DENTAL BONE SCREWS; DENTAL BONE SCREWS, CONTINUING; SCREW SYSTEM TX; SCREW SYSTEM TX PROFESSIONAL

K132212 · Hager& Meisinger GmbH · DZL · Feb 14, 2014 · Dental

Device Facts

Record IDK132212
Device NameDENTAL BONE SCREWS; DENTAL BONE SCREWS, CONTINUING; SCREW SYSTEM TX; SCREW SYSTEM TX PROFESSIONAL
ApplicantHager& Meisinger GmbH
Product CodeDZL · Dental
Decision DateFeb 14, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.4880
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Meisinger Dental Bone Screws are intended for fixation of bone segments e.g. for horizontal or vertical augmentative grafts. Depending on indication the user choose the matching screw, prepares the pit hole and inserts the screw. The screws are intended for single use only.

Device Story

Meisinger Dental Bone Screws are titanium alloy (Grade 5) fixation devices used in dental procedures. They feature a self-tapping design with socket head or hexalobular connections for driver engagement. Available in diameters of 1.1mm, 1.3mm, and 1.5mm, and lengths ranging from 7mm to 16mm. The device is used by dental clinicians to stabilize bone segments during augmentative grafting. The clinician selects the appropriate screw, prepares a pilot hole, and inserts the screw to provide clamping force. The screws are provided non-sterile and are intended for single use only.

Clinical Evidence

Bench testing only. No clinical data was required or provided for this device.

Technological Characteristics

Material: Titanium alloy Grade 5 (ASTM F136). Design: Self-tapping bone screws with socket head or hexalobular connection. Dimensions: 1.1-1.5mm diameter, 7-16mm length. Standards: AAMI ST79, ISO 17664, ISO 11607-1, ISO 2832-3, ISO 14971, ISO 10993-1, ISO 10993-5, ISO 17665-1, ASTM F543. Non-sterile delivery.

Indications for Use

Indicated for fixation of bone segments, such as horizontal or vertical augmentative grafts, in dental patients requiring bone stabilization.

Regulatory Classification

Identification

An intraosseous fixation screw or wire is a metal device intended to be inserted into fractured jaw bone segments to prevent their movement.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # 510(k) Traditional Meisinger Dental Bone Screw K132212 ## Section #5 # 510(k) Summary ## 1. Applicant's Name and Address Hager & Meisinger GmbH Hansemannstraße 10 41468 Neuss, Germany (0049) 2131 2012-0 Phone: (0049) 2131 2012- 222 Fax: Contact Person: Wiebke Stolten Management product approval and product validation (Regulatory Affairs) ## 2. Date Prepared Date prepared: 10/22/2013 ## 3. Name of the device | Trade Name: | Meisinger Screw System TX, Screw System Professional | |----------------------|------------------------------------------------------| | Common Name: | Dental Bone Screws | | Classification Name: | Screw, Fixation, Intraosseous | | Product Code: | DZL | | Regulation No: | 872.4880 | | Class: | II | | Panel: | Dental | #### 4. Predicate Devices: | K051871 - Storz am Markt | Bone Screws | |--------------------------|---------------------| | K041887 - Synthes | Craniofacial Screws | ### 5. Device Description: The Meisinger Dental Bone Screws consist of Titanium alloy Grade 5 with 1.1mm, 1.3mm and 1.5mm diameter and with lengths between 7mm - 16mm. . {1}------------------------------------------------ # 510(k) Traditional Meisinger Dental Bone Screw K132212 # Section #5 # 510(k) Summary The bone screws have a self-fapping design and the connection between screw and screw driver is realized by a socket head or a hexalobular connection to obtain a clamping force. Dental Bone Screws are delivered non sterile. ## 6. Intended Use: The Meisinger Dental Bone Screws are intended for fixation of bone segments e.g. for horizontal or vertical augmentative grafts. Depending on indication the user choose the matching screw, prepares the pit hole and inserts the screw. The screws are intended for single use only. ## 7. Performance tests and used standards / Clinical data The following standards have been followed for the development, production, performance and safety testing of Dental Bone Screws: AAMI ST79, ISO 17664, ISO 11607-1, ISO 2832-3, ASTM F136, ISO 14971, ISO 10993-1, ISO 10993-5, ISO 17665-1, ASTM F543. A performance test has been performed with the result that the design and function are equivalent to the predicted devices. Clinical data and conclusions were not needed for this device. ## 8. Basis for substantial equivalence The Meisinger Dental Bone Screws have the same intended use and are substantially equivalent to the legally marketed predicate devices in the United States. The screws use the same biocompatible materials meeting the requirements from ASTM F136 as the predicated devices. The design is similar, at least equivalent to predicated devices: All products have comparable dimensions and design and the intended use is equivalent. The function and intended use, material, possible product have been evaluated as acceptable and equivalent to predicated devices. Based on the information provided in the summary we conclude that the Meisinger Dental Bone Screws are substantially equivalent to the legally marketed predicate devices described. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles a person embracing another person. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 February 14, 2014 Hager & Meisinger Gmbh Ms.Wiebke Stolten Regulatory Affairs Hansemannstrasse 10 Neuss, D-41468 Germany Re: K132212 > Trade/Device Name: Meisinger Screw System TX, Screw System Professional Regulation Number: 21 CFR 872.4880 Regulation Name: Bone Screw Regulatory Class: Class II Product Code: DZL Dated: November 13, 2013 Received: November19, 2013 Dear Ms. Stolten: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act {3}------------------------------------------------ #### Page 2 - Wiebke Stolten or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Image /page/3/Picture/7 description: The image shows a signature and some text. The text includes "Sincerely yours," followed by the name "Tejashri Purohit-Sheth, M.D." It also includes the title "Clinical Deputy Director" and the acronym "DAGRID." The word "FOR" is at the bottom right. Erin I. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ #### 510(k) Traditional Meisinger Dental Bone Screws #### Section #4 #### Indications for Use Statement 510(k) Number (if known): K132212 Device Name: Dental Bone Screws Indications for use: ー・・・ The Meisinger Dental Bone Screws are intended for fixation of bone segments e.g. for horizontal or vertical augmentative grafts. Depending on indication the user choose the matching screw, prepares the pit hole and inserts the screw. The screws are intended for single use only. Prescription Use X X Over-The-Counter Use (Part 21 CFR 801 Subpart D) (Part 21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) | | Mary S. Runner - PA | |--------------|---------------------| | | 2014.02.14 | | | 15:42:03 -05'00' | | Susan Ringer | | Meisinger Dental Bone Screws S10(k) Traditional Page 4.1
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...