance of the surgical instruments.<br>Yomi is intended for use in partially edentulous and<br>fully edentulous adult patients who qualify for dental<br>implants.                                                                                                                                                                                                                             | The Neocis Guidance System (NGS) is a<br>computerized navigational system intended to<br>provide assistance in both the planning (pre-<br>operative) and the surgical (intra-operative)<br>phases of dental implantation surgery. The<br>system provides software to preoperatively<br>plan dental implantation procedures and<br>provides navigational guidance of the surgical<br>instruments. The NGS is intended for use in<br>partially edentulous and fully edentulous adult<br>patients who qualify for dental implants. | Equivalent - Updated to<br>include reference to<br>robotic arm.                                                                                                                                               |
|                                                  | When Yomi Plan software is used for preplanning on<br>third party PCs, it is intended to perform the<br>planning (pre-operative) phase of dental<br>implantation surgery. Yomi Plan provides pre-<br>operative planning for dental implantation<br>procedures using the Yomi Robotic System. The<br>output of Yomi Plan is to be used with the Yomi<br>Robotic System.                                                                                                                                                                                                                                                                                                                                                                    | When Yomi Plan software is used for<br>preplanning on third party PCs, it is intended<br>to perform the planning (pre-operative) phase<br>of dental implantation surgery. Yomi Plan<br>provides pre- operative planning for dental<br>implantation procedures. The output of Yomi<br>Plan is to be used with the Neocis Guidance<br>System (NGS).                                                                                                                                                                               |                                                                                                                                                                                                               |
| Contraindications                                | Most surgical dental implant placement involves<br>three-dimensional placement into a non-visible<br>space. Two-dimensional radiographs may provide<br>important information, but they often lack the oral-<br>facial dimension that completes the picture. more-<br>detailed information from 3D x-ray imaging is<br>needed, most frequently, CBCT imaging.<br>Complications in implant dentistry are common and<br>can range from minor to severe. One of the most<br>common complications in implant dentistry is<br>implant malposition. This can result in a host of<br>complications: fenestrations, dehiscences and<br>recession, difficult restoration, poor biomechanics,<br>loss of interproximal structure, damage to adjacent | None                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Different - difference<br>validated as acceptable<br>with successful<br>completion of G210363                                                                                                                 |
|                                                  | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                               |
| Feature                                          | Yomi Robotic System with YomiPlan Go                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | NGS with Yomi Plan v2.0                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | SE Analysis                                                                                                                                                                                                   |
|                                                  | K211466                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | K202264                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                                                                                                                                                                                                               |
|                                                  | anatomy, and many others.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                               |
|                                                  | Use of the YomiPlan Go is contraindicated when the<br>intended implant location has concerns related to<br>bone volume, architecture, and proximity to critical<br>anatomic structures, such as nerves, maxillary sinus<br>or nasal floor or neurovasculature.                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                               |
| OS                                               | Windows 10                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Windows 10                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Identical                                                                                                                                                                                                     |
| PC Requirements                                  | PC with 64-bit Windows 10 OS or newer with a<br>minimum of 4 GB of RAM and a 2 GHz dual core<br>processor. Local memory (hard drive) should be a<br>minimum of 100 GB with 7200 RPM or SSD.<br>Connectivity requirements include ethernet, Wi-Fi,<br>USB, or CD drive.                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | PC with 64-bit Windows 10 OS or newer with<br>a minimum of 4 GB of RAM and a 2 GHz dual<br>core processor. Local memory (hard drive)<br>should be a minimum of 100 GB with 7200<br>RPM or SSD. Connectivity requirements<br>include ethernet, Wi-Fi, USB, or CD drive.                                                                                                                                                                                                                                                          | Identical                                                                                                                                                                                                     |
| Preplanning Steps                                | Preplanning is performed by the surgeon<br>by direct visualization of the patient's<br>anatomy and placing the drill tip at the<br>desired location and angle. The dimension<br>of the implant is identified in YomiPlan Go<br>and the trajectory established by the<br>surgeon is captured.                                                                                                                                                                                                                                                                                                                                                                                                                                              | The predicate device uses a<br>preoperative CBCT scan which is<br>uploaded to YomiPlan to support<br>a surgeon's planning of the desired<br>implant placement.                                                                                                                                                                                                                                                                                                                                                                  | Equivalent                                                                                                                                                                                                    |
| Level of Concern                                 | Moderate                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Moderate                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Identical                                                                                                                                                                                                     |
| Installation                                     | Windows Installer .msi file                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Windows Installer .msi file                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Identical                                                                                                                                                                                                     |
| Origin                                           | Proprietary                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Proprietary                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Identical                                                                                                                                                                                                     |
| OTS Software                                     | TeamViewer                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | TeamViewer                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Identical                                                                                                                                                                                                     |
| Hardware                                         | No changes (activate/use pre-existing network<br>features)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | No changes (activate/use pre-existing<br>network features)                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Identical                                                                                                                                                                                                     |
| Wireless data<br>transmission over<br>LAN        | Yes, via integrated hardware, tested according to:<br>AAMI TIR69: 2017 Technical Information<br>Report Risk management of radio-<br>frequency wireless coexistence for<br>medical devices and systems. IEEE ANSI C63.27-2017 American National<br>Standard for Evaluation of Wireless<br>Coexistence                                                                                                                                                                                                                                                                                                                                                                                                                                      | Yes, via integrated hardware, tested<br>according to:  AAMI TIR69: 2017 Technical<br>Information Report Risk<br>management of radio-frequency<br>wireless coexistence for medical<br>devices and systems. IEEE ANSI C63.27-2017 American<br>National Standard for Evaluation of<br>Wireless Coexistence                                                                                                                                                                                                                         | Identical                                                                                                                                                                                                     |
| Interface                                        | Windows based GUI                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Windows based GUI                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Identical                                                                                                                                                                                                     |
| System Accuracy /<br>Precision<br>Specifications | Accuracy: N/A for YomiPlan Go<br><br>Precision: Upper 95% Probability with 95% Model<br>Fit < 1.00 mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Accuracy: Upper 95% Probability with 95%<br>Model Fit < 1.00 mm<br><br>Precision: Upper 95% Probability with 95%<br>Model Fit < 1.00 mm                                                                                                                                                                                                                                                                                                                                                                                         | Substantially<br>Equivalent -<br>specifications for<br>precision are the same.<br>Differences<br>determined to not<br>significantly impact<br>overall SE profile of<br>YomiPlan Go as validated by successful |
| Feature                                          | Subject Device<br>Yomi Robotic System with YomiPlan Go<br>K211466                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Predicate Device<br>NGS with Yomi Plan v2.0<br>K202264                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | SE Analysis                                                                                                                                                                                                   |
|                                                  |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | completion of<br>G210363.                                                                                                                                                                                     |

