CenterMed Patient Matched Assisted Surgical Planning (ASP) System
K201353 · Centermed, Inc. · DZJ · Jun 28, 2021 · Dental
Device Facts
Record ID
K201353
Device Name
CenterMed Patient Matched Assisted Surgical Planning (ASP) System
Applicant
Centermed, Inc.
Product Code
DZJ · Dental
Decision Date
Jun 28, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.4120
Device Class
Class 2
Indications for Use
CenterMed Patient Matched Assisted Surgical Planning (ASP) System is intended for use as a software system and image segmentation system for the transfer of imaging information from a medical scanner such as a CT based system. The input data file is processed by the ASP system and the result is an output data file. This file may then be provided as digital models or used as input to a rapid prototyping portion of the system that produces physical outputs including anatomical models, surgical guides, surgical splints and surgical planning case reports for use in maxillofacial surgery. CenterMed Patient Matched ASP System is also intended as a pre-operative software tool for simulating/evaluating surgical treatment options.
Device Story
CenterMed ASP System is a software and additive manufacturing platform for maxillofacial surgical planning. Inputs include patient CT/CBCT DICOM images, teeth models, and digital photos. The system uses COTS software for image segmentation, manipulation, and surgical simulation to create digital models. These models are converted to STL files and manufactured via Selective Laser Sintering (SLS) using biocompatible Polyamide-12 (PA-12) to produce physical anatomical models, surgical guides, and splints. Surgical guides may incorporate 316L stainless steel sleeves. The system is operated by trained engineers in collaboration with physicians. Outputs include digital models, physical guides/splints/models, and surgical planning case reports. Physical products are provided non-sterile for end-user steam sterilization. The system assists surgeons in pre-operative planning, bone cutting, marking, and repositioning, potentially improving surgical precision and patient outcomes.
Clinical Evidence
No clinical data was provided. Substantial equivalence was demonstrated through bench testing, including mechanical testing (ISO 178, ISO 20795-2, ISO 527-2, ASTM D638), biocompatibility (ISO 10993 series), and sterilization validation (ANSI/AAMI/ISO 17665-1). Software validation was performed on COTS applications.
Technological Characteristics
System uses COTS software for image segmentation and surgical simulation. Physical outputs are manufactured via Selective Laser Sintering (SLS) using medical-grade Polyamide-12 (PA-12). Surgical guides may include 316L stainless steel sleeves. Devices are patient-specific, single-use, and provided non-sterile for steam sterilization (10^-6 SAL).
Indications for Use
Indicated for use in maxillofacial surgery as a software and image segmentation system to process medical scanner data (CT/CBCT) for the creation of patient-specific anatomical models, surgical guides, surgical splints, and surgical planning case reports, and as a pre-operative tool for simulating and evaluating surgical treatment options.
Regulatory Classification
Identification
A bone cutting instrument and accessories is a metal device intended for use in reconstructive oral surgery to drill or cut into the upper or lower jaw and may be used to prepare bone to insert a wire, pin, or screw. The device includes the manual bone drill and wire driver, powered bone drill, rotary bone cutting handpiece, and AC-powered bone saw.
Predicate Devices
KLS Martin Individual Patient Solutions (IPS) Planning System (K181241)
Related Devices
K181241 — KLS Martin Individual Patient Solutions (IPS) Planning System · KLS-Martin L.P. · Sep 13, 2018
K211614 — CenterMed Patient Matched Assisted Surgical Planning (ASP) System · Centermed, Inc. · Aug 4, 2022
K182789 — KLS Martin Individual Patient Solutions (IPS) Planning System · KLS-Martin L.P. · Mar 11, 2019
K212570 — OsteoPlan System · Osteomed, LLC · Feb 11, 2022
K223024 — MedCAD AccuPlan System · Medcad · Mar 7, 2023
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health and Human Services. To the right of the seal is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font below. The logo is simple and professional, and it is easily recognizable.
June 28, 2021
CenterMed, Inc. % Jash Bhayani Director of Design 226 N Wiget Ln Walnut Creek, California 94598
Re: K201353
Trade/Device Name: CenterMed Patient Matched Assisted Surgical Planning (ASP) System Regulation Number: 21 CFR 872.4120 Regulation Name: Bone Cutting Instrument and Accessories Regulatory Class: Class II Product Code: DZJ, LLZ Dated: May 25, 2021 Received: May 26, 2021
Dear Jash Bhayani:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K201353
Device Name
CenterMed Patient Matched Assisted Surgical Planning (ASP) System
Indications for Use (Describe)
CenterMed Patient Matched Assisted Surgical Planning (ASP) System is intended for use as a software system and image segmentation system for the transfer of imaging information from a medical scanner such as a CT based system. The input data file is processed by the ASP system and the result is an output data file. This file may then be provided as digital models or used as input to a rapid prototyping portion of the system that produces physical outputs including anatomical models, surgical splints and surgical planning case reports for use in maxillofacial surgery. CenterMed Patient Matched ASP System is also intended as a pre-operative software tool for simulating/evaluating surgical treatment options.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW. *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number. "
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "centermed" in a stylized, sans-serif font. The word is in a dark red color, and there is a horizontal line above the word. There is a registered trademark symbol to the right of the word.
