Air Motor (Model name: FX204 M4); Air Motor (Model name: M205)

K233308 · Nakanishi, Inc. · EFB · Apr 25, 2024 · Dental

Device Facts

Record IDK233308
Device NameAir Motor (Model name: FX204 M4); Air Motor (Model name: M205)
ApplicantNakanishi, Inc.
Product CodeEFB · Dental
Decision DateApr 25, 2024
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.4200
Device ClassClass 1
AttributesTherapeutic

Indications for Use

Air Motor is intended for the following application(s): Driving the handpiece for the tooth restoration, prophylaxis, and root canal preparation.

Device Story

Air Motor (Models FX204 M4 and M205) converts compressed air from dental units into rotary motion; transmits rotation to straight or geared angle handpieces for tooth restoration, prophylaxis, and root canal preparation. M205 includes internal water spray mechanism. Used in dental clinics by dentists/dental staff. Output is mechanical rotation driving dental instruments. Benefits include efficient power transmission for dental procedures. No software or electronic processing involved.

Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing, including ISO 14457 performance testing, sterilization validation, cleaning validation, and biocompatibility testing per ISO 10993-1 and ISO 7405.

Technological Characteristics

Pneumatically driven air motor; FX204 M4 (no spray), M205 (internal spray). Materials: Fluor rubber, stainless steel, aluminum alloy. Connections: ISO 3964 (handpiece), ISO 9168 (hose). Sterilization: Autoclave (AAMI/ANSI/ISO 17665-1). No software or electronic components.

Indications for Use

Indicated for dental professionals performing tooth restoration, prophylaxis, and root canal preparation using air-powered handpieces.