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Neocis. The logo consists of the word "neocis" in a sans-serif font, with the "neo" in a golden color and the "cis" in a light gray color. Above the word is a gray line that extends horizontally, and a vertical line that intersects the horizontal line at the "o" in "neocis".

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Neocis. The word "neocis" is written in a sans-serif font, with the letters in yellow. Above the word is a gray horizontal line that extends past the letters on both sides. A gray vertical line intersects the horizontal line, forming a crosshair-like symbol above the "cis" portion of the word.

To support the difference in technological characteristics of preoperative planning without the need to upload a CT image substantial equivalence of this characteristic is comparable to Orthosoft Navitrack System-Optical TKR CT-Less System (K021760):

| Yomi Robotic System with YomiPlan Go VS Reference Device |                                                                                                                                      |                                                                                                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                           |
|----------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Features Related to<br>Preplanning the                   | Yomi Robotic System<br>with YomiPlan Go                                                                                              | Orthosoft CT-Less                                                                                                                                                                                                                                                                                                                                                                         | Why Differences Do Not<br>Affect SE                                                                                                                                                                                                       |
| Trajectory/Implant Position                              | None entered in the software.<br>The surgeon may optionally<br>take an X-ray or CT and use<br>that mentally during the<br>procedure. | None entered in the software.<br>The surgeon may optionally<br>take an X-ray or CT and use<br>that mentally during the<br>procedure.                                                                                                                                                                                                                                                      | SAME                                                                                                                                                                                                                                      |
| Preoperative Imaging                                     | None entered in the software.<br>The surgeon may optionally<br>take an X-ray or CT and use<br>that mentally during the<br>procedure. | None entered in the software.<br>The surgeon may optionally<br>take an X-ray or CT and use<br>that mentally during the<br>procedure.                                                                                                                                                                                                                                                      | SAME                                                                                                                                                                                                                                      |
| Intraoperative Planning                                  | No CT is used for guidance<br>during surgery                                                                                         | No CT is used for guidance<br>during surgery                                                                                                                                                                                                                                                                                                                                              | SAME                                                                                                                                                                                                                                      |
| Image Registration and<br>Fiducial Markers               | There is no image registration<br>and there are no fiducial<br>markers because there is no<br>image to register to.                  | There is no image registration<br>and there are no fiducial<br>markers because there is no<br>image to register to.                                                                                                                                                                                                                                                                       | SAME                                                                                                                                                                                                                                      |
| Intraoperative Data<br>Gathering for Planning            | The surgeon observes the<br>physical location, e.g.,<br>neighboring teeth, bone (if<br>flapped), soft tissue, etc.                   | Since the surgeon cannot<br>visibly observe all the<br>anatomical landmarks<br>necessary for planning, they<br>capture coordinate system<br>elements through point<br>digitization, e.g. hip center,<br>knee landmarks, etc., to guide<br>their planning. The data<br>gathered is mapped to a<br>generic image, but this is still<br>very dissimilar from what the<br>actual CT would be. | Equivalent - Neither system<br>incorporates preoperative<br>patient specific imaging for<br>planning. Both rely on patient<br>specific data either observed<br>or captured with calculations<br>to finalize the plan<br>intraoperatively. |

