X-Guide® Surgical Navigation System

K200662 · X-Nav Technologies, LLC · PLV · May 21, 2020 · Dental

Device Facts

Record IDK200662
Device NameX-Guide® Surgical Navigation System
ApplicantX-Nav Technologies, LLC
Product CodePLV · Dental
Decision DateMay 21, 2020
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 872.4120
Device ClassClass 2

Indications for Use

The X-Guide® Surgical Navigation System is a computerized navigational system intended to provide assistance in both the preoperative planning phase and intra-operative surgical phase of dental implantation procedures. The system provides software to preoperatively plan dental implantation procedures and provides navigational guidance of the surgical instruments. The device is intended for use for partially edentulous and edentulous adult and geriatric patients who need dental implants as part of their treatment plan.

Device Story

Electro-optical surgical navigation system; assists in preoperative planning and intra-operative guidance for dental implantation. Inputs: CT scan data; patient anatomy. System uses EDX Bone Screws (titanium) and EDX Tracker Arms (stainless steel) to affix tracking patterns to patient anatomy (mandible/maxilla). Tracker arms provide registration stability via spike design and clamping mechanism between bone screw and nut. Used in dental clinics by surgeons; provides real-time navigational guidance for surgical tool placement. Benefits: improved accuracy in implant positioning; reduced interference with soft tissue. Output: visual guidance on surgical display for instrument navigation.

Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing, including mechanical property assessment (deflection/stability), biocompatibility evaluation (leveraging predicate data per ISO 10993-1), and sterilization validation (ISO 17665-1).

Technological Characteristics

Electro-optical navigation system. Components: Ti6AL4V (Grade 5) titanium bone screws (ASTM F136, ASTM F543) and 316L stainless steel tracker arms. Features: self-drilling/self-tapping screws, passivated tracker arms, steam sterilization. Connectivity: integrates CT data for surgical planning. Software: computerized navigation.

Indications for Use

Indicated for partially edentulous and edentulous adult and geriatric patients requiring dental implants as part of their treatment plan.