| K201353 | |
|-------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <b>Submitter Information</b> | |
| <b>Submitter:</b> | CenterMed, Inc. |
| <b>Submitter Address:</b> | 226 N Wiget Ln, |
| | Walnut Creek, CA 94598 |
| <b>Contact Person:</b> | Jash Bhayani |
| <b>Contact Title:</b> | Director of Design |
| <b>Phone number:</b> | 855-840-8823 |
| <b>E-mail address:</b> | Jash.Bhayani@centermed.com |
| <b>Date Prepared:</b> | May 25, 2021 |
| | The content is prepared based on the requirements of 21<br>CFR 807.92 |
| <b>Submission Information</b> | |
| <b>Trade Name:</b> | CenterMed Patient Matched Assisted Surgical Planning<br>(ASP) System |
| <b>Common Name:</b> | System for the creation of patient specific anatomical<br>models, surgical splints, surgical guides and surgical<br>planning case reports |
| <b>Classification Name:</b> | Bone Cutting Instruments and Accessories<br>(21 CFR 872.4120); |
| | |
| <b>Regulatory Class:</b> | Class II |
| <b>Product Code:</b> | DZJ, LLZ |
| <b>Predicate Device:</b> | KLS Martin Individual Patient Solutions (IPS) Planning<br>System (K181241) |
| <b>Device Description:</b> | CenterMed Patient Matched Assisted Surgical Planning<br>(ASP) System is a combination of software design and<br>additive manufacturing for customized virtual pre-surgical<br>treatment planning in maxillofacial reconstruction and<br>orthognathic surgeries. The system processes patients'<br>imaging data files obtained from the surgeons for treatment<br>planning and outputs various patient-specific products (both<br>physical and digital), including surgical guides, anatomical<br>models, surgical splints, and surgical planning case reports.<br>The physical products (surgical guides, anatomical models<br>and surgical splints) are manufactured with |
| | |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the word "centermed" in a stylized, sans-serif font. The word is in a dark red color. There is a horizontal line above the word, and a circled "R" symbol is in the upper right corner of the image.
biocompatible polyamide (PA-12) using additive manufacturing (Selective Laser Sintering).
Indications for Use: CenterMed Patient Matched Assisted Surgical Planning (ASP) System is intended for use as a software system and image segmentation system for the transfer of imaging information from a medical scanner such as a CT based system. The input data file is processed by the ASP system and the result is an output data file. This file may then be provided as digital models or used as input to a rapid prototyping portion of the system that produces physical outputs including anatomical models, surgical guides, surgical splints and surgical planning case reports for use in maxillofacial surgery. CenterMed Patient Matched ASP System is also intended as a pre-operative software tool for simulating/evaluating surgical treatment options.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the Centermed logo in a dark red color. The logo features a stylized font with a horizontal line above the letters. A registered trademark symbol is located to the right of the word.
| 1 Abbreviations | | |
|-----------------|--------------|-------------------------------------------------------------|
| | Abbreviation | Definition |
| | ASP | Assisted Surgical Planning |
| | CBCT | Cone Beam Computed Tomography |
| | COTS | Commercially off-the-shelf |
| | CT | Computer Tomography |
| | DICOM | File Format; Digital Imaging and Communications in Medicine |
| | IPS | Individual Patient Solutions |
| | PA-12 | Polyamide 12 |
| | SAL | Sterility Assurance Level |
| | SLA | Stereolithography |
| | SLS | Selective Laser Sintering |
| | STL | File format used in 3D printing |
# 2 General Workflow
The general workflow of CenterMed Patient Matched Surgical Planning System begins with software simulation of patients' imaging data files obtained from the surgeon for treatment planning. Products (surgical guides, surgical splints and anatomical models) are designed based on these imaging data files, converted to 3D STL files, and sent to manufacturing to be preprocessed, printed, and post-processed. Simultaneously, a surgical planning case report is created based on the same imaging data files as well as the STLs of the designed products. The completed output of the system includes the surgical planning case report, anatomical models, and corresponding surgical guides and/or surgical splints. These are packaged, labeled, and sent to the surgeon clean and non-sterile. The surgeon or relevant hospital staff will sterilize the surgical guides, surgical splints and anatomical models before use in the operating room. The workflow is shown in Figure 1.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for Centermed. The logo is in a dark red color and features a stylized font. There is a line above the word "Centermed" and a registered trademark symbol to the right of the word.