### Benefit / Risk Analysis

The FDA Guidance Document entitled "Benefit-Risk Factors to Consider When Determining Substantial Equivalence in Premarket Notifications (510(k)) with Different Technological Characteristics" was reviewed to prepare the following benefit/risk discussion in order to support the claim of substantial equivalence.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for Neocis. The word "neocis" is written in a sans-serif font, with the letters in yellow. Above the word is a gray line that extends horizontally, and there is a gray crosshair above the "i" in "neocis". The logo is simple and modern.

| Primary Differences<br>Between Predicate<br>and Subject Device | 1. Planning: Planning is based on the user's own direct visual examination of the patient's anatomy guided by the Yomi<br>Guide Arm. Predicate device planning is based on both a patient-specific CT scan and direct visual examination.<br>2. Osteotomy: With YomiPlan Go, there is no CT scan to reference while performing the osteotomy. The predicate device<br>presents the CT scan to the user which can then be used as reference during the osteotomy. All other aspects of<br>performing the osteotomy are the same when comparing YomiPlan Go to the predicate device.<br>3. YomiPlan Go IDE data from G210363 showed that the average angular deviation was higher with YomiPlan Go. The<br>study also showed that the range of angular deviations from the IDE was within the range of angular deviations from<br>the Varga, et al study for freehand (see Varga article reference below).                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
|----------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Assessment of<br>Benefits                                      | Benefits compared to the predicate device:<br>1. YomiPlan Go can enable a faster workflow and greater adoption of guided surgery.<br>2. YomiPlan Go does not require a pre-operative CT scan therefore clinician and staff and patient exposure to x-rays is<br>reduced.<br>Benefits compared to freehand surgery:<br>1. YomiPlan Go provides a depth stop. This prevents over drilling, which can happen accidentally in freehand surgery.<br>2. By holding the drill steady, YomiPlan Go prevents skiving, i.e. slipping off of a knife-edge ridge of bone or slipping into a<br>extraction socket or into soft bone.<br>3. It enables precision throughout repeated drill stages of multiple diameter drill bits. YomiPlan Go guides the user to the<br>same position and angle even after switching drill bit diameters as they proceed through drilling the osteotomy. This<br>precision is not easy to achieve with a freehand surgery technique.<br>4. YomiPlan Go enables parallelism.<br>5. YomiPlan Go provides guidance without the workflow burdens of CT image capture, registration, and pre-planning.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Assessment of<br>Risks Related to<br>Differences               | 1. Use of Yomi Active does present some risk associated with lack of use of a CT scan in the identification and protection of<br>critical anatomical structures. Potential harms related to Yomi Active includes potential damage to the<br>maxillary/mandibular nerves (temporary or permanent in duration depending on case). However, the risks associated<br>with not using a CT scan uploaded to the Yomi System are considered to be similar to current freehand technique. Since<br>freehand is considered by many to be the standard of care today in healthy patients with adequate bone, the risk is<br>considered acceptable.<br>2. Critical structures are still protected with YomiPlan Go constraining the drill position and angle to match the plan set by<br>the surgeon. It creates a depth stop and prevents skiving into unintended areas. The difference is that the plan does not<br>use CT reference in the software, but the surgeon still has the ability to observe physically where they want the implant<br>to be placed, and they also have the option to capture a pre-operative X-ray or CT and use it for "mental navigation" of<br>the planning.<br>3. Any false sense of accuracy and appropriate position of the implant that the user may infer from the use of Yomi<br>with CT-based guidance is mitigated with labeling e.g., instructions for use update to include appropriate warnings,<br>marketing materials to also clearly differentiate between benefits and risks associated with using YomiPlan Go and<br>Yomi CT-Guided, and a training program which includes materials outlining the above as well.<br>4. The difference in average angular deviation was found to be acceptable as evidenced by the CRF feedback from the<br>(3) investigators after their completion of each case and by the (1) independent 3rd party clinical review of all 15             cases from the IDE i.e., feedback from the investigators that they had no vital anatomical structure concerns after             completion of each procedure. |