Regulatory Classification

Identification

A bone cutting instrument and accessories is a metal device intended for use in reconstructive oral surgery to drill or cut into the upper or lower jaw and may be used to prepare bone to insert a wire, pin, or screw. The device includes the manual bone drill and wire driver, powered bone drill, rotary bone cutting handpiece, and AC-powered bone saw.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. May 21, 2020 X-NAV Technologies, LLC. Fred Cowdery VP- Regulatory and Ouality Compliance 1555 Bustard Road, Suite 75 Lansdale, Pennsylvania 19446 Re: K200662 Trade/Device Name: X-Guide® Surgical Navigation System Regulation Number: 21 CFR 872.4120 Regulation Name: Bone Cutting Instrument and Accessories Regulatory Class: Class II Product Code: PLV Dated: April 20, 2020 Received: April 24, 2020 Dear Fred Cowdery: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, for Srinivas "Nandu" Nandkumar, Ph.D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K200662 Device Name X-Guide® Surgical Navigation System Indications for Use (Describe) The X-Guide® Surgical Navigation System is a computerized navigational system intended to provide assistance in both the preoperative planning phase and intra-operative surgical phase of dental implantation procedures. The system provides software to preoperatively plan dental implantation procedures and provides navigational guidance of the surgical instruments. The device is intended for use for partially edentulous and edentulous adult and geriatric patients who need dental implants as part of their treatment plan. | Type of Use (Select one or both, as applicable) | | |----------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------| | <span style="font-family: sans-serif;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: sans-serif;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) | # CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image is a logo for X-NAV Technologies. The "X" in X-NAV is stylized with green, blue, and gray horizontal lines. The rest of the text is in a dark blue color, and the word "TECHNOLOGIES" is in gray and located below the X-NAV text. # 510(k) Summary (As required by Section 807.92) | Date Prepared: | May 20, 2020 | |--------------------------------|----------------------------------------------------------------------------------------------------------------------------| | Applicant: | X-NAV Technologies, LLC<br>1555 Bustard Road, Suite 75<br>Lansdale, PA. 19446 | | Contact Person: | Fred Cowdery<br>Vice President - Regulatory and Quality Compliance<br>Ph 267.436.0414<br>Email: fred.cowdery@x-navtech.com | | Device Trade/Proprietary Name: | X-Guide® Surgical Navigation System | Device Name: Common / Usual: Surgical Navigation System | ClassificationName: | 21 CFR 872.4120 (Bone Cutting Instrument and Accessories) | |---------------------------|-----------------------------------------------------------| | Regulatory Class: | II | | Product Code: | PLV | | Primary Predicate Device: | X-Guide® Surgical Navigation System (K150222) | | Reference Device: | KLS Martin Individual Patient Solutions (K163579) | # Indications for Use: The X-Guide® Surgical Navigation System is a computerized navigational system intended to provide assistance in both the preoperative planning phase and intra-operative surgical phase of dental implantation procedures. The system provides software to preoperatively plan dental implantation procedures and provides navigational guidance of the surgical instruments. The device is intended for use for partially edentulous and edentulous adult and geriatric patients who need dental implants as part of their treatment plan. {4}------------------------------------------------ Image /page/4/Picture/0 description: The image is a logo for X-NAV Technologies. The logo features a stylized "X" made up of green, blue, and gray bars. To the right of the "X" is the word "-NAV" in blue, and below the entire logo is the word "TECHNOLOGIES" in gray. #### Device Description: The X-Guide® Surgical Navigation Systemis anelectro-optical dental implantation procedures by providing the surgeon with accurate surgical tool placement and guidal plan built upon Computed Tomographic (CT scan) data. The X-Guide® Surgical Navigation System is currently cleared (K150222) with Bone Screws which are intended to affix an Edentulous Clin to an edentulous patients anatomy. The Edentulous clip is necessary to attach tracking patterns to facilitate the navigation and tracking process, and is calibrated to the patient anatomy and CT. As an alternative, the proposed EDX Bone Screws and EDX Nut may be used to secure tracking patterns to edentulous patient anatomy. The proposed EDX Bone Screws are a titanium screws which are tapered