The Neocis Guidance System (NGS) is a computerized navigational system intended to provide assistance in both the planning (pre-operative) and the surgical (intra-operative) phases of dental implantation surgery. The system provides software to preoperatively plan dental implantation procedures and provides navigational guidance of the surgical instruments. The NGS is intended for use in partially edentulous and fully edentulous adult patients who qualify for dental implants.
Device Story
NGS is a computerized dental stereotaxic instrument for planning and guiding dental implant surgery. Input: CBCT scan for virtual planning. Operation: Surgeon plans implant placement in software; system uses a Patient Tracker (PT) and Edentulous Patient Splint (EPS) or Chairside Patient Splint (CPS) to track patient movement in real-time. The system provides haptic feedback by constraining a standard powered bone-cutting instrument to the pre-operative plan; it automatically adjusts the cutting angle, position, and depth if the patient moves. Used in dental clinics by surgeons. Output: Physical, visual, and audible guidance during osteotomy. Benefits: Minimally invasive, flapless procedures with high accuracy; improved safety via real-time motion compensation.
Clinical Evidence
Prospective, two-center, non-randomized, non-controlled IDE feasibility study (G190282). Sample: 10 adult patients (male, partially or fully edentulous), 67 total implants. Endpoints: usability, safety, effectiveness. Results: No adverse events or usability concerns reported. Implant location accuracy (pre-op plan vs. post-op CT) met system specifications. Bench testing included EPS deflection, kinematic mount repeatability, and total system accuracy on typodonts.
Technological Characteristics
Materials: Ixef HC-1022, stainless steel, titanium alloy. Sensing: Force/Torque (F/T) sensors (+/- 30 N, +/- 2 Nm). Connectivity: Networked/standalone. Sterilization: Steam (ISO 17665-1). Software: Moderate level of concern. Accuracy: RMS < 1 mm (lateral/depth), RMS < 6.0° (angular). Power: 120VAC/60Hz. Class I equipment.
Indications for Use
Indicated for partially edentulous and fully edentulous adult patients requiring dental implants. Contraindicated for patients with insufficient bone quality to retain the Edentulous Patient Splint (EPS) or history of jaw/TMJ pain.
Regulatory Classification
Identification
A bone cutting instrument and accessories is a metal device intended for use in reconstructive oral surgery to drill or cut into the upper or lower jaw and may be used to prepare bone to insert a wire, pin, or screw. The device includes the manual bone drill and wire driver, powered bone drill, rotary bone cutting handpiece, and AC-powered bone saw.
Stryker Mandibular Fracture and Reconstruction Fixation System, Bone Fixation Screws (K014263)
Related Devices
K200348 — Neocis Guidance System (NGS) with Fiducial Array Splint (FAS) · Neocis, Inc. · May 8, 2020
K203401 — Neocis Guidance System (NGS) with Patient Splints (EPS) · Neocis, Inc. · Dec 15, 2020
K173402 — Neocis Guidance System (NGS) with Chairside Splint · Neocis, Inc. · Feb 22, 2018
K191605 — Neocis Guidance System (NGS) · Neocis, Inc. · Sep 11, 2019
K161399 — Neocis Guidance System · Neocis, Inc. · Dec 22, 2016
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font size below.
July 28, 2020
Neocis Inc. Thomas E. Claiborne, Ph.D. Regulatory Affairs Manager 2800 Biscayne Blvd. Suite 600 Miami, Florida 33137
Re: K200805
Trade/Device Name: Neocis Guidance System (NGS) with Patient Splints Regulation Number: 21 CFR 872.4120 Regulation Name: Bone Cutting Instrument and Accessories Regulatory Class: Class II Product Code: PLV Dated: April 28, 2020 Received: April 29, 2020
Dear Thomas E. Claiborne:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and probibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Srinivas "Nandu" Nandkumar, Ph.D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K200805
Device Name
Neocis Guidance System (NGS) with Patient Splints
Indications for Use (Describe)
The Neocis Guidance System (NGS) is a computerized navigational system intended to provide assistance in both the planning (pre-operative) and the surgical (intra-operative) phases of dental implantation surgery. The system provides software to preoperatively plan dental implantation procedures and provides navigational guidance of the surgical instruments. The NGS is intended for use in partially edentulous adult patients who qualify for dental implants.
