PS SERIES WITH FIBER LIGHT GUIDE AND HEADLIGHT, MODELS PS200, PS250, PS500
Applicant
Photonic Optische Gerate GmbH & Cokg
Product Code
EBA · Dental
Decision Date
Feb 12, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.4630
Device Class
Class 1
Attributes
3rd-Party Reviewed
Indications for Use
The AC-powered dental operating light, which can be attached via optical fibers to a headlight, is a device intended to illuminate oral structures and operating areas. The device is intended for Prescription Use only!
Device Story
Plated culture medium; RPMI 1640 base supplemented with 35g/L MOPS, 2% glucose, and agar. Used with antibiotic gradient-based systems (preformed rectangular plastic strips impregnated with antifungal agents). Inoculated agar plate receives strip; agent releases into agar matrix creating exponential concentration gradient. After 24-48 hours aerobic incubation at 35°C, symmetrical inhibition ellipse forms around strip. Intersection of zone edge and strip indicates Minimum Inhibitory Concentration (MIC). Used in clinical microbiology laboratories by trained personnel. Output is MIC value (μg/ml) used by clinicians to determine antifungal susceptibility of Candida isolates, aiding therapeutic decision-making.
Clinical Evidence
No clinical data provided. Bench testing only; device meets international safety standards including IEC 60601-1, IEC 60601-1-2, IEC 60601-2-18, IEC 60601-2-41, UL2601-1, and FCC requirements.
Technological Characteristics
Plated medium: RPMI 1640 base, 35g/L MOPS, 2% glucose, agar. Principle: Dilution/diffusion gradient. Energy: Passive diffusion. Dimensions: 5 x 60 mm strip. Sterilization: Not specified. Software: None.
Indications for Use
Indicated for use with antibiotic gradient systems for quantitative determination of susceptibility to fluconazole, itraconazole, and flucytosine when testing Candida species.
Regulatory Classification
Identification
A dental operating light, including the surgical headlight, is an AC-powered device intended to illuminate oral structures and operating areas.
K061622 — PS SERIES · Photonic Optische Gerate Gesmbh & Cokg · Mar 5, 2007
K163185 — OptiLux LED Illuminator · Isolux, LLC · Nov 21, 2016
K032283 — DESIGNS FOR VISION FIBEROPTIC LIGHT · Designs For Vision, Inc. · Aug 20, 2003
K053227 — LUXENON 150-300 WATT MEDICAL FIBEROPTIC LIGHT SOURCE ILLUMINATING HEADLIGHTS, SPOTLIGHTS, REFRACTORS AND ENDOSCOPES · Medical Illumination International, Inc. · Jan 13, 2006
K042382 — DUAL LED LITE ENGINE WITH ILLUMINATION LED CABLE COUPLED TO A LED LITE HEAD · Med-General Technologies, LLC · Nov 30, 2004
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number:
K072827
B. Purpose for Submission:
To obtain substantial equivalence determination for the RPMI 1640 Agar with MOPS and 2% Glucose Antifungal Susceptibility Test Medium
C. Measurand:
Susceptibility of fungal isolates
D. Type of Test:
Antifungal Susceptibility Test Medium for use with antibiotic gradient-based systems
E. Applicant:
Remel, Inc.
F. Proprietary and Established Names:
Susceptibility Test Medium, RPMI 1640 Agar with MOPS and 2% Glucose
G. Regulatory Information:
1. Regulation section:
21 CFR Part 866.1700 Culture medium for antimicrobial susceptibility tests
2. Classification:
Class II
3. Product code:
MJE - Culture media, antifungal susceptibility test
4. Panel:
{1}
83 Microbiology
1. Intended use(s):
RPMI 1640 Agar with MOPS and 2% Glucose is a plated medium recommended for use with antibiotic gradient-based systems for quantitative determination of susceptibility to antifungal agents when testing *Candida* species directly from colonies grown on nonselective media.
2. Indication(s) for use:
RPMI 1640 Agar with MOPS and 2% Glucose antifungal susceptibility test media is indicated for use with antibiotic gradient systems for quantitative determination of susceptibility to fluconazole, itraconazole, and flucytosine when testing *Candida* species.
3. Special conditions for use statement(s):
Prescription use only.
For antifungal testing, due to trailing effect, minimum inhibitory concentrations (MIC’s) should be read at approximately 90% inhibition of growth, ignoring faint hazes and minute colonies for flucytosine and 80% inhibition for fluconazole and itraconazole.