and vary in diameters of 2.7 mm and 3.2 mm wth thread lengths of 7mm and stud lengths of 4 mm, and 16 mm. In addition to the dimensional design changes, the screws have been modified with an M5 threading to accept the EDX Nut. The addition of M5 the Bone Screw is necessary to secure the tracking patterns to the patient. The 2.7 mm diameter EDX Bone Screws are typically placed to secure the EDX Tracker Arm. Should the bone be too soft resulting in a loose Bone Screw, a large diameter "rescue" EDX Bone Screw (3.2 mm diameter) can be placed. All of the EDX Bone Screws and EDX Bone Nuts are manufactured using a Ti6AL4V (Grade 5) alloy per ASTM F136 and adhere to the requirements of ASTM F543. The EDX Bone Screws are intended to be removed from the conclusion of the surgical implant procedure. The EDX Bone Screws are single use devices, sold in a non-sterile state and intended to be sterilized by the end user prior to use. The body of the proposed EDX Tracker Arms a stanless steel, which is a harder material providing more stability / rigidity in the Tracker Arm. A passivation coating is added to prevent oxidation of the during steam sterilization. The EDX Tracker Arms include a spike intended to added stability in EDX Tracker Arm registration. The body of the proposed EDX Tracker Arms are contoured in a variety of geometric shapes intended to minimize interferences with patient soft tissue by positioning on the Mandible and Maxilla in the posterior positions of the oral cavity. The EDX Tracker Arms are distributed in a non-sterile state, and intended to be sterilized by the end user prior to use. {5}------------------------------------------------ Image /page/5/Picture/0 description: The image is a logo for X-NAV Technologies. The logo features a stylized "X" made of green and gray horizontal bars, followed by the text "-NAV" in dark blue. Below the main logo is the word "TECHNOLOGIES" in a smaller, lighter gray font. # Proposed Components: | X-Nav Part Number | Description | |-------------------|--------------------------------------------------| | P010303 | Bone Screw, Length: 4mm; Diameter: 2.7mm | | P010304 | Bone Screw, Length: 8mm; Diameter: 2.7mm | | P010305 | Bone Screw, Length: 16mm; Diameter: 2.7mm | | P010306 | Bone Screw, Length: 4mm; Diameter: 3.2mm | | P010307 | Bone Screw, Length: 8mm; Diameter: 3.2mm | | P010308 | Bone Screw, Length: 16mm; Diameter: 3.2mm | | P010869 | Tracker Arm Assembly; EDX, Posterior Upper Right | | P010870 | Tracker Arm Assembly; EDX, Posterior Upper Left | | P010871 | Tracker Arm Assembly; EDX, Anterior Upper Right | | P010872 | Tracker Arm Assembly; EDX, Anterior Upper Left | | P010873 | Tracker Arm Assembly; EDX, Posterior Lower Right | | P010874 | Tracker Arm Assembly; EDX, Posterior Lower Left | | P010875 | Tracker Arm Assembly; EDX, Anterior Lower Right | | P010876 | Tracker Arm Assembly; EDX, Anterior Lower Left | | P010660 | Bone Screw Nut, EDX | {6}------------------------------------------------ Image /page/6/Picture/0 description: The image is a logo for X-NAV Technologies. The logo features a stylized "X" made of green, blue, and gray bars, followed by the text "-NAV" in large, bold, blue letters. Below the main logo is the word "TECHNOLOGIES" in smaller, gray letters. # Comparison to Predicate and Reference Devices: As described in the tables below, the application System is substantially equivalent in intended use, design and physician use to the predicate device X-Guide® System (K15022). The application device is substantially equivalent in scientific technology to the reference device KLS Martin Drill Free® MMF Screw. | | Application Device | Primary Predicate Device | Reference Device | | |-----------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Use<br>Specifications | X-Guide® Surgical<br>Navigation System | X-Guide® Surgical<br>Navigation System<br>(K150222) | KLS Martin<br>Individual Patient<br>Solutions (K163579) | Justification of<br>Differences | | Indications for<br>Use | The X-Guide® Surgical Navigation<br>System is a computerized<br>navigational system intended to<br>provide assistance in both the<br>preoperative planning phase and the<br>intra- operative surgical phase of<br>dental implantation procedures.<br>The system provides software to<br>preoperatively plan dental<br>surgical procedures and provides<br>navigational guidance of the<br>surgical instruments. The device<br>is intended for use for partially<br>edentulous and edentulous adult<br>and geriatric<br>patients who need dental<br>implants as part of their<br>treatment plan. | The X-Guide® Surgical Navigation<br>System is a computerized navigational<br>system intended to provide assistance in<br>both the preoperative planning phase and<br>the intra-operative surgical phase of<br>dental implantation procedures.