Type of Use (Select one or both, as applicable)
| <div> <span></span>Prescription Use (Part 21 CFR 801 Subpart D) </div> |
|---------------------------------------------------------------------------|
| <div> <span></span>Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Neocis. The logo consists of the word "neocis" in a sans-serif font, with the letters in a golden color. Above the word is a horizontal gray line that extends over the "neo" part of the word. A vertical gray line intersects the horizontal line above the "c" in "neocis", forming a plus sign with a small circle at the intersection.
#### 510(k) Summary
#### I. Submitter
Neocis Inc. 2800 Biscayne Blvd. Suite 600 Miami, FL 33137 Tel: 1-855-9NEOCIS
Contact Person: Thomas E. Claiborne, Ph.D., Regulatory Affairs Manager Date Prepared: July 27, 2020
#### II. Device
| Trade Name: | Neocis Guidance System (NGS) with Patient Splints |
|----------------------|-----------------------------------------------------------|
| Common Name: | Dental Stereotaxic Instrument |
| Classification Name: | Bone cutting instrument and accessories (21 CFR 872.4120) |
| Classification: | Class II |
| Product Code: | PLV |
#### III. Predicate Devices
- X-Guide Surgical Navigation System (K150222) ●
#### IV. Reference Devices
- Neocis Chairside Splint (K173402)
- . Dentsply Sirona Cerec Guides (K190059)
- . Stryker Mandibular Fracture and Reconstruction Fixation System, Bone Fixation Screws (K014263)
### V. Indications for Use
The Neocis Guidance System (NGS) is a computerized navigational system intended to provide assistance in both the planning (pre-operative) and the surgical (intra-operative) phases of dental implantation surgery. The system provides software to preoperatively plan dental implantation procedures and provides navigational guidance of the surgical instruments. The NGS is intended for use in partially edentulous and fully edentulous adult patients who qualify for dental implants.
### VI. Device Description
The Neocis Guidance System (NGS) (K161399) is a dental stereotaxic instrument (Product Code PLV) and a powered surgical device for bone cutting (21 CFR 872.4120). The Neocis Guidance System (NGS) is a computerized navigational system intended to provide assistance in both the planning (pre-operative) and the surgical (intra-operative) phases of dental implantation surgery. The system provides precise and accurate navigational guidance of surgical instruments, with regard to planning in dental implantation procedures. The system allows the user to plan the surgery virtually in software using a cone beam computed tomography (CBCT) scan of the patient, and the plan is used by a guidance system to provide physical, visual, and audible feedback to the surgeon during the implant site preparation. The holds and guides a standard FDA-cleared powered bone cutting instrument.
The implant process occurs in two phases. First, the dental surgical procedure with the planning software. A virtual implant is placed at the desired location in the CT scan, allowing the dental surgeon to avoid interfering with critical anatomical structures during implant surgery. Second, when the implant plan is optimally positioned, the NGS provides accurate guidance of the dental surgical instruments according to the pre-operative plan. The NGS can be used for flapless dental implant procedures, which is a type of minimally invasive surgical approach. The NGS provides haptic feedback to the surgeon by constraining the bone cutting instrument to the plan. This allows the surgeon to feel resistance to attempts at motions that may deviate from the plan.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for Neocis. The logo consists of the word "neocis" in a sans-serif font, with the letters in a light orange color. Above the word is a thin gray horizontal line that extends beyond the letters on both sides. A thin gray vertical line intersects the horizontal line at a small gray circle above the "i" in "neocis".
