← Product Code [ELC](/submissions/DE/subpart-e%E2%80%94surgical-devices/ELC) · K240707

# Ultrasonic Scaler (WD-JY-B002,WD-JY-W001,WD-JY-R003,WD-JY-G004,WD-JY-B003,WD-JY-C006,WD-JY-M0014, WD-JY-N0015,WD-JY-N0016,WD-JY-N0017,WD-JY-E007, WD-JY-F008,WD-JY-A005,WD-JY-H009,WD-JY-I0011, WD-JY-K0012,WD-JY-K0013,WD-JY-K0014, WD-JY-L0013,WD-JY-L0014,WD-JY-L0015, WD-JY-L0016,WD-JY-L0017) (K240707)

_Shenzhen Micro Electric Intelligence Co., Ltd. · ELC · Jul 2, 2024 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/DE/subpart-e%E2%80%94surgical-devices/ELC/K240707

## Device Facts

- **Applicant:** Shenzhen Micro Electric Intelligence Co., Ltd.
- **Product Code:** [ELC](/submissions/DE/subpart-e%E2%80%94surgical-devices/ELC.md)
- **Decision Date:** Jul 2, 2024
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.4850
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Therapeutic

## Indications for Use

Ultrasonic Scaler is used to remove dental plaque from teeth and remove stains on teeth during dental cleaning.

## Device Story

Ultrasonic Scaler is a dental device used for plaque and stain removal. It consists of an electronic oscillation circuit that generates ultrasonic frequency electric pulse waves. These pulses are amplified and transmitted to a transducer, converting electrical energy into mechanical energy. This mechanical vibration is delivered to a working head (scaler tip) to crush calculus and loosen tartar. The device is intended for use by dental professionals during cleaning and periodontal therapy. The operator controls the device, and output is monitored via an external display driver circuit with LED indicators. The device aids in reducing tooth decay and cleaning gingival areas. Benefits include efficient removal of adherent plaque and food debris from teeth and gingival lines.

## Clinical Evidence

Bench testing only. No clinical data provided. Performance verified through electrical safety, EMC, and biocompatibility testing (ISO 10993-5, ISO 10993-10, ISO 10993-23). Compliance with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 62471, and IEC 80601-2-60 confirmed.

## Technological Characteristics

Ultrasonic transducer-based mechanical vibration; 3.7VDC power supply; 2.8W output power. Biocompatibility per ISO 10993-5 and ISO 10993-10. Electrical safety per IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 62471, and IEC 80601-2-60. LED-based display mode. Cleaning method involves boiling water.

## Regulatory Identification

An ultrasonic scaler is a device intended for use during dental cleaning and periodontal (gum) therapy to remove calculus deposits from teeth by application of an ultrasonic vibrating scaler tip to the teeth.

## Predicate Devices

- Ultrasonic scaler ([K163414](/device/K163414.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services. To the right of the symbol, there is the FDA logo in blue, with the words "U.S. FOOD & DRUG" above the word "ADMINISTRATION". The logo is simple and professional, reflecting the FDA's role in regulating food and drugs.

July 2, 2024

Shenzhen Micro Electric Intelligence Co., Ltd. Fengmeng Han General Manager 301, Building A5, No. 13 Baonan Road, Longgang Community Longgang District Shenzhen, Guangdong 518100 CHINA

Re: K240707

Trade/Device Name: Ultrasonic Scaler (WD-JY-B002,WD-JY-W001,WD-JY-R003,WD-JY-G004.WD-JY-B003.WD-JY-C006.WD-JY-M0014. WD-JY-N0015.WD-JY-N0016.WD-JY-N0017.WD-JY-E007. WD-JY-F008.WD-JY-A005.WD-JY-H009,WD-JY-I0011, WD-JY-K0012,WD-JY-K0013,WD-JY-K0014, WD-JY-L0013,WD-JY-L0014,WD-JY-L0015, WD-JY-L0016,WD-JY-L0017) Regulation Number: 21 CFR 872.4850 Regulation Name: Ultrasonic Scaler Regulatory Class: Class II Product Code: ELC Dated: June 3, 2024 Received: June 3, 2024

Dear Fengmeng Han:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}------------------------------------------------

Sincerely,

# Michael E. Adjodha -S

Michael E. Adjodha, MChE, RAC, CQIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

## Indications for Use

510(k) Number (if known) K240707

#### Device Name

Ultrasonic Scaler (WD-JY-B002,WD-JY-R003,WD-JY-R003,WD-JY-G003,WD-JY-C006,WD-JY-M0014, WD-JY-N0015,WD-JY-N0016,WD-JY-N0017,WD-JY-E007, WD-JY-F008,WD-JY-H009,WD-JY-I0011, WD-JY-K0012,WD-JY-K0013,WD-JY-K0014, WD-JY-L0013,WD-JY-L0015, WD-JY-L0015, WD-JY-L0016,WD-JY-L0017)

#### Indications for Use (Describe)

Ultrasonic Scaler is used to remove dental plaque from teeth and remove stains on teeth during dental cleaning.

| Type of Use (Select one or both, as applicable)                                         |                                                                                        |
|-----------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|
| <span style="font-family: Arial;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: Arial;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

# K240707_510(k) summary

#### Submitter l

### Device Submitter

Shenzhen Micro Electric Intelligence Co., Ltd.