### VIII. Nonclinical Performance Testing

This submission includes certain nonclinical performance testing including verification (software run-through, repeatability) and end user validation to confirm that users' needs have been satisfied while using Yomi with YomiPlan Go. End user validation was performed by a surgeon in a simulated use environment on typodonts.

#### IX. Clinical Testing

A clinical study (G210363) was performed in order to support substantial equivalence of YomiPlan Go to the predicate device. The primary objective of this study was to evaluate human factors related issues with YomiPlan Go during the preoperative and intraoperative use with the

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for Neocis. The logo features the word "neocis" in a sans-serif font, with the letters in yellow. Above the word is a gray horizontal line that extends to the left and right of the word. There is a gray vertical line that intersects the horizontal line at the center of the word, forming a cross shape.

Yomi Robotic System. The secondary objective focused on evaluating accuracy of the YomiPlan Go. This included a comparison of pre-operative CBCT scans on which implants are planned in ideal locations with postoperative CBCT scans capturing the actual placement of implants placed clinically with the YomiPlan Go. It also included a third-party oral surgeon's assessment regarding safety to ensure no critical structures are damaged, an evaluation for parallelism for implants that are to be placed parallel to one another, and an assessment of restorability. The information from this study was used to guide device modifications with particular focus on informing changes to the device labeling/instructions for use to ensure appropriate use of the YomiPlan Go.

The investigation was a prospective, multi-center, single-arm study, in which partially or fully edentulous patients had as part of their treatment single or multiple implant placement in the maxilla or the mandible. The subjects enrolled in the study had to satisfy certain inclusion criteria:

- Age more than 18 years at inclusion
- Signed informed consent
- . Partially or fully edentulous patients that will have implant placement in the maxilla or the mandible as part of their treatment
- . Must have sufficient bone height and width appropriate for dental implant surgery as evaluated on a Cone Beam Computed Tomography (CBCT) survey

A total of 44 of implants were placed in fifteen (15) patient arches with Yomi Active. Each of the (5) sites had (3) partial and (2) full arch dental restorations performed by each investigator. The study was performed by general dentists. A 31d party oral surgeon performed an independent clinical assessment of all cases.

Each investigator provided subjective data through completion of quantitative and open-ended subjective questionnaires. There were also no serious adverse events during the conduct of this clinical study. Overall, according to the qualitative survey, the investigators determined that the YomiPlan Go met its design and performance requirements.

An adverse event did occur with one subject where the implant was transported into the sinus during hand torquing. The surgeon placed another implant to complete the dental restoration at the tooth site and completed an intervention on a later date to remove the implant. The surgeon documented on the CRF the severity of this event as "minor" and an unlikely relationship to Yomi Active. Per the surgeon, the robot was able to maintain the reference point for the osteotomy protocol that followed the initial osteotomy. After further Neocis investigation, root cause of this event was related to user error due to the surgeon hand torquing the implant deeper.

YomiPlan Go accuracy was also assessed by comparing the preoperative plans to their respective post-operative implant placement. The precision, depth and lateral error of the YomiPlan Go is equivalent to the predicate. Angular deviations were higher with YomiPlan Go than with the predicate device (see summary table below).

| Comparison Metric | Raw Data of Implant Placements Grouped by Patient, Absolute Values,<br>reported in Mean/StdDev |                    |                      |
|-------------------|------------------------------------------------------------------------------------------------|--------------------|----------------------|
|                   | Depth Error (mm)                                                                               | Lateral Error (mm) | Angular Error (degs) |

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the logo for Neocis. The logo consists of the word "neocis" in a sans-serif font, with the letters in yellow. A gray line is above the word. To the right of the word is a gray circle with a vertical and horizontal line through it.

| Investigator Plan on PreOp CT vs Investigator<br>PostOp CT of Final Implant Placement | 1.38 +/- 0.71 | 2.04 +/- 1.59 | 12.91 +/- 7.56 |
|---------------------------------------------------------------------------------------|---------------|---------------|----------------|
|---------------------------------------------------------------------------------------|---------------|---------------|----------------|

These results were found to be within a similar range to freehand accuracy data. However, accuracy is not a specification or an advantage of YomiPlan Go, whereas precision is a specification for the system and is supported by benchtop verification testing.