The patient tracking portion of the NGS is comprised of linkages from the NGS, which include the Chairside Patient Splint (CPS) (K173402), the End Effector (EE) and the Patient Tracker (PT). The Patient Splint is attached to the contralateral side of the patient's mouth over stable teeth. The CPS is placed on the patient using on-label dental materials (K182776) prior to the presurgical CBCT scan. A Fiducial Array (FA) with radio-opaque fiducial markers is placed on the CPS prior to the CBCT scan so the virtual plan can be related to the physical space of the system using the markers. The PT is an electromechanical feedback system that is connected to the CPS on the patient, which relays information to the control software in order to track patient movement. If patient movement occurs during the surgical procedure, the system will respond by altering the prescribed surgical cutting angle, position, and depth to accommodate the patient movement, which will maintain the accuracy of the osteotomy.
The subject of this submission is our Edentulous Patient Splint (EPS). The EPS is affixed to the anterior mandible or maxilla using standard bone screws. Like the CPS, the EPS serves as rigid connection to the patient for robotic tracking of the patient during the procedure. The EPS is intended for use in partially edentulous and fully edentulous adult patients.
### VII. Comparison of Technological Characteristics
This submission includes changes to the indications for use and a new patient splint for use in partially edentulous and fully edentulous patients. There are no changes to the NGS hardware or software in this submission. Since the NGS (K161399) was previously found to be substantially equivalent to K150222, we refer to our prior clearance for a full technological comparison. We have selected K150222 again as a predicate device because we now have technology that allows us to match its performance requirements for fully edentulous patients, which is the focus of this submission. The differences introduced by this modification are detailed in Table 1: Comparison of technological characteristics to the predicates below.
{5}------------------------------------------------
Image /page/5/Picture/6 description: The image shows the logo for Neocis. The logo features the word "neocis" in a sans-serif font, with the letters in a golden color. Above the word is a gray horizontal line, and to the right of the word is a gray vertical line with a circle at the top.
| Technologic<br>al<br>Characterist<br>ics | NGS with<br>Patient<br>Splints<br>Subject<br>Device | X-Guide Surgical<br>Navigation System<br>Predicate Device<br>(K150222) | Reference<br>Device-CPS<br>K173402 | Reference<br>Device-<br>Sleeves<br>K190059 | Reference<br>Device-<br>Screws<br>K014263 | SE Analysis |
|----------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------|-------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use (IFU) | The Neocis<br>Guidance<br>System (NGS)<br>is a<br>computerized<br>navigational<br>system<br>intended to<br>provide<br>assistance in<br>both the<br>planning (pre-<br>operative) and<br>the surgical<br>(intra-<br>operative)<br>phases of<br>dental<br>implantation<br>surgery. The<br>system<br>provides<br>software to<br>preoperatively<br>plan dental<br>implantation<br>procedures and<br>provides<br>navigational<br>guidance of the<br>surgical<br>instruments.