DSM Biomedical Dental Bone Graft Plus

K193212 · Dsm Biomedical · NPM · Sep 10, 2020 · Dental

Device Facts

Record IDK193212
Device NameDSM Biomedical Dental Bone Graft Plus
ApplicantDsm Biomedical
Product CodeNPM · Dental
Decision DateSep 10, 2020
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.3930
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The DSM Biomedical Dental Bone Graft Plus is indicated for: - Augmentation or reconstructive treatment of the alveolar ridge - Filling of infrabony periodontal defects - Filling of defects after root resection, apicoectomy, and cystectomy - Filling of extraction sockets to enhance preservation of the alveolar ridge - Elevation of the maxillary sinus floor - Filling of periodontal defects in conjunction with products intended for Guided Tissue Regeneration (GTR) and Guided Bone Regeneration (GBR) - Filling of peri-implant defects in conjunction with products intended for Guided Bone Regeneration (GBR)

Device Story

Non-pyrogenic porous bone mineral and collagen matrix; composed of anorganic porcine bone granules and bovine collagen formed into small cylinders. Used in periodontal, oral, and maxillofacial surgery to facilitate bone formation. Supplied sterile via gamma irradiation for single-use. Clinician implants device into bone defects; acts as osteoconductive scaffold to support natural bone regeneration. Benefits patient by enhancing bone volume and structural integrity in oral sites.

Clinical Evidence

Bench testing included material characterization, stability, and packaging validation (ISO 11607-1). Biocompatibility testing per ISO 10993-1 (cytotoxicity, sensitization, irritation, systemic toxicity, genotoxicity, implantation) and pyrogenicity (USP <85>) confirmed safety. Animal testing in a canine intrabony defect model demonstrated bone healing performance equivalent to the predicate device.

Technological Characteristics

Osteoconductive hydroxyapatite bone mineral (porcine-derived) with bovine collagen. Form: cylinder. Dimensions: 100mg, 250mg, 500mg. Porosity: 73.42-77.26%. Sterilization: Gamma irradiation (ISO 11137-1/2). Biocompatibility: ISO 10993-1. Animal source compliance: ISO 22442-2/3.

Indications for Use

Indicated for human oral and periodontal surgery requiring alveolar ridge augmentation, infrabony defect filling, root resection/apicoectomy/cystectomy defect filling, extraction socket preservation, maxillary sinus floor elevation, and GTR/GBR-assisted periodontal or peri-implant defect filling.

Regulatory Classification

Identification

Bone grafting material is a material such as hydroxyapatite, tricalcium phosphate, polylactic and polyglycolic acids, or collagen, that is intended to fill, augment, or reconstruct periodontal or bony defects of the oral and maxillofacial region.