### Supporting Clinical References X.

The angular accuracy of the YomiPlan Go is higher than the predicate. To support the difference in performance the following articles provide support regarding performance between YomiPlan Go as compared to freehand technique:

- . Dynamic Navigation for Dental Implant Surgery, Penchal, et al, Oral Maxillofac Surq Clin North Am, Nov, 2019 – In line with this article, YomiPlan Go provides a better balance of guidance and workflow disruption. YomiPlan Go provides guidance without the workflow burdens of CT image capture, image registration, and pre-planning.
- . Digital Implant Planning and Guided Implant Surgery - Workflow and Reliability, Schubert, et al, BR Dent J, Jan 2019 – This article discussed ideal workflows and improved reliability during guided implant surgery. In line with this article, YomiPlan Go does not require a pre-operative CT scan therefore clinician and staff and patient exposure to xrays is avoided.
- . Evaluating the Health Economic Implications and Cost-Effectiveness of Dental Implants: A Literature Review, Vogel, et al, Int J Oral Maxillofac Implants, Mar-April 2013 – This article discusses the positive economic implications of more patients having access to dental implant surgery. Since YomiPlan Go does not require a CT scan, general dentists who perform this procedure in a freehand manner, their patients, and thus in general health economics, could all benefit with YomiPlan Go.
- . Factors Influencinq the Accuracy of Freehand Implant Placement: A Prospective Clinical Study, Schnutenhaus S, et al, Dent J (Basel), May, 2021 - Risks associated with YomiPlan Go can be considered similar to current freehand technique. Since freehand is considered by many to be the standard of care today in healthy patients with adequate bone, the risk associated with YomiPlan Go is considered acceptable. This article supports this conclusion. Schnutenhaus S, et al, also included a reference to a randomized clinical study by Varga E, et al (i.e., Varga E., Jr., Antal M., Major L., Kiscsatari R., Braunitzer G., Piffko J. Guidance means accuracy: A randomized clinical trial on freehand versus guided dental implantation. Clin. Oral Implant Res. 2020), where the deviations resulting from freehand implant placements and the various implant placements with the aid of drilling templates were compared. They found an average angular deviation of 7.03° in a freehand procedure with a range of 0.7–21.3°.
- Image Guidance for Implants Improves Accuracy and Predictability, Compendium of CE in Dentistry, Ganeles, et al, Nov/Dec 2011 – The article discussed how surgeons have the option to capture a pre-operative X-ray or CT and use it for "mental navigation" of the planning. This is similar to how a preoperative image would be used with YomiPlan Go.

#### XI. Conclusion

The substantial equivalence discussion, risk profile comparisons, and testing described above demonstrate that YomiPlan Go and the predicate device have many similarities and can be

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows the logo for Neocis. The word "neocis" is written in a sans-serif font in yellow. Above the word is a gray line that extends to the left and right of the word. There is a gray circle above the "c" in "neocis" with a vertical line extending upwards and a horizontal line extending to the right.

considered similar. There were some differences noted with the clinical test data such as the difference in the performance of the predicate (angular deviation of 3.3°) to the performance of the subject device (angular deviation of 12°) and performance of freehand (angular deviation range of up to 21°). These differences were found to be acceptable as evidenced by the CRF feedback from all (3) investigators after their completion of each case and by the (1) independent 3rd party clinical review of all 15 cases from the IDE i.e., feedback from the investigators was that they had no vital anatomical structure concerns after completion of each procedure. Also, when weighing the benefits (e.g., faster workflows when YomiPlan Go is indicated; less radiation exposure) against any risks associated with YomiPlan Go (e.g., less accuracy and pre-planning of critical structures, does not require the use of a preloaded CT scan onto the Yomi system for planning), it was concluded that YomiPlan Go is substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/DE/subpart-e%E2%80%94surgical-devices/PLV/K211466](https://fda.innolitics.com/submissions/DE/subpart-e%E2%80%94surgical-devices/PLV/K211466)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