301, Building A5,No. 13 Baonan Road, Longgang Community, Longgang District, Shenzhen

### Contact Person

FANGMENG HAN General Manager Phone: 86-755-23310679 E-mail: 360721873@qq.com Prepare Date: July 1, 2024

#### Device =

| Trade Name of Device: | Ultrasonic Scaler  |
|-----------------------|--------------------|
| Common Name           | Ultrasonic Scaler  |
| Regulation Number:    | 21 CFR 872.4850    |
| Regulation Name:      | Scaler, Ultrasonic |
| Regulatory Class:     | II                 |
| Product code:         | ELC                |
| Review Panel:         | Dental             |

# III Predicate Devices

| 510(k) Number      | K163414                    |
|--------------------|----------------------------|
| Device Name        | Ultrasonic scaler          |
| Manufacturer       | NANNING VV DENTAL CO., LTD |
| Regulation Number: | 21 CFR 872.4850            |
| Regulatory Class:  | II                         |
| Product code:      | ELC                        |
| Review Panel:      | Dental                     |

# IV Device description

A device intended for use during dental cleaning and periodontal (gum) therapy to remove adherent plaque and food debris from the teeth to reduce tooth decay, potential stains in the gingival groove, the back of the tooth and the tooth and associated gingival line.

The electronic oscillation circuit generates ultrasonic frequency electric pulse wave, which is amplified to a certain intensity and conveyed to the transducer. The electrical energy is

{5}------------------------------------------------

converted into mechanical energy, and the working head tool is stimulated to produce vibration of the same frequency, so as to achieve the purpose of crushing calculus and loosening tartar.

#### > Indications for use

Ultrasonic Scaler is used to remove dental plaque from teeth and remove stains on teeth during dental cleaning.

# VI Comparison of technological characteristics with the predicate devices

The Ultrasonic Scaler has the same intended use, technology, design and performance specifications are either identical or substantially equivalent to existing legally marketed predicate devices. The differences between the Ultrasonic Scaler and predicate devices do not alter suitability of the proposed device for its intended use.

| Item                            | Proposed Device                                                                                                                                                        | Predicate Device<br>(K163414)                                                                                                                                                                                                                                           | Comparison |
|---------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Manufacturer                    | Shenzhen Micro Electric<br>Intelligence Co., Ltd.                                                                                                                      | NANNING VV DENTAL<br>CO., LTD.                                                                                                                                                                                                                                          | /          |
| Regulation No.                  | 21 CFR 872.4850                                                                                                                                                        | 21 CFR 872.4850                                                                                                                                                                                                                                                         | Identical  |
| Class                           | II                                                                                                                                                                     | II                                                                                                                                                                                                                                                                      | Identical  |
| Product Code                    | ELC                                                                                                                                                                    | ELC                                                                                                                                                                                                                                                                     | Identical  |
| Indication for<br>use           | Ultrasonic Scaler is used<br>to remove dental plaque<br>from teeth and remove<br>stains on teeth during<br>dental cleaning.                                            | The ultrasonic scaler is<br>intended to:<br>1. Remove supra and sub<br>gingival calculus<br>deposits and stain from<br>the teeth;<br>2. Carry out periodontal<br>pocket lavage with<br>simultaneous ultrasonic tip<br>movement;<br>3. Clean and irrigate root<br>canals | Identical  |
| Operation<br>Controls           | Cleaning and disinfection<br>method of tooth cleaning<br>tip: after each use, scald                                                                                    | Scaler tip: autoclaved<br>under high temperature and<br>high pressure                                                                                                                                                                                                   | Different  |
| Item                            | Proposed Device                                                                                                                                                        | Predicate Device<br>(K163414)                                                                                                                                                                                                                                           | Comparison |
|                                 | the tooth cleaning tip with<br>boiling water for five<br>minutes, then wipe it dry<br>and store it for next use                                                        |                                                                                                                                                                                                                                                                         |            |
| Display Modes                   | Corresponding signal of<br>external display driver<br>circuit lights up<br>corresponding LED.                                                                          | Corresponding signal of<br>external display driver<br>circuit lights up<br>corresponding LED.                                                                                                                                                                           | Identical  |
| Measurement<br>Items            | remove adherent plaque<br>from the teeth and<br>potential stains in the<br>gingival groove, the back<br>of the tooth and the tooth<br>and associated<br>gingival line. | Remove calculus or plaque<br>on surface of teeth in oral<br>cavity, cleaning and<br>irrigation of root canals.                                                                                                                                                          | Similar    |
| Power Supply                    | Input Voltag: 3.7VDC 50HZ<br>Output Power: 2.8W                                                                                                                        | Input Voltage: 30VDC 50HZ<br>Output Power: 3-20W                                                                                                                                                                                                                        | Different  |
| Operating<br>conditions         | Temperature: 0-35°C<br>Relative Humidity:<br><80%R.H<br>Pressure: 70-106kpa                                                                                            | Temperature: 5°C-40°C<br>Relative Humidity: 30%--<br>80%<br>Pressure: 50kpa-106kpa                                                                                                                                                                                      | Different  |
| Storage<br>conditions           | Temperature: 0°C - +40°C<br>Relative Humidity:<br><80%R.H.<br>Pressure: 70-106kpa                                                                                      | Temperature: -10°C -+50°C<br>Relativehumidity: ≤ 80%<br>Pressure: 50kpa-106kpa                                                                                                                                                                                          | Different  |
| Tips of<br>ultrasonic<br>scaler | Mechanical vibration<br>transmitted from the<br>handpiece is converted to<br>the vibration of the scaler<br>tip, then removes calculus<br>or plaque from the teeth.    | Mechanical vibration<br>transmitted from the<br>handpiece is converted to<br>the vibration of the scaler<br>tip, then<br>removes calculus or plaque<br>from the<br>teeth                                                                                                | Different  |
| Item                            | Proposed Device                                                                                                                                                        | Predicate Device<br>(K163414)                                                                                                                                                                                                                                           | Comparison |
| Electrical<br>Safety            | IEC 60601-1                                                                                                                                                            | IEC 60601-1                                                                                                                                                                                                                                                             | Identical  |
| EMC                             | IEC 60601-1-2                                                                                                                                                          | IEC 60601-1-2                                                                                                                                                                                                                                                           | Identical  |
| Performnce                      | IEC 60601-1-11                                                                                                                                                         | IEC 61205                                                                                                                                                                                                                                                               | Different  |
| Biocompatibility                | ISO 10993-5, ISO 10993-<br>10                                                                                                                                          | ISO 10993-5, ISO 10993-<br>10                                                                                                                                                                                                                                           | Identical  |
| Software Level<br>of Concern    | Moderate                                                                                                                                                               | Moderate                                                                                                                                                                                                                                                                | Identical  |

