K203706 · Guilin Woodpecker Medical Instrument Co., Ltd. · EBW · Jan 13, 2022 · Dental
Device Facts
Record ID
K203706
Device Name
Dental Electric Motor
Applicant
Guilin Woodpecker Medical Instrument Co., Ltd.
Product Code
EBW · Dental
Decision Date
Jan 13, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.4200
Device Class
Class 1
Attributes
Therapeutic
Indications for Use
The MT2 dental electric motor consists of main unit, motor, motor tail, power adapter and power cord. It is intended for use in general dental applications such as: cutting a tooth for cavity preparation, crown preparation, crown finishing, inlay, filing, polishing, prophylaxis and endodontic treatment. The Dental Electric Motor can be used with ISO E-type a straight, right-angle or contra-angle handpiece attachment of equal, gear-reducing, or gear increasing speed. The contra-angle handpieces (model: WJ-15, WJ-15L) are driven by a micro-motor to rotate at a specified speed, so as to drive the dental bur. It is applicable to drilling and grinding in dental surgery.
Device Story
MT2 Dental Electric Motor is a control unit system comprising a main unit, DC electric micromotor, motor tail, power adapter, and power cord. It connects to an AC power supply and drives a DC micromotor; speed and on/off functions are regulated via a dental unit foot control (not included). The device accepts ISO E-type straight, right-angle, or contra-angle handpieces (e.g., WJ-15, WJ-15L) to drive dental burs for cutting, grinding, and polishing. Used in dental clinics by dental professionals. The system provides low-voltage DC power to the handpiece, which may include LED lighting. The output is mechanical rotation of the dental bur, enabling the clinician to perform restorative and surgical dental procedures. Benefits include precise control of motor speed and torque for various dental applications.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing, including electrical safety (IEC 60601-1, IEC 80601-2-60), EMC (IEC 60601-1-2), biocompatibility (ISO 10993-5, ISO 10993-10), and sterilization validation (ISO 17665-1).
Technological Characteristics
System includes main control unit, DC micromotor (aluminum alloy exterior), and power adapter (100-240Vac input, 30V DC output). Supports ISO 3964 E-type handpieces. Features LED lighting, forward/reverse rotation, and 2,000-40,000 rpm speed range. Sterilization via steam (autoclave) per ISO 17665-1. Software is moderate level of concern.
Indications for Use
Indicated for general dental applications including cavity preparation, crown preparation/finishing, inlay, filing, polishing, prophylaxis, endodontic treatment, and dental surgery drilling/grinding. Used with ISO E-type handpieces.
Regulatory Classification
Identification
A dental handpiece and accessories is an AC-powered, water-powered, air-powered, or belt-driven, hand-held device that may include a foot controller for regulation of speed and direction of rotation or a contra-angle attachment for difficult to reach areas intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.
K990954 — ELEC-MATE, MODELS EX-6 SET, EX-6L SET, EX-30 SET · Nsk Nakanishi, Inc. · Jun 9, 1999
K032572 — A-DEC/W&H ELECTRIC MOTOR, MODEL EA-50LT · A-Dec, Inc. · Dec 3, 2003
K202075 — Dental Electric Motor · Foshan Cicada Dental Instrument Co, Ltd. · Mar 11, 2022
K992648 — ROTARY MASTER · J. Morita USA, Inc. · Jan 31, 2000
K040153 — I-TEC PORTABLE DENTAL OPERATORY · Indigenous Peoples Technology and Education Center · May 21, 2004
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 13, 2022
Guilin Woodpecker Medical Instrument Co., Ltd. % Field Fu Senior Consultant Shenzhen Joyantech Consulting Co., Ltd. Block A, Zhongguan Times Square, Liuxian Avenue, Xili Town, Nanshan District Shenzhen, Guangdong CHINA
Re: K203706
Trade/Device Name: Dental Electric Motor Regulation Number: 21 CFR 872.4200 Regulation Name: Dental Handpiece And Accessories Regulatory Class: Class I, reserved Product Code: EBW, EGS Dated: December 7, 2021 Received: December 20, 2021
Dear Field Fu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K203706
Device Name Dental Electric Motor
#### Indications for Use (Describe)
The MT2 dental electric motor consists of main unit, motor tail, power adapter and power cord. It is intended for use in general dental applications such as: cutting a tooth for cavity preparation, crown finishing, inlay, filing, polishing, prophylaxis and endodontic treatment. The Dental Electric Motor can be used with ISO E-type a straight, right-angle or contra-angle handpiece attachment of equal, gear-reducing, or gear increasing speed.
