K211358 · Guilin Woodpecker Medical Instrument Co., Ltd. · EBW · Nov 21, 2022 · Dental
Device Facts
Record ID
K211358
Device Name
Implanter incl. Accessories
Applicant
Guilin Woodpecker Medical Instrument Co., Ltd.
Product Code
EBW · Dental
Decision Date
Nov 21, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.4200
Device Class
Class 1
Attributes
Therapeutic
Intended Use
Mechanical drive unit with coolant supply for transmission instruments with ISO 3964 (DIN13940) compatible coupling system, for use in dental surgery, implantology and maxillofacial surgery (CMF) for treatment of dental hard tissue.
Device Story
Implanter is a mechanical drive unit for dental surgery, implantology, and maxillofacial surgery. It consists of a control unit with a capacitive touch display, a motor with cable (SPM58L), a wired foot control (MF4), and surgical contra-angle handpieces (WP-1L/WP-1). The device provides coolant supply for transmission instruments with ISO 3964 coupling. Users select from eight programs (Positioning, Hole-drilling, Hole-Broadening, Tapping, Implanting, Lock Abutment Screw, User Defined, Cleaning) via foot control or touch display. The system controls motor speed and torque to assist clinicians in dental hard tissue procedures. Output is delivered through the handpiece to the surgical site. The device is intended for use in dental practices and clinics by dental professionals. It benefits patients by providing controlled, precise mechanical power for implant placement and related surgical tasks.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated through bench testing, including electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), handpiece function/lifecycle (ISO 14457), thermal safety (IEC 62471), biocompatibility (ISO 10993-1, -5, -10; ISO 7405), and reprocessing validation.
Technological Characteristics
Mechanical drive unit; 120W max power; 300-40,000 rpm speed range. Materials: stainless steel, aluminum alloy, silicone, chromium-plated/coated steel/brass. ISO 3964 coupling. Wired foot control. Sterilization: dynamic-air-removal (132°C/4 min). Software: Moderate level of concern, compliant with IEC 62304.
Indications for Use
Indicated for use in dental surgery, implantology, and maxillofacial surgery (CMF) for treatment of dental hard tissue in patients requiring such procedures.
Regulatory Classification
Identification
A dental handpiece and accessories is an AC-powered, water-powered, air-powered, or belt-driven, hand-held device that may include a foot controller for regulation of speed and direction of rotation or a contra-angle attachment for difficult to reach areas intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 21, 2022
Guilin Woodpecker Medical Instrument Co., Ltd. % Fu Ailing Consultant Shenzhen Joyantech Consulting Co., Ltd. 1713A, 17th Floor, Block A, Zhongguan Times Square Liuxian Avenue, Xili Town Nashan District Shenzhen, Guangdong 518055 CHINA
Re: K211358
Trade/Device Name: Implanter incl. Accessories Regulation Number: 21 CFR 872.4200 Regulation Name: Dental Handpiece And Accessories Regulatory Class: Class I, reserved Product Code: EBW Dated: October 25, 2022 Received: October 25, 2022
Dear Fu Ailing:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Michael E. Adjodha -S
Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K211358
Device Name Implanter incl. Accessories
Indications for Use (Describe)
Mechanical drive unit with coolant supply for transmission instruments with ISO 3964 (DIN13940) compatible coupling system, for use in dental surgery, implantology and maxillofacial surgery (CMF) for treatment of dental hard tissue.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Version:A/0
Product: Implanter
# 510(k) Summary — K211358
This 510(k) Summary is submitted according to the requirements of 21 CFR §807.92.