<br>The NGS is<br>intended for<br>use in partially<br>edentulous and<br>fully edentulous<br>adult patients<br>who qualify for<br>dental<br>implants. | The X-Guide(R)<br>Surgical Navigation<br>System is a<br>computerized<br>navigational system<br>intended to provide<br>assistance in both the<br>preoperative planning<br>phase and intra-<br>operative surgical<br>phase of dental<br>implantation<br>procedures. The<br>system provides<br>software to<br>preoperatively plan<br>dental implantation<br>procedures and<br>provides navigational<br>guidance of the<br>surgical instruments.<br>The device is intended<br>for use for partially<br>edentulous and<br>edentulous adult and<br>geriatric patients who<br>require dental<br>implants as part of<br>their treatment plan. | The Neocis<br>Guidance<br>System<br>(NGS) is a<br>computerize<br>d<br>navigational<br>system<br>intended to<br>provide<br>assistance in<br>both the<br>planning<br>(pre-<br>operative)<br>and the<br>surgical<br>(intra-<br>operative)<br>phases of<br>dental<br>implantation<br>surgery. The<br>system<br>provides<br>software to<br>preoperative<br>ly plan<br>dental<br>implantation<br>procedures<br>and<br>provides<br>navigational<br>guidance of<br>the surgical<br>instruments. | N/A | N/A | Added condition<br>and patient<br>population from<br>predicate. |
| Technologic<br>al<br>Characterist<br>ics | NGS with<br>Patient<br>Splints<br>Subject<br>Device | X-Guide Surgical<br>Navigation System<br>Predicate Device<br>(K150222) | Reference<br>Device-CPS<br>K173402 | Reference<br>Device-<br>Sleeves<br>K190059 | Reference<br>Device-<br>Screws<br>K014263 | SE Analysis |
| Contraindication<br>s | The Neocis<br>Guidance<br>System is not<br>intended for<br>use with<br>patients that<br>have<br>insufficient<br>bone to retain<br>a Neocis<br>Edentulous<br>Patient Splint<br>(EPS) rigidly<br>throughout a<br>surgical<br>procedure.<br>The EPS<br>should not be<br>affixed to<br>patients that<br>exhibit:<br>•Patients<br>with<br>insufficient<br>bone quality<br>•Patients<br>with a<br>history of<br>jaw or TMJ<br>pain | Medical conditions<br>which contraindicate<br>the use of X-Guide<br>and its associated<br>applications include<br>any medical<br>conditions which may<br>contraindicate the<br>medical procedure<br>itself.<br>Only for those where<br>dental implants are<br>appropriate & patient<br>is healthy.<br>• Not for use with<br>patients less than<br>21 years of age.<br>• Not for use with<br>photosensitive<br>epileptic patients.<br>Patient sensitivity<br>may be caused<br>from the LEDs. | The<br>Neocis<br>Guidance<br>System is<br>not<br>intended<br>for use<br>with<br>patients<br>that have<br>insufficien<br>t bone to<br>retain a<br>Neocis<br>Edentulou<br>s Patient<br>Splint<br>(EPS)<br>rigidly<br>throughout<br>a surgical<br>procedure.<br>The EPS<br>should not<br>be affixed<br>to patients<br>that<br>exhibit:<br>Patients<br>with<br>insufficien<br>t bone<br>quality<br>Patients<br>with a<br>history of<br>jaw or TMJ<br>pain<br>Patients<br>with<br>allergies to<br>acrylates | N/A | N/A | Adjusted for<br>EPS.<br>Acrylic-like<br>dental materials<br>are not used with<br>the EPS.<br>Similar to X-<br>Guide. |
| Patient<br>Contacting<br>Materials | • Ixef®-<br>HC-1022<br>• Stainless<br>Steel<br>• Titanium<br>Alloy | • Stainless<br>Steel<br>• Titanium<br>Alloy | IXEF | Stainless<br>Steel | Titanium | Addition of<br>predicate<br>materials |
| NGS Power<br>Supply | 120VAC/60 Hz | 120VAC/60Hz | Same as the<br>subject<br>device | N/A | N/A | Same |