<br>The system provides software to<br>preoperatively plan dental surgical<br>procedures and provides navigational<br>guidance of the surgical instruments. The<br>device is intended for use for partially<br>edentulous and edentulous adult and<br>geriatric<br>patients who need dental implants as<br>part of their treatment plan. | KLS Martin Individual Patient Solutions<br>implant devices are intended for use in<br>the stabilization and fixation of<br>mandibular fractures and mandibular<br>reconstruction. | The proposed device and Predicate<br>device share the same intended use.<br>Although the reference device has a<br>different intended use, it was chosen to<br>show there are cleared devices of the<br>same diameter. | | Product Code | PLV | PLV | DZL | | | | Application Device | Primary Predicate Device | Reference Device | | | Technology / Performance<br>Characteristics - Bone Screws | X-Guide® Surgical<br>Navigation System (K200662) | X-Guide® Surgical<br>Navigation System (K150222) | KLS Martin Individual<br>Patient Solutions (K163579) | Justification of Differences | | Material | Titanium Ti-6AL-4V (Grade 5) | Titanium Ti-6AL-4V (Grade 5) | Titanium Ti-6AL-4V | No Difference | | Screw Type | Self- Drilling<br>Self-Tapping | Self- Drilling<br>Self-Tapping | Self- Drilling<br>Self-Tapping | No Difference | | Manufacturing Method | Traditional (Subtractive) | Traditional (Subtractive) | Traditional (Subtractive) | No Difference | | Sterilization | Non-sterile (Steam) | Non-sterile (Steam) | Non-sterile (Steam) | No Difference | | Anatomical Sites | Mandible and Maxilla | Mandible and Maxilla | Mandible and Maxilla | No Difference | | Diameter | 2.7 mm, 3.2 mm | 2.0 mm, 2.7 mm | 2.0 mm – 3.2mm | No Difference | | Thread Length | 7 mm | 4, 6, 8, 10, mm | 5 mm – 22 mm | Application Device Thread Length is<br>within range of Predicate and<br>Reference devices. | | Thread Pitch | 2.7mm diameter: 1.20 mm<br>3.2mm diameter: 1.23 mm | 2.0mm Diameter: 0.80 mm<br>2.7 mm Diameter: 1.20 mm | Documentation does not list<br>thread pitch | 2.7mm pitch is based off of<br>predicate. 3.2 pitch is derived from<br>Table A5.2 listed in ASME F543-17,<br>Metallic Medical Bone Screws. | | Stud (Shaft) Length | 4,8 16 mm | N/A | N/A | Stud offsets EDX Tracker Arm from<br>bone and tissue, allowing the tracker<br>arm to be clamped between the EDX<br>Nut and Bone Screw providing<br>increased mounting stability.<br>Predicate and reference devices<br>clamp Edentulous Clip or Bone Plate<br>directly to the bone which may have<br>curvatures. | | Biocompatibility | Biocompatible in accordance<br>with ISO 10993-1 | Biocompatible in accordance<br>with ISO 10993-1 | Biocompatible in accordance<br>with ISO 10993-1 | No Difference | | Single Use | Yes | Yes | Yes | No Difference | | Shelf Life | None | None | Unknown | No Difference | | Surface Casting | Anodizing | Anodizing | Anodizing | No Difference | | | Application Device (K200662) | Primary Predicate Device | | | | Technology / Performance<br>Characteristics - Tracker Arm | X-Guide® Surgical Navigation<br>System<br>(EDX Tracker Arms) | X-Guide® Surgical<br>Navigation System<br>(K150222)<br>(Predicate Tracker Arms) | Justification of Differences | | | Material | Stainless Steel (316L) | Aluminum 6061-T6 (Grade 5),<br>316L, | Stainless Steel has stronger material properties and is therefore a more<br>rigid material than Aluminum. | | | Manufacturing Method | Traditional (Subtractive) | Traditional (Subtractive) | No Difference | | | Sterilization | Non-sterile (Steam) | Non-sterile (Steam) | No Difference | | | Anatomical Sites | Mandible and Maxilla (8<br>Versions) | Mandible and Maxilla (3<br>Versions) | Additional arm shapes to avoid interference with patient soft tissue<br>and surgical instruments. | | | Spikes | Spike is between mounting holes<br>on tracker arm to provide stability | Spikes are underneath E-Clip<br>to provide stability | No Difference | | | Attachment Method | Clamped between EDX Nut and<br>EDX Bone Screw | Tracker arm is attached to E-<br>clip via tracker arm screw. E-<br>Clip is clamped between bone<br>screw and patient bone. | EDX mounting technique provides improved mounting stability,<br>allowing the tracker arm to be clamped between the EDX Nut and bone<br>screw and offset from soft tissue. | | | Stud (Shaft) Length | 4, 8, 16 mm | Not Applicable | EDX Bone Screw shaft offsets Tracker Arm above bone and tissue,<br>allowing the tracker arm to be clamped between the EDX Nut and bone<br>screw. | | | Biocompatibility | Biocompatible in accordance with<br>ISO 10993-1 | Biocompatible in accordance with ISO 10993-1 | No Difference | | | Single Use | No | No | No Difference | | | Shelf Life | None | None | Similar | | | Surface Casting | Passivation | Anodizing/Passivation | Predicate