# 5.1 Administrative Information
| Date Summary prepared | November 17, 2022 |
|--------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer information | Submitter's Name: Guilin Woodpecker Medical Instrument<br>Co., Ltd.<br>Address: Information Industrial Park, Guilin National High-<br>Tech Zone, Guilin, Guangxi, 541004, P.R. China<br>Contact person: Yang Yunfeng<br>TEL: +86-773-2350532<br>FAX: +86-773-5822450<br>Mail: ipr@glwoodpecker.com |
| Submission Correspondent | Company Name: Shenzhen Joyantech Consulting Co., Ltd.<br>Address: Room 1713A, 17 Floor, Block A, Zhongguan Times<br>Square, Liuxian Avenue, Xili Town, Nanshan<br>District, Shenzhen, Guangdong, 518055, China<br>Contact person: Ms. Fu Ailing<br>E-Mail: aileen@cefda.com<br>Image: [Logo] |
| Establishment registration<br>number | 3005581016 |
# 5.2 Device Information
| Type of 510(k) submission: | Traditional |
|----------------------------|--------------------------------------|
| Common Name: | Controller, Foot, Handpiece and Cord |
| Trade Name: | Implanter incl. Accessories |
| Model: | Implanter |
| Classification name: | Dental Handpiece and Accessories |
{4}------------------------------------------------
Product: Implanter
Version:A/0
| Review Panel: | Dental |
|--------------------|-----------------|
| Product Code: | EBW |
| Device Class: | I |
| Regulation Number: | 21 CFR 872.4200 |
## 5.3 Predicate Device
#### Predicate Device
| Sponsor: | W&H DENTALWERK BÜRMOOS GMBH |
|----------------|--------------------------------------|
| Device: | Implantmed SI-1015 incl. Accessories |
| 510(K) Number: | K161957 |
#### Reference Device
| Sponsor: | W&H Dentalwerk Buermoos GmbH |
|----------------|----------------------------------------------------------------------------------------|
| Device: | ImplantMED SI-915 (115V Version) ImplantMED SI-923<br>(230V Version) Incl. Accessories |
| 510(K) Number: | K052741 |
#### 5.4 Device Description
The Implanter is intended for use in dental surgery, implantology and maxillofacial surgery (CMF) for treatment of dental hard tissue.
The submission consists of:
- the control unit
- a motor with cable with light (sterilizable: SPM58L)
- a wired foot control (MF4)
- and as an attachment the surgical contra-angle handpieces (with LED:WP-1L/without LED:WP-1)
{5}------------------------------------------------
| Guilin Woodpecker | 009_510(k) Summary |
|--------------------|--------------------|
| Product: Implanter | Version: A/0 |
Eight different programs can be selected by the user. Switching between these programs is performed by foot control or via touch display.
The programs include Positioning, Hole-drilling, Hole-Broadening, Tapping, Implanting, Lock the Abutment Screw, User Defined Mode and Cleaning.
The control unit is intended to be used with the motor SPM58L.
To run the Implanter according to its intended use, Guilin Woodpecker provides two different surgical handpieces.
| Type | Transmission ratio | Max. speed | Max. torque |
|-------|--------------------|------------|-------------|
| WP-1L | 20:1 | 40,000 | 80 Ncm |
| WP-1 | 20:1 | 40,000 | 80 Ncm |
## 5.5 Indications for Use
lmplanter (incl.SPM58L):
Mechanical drive unit with coolant supply for transmission instruments with ISO 3964 (DIN13940) compatible coupling system, for use in dental surgery, implantology and maxillofacial surgery (CMF) for treatment of dental hard tissue.
# 5.6 Indication for Use and Technological Characteristics of the Subject Device Compared to the Predicate Device
{6}------------------------------------------------
# Table 1 Comparison of Indications for Use and Technological Characteristics Between the Subject Device and the Predicate Device
| General Specification Implanter incl. Accessories | | | | | | |
|---------------------------------------------------|--------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| ID | Comparison<br>Item | Proposed Device<br>Implanter incl.<br>Accessories | Predicate Device<br>Implantmed SI-1015<br>incl. Accessories<br>K161957 | Reference Device<br>ImplantMED SI-915 (115V<br>Version) ImplantMED SI-<br>923 (230V Version) Incl.<br>Accessories<br>K052741 | Difference Between the<br>Proposed Device and the<br>Predicate Device | |
| 1 | Picture | Image: Proposed Device Implanter incl. Accessories | Image: Predicate Device Implantmed SI-1015 incl. Accessories K161957 | Image: Reference Device ImplantMED SI-915 (115V Version) ImplantMED SI-923 (230V Version) Incl. Accessories K052741 | - | |
| 2 | Name | Implanter | Implantmed SI-1015 | Implantmed SI-915/923 | - | |
| 3 | Manufacturer | Guilin Woodpecker<br>Medical Instrument Co.,<br>Ltd. | W&H Dentalwerk<br>Bürmoos GmbH | W&H Dentalwerk Bürmoos<br>GmbH | - | |