This product is for *In Vitro* diagnostic use and should be used by properly trained individuals.
4. Special instrument requirements:
N/A
I. Device Description:
RPMI 1640 is a modified plated media developed from the RPMI 1630 media series, which is supplemented with 35 grams of morpholinepropanesulfonic acid (MOPS) per liter of broth media, 2% glucose and agar. Antifungal agents are impregnated on a preformed rectangular plastic strip (5 x 60 mm). One side of the strip carries a letter code designating the identity of the antifungal agent and is calibrated in terms of MIC values in micrograms/ml (μg/ml). A predefined and exponential gradient of the dried and stabilized antifungal agent, covering a continuous concentration range across 15 two fold dilutions of a conventional MIC method, is immobilized on the opposite surface of the carrier strip. The concentration maximum is at the opposite end from the concentration minimum. When applied to an inoculated RPMI-1640 agar plate, the agent is immediately released into the agar matrix beneath the carrier. After aerobic incubation at 35°C from 24 to 48 hours a symmetrical inhibition ellipse
2
{2}
centered along the carrier is seen. The zone edge intersects the strip at the minimum inhibitory concentration (MIC) value given in micrograms/ml (μg/ml).
# J. Substantial Equivalence Information:
1. Predicate device name Mueller Hinton Agar with $2\%$ NaCl
2. Predicate 510(k) number K960313
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Atmosphere | Aerobic | Same |
| Test medium | Gradient strip(s) placed on solid agar | Same |
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Incubation time and temperature | 35°C for 24 hours with MIC confirmation at 36 - 48 hours | 30 - 35°C for 24 hours |
| Inoculum concentration | Candida species, suspension adjusted to 0.5 McFarland standard, or 1 McFarland standard for mucoid strains | Staphylococcus species suspension adjusted to 0.5 McFarland standard |
| Inoculum technique | Plate swabbed 6 times for confluent growth | Plate swabbed 3 times for confluent growth |
| Interpretation | MIC's should be read at approximately 90% inhibition of growth, ignoring faint hazes and minute colonies for Flucytosine and 80% inhibition for Fluconazole and Itraconazole. | MIC is read at the end point where there is complete inhibition by oxacillin |
# K. Standard/Guidance Document Referenced (if applicable):
M27-A2, "Reference Method for Broth Dilution Antifungal Susceptibility Testing of
{3}
Yeasts; Approved Standard – Second Edition”, NCCLS, 2002.
L. Test Principle:
RPMI 1640 was developed by Moore et al. at Roswell Park Memorial Institute (RPMI). The formulation is based on the RPMI-1630 media series which utilizes a bicarbonate buffering system and differing amounts of amino acids and vitamins. RPMI-1640 (liquid) medium is widely used in cell culture and more recently as the reference method for antifungal broth microdilution recommended by the Clinical Laboratory Standards Institute (CLSI). This formulation of RPMI 1640 is supplemented with 35 grams of morpholinepropanesulfonic acid (MOPS) per liter of broth media, 2% glucose and agar. The antibiotic gradient-based method is based on a combination of the concepts of both dilution and diffusion principles for susceptibility testing. Antibiotic/antifungal agents are impregnated on a preformed plastic strip, which, when applied to an inoculated agar plate, are immediately released into the agar matrix. A continuous and exponential gradient of antibiotic/antifungal concentration is created beneath the carrier. After incubation at 35°C a symmetrical inhibition ellipse centered along the carrier is seen. The zone edge intersects the strip at the minimum inhibitory concentration (MIC) value given in micrograms/ml (μg/ml).
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility of QC organisms:
Three different lots of Etest®/RPMI-1640 media agar with MOPS and 2% Glucose were tested in triplicate for three consecutive days at one internal site and at 3 external clinical trial sites for intra-site and inter-site reproducibility. There were 648 results generated at 24 hours, and a total of 729 results generated at 48 hours, respectively, using three fungal QC ATCC organisms, at the external clinical sites. A total of 72 results at 24 hours and 81 results at 48 hours were obtained at the internal site. Fluconazole and Itraconazole MIC results demonstrated >95% reproducibility. Flucytosine MIC's demonstrated >95% reproducibility after 24 hours of incubation. There were 10 C. albicans ATCC 90028 MIC values that read one dilution above the expected range after 48 hours of incubation, only at one external study site. There were 12 MIC's at the same site that could be read as either 2 or 4 μg/ml. For consistency all 12 readings were analyzed at 4 μg/ml, which were within the expected MIC range. The Reproducibility Study is summarized in the tables
{4}
below.