The contra-angle handpieces (model: WJ-15, WJ-15L) are driven by a micro-motor to rotate at a specified speed, so as to drive the dental bur. It is applicable to drilling and grinding in dental surgery.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| <span style="font-family: Arial, sans-serif;"> <span style="font-size: 10pt;"> <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </span> | |
| Over-The-Counter Use (21 CFR 801 Subpart C) | |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
# K203706
This summary of 510(K) safety and effectiveness information is submitted As Required by requirements of SMDA and 21 CFR §807.92.
# 5.1 Administrative Information
| Date of Summary<br>prepared | January 11, 2022 |
|--------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer<br>information | Company title:<br>Guilin Woodpecker Medical Instrument Co., Ltd. |
| | Company address:<br>Information Industrial Park, Guilin National High-<br>Tech Zone, Guilin, Guangxi, 541004, China<br>Contact person: Jiakang Ning<br>Phone: +86 7735873198<br>Fax: +86-7735822450<br>E-mail:zmnbg03@glwoodpecker.com |
| Submission<br>Correspondent | Shenzhen Joyantech Consulting Co., Ltd.<br>Address: 1713A, 17th Floor, Block A, |
| Image: [logo]卓远天成 | Zhongguan Times Square, Liuxian Avenue, Xili<br>Town, Nanshan District, Shenzhen, Guangdong<br>Province, China.<br>Contact person: Mr. Field Fu<br>E-Mail:field@cefda.com; |
| Establishment<br>registration number | 3005581016 |
# 5.2 Device Information
| Type of 510(k) submission: | Traditional |
|----------------------------|--------------------------------------|
| Trade Name: | Dental Electric Motor |
| Model: | MT2 |
| Classification name: | Controller, Foot, Handpiece and Cord |
| Regulation Name: | Dental Handpiece and Accessories |
{4}------------------------------------------------
| Review Panel: | Dental |
|--------------------------|----------|
| Product Code: | EBW |
| Additional Product Code: | EGS |
| Device Class: | I |
| Regulation Number: | 872.4200 |
### 5.3 Predicate Device Information
| Primary predicate device | |
|--------------------------|----------------------------------|
| Sponsor: | Nakanishi, Inc. |
| Device: | A-dec NLZ electric motor system |
| 510(K)Number: | K163131 |
| Reference device | |
| Sponsor: | Nakanishi, Inc. |
| Device: | General Cutting Contra Handpiece |
| 510(K)Number: | K182999 |
### 5.4 Device Description
The MT2 dental electric motor consists of main unit, motor, motor tail, power adapter and power cord. It is intended for use in general dental applications such as: cutting a tooth for cavity preparation, crown preparation, crown finishing, inlay, filing, polishing, prophylaxis and endodontic treatment, with use of a straight, rightangle or contra-angle handpiece attachment of equal, gear-reducing, or gear increasing speed.
Dental Electric Motor (model: MT2) is a control unit, which is connected to an AC power supply and a handpiece hose, that drives a DC electric micromotor, and turns on or off or regulates the speed of the motor by the foot control of the dental unit. The foot switch is not included in the control unit. It is used together with a handpiece attachment. It provides a low voltage DC power supply for the handpiece attachment with light.
The device is for prescription use.
{5}------------------------------------------------
### 5.5 Indications for Use
The MT2 dental electric motor consists of main unit, motor, motor tail , power adapter and power cord. It is intended for use in general dental applications such as: cutting a tooth for cavity preparation, crown preparation, crown finishing, inlay, filing, polishing, prophylaxis and endodontic treatment. The Dental Electric Motor can be used with ISO E-type a straight, right-angle or contra-angle handpiece attachment of equal, gear-reducing, or gear increasing speed.
The contra-angle handpieces (model: WJ-15, WJ-15L) are driven by a micro-motor to rotate at a specified speed, so as to drive the dental bur. It is applicable to drilling and grinding in dental surgery.