Regulatory Classification

Identification

A dental handpiece and accessories is an AC-powered, water-powered, air-powered, or belt-driven, hand-held device that may include a foot controller for regulation of speed and direction of rotation or a contra-angle attachment for difficult to reach areas intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text. April 25, 2024 NAKANISHI INC. % Takahiro Haruyama President Globizz Corpration 1411 W. 190th St. Suite 200 Gardena, California 90248 Re: K233308 Trade/Device Name: Air Motor (Model name: FX204 M4); Air Motor (Model name: M205) Regulation Number: 21 CFR 872.4200 Regulation Name: Dental Handpiece And Accessories Regulatory Class: Class I, reserved Product Code: EFB Dated: March 7, 2024 Received: March 13, 2024 Dear Takahiro Haruyama: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). {2}------------------------------------------------ Sincerely, # Michael E. Adjodha -S Michael E. Adjodha, MChE, RAC, CQIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ### Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. Submission Number (if known) K233308 Device Name Air Motor (Model name: FX204 M4) Air Motor (Model name: M205 ) Indications for Use (Describe) Air Motor is intended for the following application(s): Driving the handpiece for the tooth restoration, prophylaxis, and root canal preparation. Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for NSK, a company that specializes in bearings and automotive products. The logo is white text on a blue background. The letters "NSK" are in a bold, sans-serif font. The logo is simple and recognizable. # 510(k)#: K233308 510(k) Summary #### Contact Details Applicant Name: NAKANISHI INC. Applicant Address: 700 Shimohinata Kanuma Tochigi, 322-8666, Japan Applicant Contact Telephone: +81-289-64-7277 Applicant Contact: Masaaki Kikuchi Applicant Contact Email: nskra@nsk-nakanishi.co.jp #### Device Name Device Trade Name: Air Motor (Model name: FX204 M4) Air Motor (Model name: M205) Common Name: Dental handpiece and accessories Classification Name: Handpiece, Air-Powered, Dental Regulation Number: 872.4200 Device Class: I Product Codes: EFB #### Legally Marketed Predicate Devices Predicate Device #: K143704 Predicate Trade Name: Advanced Air System Product Code: EFB Reference Device #: K162926 Predicate Trade Name: W&H Air-Powered Handpieces And Handpiece Attachments Product Code: EFB #### Device Description Summary The air motor converts the compressed air supplied from the dental unit into rotary motion. The rotation is transmitted to instruments for tooth restoration, prophylaxis, and root canal preparation connected to straight and geared angle handpieces. #### Intended Use/ Indications for Use Air Motor is intended for the following application(s): Driving the handpiece for the tooth restoration, prophylaxis, and root canal preparation. # NAKANISHI INC. {5}------------------------------------------------ Comparison of Technological Characteristics with the Predicate Device-1 ### DEVICE AND PREDICATE COMPARISON TABLE-1 ### For Air Motor, FX204 M4 (model without spray) | Descriptive Information | Subject Device | Primary Predicate Device | Reference Device | Judgement | |---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Device Name | Air Motor, FX204 M4 | Advanced Air System<br>(RM-25 L RM) | Air-Powered Handpieces<br>and Handpiece<br>Attachments<br>(AM-25 RM: model w/o<br>spray) | | | K number | K233308 | K143704 | K162926 | | | Indication for Use | Air Motor is intended for<br>the following<br>application(s): Driving the<br>handpiece for the tooth<br>restoration, prophylaxis,<br>and root canal<br>preparation. | Pneumatic drive system<br>for dental handpieces and<br>dental air motors, which is<br>intended to be used in<br>general dental applications<br>such as: removal of<br>decayed materials, cavities<br>and crown preparations,<br>removal of filings, finishing<br>of tooth and restoration<br>surfaces, polishing,<br>prophylaxis and<br>endodontics | The turbine handpiece is<br>intended for the following<br>applications: Removal of<br>decayed materials, cavities<br>and crown preparation,<br>removal of fillings,<br>finishing of tooth and<br>restoration surfaces.<br>The dental<br>handpiece/contra-angle is<br>intended for the following<br>applications: Removal of<br>decayed materials, cavities<br>and crown cement,<br>removal of fillings,<br>finishing and polishing of<br>tooth and restoration<br>surfaces. | Same: all devices are<br>intended for general<br>dental applications<br>including restoration,<br>dental cleaning<br>(prophylaxis), and root<br>canal preparation | | Device Description | The air motor converts<br>the compressed air<br>supplied from the dental | The Advanced Air System<br>is intended for dental<br>transmission instruments | The Air-powered<br>Handpieces contain pipes<br>inside which supply air for | Same: all devices use air<br>motors, are<br>pneumatically driven | | | | | | | | unit into rotary motion. The rotation is transmitted to instruments for tooth restoration, prophylaxis, and root canal preparation connected to straight and geared angle handpieces. | used in the field of preventive dentistry, restorative dentistry such as cavity preparation and prosthodontics such as crown preparation and endodontics. The system consists of the control unit (AC-1.0), which is designed to be built in a dental chair. As an attachment the air-driven handpiece (RK-97 L, RG- 97 L, and/or RK-94 L) or the air motor (RM-25 L RM) can be used. The air-driven handpiece is equipped with a speed sensor, which is connected to a control module that regulates the applied air pressure to the handpiece through a proportional valve in order to maintain constant speed throughout the dental treatment | driving the turbine's impeller as well as supply air and water for creating spray exhausting through 3, 4 or 5 ports located in the handpiece's head and intended for cooling the working area as well as the rotating bur. Depending on the type of handpiece – the handle might contain an integrated circuit board and wire bundles, for supplying current and voltage to a head sided assembled LED light source. The Handpiece Attachments also contain pipes inside which