Image /page/6/Figure/2 description: This image is a flowchart that outlines the steps involved in a surgical treatment planning process. The process begins with input, which includes CT, CBCT, DICOM images, teeth models, and patient's digital photos. The next steps are design, pre-processing, printing, post-processing, and output, which leads to the end user sterilization.
Figure 1: General Workflow of CenterMed Patient Matched ASP System
## 3 Technological Characteristics / Substantial Equivalence Discussion
Both the subject device and predicate device have the same indications for use. The subject and predicate devices are intended for use as a software system and image segmentation system for the transfer of imaging information from a medical scanner such as a CT based system. The input data file is processed by the system and the result is an output data file that may then be provided as digital models or used as input to a rapid prototyping portion of the system that produces physical outputs, including anatomical models, surgical guides, surgical splints and surgical planning case reports for use in maxillofacial surgery. The subject device and predicate device are also both intended as a pre-operative software tool for simulating/ evaluating surgical treatment options.
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for Centermed. The logo is in a dark red color and features a stylized font. A horizontal line runs above the text. A registered trademark symbol is located in the upper right corner of the logo.
### 4 Similarities to Predicate
Both the subject device and predicate device fit the same premarket regulation and are identical in conditions of use.
Both the subject device and predicate device share the same fundamental technologies. They both use a combination of software and hardware. Commercially off-the-shelf (COTS) software systems are used for image transfer, manipulation, surgical simulation as well as digital model creation of patient specific anatomical models, surgical splints and surgical guides. These software systems are intended to be operated by well-trained engineers and the outputs are evaluated by physicians. The additive manufacturing hardware is used to manufacture the physical models from the digital models.
Both the subject device and predicate device share fundamental technologies as follows:
- Both use medical imaging data, such as CT scans, in DICOM format as input data file;
- Both have the same system outputs, including anatomical models, surgical splints and . surgical guides;
- Both generate digital models and surgical planning case reports to assist physicians in . how to use the products during clinical operation;
- Both use the same materials for surgical guides (Polyamide-12); .
- Both use additive manufacturing techniques for product manufacturing;
- Both use the same manufacturing methods for printing surgical guides (Selective Laser Sintering (SLS));
- . Both are passive medical devices;
- Both are patient-specific devices;
- Both are provided for single-use; ●
- Both are provided non-sterile with an end-user sterilization method of steam sterilization . with the validated assurance level of 10to SAL;
- . Both use temporary screws for surgical guide fixation to temporarily fix the surgical guide to the target bone surface for more precise guidance of the osteotomy/repositioning position;
- Both devices require trained employees/engineers who utilize the software applications . to manage data and work with the physician to create the virtual surgical plan;
- Both devices need the physician to provide input for surgical planning and interactive . feedback through viewing the digital models of system outputs that are modified by the trained employee/engineer during the planning session.
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the logo for Centermed. The logo is in a dark red color and features a stylized font. A horizontal line runs above the text. A registered trademark symbol is located in the upper right corner of the logo.
## 5 Differences from Predicate
- Subject device creates anatomical models using Polyamide-12 (PA-12) material with a . SLS manufacturing process, whereas the predicate device uses resin material with an SLA manufacturing process.
- . Compared to that of the predicate device, the subject device's input data also includes patients' teeth models (either physical or digital models) and patients' digital photos.
- For the predicate device, manufacturing materials include epoxy/resin and acrylic for . anatomical models, methacrylate for splints, and polyamide, Titanium Alloy (Ti-6A1-4V) and CP Titanium for guides. The subject device only uses PA-12 to manufacture all anatomical models, surgical guides, and surgical splints.
- . The predicate device provides two kinds of surgical guides, one made with polyamide and the other made with Titanium Alloy (Ti-6Al-4V) and CP Titanium. The subject device only uses surgical guides made of polyamide, along with medical grade 316L stainless steel sleeves. Surgical splints of the subject device are made with the same manufacturing process as the surgical guides, both using polyamide as the raw material.
- . The predicate device guides are used for bone cutting or marking, while those of the subject device are provided in three types: bone cutting, bone marking, and bone repositioning.