{6}------------------------------------------------

510(k) Summary

{7}------------------------------------------------

### Discussion

Similarities:

Measurement Items is similar.

Proposed Device: remove adherent plaque and food debris from the teeth to reduce tooth decay, potential stains in the gingival groove, the back of the tooth and the tooth and associated gingival line.

Predicate Device: Remove calculus or plaque on surface of teeth in oral cavity, cleaning and irrigation of root canals.

## Difference:

Design:Operation Controls, Power Supply, Operating conditions and Storage conditions is slight different. However, Proposed Device has passed the Electrical Safety and EMC tests. Performance: the requirement of performance standards is slight different. However, Proposed Device has passed the performance test.

- The difference does not raise additional questions for safety and effectiveness. . Performance testing including biocompatibility evaluation has been performed on the final finished device which includes all construction materials and color additives.
In conclusion, there is no substantial difference between our products and similar products currently marketed in the United States.

# VII Performance data

The following performance data were provided in support of the substantial equivalence determination.

# Biocompatibility testing

Biocompatibility of the Ultrasonic Scaler was evaluated in accordance with ISO 10993-1:2018 for the body contact category of "Surface Device - Mucosal Membrane" with a contact duration of "Limited (< 24 hours)". The following tests were performed, as recommended: Cytotoxicity ISO 10993-5: 2009

{8}------------------------------------------------

Skin sensitization Intracutaneous reactivity

ISO 10993-10: 2021 ISO 10993-23: 2021

Non-Clinical Testing:

A battery of tests was performed to verify that the proposed device met all design specification. The test result demonstrated that the proposed device complies with the following standards: Electrical safety and electromagnetic compatibility

IEC 60601-1:2005 + AMD1:2012 + AMD2:2020 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance

IEC 60601-1-2:2015 Medical electrical equipment -Part 1-2: General requirements for basic safety and essential performance-Collateral Standard: Electromagnetic disturbances -Requirements and tests

IEC 60601-1-11:2015+AMD1:2020 Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment

IEC 62471:2006 Photobiological safety of lamps and lamp systems

IEC 80601-2-60:2019 Medical electrical equipment - Part 2-60: Particular requirements for the basic safety and essential performance of dental equipment

#### VIII Conclusion

Conclusions drawn from non-clinical trials indicate that the subject device is as safe and effective as the legally marketed predicted K163414 ultrasonic scale and performs as well or better than the K163414 ultrasonic scaler.

---

**Source:** [https://fda.innolitics.com/submissions/DE/subpart-e%E2%80%94surgical-devices/ELC/K240707](https://fda.innolitics.com/submissions/DE/subpart-e%E2%80%94surgical-devices/ELC/K240707)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