supply air and water for spray exhausting and for cooling the working area as well as the rotating bur. Depending on the type of handpiece – the handle might contain an integrated circuit board and wire bundles, for supplying current and voltage to a head-sided assembled LED light source. | and used for driving dental handpieces. | | | Supply air | 0.20 – 0.25 MPa | 0.52 – 0.59 MPa | unknown | Different<br>The subject device has | | | | | | lower pressure | | Optic lighting | no | yes | no | Same with Air-Powered<br>Handpieces and<br>Handpiece Attachments | | Speed range | $19,800 - 24,200 min^{-1}$<br>(0.25MPa) | $2,000 – 20,000 rpm$ | $25,000 +/-10%$ | Similar<br>All devices have similar<br>upper speed range | | Max. air consumption | < 66NL/min | 1.5 l/s (3.2 cfm) | 42-50NI/min | Similar with Air-<br>Powered Handpieces<br>and Handpiece<br>Attachments. | | Weight | 61.7g | 39 g | 73g | Similar weight to Air-<br>Powered Handpieces<br>and Handpiece<br>Attachments | | Performance Testing | Performance: ISO<br>14457:2017<br>Sterilization:<br>AAMI/ANSI/ISO 17665-<br>1:2006<br>Cleaning: Guidance for<br>the Reprocessing<br>Medical Devices in Health<br>Care Settings: Validation<br>Methods and Labeling<br>Guidance for Industry and<br>Food and Drug<br>Administration Staff. | Electrical Safety Tests: IEC<br>60601-1:2005<br>Electromagnetic<br>Compatibility: IEC 60601-<br>1-1:2007<br>Product testing of<br>handpiece function and<br>life cycle testing: ISO<br>14457:2012<br>Software validation: IEC<br>62304:2006<br>Usability validation: IEC<br>62366:2007<br>Thermal safety: IEC<br>62471:2006 | Product testing of the<br>handpieces function and<br>life cycle testing: ISO<br>14457:2012<br>EMC Testing: IEC 60601-1-<br>2:200<br>Thermal safety: IEC<br>62471:2006 | Similar<br>All the devices passed<br>the ISO 14457<br>Performance testing.<br>Relevant testing to<br>predicate devices.<br>• ISO 14457 2017<br>Conformity<br>confirmation report<br>• Drop test report<br>• Reprocessing<br>resistance test report<br>• Water leakage and<br>water ingress test<br>report<br>• Statistical calculation<br>of sample size | | | | | | | | | | | | • Equivalence<br>confirmation for<br>package test report<br>• Vibration test report<br>• Sterilization<br>validation report<br>• Conformity<br>assessment to the<br>product family group for<br>sterilization<br>• Product family group<br>conformity confirmation<br>on cleaning validation<br>• Cleaning validation | | | | | | report | | | | | | • Cleaning validation of<br>accessories | | Biocompatibility | ISO 7405: 2018<br>AAMI/ANSI/ISO 10993-1:<br>2018 | ISO 7405:2008<br>ISO 10993-5:2009 | ISO 7405:2008<br>ISO 10993-5:2009<br>Evaluation of<br>biocompatibility is based<br>upon the fact that patient<br>contacting materials in the<br>subject handpiece are<br>identical to those in the<br>previously cleared<br>predicates<br>(K143704 and K070663),<br>which as handpieces and<br>handpiece attachments,<br>present the same level and<br>duration of contact. In | <i>Similar</i><br>Biocompatibility testing<br>confirms compliance<br>with ISO 10993-1 and<br>supports substantial<br>equivalence.<br>Relevant testing to<br>predicate devices.<br>• Biological safety<br>evaluation report<br>• Confirmation of<br>effectiveness and<br>biological safety of<br>accessories | | | | | addition, Cytotoxicity<br>Testing per EN<br>ISO 10993-5:2009 was<br>performed. This evaluation<br>meets the requirements of<br>ISO 7405:2008 for<br>preclinical evaluation of<br>biocompatibility of dental<br>devices. | • MSDS | | Clinical Testing | NO | NO | NO | Same | | Air/water port | No spray | No spray | No Spray | Same | | Dimensions | 78.3 x φ20 mm | 87,87 x φ20 mm | Unknown | Similar dimensions with<br>Advanced Air System.<br>ISO 14457 performance<br>testing supports<br>substantial equivalence. | | Patient-contacting portions | Direct contact: No<br>Indirect contact: No | Direct contact: Unknown<br>Indirect contact: No | Direct contact: Unknown<br>Indirect contact: No | Similar<br>Conformity to ISO<br>10993-1 supports<br>substantial equivalence. | | Handpiece Connection | ISO 3964 | Unknown | ISO 3964 | Same with Air-Powered<br>Handpieces and<br>Handpiece<br>Attachments. | | Hose Connection | ISO 9168 | Unknown | ISO 9168 | Same with Air-Powered<br>Handpieces and<br>Handpiece<br>Attachments. | | Lubricant | PANA SPRAY Plus<br>(K163483) | Unknown | W&H Service Oil F1, MD-<br>400<br>W&H Assistina | Similar with Air-<br>Powered Handpieces<br>and Handpiece<br>Attachments.<br>Conformity to ISO<br>10993-1 supports<br>substantial equivalence. | {6}------------------------------------------------ {7}------------------------------------------------ {8}------------------------------------------------ NSK {9}------------------------------------------------ {10}------------------------------------------------ #### Discussion: Intended use: same. The intended use for the Air Motor FX204 M4 and its predicate device Advanced Air System RM-25 U RM is for general dental applications. The principle of the operating technology: same (pneumatically driven). There are differences in the two devices features, such as "supply air pressure". To verify if the subject device affects its efficiency, we compared the subject device and the reference device under air supply pressures, reflecting the and reference device. The performance gap at the lower air supply pressure limit was relatively small. Despite this, the product's performance as an air motor remains sufficient for power transmission to attachments and grinding teeth (Test Report_Equivalence Check of Similar Devices). In terms of performance, it is equivalent to the reference device. Therefore, we conclude that the subject device Air Motor FX204 M4 and its predicate devices Advanced Air System RM-25 L RM and Air-Powered Handpieces and Handpiece Attachments AM-25 RM are substantially equivalent. {11}------------------------------------------------ # DEVICE AND PREDICATE COMPARISON TABLE-2 ### For Air Motor, M205 (Internal Spray equipped) | Descriptive Information | Subject Device | Primary Predicate Device | Reference Device | Judgement | |--------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------…
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...