- . The predicate device uses previously cleared fixation screws which range from 1.5mm -2.7mm diameter, with length of 4mm - 22mm. The subject device recommends a range of standard screw sizes to the doctor (diameter 1.5mm - 2.1mm, length 5mm - 22mm) (Standard screw sizes determined by manufacturer catalogs of medical grade screws for maxillofacial surgeries).
- The predicate device utilizes four (4) commercially off-the-shelf (COTS) software . applications for image segmentation and processing, whereas the subject device utilizes three (3) commercially off-the-shelf (COTS) software applications for image segmentation and processing, and two (2) commercially off-the-shelf (COTS) software for manufacturing.
- The predicate device has one contraindication that is not present in the subject device. . "Bone tumors located in the implant base region". The subject device does not include implants and therefore this contraindication is not relevant to the subject device. All other contraindications are identical to the predicate device.
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image shows the word "Centermed" in a stylized, sans-serif font. The word is in a dark red color. There is a registered trademark symbol to the right of the word.
# Table 1: Subject and Predicate Device Comparison
| | Subject Device | Predicate Device |
|--------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Characteristics | CenterMed Patient Matched Assisted<br>Surgical Planning (ASP) System | KLS Martin IPS Planning System (K181241) |
| Product Code | DZJ, LLZ | DZJ, LLZ |
| Classification | 21 CFR 872.4120, Class II<br>21 CFR 892.2050, Class II | 21 CFR 872.4120, Class II<br>21 CFR 892.2050, Class II |
| Common Name | System for the creation of patient specific<br>anatomical models, surgical splints, surgical<br>guides and surgical planning case reports | System for the creation of patient specific<br>anatomical models, cutting/marking guides,<br>splints, and case reports |
| Indications for<br>use | CenterMed Patient Matched Assisted Surgical<br>Planning (ASP) System is intended for use as a<br>software system and image segmentation system<br>for the transfer of imaging information from a<br>medical scanner such as a CT based system. The<br>input data file is processed by the ASP system<br>and the result is an output data file. This file<br>may then be provided as digital models or used<br>as input to a rapid prototyping portion of the<br>system that produces physical outputs including<br>anatomical models, surgical guides, surgical<br>splints and surgical planning case reports for use<br>in maxillofacial surgery. CenterMed Patient<br>Matched ASP System is also intended as a pre-<br>operative software tool for simulating/evaluating<br>surgical treatment options. | The KLS Martin Individual Patient Solutions<br>(IPS) Planning System is intended for use as a<br>software system and image segmentation system<br>for the transfer of imaging information from a<br>medical scanner such as a CT based system. The<br>input data file is processed by the IPS Planning<br>System and the result is an output data file that<br>may then be provided as digital models or used as<br>input to a rapid prototyping portion of the system<br>that produces physical outputs including<br>anatomical models, guides, splints, and case<br>reports for use in maxillofacial surgery. The IPS<br>Planning System is also intended as a pre-<br>operative software tool for simulating/evaluating<br>surgical treatment options. |
| | | |
| Contraindications | 1. Active infections (obvious, or clinically<br>apparent).<br>2. Hypersensitivity to foreign bodies.<br>3. Circulatory problems, systematic diseases, or<br>metabolic disorders.<br>4. Insufficient or inadequate bone tissue.<br>5. Secondary diseases such as degenerative<br>processes that may have negative influences.<br>6. Interventions carried out in a non-sterile<br>environment<br>7. Regions exposed to inappropriate forces or<br>excessive weight loads<br>8. Patients unwilling to follow instructions<br>during the postoperative phase due to their<br>mental, neurological, or physical condition.<br>9. Obvious drug or alcohol abuse<br>10. Significant changes to the patient's anatomy<br>have occurred since the medical scan used for<br>planning purposes was obtained. | 1. Obvious infections.<br>2. Hypersensitivity to foreign bodies.<br>3. Circulatory problems, systemic diseases,<br>and metabolic disorders.<br>4. Insufficient or inadequate bone tissue.<br>5. Secondary diseases such as degenerative<br>processes that may negatively influence<br>the healing process.<br>6. Interventions carried out in a non-sterile<br>environment (e.g. paranasal sinuses).<br>7. Regions exposed to inappropriate forces<br>or excessive weight loads.<br>8. Patients unwilling or unable to follow<br>instructions during the postoperative<br>phase due to their mental, neurological, or<br>physical condition.<br>9. Bone tumors located in the implant base<br>region.<br>10. Obvious drug or alcohol abuse.<br>11. Significant changes to the patient's<br>anatomy have occurred since the medical<br>scan used for planning purposes was<br>obtained. |