| 24 hours | Fluconazole | Itraconazole | Flucytosine | Total |
| --- | --- | --- | --- | --- |
| External sites | 162/162* | 243/243 | 243/243 | 648/648 |
| Internal site | 18/18 | 27/27 | 27/27 | 72/72 |
| Total | 180/180 | 270/270 | 270/270 | 720/720 |
| Percentage in Range | 100%** | 100% | 100% | 100% |
| 48 hours | | | | |
| External sites | 243/243 | 242/243 | 221/243 | 706/729 |
| Internal site | 27/27 | 27/27 | 27/27 | 81/81 |
| Total | 270/270 | 269/270 | 248/270 | 787/810 |
| Percentage in Range | 100% | 99.6% | 91.9% | 97.2% |
* Number of readings in expected MIC range / total number of readings
**Percentage of readings within the expected MIC range
The results for which the Flucytosine MIC values were above the antifungal gradient recommended QC ranges were obtained all those same runs. The subjective reading of antifungal susceptibility test results due to the trailing effect of Flucytosine at 48 hours of incubation was due to operator-related interpretation.
Although the number of reproducible Flucytosine results at 48 hours was below the expected level at one site, the combined reproducibility for all three antifungal agents at the three external sites and the one internal site was greater than 95%. Therefore, the combined reproducibility data are acceptable.
b. Linearity/assay reportable range:
N/A
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
The CLSI recommended QC organisms Candida parapsilosis ATCC 22019 and Candida krusei ATCC 6258 were tested at each clinical trial site using RPMI-1640 media agar with MOPS and 2% Glucose, and the antifungal gradient-based method. The CLSI and Etest® QC expected ranges for yeast microdilution testing for each QC organism are provided below.
Expected QC Ranges for Yeast Microdilution Testing in µg/ml
| Agent | C.parapsilosis
ATCC 22019
CLSI | C. parapsilosis
ATCC 22019
Etest® | C. krusei
ATCC 6258
CLSI | C. krusei
ATCC 6258
Etest®** | C. albicans
ATCC 90028
Etest® |
| --- | --- | --- | --- | --- | --- |
| Fluconazole | | | | | |
| 24 hours | 0.5 - 4 | 1 - 4 | 8 - 64 | N/A | 0.125-0.5 |
| 48 hours | 1 - 4 | 1 - 4 (8)* | 16 - 128 | 128 - >256 | 0.125-0.5 |
| | | | | | |
| Itraconazole | | | | | |
{5}
| 24 hours | 0.12 - 0.5 | 0.032 - 0.25 | 0.12 - 1 | 0.125 - 0.5 | 0.032 - 0.125 |
| --- | --- | --- | --- | --- | --- |
| 48 hours | 0.12 - 0.5 | 0.064 - 0.25 | 0.25 - 1 | 0.25 - 1 | 0.064 - 0.25 |
| | | | | | |
| Flucytosine*** | | | | | |
| 24 hours | 0.06 - 0.25 | 0.064 - 0.25 | 4 - 16 | ≥ 32 | 0.125 - 1 |
| 48 hours | 0.12 - 0.5 | 0.064 - 0.25 | 8 - 32 | ≥ 32 | 0.5 - 2 |
* Occasionally, isolated colonies may grow up to 8 μg/ml.
** C. krusei are assumed to be intrinsically resistant to azoles. CLSI Expected Range QC differs from Etest.
C. krusei was removed from the current Etest Package Insert (Table 2) because the QC ranges/values are off-range. This isolate is no longer recommended as a useful QC strain for Fluconazole or Flucytosine by AB Biodisk.