Device is Anodized. Application Device is passivated. Both<br>methods are acceptable industry standards to prevent corrosion. | | #### Table 1: Summary of Technological Characteristics Comparison {7}------------------------------------------------ Image /page/7/Picture/0 description: The image is a logo for X-NAV Technologies. The logo features a stylized "X" made of green, blue, and gray horizontal bars. The words "-NAV" are in bold, dark blue letters, and the word "TECHNOLOGIES" is in light gray letters below the "X-NAV". {8}------------------------------------------------ Image /page/8/Picture/0 description: The image is a logo for X-NAV Technologies. The logo has a stylized X on the left side, with green, blue, and gray colors. The letters NAV are in blue, and the word TECHNOLOGIES is in gray below the X-NAV. {9}------------------------------------------------ Image /page/9/Picture/0 description: The image is a logo for X-NAV Technologies. The logo features a stylized "X" made of blue, green, and gray horizontal bars. To the right of the "X" is the word "NAV" in large, bold, dark blue letters. Below the "X-NAV" portion of the logo is the word "TECHNOLOGIES" in smaller, light gray letters. Risk Management was performed in compliance with ISO 14971:2012 and includes FMEA analysis to review the following: - risks associated with the use, usability and performance of the device (HFFMEA and Risk Analysis) . - the risks associated with and specific to the design aspects of the device (Risk Analysis) . - Risks associated with biocompatibility, cleaning and sterilization Additional warning information was added to the product labeling to mitigate any residual risks. # Performance Testing - Non-Clinical: No performance standards have been established under Section 514 of the Food, Drug and Cosmetic Act for Dental Stereotaxic Instruments. - 1. Mechanical Properties: The predicate devices Bone Screws were tested in a STM E543-17. A review of the test results and comparison of the mechanical properties of the predicate device Bone Screws and the EDX Bone Screws are substantialy equival are substantially equivalent to the predicate device Bone Screws. Additional mechanical testing was conducted on the subject device to assess comparative deflection and stability of the EDX assembly under load conditions. - 2. Biocompatibility: Per FDA Guidance document 1811, "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process", section IV.A, biocompatibility testing is not necessary 'if the sponsor is able to document the use of a particular material in a legally-marketed predicate device with comparable tissue exposure, and is able to explain why manufacturing is not expected to adversely impact biocompatibility." The EDS Bone Screws, EDX Bone Screw Nut, and EDX Tracker Arms were deemed to be equivalent in material, process, and tissue contact to the predicate device. Therefore, this submission is results of Biocompatibility Testing (Cytotoxicity, Sensitization, and Irritation) conducted on our predicate device (K150222) to demonstrate biocompatibility compliance for the proposed components. - 3. Sterilization Testing: The EDX Bone Screws are considered geometrically similar to the predicate X-Guide® Surgical Navigation Systems, which were successfully tested and determined to meet a sterility assurance level (SAL) of ≤10-6 using the biological indicator (BI) overkill method. Therefore, this submission is results of cleaning and sterliization validation conducted per ISO 17665-1 and FDA Reprocessing Guidance Document to demonstrate compliance for the subject components. #### Performance Testing - Clinical Testing: Clinical testing was not necessary for the substantial equivalence determination. {10}------------------------------------------------ Image /page/10/Picture/0 description: The image shows the logo for X-NAV Technologies. The X is made up of green, blue, and gray horizontal lines. The NAV is in large, bold, blue letters. The word TECHNOLOGIES is in smaller, gray letters below the X-NAV. #### Substantial Equivalence Conclusion: Substantial equivalence between the predicate and the proposed EDX Bone Screw, EDX Tracker Arms, and EDX Bone Screw Nut has been demonstrated throughout this submission by performance and bench testing, companson of intended clinical and similarity in materials. This submission has demonstrated that differences in materials, geometry, or application methods between the predicate and reference devices do not raise any new issues of substantial equivalence. Additionally, the proposed changes within this Special 510(k) are substantially equivalent to the following: - Same indications for Use . - Sterilized in accordance with validated methods ● - Biocompatibility ●
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...