# 5.6 Technological characteristics of the subject device compared to the predicate device
| Item | Subject Device | Primary Predicate<br>Device(K163131) | Reference<br>device(K182999) | Remarks |
|-------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------|
| Product<br>Code | EBW, EGS | EBW | EGS | Identic<br>al |
| Regulation<br>No. | 872.4200 | 872.4200 | 872.4200 | Identic<br>al |
| Class | I | I | I | identic<br>al |
| Indication<br>for use | The MT2 dental electric<br>motor consists of main<br>unit, motor, motor tail,<br>power adapter and<br>power cord. It is<br>intended for use in<br>general dental<br>applications such as:<br>cutting a tooth for cavity<br>preparation, crown<br>preparation, crown<br>finishing, inlay, filing,<br>polishing, prophylaxis<br>and endodontic<br>treatment. The Dental<br>Electric Motor can be<br>used with ISO E-type a<br>straight, right-angle or | The A-dec NLZ<br>electric motor<br>system is comprised<br>of a control unit that<br>drives a direct<br>current (DC) electric<br>micromotor that is<br>activated by means<br>of a foot control. It is<br>intended for use in<br>general dental<br>applications such<br>as: cutting a tooth<br>for cavity<br>preparation. Crown | The General<br>Cutting Contra<br>Handpiece is<br>powered by either<br>an air-motor or<br>electric<br>micromotor for<br>use in general<br>dentistry. The<br>device is intended<br>for cutting and<br>grinding teeth,<br>cavity<br>preparations,<br>tooth and crown | Simila<br>r<br>(Discu<br>ssion<br>1) |
| Item | Subject Device | Primary Predicate Device(K163131) | Reference device(K182999) | Remarks |
| | contra-angle handpiece<br>attachment of equal,<br>gear-reducing, or gear<br>increasing speed.<br>The contra-angle<br>handpieces (model:WJ-<br>15, WJ-15L) are driven<br>by a micro-motor to<br>rotate at a specified<br>speed, so as to drive<br>the dental bur. It is<br>applicable to drilling<br>and grinding in dental<br>surgery. | preparation, crown<br>finishing, inlay,<br>filing, polishing,<br>prophylaxis and<br>endodontic<br>treatment, with use<br>of a straight, right-<br>angle or contra-<br>angle ISO E-type<br>handpiece<br>attachment of equal,<br>gear-reducing, or<br>gear-increasing<br>speed. | preparations,<br>finishing and<br>trimming teeth<br>and filling<br>materials and<br>removal of crowns<br>and filling<br>materials. | |
| Dental electric Motor | | | | |
| Drive | Electric-micromotor | Electric-micromotor | NA | Identical |
| Components | Main unit, DC adapter,<br>DC cord, motor, motor<br>tail | Motor controller,<br>Electric micromotor,<br>Motor tubing | NA | Identical |
| Power<br>source | AC/DC Adapter:<br>Power supply<br>input:100-240Vac,<br>50/60Hz, 2.5A<br>Power supply output:<br>DC30V, 3.0A | --- | NA | --- |
| Size | Main unit:<br>W165.5 x D129.7 x<br>H77.6mm<br>Motor:<br>DΦ22 x H76.7mm<br>Motor tail:<br>Length: 1800mm | Motor controller:<br>D78.5 x W148.0 x<br>H43.0mm<br>Motor:<br>Length: 31mm,<br>Diameter:<br>Φ20.1(Front),<br>Φ22.1(Rear)<br>Motor tubing:<br>Length:<br>NLZ CDAS:<br>1600mm<br>NLZ CDAI: 2080mm | NA | Different<br>(Discussion 2) |
| Material of<br>Motor<br>exterior | Aluminum alloy | Titanium | NA | Different<br>(Disc |
| Item | Subject Device | Primary Predicate<br>Device(K163131) | Reference<br>device(K182999) | Remarks |
| Light | LED | LED | NA | Identical |
| Rotation<br>Speed | 2,000-40,000 rpm | 1,000 - 40,000 rpm | NA | Similar(Discussion 2) |
| Rotation<br>direction | Forward and reverse | Forward and<br>reverse | NA | Identical |
| Coolant<br>mechanism | Coolant air | Coolant air | NA | Identical |
| Available<br>Handpiece<br>type | E-type (ISO 3964) | E-type (ISO 3964) | NA | Identical |
| Usage<br>environment | Temperature: 5-40°C,<br>Humidity: 30-75% | Temperature: 0 -<br>40°C, Humidity: 30-75% | NA | Similar(Discussion 2) |
| Handpiece | | | | |
| Operational<br>modes | Air-power | NA | Air-power | Identical |
| Optical<br>fiber | WJ-15: without light<br>WJ-15L: with light | NA | With light, without<br>light | Identical |
| Type of<br>chuck | Mechanical type chuck | NA | Mechanical type<br>chuck | Identical |
| Coupling<br>dimensions | Middle (ISO 3964-<br>2016) | NA | ISO 3964 (EN<br>ISO<br>3964) Standard<br>Coupling | Identical |