| Type of<br>Protection<br>against Electric<br>Shock | Class I<br>Equipment | Class I Equipment | Same as the<br>subject<br>device | N/A | N/A | Same |
| Technologic<br>al<br>Characterist<br>ics | NGS with<br>Patient<br>Splints<br>Subject<br>Device | X-Guide Surgical<br>Navigation System<br>Predicate Device<br>(K150222) | Reference<br>Device-CPS<br>K173402 | Reference<br>Device-<br>Sleeves<br>K190059 | Reference<br>Device-<br>Screws<br>K014263 | SE Analysis |
| Equipment<br>Suitable for<br>use in the<br>presence of<br>Flammable<br>Mixtures? | No | No | N Same as<br>the subject<br>device /A | N/A | N/A | Same |
| Electrical<br>Safety | ANSI AAMI<br>ES60601-<br>1:2005/(R)2012<br>and A1:2012,<br>C1:2009/(R)20<br>12 and<br>A2:2010/(R)20<br>12<br>(Consolidated<br>Text) Medical<br>electrical<br>equipment -<br>Part 1: General<br>requirements<br>for basic safety<br>and essential<br>performance<br>(IEC 60601-<br>1:2005, MOD) | ANSI AAMI<br>ES60601-<br>1:2005/(R)2012 and<br>A1:2012,C1:2009/(R)<br>2012 and<br>A2:2010/(R)2012<br>(Consolidated Text)<br>Medical electrical<br>equipment - Part 1:<br>General requirements<br>for basic safety and<br>essential performance<br>(IEC 60601-1:2005,<br>MOD) | Same as the<br>subject<br>device | N/A | N/A | Same |
| Electromagneti<br>c Disturbances | IEC 60601-1-2<br>Edition 4.0<br>2014-02<br>Medical<br>electrical<br>equipment -<br>Part 1-2:<br>General<br>requirements<br>for basic safety<br>and essential<br>performance -<br>Collateral<br>Standard:<br>Electromagneti<br>c disturbances .<br>Requirements<br>and tests | IEC 60601-1-2<br>Edition 4.0 2014-02<br>Medical electrical<br>equipment - Part 1-2:<br>General requirements<br>for basic safety and<br>essential performance<br>- Collateral Standard:<br>Electromagnetic<br>disturbances -<br>Requirements and<br>tests | Same as the<br>subject<br>device | N/A | N/A | Same |
| Ingres<br>S<br>Protec<br>tion | IPX0 | IPX2 | Same as the<br>subject<br>device | N/A | N/A | The X-Guide has<br>a higher IPX<br>rating. NGS is 0<br>because of our<br>draping. |
| Mode<br>of<br>Opera<br>tion | Continuous<br>Operation | Continuous Operation | Same as the<br>subject<br>device | N/A | N/A | Same |
| System<br>Lateral<br>Accuracy | RMS < 1 mm | RMS < 1 mm | Same as the<br>subject<br>device | N/A | N/A | Same |
| Technologic<br>al<br>Characterist<br>ics | NGS with<br>Patient<br>Splints<br>Subject<br>Device | X-Guide Surgical<br>Navigation System<br>Predicate Device<br>(K150222) | Reference<br>Device-CPS<br>K173402 | Reference<br>Device-<br>Sleeves<br>K190059 | Reference<br>Device-<br>Screws<br>K014263 | SE Analysis |
| System<br>Depth<br>Accuracy | RMS < 1 mm | RMS < 1 mm | Same as the<br>subject<br>device | N/A | N/A | Same |
| System<br>Angular<br>Accuracy | RMS < 6.0° | --- | Same as the<br>subject<br>device | N/A | N/A | Same |
| CT Scan<br>Quality<br>Requirement<br>S | 0.3 mm<br>Voxel, 0.3<br>mm Slice<br>Thickness,<br>Matrix 512<br>x 512, Full<br>13 cm 21<br>sec. Multi 2<br>DICOM<br>format. | --- | Same as the<br>subject<br>device | N/A | N/A | NGS has tighter<br>tolerances than<br>X-Guide. |
| F/T Sensor<br>Force<br>Measurement<br>Range | +/- 30 N | N/A | Same as the<br>subject<br>device | N/A | N/A | The X-Guide<br>does not use F/T<br>sensors |
| F/T Sensor<br>Torque<br>Measurement<br>Ra nge | +/- 2 Nm | N/A | Same as the<br>subject<br>device | N/A | N/A | The X-Guide<br>does not use F/T<br>sensors |