| | Product: Implanter | | | | | Version:A/0 |
| 4 | Where Used | Dental practice, dental clinic | Dental practice, dental clinic | Dental practice, dental clinic | None | |
| 5 | Indication for Use | Mechanical drive unit with coolant supply for transmission instruments with ISO 3964 (DIN13940) compatible coupling system, for use in dental surgery, implantology and maxillofacial surgery (CMF) for treatment of dental hard tissue. | Mechanical drive unit with coolant supply for transmission instruments with ISO 3964 (DIN13940) compatible coupling system, for use in dental surgery, implantology and maxillofacial surgery (CMF) for treatment of dental hard tissue. | Mechanical drive unit with coolant supply for transmission instruments with coupling system according to ISO 3964 (Din 13.940) The equipment is a drive unit for use in dental surgery, implantology and maxilla-facial surgery for treatment of dental hard tissue. | None | |
| 6 | Control Unit | Main dimensions:<br>276x267x110 mm<br><br>Front panel:<br>Display with capacitive touch<br><br>Programs:<br>8 programs for various | Main dimensions:<br>100x262x291 mm<br><br>Front panel:<br>Display with capacitive touch<br><br>Programs:<br>5 programs for various | Main dimensions:<br>90x252x254mm<br><br>Front panel:<br>Graphical display without backlighting<br><br>Programs:<br>5 programs for various | Yes.<br>For Guilin Woodpecker Medical Instrument Co., Ltd., there are 8 programs:<br>1. Positioning,<br>2. Hole-drilling,<br>3. Hole-Broadening,<br>4. Tapping, | |
{7}------------------------------------------------
009_510(k) Summary
{8}------------------------------------------------
| Product: Implanter | | | | Version:A/ | | |
|--------------------|-------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| | stages of implantology | stages of implantology | stages of implantology | 5. Implanting, | | |
| | Irrigation: | Irrigation: | Irrigation: | 6. Lock the abutment screw, | | |
| | 110ml/min. | 100ml/min. | 100ml/min. | 7. User defined Mode, | | |
| | | | | 8. Cleaning | | |
| | | | | For W&H Dentalwerk Bürmoos<br>GmbH, there are | | |
| | Irrigation Tubing can be<br>inserted ergonomically on<br>the unit's side face | Irrigation Tubing can be<br>inserted ergonomically on<br>the unit's side face | Irrigation Tubing can be<br>inserted on the unit's front<br>face | 5 programs: | | |
| | | | | Drill function, | | |
| | | | | Drill function, | | |
| | | | | Drill function, | | |
| | | | | Thread-cutter function, | | |
| | | | | Implant insertion. Similar<br>functions include: | | |
| | | | | 2 vs a or b or c; | | |
| | | | | 4 vs d; | | |
| | | | | 5 vs e. | | |
| | | | | The additional programs | | |
| | | | | included for the proposed<br>device include 1,3,6,7 and 8. | | |
| | | | | 1 – Includes set ranges of<br>speed and torque for | | |
| | | | | positioning of implant<br>placement. | | |
| Product: Implanter | | | | Version:A/C | | |
| | | | | 3 – Includes set ranges of<br>speed and torque for hole-<br>broadening procedures.<br>6 – Includes set ranges of<br>speed and torque for<br>placement of the abutment<br>screw.<br>7 – User defined mode for<br>use with a straight handpiece<br>with ball drill bit.<br>8 – Used to adjust the water<br>capacity.<br>Differences between<br>functions are supported by<br>performance testing. | | |
| 7 | Foot Control | Main dimension:<br>249.7x181.7x160.8<br><br>Features:<br>4 buttons for pump on/off<br>Forward/reverse<br>Change programs | Main dimension:<br>154x202x210<br><br>Features:<br>4 buttons for pump on/off<br>Forward/reverse<br>Change programs | Main dimension:<br>215x190x40 (without<br>cable)<br><br>Features:<br>4 buttons for pump on/off<br>Forward/reverse<br>Change programs | Yes. The power supply of<br>the proposed device is<br>performed via wired foot<br>control not via<br>wireless control. The<br>reference device also<br>includes a wired foot<br>control. | |
| Product: Implanter | | | | | | Version:A/0 |
| | | Motor control (on/off and variable) | Motor control (on/off and variable) | Motor control (on/off and variable) | | |
| | | Power supply:<br>Wired<br>Via cable | Power supply:<br>Wireless<br>Via 3xAA batteries | Power supply:<br>wired<br>via cable | | |
| 8 | Motor with<br>Cable | Length:<br>110 mm<br>With LED contacts<br>without LED contacts | Length:<br>71.65 mm<br>With LED contacts | Length:<br>75 mm<br>Without LED contacts | Yes. The motor is with LED<br>contacts or without LED<br>contacts. The predicate<br>device includes LED<br>contacts and the reference<br>device includes no LED<br>contacts. | |