| Material of<br>handpiece | Copper, Stainless steel | NA | Titanium | Different<br>(Discussion<br>2) |
| Shanks for<br>rotary<br>instruments | Type3,Φ1.6mm drill<br>( ISO 1797-1-2011) | NA | Type 1, Type 3 | Identical |
| Item | Subject Device | Primary Predicate<br>Device(K163131) | Reference<br>device(K182999) | Remarks |
| Gear Ratio<br>Max<br>rotation<br>speed<br>(handpiece) | 1:5 Increasing-<br>200,000 min-1 | NA | 16:1 Reduction-<br>2,500 min-1<br>10:1 Reduction-<br>4,000 min-1<br>4:1 Reduction-<br>10,000 min-1<br>1:1 Direct Drive-<br>40,000 min-1<br>1:5 Increasing-<br>200,000 min-1 | Identical |
| Lubricant | NSK Pana-Spray<br>(K052700) | NA | NSK PANA<br>SPRAY<br>Plus (K163483) | Identical |
| Sterilization | Sterilized by user<br>(Steam sterilization) | Sterilized by user<br>(Steam sterilization) | Sterilized by user<br>(Steam<br>sterilization) | Identical |
| Sterilization<br>Validation<br>Standard | ANSI/AAMI/ISO<br>17665-1:2006 | ANSI/AAMI/ISO<br>17665-1:2006 | ANSI/AAMI/ISO<br>17665-1:2006 | Identical |
### Comparison to predicate device:
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
Discussion 1: The predicate device is only a motor system. while the subject device includes a motor system and handpiece. Therefore, the intended use of the two is a little different. In order to be equivalent, we have added a reference device(K182999).
Discussion 2: In the comparison table, there are some differences between the subject device and the predicate device, such as the size and the materials of the considered that the differences did not affect the identical principles of operation between the subject device and the predicate device. And the differences would not cause new risk to users and patients. Performance testing including Biocompatibility testing, Electric safety and EMC testing, bench testing was performed in order to demonstrate substantial equivalence to the predicate device. Also, sterilization validation and software validation were performed in accordance with FDA Guidance. The subject device met all the requirements of the standards. Therefore, the subject device demonstrated substantial equivalence to the predicate device.
Note1: DC adaptor meets IEC 60601-1 standard.
## 5.7 Brief discussion of the nonclinical tests
The subject device conforms to the following standards:
IEC 60601-1:2005+AM1:2012 Medical Electrical Equipment - Part 1: General
{9}------------------------------------------------
Requirements For Basic Safety And Essential Performance.
IEC 60601-1-2:2014 Medical Electrical Equipment - Part 1-2: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Disturbances - Requirements And Tests
IEC 80601-2-60 : 2019 Medical Electrical Equipment - Part 2-60: Particular Requirements For The Basic Safety And Essential Performance Of Dental Equipment
ISO 10993-5:2009 Biological Evaluation Of Medical Devices - Part 5: Tests For In Vitro Cytotoxicity.
ISO 10993-10:2010 Biological Evaluation of Medical Devices- Part 10: Tests for Irritation and Skin Sensitization.
ISO 14457: 2017 Dentistry- Handpieces and motors.
Software Documentation of a MODERATE level of concern was provided per FDA Software Guidance titled: "Guidance for Industry and FDA Staff- Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices.
Cleaning and Sterilization Validation of the Contra-Angle and Motor and High-Level Disinfection Validation for the Motor Tubing were provided per FDA Reprocessing Guidance titled: "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling - Guidance for Industry and Food and Drug Administration Staff.
## 5.8 Brief discussion of clinical tests
N/A.
#### 5.9 Other information (such as required by FDA guidance/Test)
The FDA Dental Handpiece Guidance titled: "Guidance for Industry and FDA Staff - Dental Handpieces - Premarket Notification [510(k)] Submissions," available at https://www.fda.gov/media/71432/download.
{10}------------------------------------------------
## 5.10 Conclusions
The differences between subject device and predicate/reference devices do not affect the safety and effectiveness of the subject devices.
Thus, the subject device is substantially equivalent to the predicate device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.