| F/T Sensor<br>Single Axis<br>Force Overload<br>Limit | 200 N | N/A | Same as the<br>subject<br>device | N/A | N/A | The X-Guide<br>does not use F/T<br>sensors |
| F/T Sensor<br>Single Axis<br>Torque<br>Overload Limit | 20 Nm | N/A | Same as the<br>subject<br>device | N/A | N/A | The X-Guide<br>does not use F/T<br>sensors |
| Upper limit<br>specification<br>for Guidance<br>Arm<br>Translation<br>Speed | 1.25 m/s | N/A | Same as the<br>subject<br>device | N/A | N/A | The X-Guide<br>does not have a<br>guidance arm. |
| Storage<br>Requirem<br>ents | Store powered<br>at Room<br>Temperature<br>(68°F to 76°F<br>or 20°C to<br>24.4°C) and<br>standard<br>ambient<br>humidity (5%<br>to 95%) in a<br>dust free,<br>clean<br>environment. | -20 to 60 °C, 10-95%<br>humidity | Same as the<br>subject<br>device | N/A | N/A | Same |
| Technologic<br>al<br>Characterist<br>ics | NGS with<br>Patient<br>Splints<br>Subject<br>Device | X-Guide Surgical<br>Navigation System<br>Predicate Device<br>(K150222) | Reference<br>Device-CPS<br>K173402 | Reference<br>Device-<br>Sleeves<br>K190059 | Reference<br>Device-<br>Screws<br>K014263 | SE Analysis |
| Patient Tracking | Physical<br>linkage to<br>patient via<br>Patient<br>Tracker and<br>Kinematic<br>Mount<br>connected to:<br>· EPS. or<br>• CPS<br>(K173402<br>) | Stereo-LED Optical<br>tracking of:<br>• E-clip<br>• Tracker Arm<br>· X-Corner Patient<br>Tracker | Same as the<br>subject<br>device with<br>CPS | N/A | N/A | Addition of EPS<br>to NGS<br>workflow, which<br>is SE to the X-<br>Guide E-clip. |
| Affixation of<br>tracking<br>technology to<br>patient | · EPS<br>with<br>sleeves<br>and<br>bone<br>screws<br>· CPS<br>with<br>dental<br>materia<br>ls<br>(K1827<br>76) | Metal strip<br>(stainless steel<br>or titanium bone<br>fixation plate)<br>and titanium<br>bone screws | Dental<br>Materi<br>als | Slee<br>ves | Screws<br>and<br>screwdri<br>ver | Same<br>methods for<br>subject<br>device and<br>predicate<br>device<br>using<br>reference<br>devices for<br>implementa<br>tion |
| Patient<br>attachment<br>removal | · CPS can<br>be<br>removed<br>manually<br>or by<br>powered<br>cutting<br>tool<br>• EPS is<br>removed<br>by<br>unscrewi<br>ng the<br>bone<br>screws | The metal strip is<br>removed by<br>unscrewing the bone<br>screws | Standard<br>dental<br>techniques | N/A | Use of<br>screwdriver | Addition of<br>predicate<br>removal<br>techniques and<br>reference device<br>implementation |
| Fiducials | Fiducial<br>Array (FA)<br>attached to<br>splint | X-clip fiducial<br>attached to patient's<br>teeth or organic<br>fiducials in patient's<br>bone | Same as the<br>Subject<br>Device | N/A | N/A | NGS Fiducial is<br>not patient<br>contacting |
| Kinematic<br>mount | Integrated<br>into the splint | E-Clip tracker arm<br>slot integrated into<br>metal strip (bone<br>plate) | KM as<br>separate<br>part | N/A | N/A | Similar<br>implementation |
| Biocompatibility | Yes (ISO<br>10993-1, -5, -<br>10, -12) | Yes (ISO 10993-1, -5,<br>-10, -11,-12) | Yes (ISO<br>10993-1, -5,<br>-10, -12) | Yes (ISO<br>10993-1, -<br>5, -10, -<br>12) | Yes (ISO<br>10993-1, -5, -<br>10, -12) | Same |
| Technologic<br>al<br>Characterist<br>ics | NGS with<br>Patient<br>Splints<br>Subject<br>Device | X-Guide Surgical<br>Navigation System<br>Predicate Device<br>(K150222) | Reference<br>Device-CPS<br>K173402 | Reference<br>Device-<br>Sleeves<br>K190059 | Reference<br>Device-<br>Screws<br>K014263 | SE Analysis |
| Sterili…
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.