| 9 | Module | / | W&H SI-SQ:<br>Connection via USB-<br>cable | Osstell ISQ:<br>Stand-alone device | Yes. The proposed device<br>does not include a module.<br>However, performance<br>testing has been included<br>demonstrating its use. | |
{9}------------------------------------------------
{10}------------------------------------------------
{11}------------------------------------------------
009_510(k) Summary
| Version:A/<br>Product: Implanter | | | | | | |
|----------------------------------------------------------------------------------|------------------------------------|---------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Technical Data/Functions Implanter incl. Accessories | | | | | | |
| ID | Comparison<br>Item | Proposed Device<br>Implanter incl.<br>Accessories | Predicate Device<br>Implantmed SI-1015<br>incl. Accessories | Reference Device<br>ImplantMED SI-915 (115V<br>Version) ImplantMED SI-<br>923 (230V Version) Incl.<br>Accessories | Difference Between the<br>Proposed Device and the<br>Predicate Device | |
| 10 | Max.<br>Mechanical<br>output power | 120 W | 80 W | 70 W | Yes. Due to different design<br>scheme, the proposed<br>device has the higher<br>power, but this does not<br>lead to any negative effect<br>regarding substantial<br>equivalence because of the<br>constructive power<br>reserves of the motor (no<br>big difference between<br>5.5 Ncm - 6.2 Ncm for the<br>motor). | |
| 11 | Torque at the<br>motor | 5.5 Ncm | 6.2 Ncm | 5.5 Ncm | Yes. The proposed device<br>offers a lower maximum<br>level of torque output at the<br>motor: 5.5 Ncm. However,<br>this is in alignment with the<br>reference device. | |
| Version:A/0<br>Product: Implanter | | | | | | |
| 12 | Speed range<br>of motor | 300 - 40,000 rpm | 200 - 40,000 rpm | 200 - 40,000 rpm | None | |
| 13 | Supply<br>voltage | 100-120 VAC | 100-130 VAC | 100-130 VAC | None | |
| 14 | Rated<br>current | 1.5-2 A | 0.3-1.6 A | 0.2-1.6 A | Yes. This does not affect<br>substantial equivalence<br>according to performance<br>testing. | |
| 15 | Frequency | 50-60 Hz | 50-60 Hz | 50-60 Hz | None | |
| Material | | | | | | |
| ID | Comparison<br>Item | Proposed Device<br>Implanter incl.<br>Accessories | Predicate Device<br>Implantmed SI-1015<br>incl. Accessories | Reference Device<br>ImplantMED SI-915 (115V<br>Version) ImplantMED SI-<br>923 (230V Version) Incl.<br>Accessories | Difference Between the<br>Proposed Device and the<br>Predicate Device | |
| 16 | Control unit<br>housing | Plastic material | Plastic material | Plastic material | None | |
| 17 | Tubing outer<br>shealth | Customer specific | Customer specific | Customer specific | None | |
| Product: Implanter Version:A/C | | | | | | |
| 18 | Motor with<br>cable | Stainless Steel,<br>Aluminum Alloy (Motor)<br>and Silicone, Silicone<br>dioxide,<br>Polydimethylsiloxane<br>(Cable) | Stainless steel | Stainless steel | Yes. This does not raise any<br>additional questions regarding<br>substantial equivalence<br>according to testing<br>conducted per ISO 10993. | |
| 19 | Surgical<br>handpieces | Chromium plated steel<br>and chromium coated<br>brass | Chromium coated steel<br>and chromium coated<br>brass | Chromium coated steel<br>and chromium coated<br>brass | Yes. This does not raise any<br>additional questions<br>regarding substantial<br>equivalence according to<br>testing conducted per ISO<br>10993. | |
| 20 | Module | / | Stainless steel | Stainless steel | Yes. The proposed device<br>does not include a module.<br>Therefore, material<br>composition is not applicable. | |
| Hygiene/Maintenance<br>Surgical Handpieces WP-1L/WP-1, Motors SPM58L, and others | | | | | | |
| Product: Implanter Version:A/0 | | | | | | |
| ID | Comparison<br>Item | Proposed Device<br>Implanter incl.<br>Accessories | Predicate Device<br>Implantmed SI-1015<br>incl. Accessories | Reference Device<br>ImplantMED SI-915 (115V<br>Version) ImplantMED SI-<br>923 (230V Version) Incl.<br>Accessories | Difference Between the<br>Proposed Device and the<br>Predicate Device | |
| 21 | Lubrication | After max. 30 minutes of<br>use<br>SPM58L:<br>No lubrication is needed | After max. 30 minutes of<br>use<br>EM-19 LC/EM-19:<br>No lubrication is needed | After max. 30 minutes of<br>use<br>Motor with cable:<br>No lubrication is needed | None | |
| 22 | Cleaning | Rinse under distilled<br>water or deionized water<br>(<45 °C/113°F) with a<br>soft cloth | Rinse under<br>demineralized water (< 38 °C/100°F) with aid of<br>brush | Rinse under demineralized<br>water (< 38°C/100°F) with<br>aid of brush | Yes. This does not affect<br>substantial equivalence<br>according to reprocessing<br>validation. | |