***Flucytosine: Interpretive criteria and recommended QC organisms as described in CLSI document M27- are the same as those referenced in the FDA approved drug label updated on 08/10/2006. $\mathrm{N / A} =$ Not applicable
d. Detection limit:
N/A
e. Analytical specificity:
N/A
f. Assay cut-off:
N/A
# 2. Comparison studies:
# a. Method comparison:
A total of 16 fungal clinical isolates with known MIC results were tested once at two clinical sites with the RPMI 1640 agar with MOPS and $2\%$ Glucose, with gradient-based methodology. The original test method used at Site 1 was broth microdilution, prepared according to the CLSI reference method. The original test method at Site 2 was the antibiotic gradient method, tested on in-house RPMI agar plates prepared at the site. This data is presented for comparative purposes only and is not included in the reproducibility study calculations. The summary of the results are presented in the table below.
| Additional Clinical Isolates | N | % EA | % CA* |
| --- | --- | --- | --- |
| Fluconazole | | | |
| Site 1 | 6 | 100 | 100 |
| Site 2 | 10 | 100 | 90 |
{6}
| Total | 16 | 100 | 93.8 |
| --- | --- | --- | --- |
| | | | |
| Itraconazole | | | |
| Site 2 | 10 | 90 | 80 |
| Total | 10 | 90 | 80 |
| | | | |
| Flucytosine | | | |
| Site 1 | 6 | 83.3 | 100 |
| Site 2 | 10 | 90 | 90 |
| Total | 16 | 87.5 | 93.8 |
| Combined Total | | 92.9 | 90.5 |
* Interpretation criteria per CLSI document M27-A2
antifungal
Essential Agreement (EA) was calculated when the results for the drug MIC were within +/- two doubling dilutions for each agent.
Glucose MIC results
The overall correlation between RPMI 1640 agar with MOPS and 2% using gradient-based methodology MIC's with previously known obtained using two different test methods was good.
There were 7 results that did not correlate, with six of the seven results observed at one site (Site 2). A summary of the results that did not correlate is provided in the table below.
| | Site | Previously known MIC | CLSI Interpretation | Etest® MIC | CLSI Interpretation |
| --- | --- | --- | --- | --- | --- |
| Isolate #5 | 1 | Flucytosine - 0.023 | S | 0.25 | S |
| Isolate #1 | 2 | Fluconazole - 32 | SDD | 128 | R |
| Isolate #7 | 2 | Itraconazole - 1 | R | 32 | R |
| Isolate #8 | 2 | Itraconazole - 0.06 | S | 0.25 | SDD |
| Isolate #10 | 2 | Itraconazole - 0.12 | S | 0.5 | SDD |
| Isolate #6 | 2 | Flucytosine - 16 | I | 32 | R |
| Isolate #9 | 2 | Flucytosine - 0.06 | S | 0.01 | S |
S = Susceptible; SDD = Susceptible-Dose Dependent; I = Intermediate; R = Resistant
All 7
Three of the seven results remained within CA. However, there was a trend for gradient-based methodology results to show slightly higher MIC's than the predetermined MIC. This observation may be due to the well documented trailing
{7}
effect and operator related interpretation associated with antifungal susceptibility testing using this testing method.
The combined performance for the three antifungal agents tested at two clinical sites demonstrated an EA of 92.9% and a CA of 90.5%, which are acceptable.
## Detection of non-susceptible clinical isolates
The primary role of antifungal susceptibility testing is to identify resistant isolates among susceptible populations. Of the 24 total Candida species evaluated during the reproducibility study, there were 11 isolates that demonstrated resistance to one or more antifungal agent, both internally and at one clinical site (Site 2).
There were a total of 18 different resistance phenotype combinations detected. The gradient-method detected 100% (18/18) of the non-susceptible isolates. There were 17 of 18 (94.4%) Candida species isolates grouped into the appropriate interpretive category, as either SDD or R. All non-susceptible isolates were identified, the data are acceptable and the performance was good.
b. Matrix comparison:
N/A
3. Clinical studies:
a. Clinical Sensitivity:
N/A
b. Clinical specificity:
N/A
c. Other clinical supportive data (when a. and b. are not applicable):
4. Clinical cut-off:
N/A
5. Expected values/Reference range for Candida species:
| CLSI Interpretive Criteria Category | Fluconazole | Itraconazole | Flucytosine** |
| --- | --- | --- | --- |
| Susceptible | ≤ 8 | ≤ 0.125 | ≤ 4 |
| Susceptible – Dose Dependent* | 16 – 32 | 0.25 – 0.5 | - |
| Intermediate | - | - | 8 – 16 |
| Resistant | ≥ 64 | ≥ 1 | ≥ 32 |
*Susceptible – Dose Dependent requires achieving maximum possible blood level for azoles.
**Flucytosine: Interpretive criteria and recommended QC organisms as described in CLSI document M27-A2 are the same as those referenced in the FDA approved drug label updated on 08/10/2006.
{8}
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The information submitted in this premarket notification is complete and supports a substantial equivalence determination.
9
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.