| Clinical<br>Application | Maxillofacial surgeries | Maxillofacial surgeries |
| Prescription Use | Yes, intended to be used by physicians, not for<br>use by patients. | Yes, intended to be used by physicians, not for use<br>by patients. |
| Energy<br>used/delivered | Passive | Passive |
| 10(k) Summary | | |
| System Inputs | CT, CBCT DICOM images,<br>Teeth models (stone models, STL file or CBCT<br>of mouth/teeth) if applicable, Patients' digital<br>photos if applicable | CT, CBCT DICOM images |
| System Outputs | Anatomical models, Surgical splints, Surgical<br>guides, Surgical planning case reports | Anatomical models, Splints,<br>Guides, Case reports |
| Materials | Anatomical models: Medical Grade Polyamide<br>(PA-12)<br>Surgical splints: Medical Grade Polyamide (PA-<br>12)<br>Surgical guides:<br>3D Printed Part: Medical Grade Polyamide (PA-<br>12)<br>Sleeve: Medical Grade Stainless Steel 316L | Anatomical models: Epoxy/Resin, Acrylic<br>Splints: Methacrylate<br>Cutting/Marking guides: Polyamide, Titanium<br>Alloy (Ti-6Al-4V), CP Titanium |
| Manufacturing<br>Method | Medical Grade Polyamide (PA-12):<br>3D (Additive; Selective Laser Sintering (SLS))<br>Medical Grade Stainless Steel 316L:<br>Traditional (Subtractive) | Epoxy/Resin, Acrylic:<br>3D (Additive; Stereolithography (SLA))<br>CP Titanium: Traditional (Subtractive)<br>Ti-6Al-4V: 3D (Additive; Selective Laser<br>Melting)<br>Polyamide: 3D (Additive; Selective Laser<br>Sintering) |
| Patient-specific<br>configuration? | Yes | Yes |
| Provided for<br>single use? | Yes | Yes |
| Provide sterile? | No | No |
| Sterilization<br>Method | Steam Sterilization | Steam Sterilization |
| 10(k) Summary | | |
| Recommended<br>Temporary Screw<br>Diameter | 1.5mm - 2.1mm | 1.5mm - 2.7mm |
| Recommended<br>Temporary Screw<br>Length | 5mm - 22 mm | 4mm -22 mm |
| Recommended<br>Temporary Screw<br>Style | Drill-Free, Tapping-Free | MaxDrive & CrossDrive<br>(Drill-Free, locking [ThreadLock Taper Screw - TLTS]) |
{10}------------------------------------------------
Image /page/10/Picture/1 description: The image contains the word "Centermed" in a stylized, red font. A horizontal line extends over the letters "enter" in the word. A registered trademark symbol is located to the right of the word.
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image contains the word "Centermed" in a stylized, red font. The letters are connected and have a modern, slightly rounded appearance. A registered trademark symbol is located to the right of the word.
{12}------------------------------------------------
Image /page/12/Picture/1 description: The image shows the word "centermed" in a stylized, sans-serif font. The word is in a dark red color. There is a horizontal line above the word, and a registered trademark symbol to the right of the word.
{13}------------------------------------------------
Image /page/13/Picture/0 description: The image shows the logo for Centermed. The logo is in a dark red color and features a stylized, sans-serif font. A horizontal line extends over the top of the letters 'enter' in the word. A registered trademark symbol is located in the upper right corner of the logo.
# 6 Device Types and Functions
Table 2 lays out the types, materials, and functions of the various devices offered in the CenterMed Patient Matched ASP System.
| Category | Material | Function |
|---------------------|------------------------|----------------------------------------------------------------------|
| Splint | • PA-12 | • Maintaining occlusion |
| | | • Positioning maxillofacial bone to intermediate or final occlusions |
| Marking Guide | • PA-12 | • Marking of maxillofacial bone |
| | • 316L Stainless Steel | |
| Cutting Guide | • PA-12 | • Cutting/drilling maxillofacial bone |
| | • 316L Stainless Steel | |
| Repositioning Guide | • PA-12 | • Positioning maxillofacial bone |
| | • 316L Stainless Steel | |
Table 2: Types and Functions of CenterMed Patient Matched Devices
# 7 Non-clinical Performance Data
The following non-clinical performance testing was performed as supportive evidence to demonstrate substantial equivalence:
- Mechanical testing
- Biocompatibility testing .
- Sterilization testing .
- . Software validation
Table 3 shows an overview of the testing performed on surgical guides, surgical splints and validation testing performed on the COTS software.
| | | | Table 3: Summarv of Non-Clinical Performance Data | |
|--|--|--|---------------------------------------------------|--|
| | | | | |
| Test performed | Test<br>description/Guidelines | Conclusion | Safety and Efficacy<br>Confirmed |
|------------------------------|----------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------…
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.