| 23 | Disinfection | Wiping disinfection using<br>disinfectant cloths | Wiping disinfection using<br>disinfectant cloths | Wiping disinfection using<br>disinfectant cloths | None | |
| 24 | Sterilization | Dynamic-air-removal<br>sterilizers:<br>270 °F (132 °C) for 4<br>minutes | Dynamic-air-removal<br>sterilizers:<br>270 °F (132 °C) for 4<br>minutes or<br>Gravity displacement<br>sterilizers:<br>270 °F (132 °C) for 15<br>minutes | Dynamic-air-removal<br>sterilizers:<br>270°F (132°C) for 4<br>minutes or<br>Gravity displacement<br>sterilizers:<br>270°F (132°C) for 15<br>minutes | None. Same Sterilization Cycle<br>Parameters for Dynamic-air-<br>removal. | |
{12}------------------------------------------------
#### Guilin Woodpecker
{13}------------------------------------------------
{14}------------------------------------------------
{15}------------------------------------------------
From above table, the proposed device and the predicate device have identical indications for use. Although there are subtle technological characteristic differences between the proposed device and the predicate device, the technological characteristic differences discussed do not affect the substantial equivalence.
## 5.7 Brief discussion of the non-clinical tests
To verify the performance requirements of the proposed device, the following tests were performed in accordance with listed standards and FDA guidance documents. Testing results demonstrate substantial equivalence.
- To verify the conformity of the proposed device with the requirements of IEC 60601-1:(Medical electrical equipment Part 1: General requirements for basic safety and essential performance).
- To verify the conformity of the proposed device with the requirements of IEC 60601-1-2:(Medical electrical equipment Part 1-2: General requirements for basic safety and essential performance Collateral Standard: Electromaqnetic compatibility).
- To verify the handpiece function and lifecycle according to ISO 14457:2017: (Dentistry-Handpieces and Motors).
- To validate the proposed device's software and its foot control's software respectively in conformity with IEC 62304 (Medical device software - Software lifecycle processes) and FDA Guidance titled "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" for a MODERATE level of concern.
- To verify the thermal safety in conformity with the standard IEC 62471:2004: (Photobiological safety of lamps and lamp systems).
- To evaluate the biocompatibility of patient contacting components of the proposed device according to the requirements ISO 10993-1 (Biological evaluation of medical devices -Part 1: Evaluation and testing within a risk management process), ISO 10993-5 (Biological evaluation of medical devices Part 5: Test for cytotoxicity), ISO 10993-10 ( Biological evaluation of medical devices - Part 10: Tests for skin sensitization ) and ISO 7405 (Dentistry - Evaluation of biocompatibility of medical devices used in dentistry).
- To verify the conformity of the proposed device with the requirements of IEC 60601-1-2:(Medical electrical equipment Part 1-2: General requirements for basic safety and essential performance Collateral Standard: Electromagnetic compatibility) and IEC 60601-1-1 (Medical electrical equipment - Part 1-1: General requirements for safety - Collateral standard: Safety requirements for medical electrical systems).
- To validate the reprocessing (Cleaning, Sterilization) according to FDA Guidance titled "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling -Guidance for Industry and Food and Drug Administration Staff,"
- To verify the conformity of the proposed device with the requirements of FDA Guidance on Dental Handpieces titled "Dental Handpieces – Premarket Notification [510(k)] Submissions.
{16}------------------------------------------------
### 5.8 Brief discussion of clinical tests
No clinical data is needed to demonstrate substantial equivalence.
### 5.9 Other information (such as required by FDA guidance/Test)
N/A.
## 5.10 Conclusions
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification, Guilin Woodpecker Medical Instrument Co., Ltd. concludes that:
- The indications for use of the proposed device is identical to that of the predicate device. .
- The technological characteristic differences between the proposed device and the predicate . device do not affect the substantial equivalence, so no new risk is raised.
- Demonstrated by the safety and performance tests, the characteristics of the proposed device . are substantially equivalent to those of the predicate device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.