Special Controls

*Classification.* (1) Class II (special controls) for bone grafting materials that do not contain a drug that is a therapeutic biologic. The special control is FDA's “Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices.” (See § 872.1(e) for the availability of this guidance document.)(2) Class III (premarket approval) for bone grafting materials that contain a drug that is a therapeutic biologic. Bone grafting materials that contain a drug that is a therapeutic biologic, such as biological response modifiers, require premarket approval. (c) *Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath. September 10, 2020 DSM Biomedical Susan Pileggi Senior Regulatory Specialist 735 Pennsylvania Drive Exton, Pennsylvania 19431 Re: K193212 Trade/Device Name: DSM Biomedical Dental Bone Graft Plus Regulation Number: 21 CFR 872.3930 Regulation Name: Bone Grafting Material Regulatory Class: Class II Product Code: NPM Dated: August 17, 2020 Received: August 19, 2020 Dear Susan Pileggi: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting- combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. for Srinivas Nandkumar, Ph.D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use #### 510(k) Number (if known) K193212 Device Name DSM Biomedical Dental Bone Graft Plus Indications for Use (Describe) The DSM Biomedical Dental Bone Graft Plus is indicated for: - Augmentation or reconstructive treatment of the alveolar ridge - Filling of infrabony periodontal defects - Filling of defects after root resection, apicoectomy, and cystectomy - Filling of extraction sockets to enhance preservation of the alveolar ridge - Elevation of the maxillary sinus floor - Filling of periodontal defects in conjunction with products intended for Guided Tissue Regeneration (GTR) and Guided Bone Regeneration (GBR) - Filling of peri-implant defects in conjunction with products intended for Guided Bone Regeneration (GBR) | Type of Use (Select one or both, as applicable) | <table><tr><td><input checked="true" type="checkbox"/> Commercial Use (including CBD extraction) </td><td><input type="checkbox"/> Personal Use (including CBD extraction) </td></tr></table> | <input checked="true" type="checkbox"/> Commercial Use (including CBD extraction) | <input type="checkbox"/> Personal Use (including CBD extraction) | |-----------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|------------------------------------------------------------------| | <input checked="true" type="checkbox"/> Commercial Use (including CBD extraction) | <input type="checkbox"/> Personal Use (including CBD extraction) | | | X | Prescription Use (Part 21 CFR 801 Subpart D) | | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for DSM, a global science-based company. The logo consists of a colorful abstract graphic on the left and the letters "DSM" in bold, blue font on the right. The graphic is made up of overlapping shapes in various colors, including blue, green, yellow, and purple, creating a sense of movement and dynamism. # K193212 # 510(k) Summary # I. SUBMITTER | Submitter: | DSM Biomedical<br>735 Pennsylvania Drive<br>Exton, PA 19341 | |-----------------|-------------------------------------------------------------| | Phone: | 484-713-2100 | | Contact Person: | Susan Pileggi | | Date Prepared: | September 10, 2020 | ### II. DEVICE | Name of Device: | DSM Biomedical Dental Bone Graft Plus | |----------------------|------------------------------------------| | Common/Usual Name: | Bone Grafting Material | | Classification Name: | Bone Grafting Material (21 CFR 872.3930) | | Device Class: | II | | Device Code: | NPM | # III. PREDICATE DEVICE Substantial equivalence is claimed to the following device: - Bio-OSS® Collagen, K122894 ● The following device is referenced within the submission: - DSM Biomedical Dental Bone Graft, K170245 ● # IV. DEVICE DESCRIPTION The DSM Biomedical Dental Bone Graft Plus is a non-pyrogenic porous bone mineral and collagen matrix for use in periodontal, oral, and maxillofacial surgery. The DSM Biomedical Dental Bone Graft Plus is composed of anorganic porcine bone granules combined with bovine collagen to form small cylinders. The device is provided in sizes ranging from 100 - 500mg. DSM Dental Bone Graft Plus is supplied sterile by gamma irradiation and is for single use only. {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the DSM company logo. The logo consists of a colorful abstract design on the left and the letters "DSM" in blue on the right. The abstract design is made up of overlapping shapes in various colors, including blue, green, orange, and purple. The letters "DSM" are in a bold, sans-serif font. ### V. Indications For USE: DSM Biomedical Dental Bone Graft Plus is indicated for: - . Augmentation or reconstructive treatment of the alveolar ridge - Filling of infrabony periodontal defects . - Filling of defects after root resection, apicoectomy, and cystectomy - Filling of extraction sockets to enhance preservation of the alveolar ridge ● - Elevation of the maxillary sinus floor - . Filling of periodontal defects in conjunction with products intended for Guided Tissue Regeneration (GTR) and Guided Bone Regeneration (GBR) - Filling of peri-implant defects in conjunction with products intended for Guided Bone . Regeneration (GBR) # VII. COMPARISON OF TECHNOLOGICAL CHARACTERSTICS WITH THE PREDICATE DEVICE The DSM Biomedical Dental Bone Graft Plus is substantially equivalent in terms of indications for use, material composition, technological characteristics, and performance characteristics to the predicate device, Bio-OSS® Collagen [Geistlich-Pharma] K122894. The subject device is substantially equivalent in terms of indications for use, technological characteristics and performance characteristics to the reference device, DSM Biomedical Dental Bone Graft K170245, but differs in material composition as the subject device contains collagen in addition to the anorganic bone material of the reference device. The devices are compared in the Table 1. {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the logo for DSM, a global science-based company. The logo consists of a colorful abstract graphic on the left and the letters "DSM" in blue on the right. The graphic is made up of overlapping shapes in various colors, including blue, green, orange, and purple. The letters "DSM" are in a bold, sans-serif font. | | Table | | 1: | |--|-------|--|----| |--|-------|--|----| | | DSM Biomedical Dental<br>Bone Graft Plus | Bio-OSS® Collagen | DSM Biomedical Dental<br>Bone Graft | |---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Characteristic | (K193212: Subject Device) | (K122894: Predicate<br>Device) | (K170245: Reference<br>Device) | | Indications for<br>Use | DSM Biomedical Dental Bone<br>Graft Plus is indicated for<br>augmentation or<br>reconstructive treatment of<br>the alveolar ridge; filling of<br>infrabony periodontal<br>defects; filling of defects<br>after root resection,<br>apicoectomy, and<br>cystectomy; filling of<br>extraction sockets to<br>enhance preservation of the<br>alveolar ridge; elevation of<br>the maxillary sinus floor;<br>filling of periodontal defects<br>in conjunction with products<br>intended for Guided Tissue<br>Regeneration (GTR) and<br>Guided Bone Regeneration<br>(GBR); filling of peri-implant<br>defects in conjunction with<br>products intended for<br>Guided Bone Regeneration<br>(GBR). | Augmentation or<br>reconstructive treatment of<br>the alveolar ridge; filling of<br>periodontal defects; filling<br>of defects after root<br>resection, apicoectomy, and<br>cystectomy; filling of<br>extraction sockets to<br>enhance preservation of the<br>alveolar ridge; elevation of<br>the maxillary sinus floor;<br>filling of periodontal defects<br>in conjunction with products<br>intended for Guided Tissue<br>Regeneration (GTR) and<br>Guided Bone Regeneration<br>(GBR); filling of peri-implant<br>defects in conjunction with<br>products intended for<br>Guided Bone Regeneration<br>(GBR). | The DSM Biomedical Dental<br>Bone Graft is indicated for<br>use in augmentation or<br>reconstructive treatment of<br>the alveolar ridge; filling of<br>infrabony periodontal<br>defects; filling of defects<br>after root resection,<br>apicoectomy, and<br>cystectomy; filling of<br>extraction sockets to<br>enhance preservation of the<br>alveolar ridge; elevation of<br>the maxillary sinus floor;<br>filling of periodontal defects<br>in conjunction with products<br>intended for Guided Tissue<br>Regeneration (GTR) and<br>Guided Bone Regeneration<br>(GBR); filling of peri-implant<br>defects in conjunction with<br>products intended for<br>Guided Bone Regeneration<br>(GBR). | | Target<br>Population | Human, oral, periodontal | Human, oral, periodontal | Human, oral, periodontal | | Material<br>Composition | Porcine-derived<br>osteoconductive<br>hydroxyapatite bone<br>mineral with bovine<br>collagen. | Bovine-derived<br>osteoconductive<br>hydroxyapatite bone<br>mineral with porcine<br>collagen | Porcine-derived<br>osteoconductive<br>hydroxyapatite bone<br>mineral | | Form | Cylinder | Block | Granules | | Dimensions | 100mg, 250mg, and 500mg | 100mg, 250mg, and 500mg | 0.5cc, 1.0cc, 2.5cc | | Cumulative Pore<br>Volume | 1.6272 - 1.8329 mL/g | 0.9524 - 1.1751 mL/g | 1.3494mL/g | | Characteristic | DSM Biomedical Dental<br>Bone Graft Plus<br>(K193212: Subject Device) | Bio-OSS® Collagen<br>(K122894: Predicate<br>Device) | DSM Biomedical Dental<br>Bone Graft<br>(K170245: Reference<br>Device) | | Pore Size<br>Distribution | 0.003 - 360.86μm | 0.003 - 346.995μm | 0.003 - 645μm | | Pore Size Range | Primarily macropores | Primarily macropores | Primarily macropores | | Porosity | 73.42 - 77.26% | 53.47 - 61.34% | 65.97 - 68.74% | | Package | Double sterile barrier<br>consisting of an inner and<br>outer tray | Double sterile barrier<br>consisting of an inner and<br>outer tray | Granules supplied in a<br>single use sterilized vial. | | Performance | Bone formation | Bone formation | Bone formation | | Biocompatible | Yes (ISO 10993-1) | Yes | Yes (ISO 10993-1) | | Reusable | Single Use Device | Single Use Device | Single Use Device | | Sterility | Sterile, SAL 10-6<br>Gamma Irradiation<br>(ISO 11137-1) | Sterile, SAL 10-6<br>Gamma Irradiation | Sterile, SAL 10-6<br>Gamma Irradiation<br>(ISO 11137-1) | | Pyrogencity | Non-pyrogenic<br>(USP <85> and ISO 10993-<br>11) | Non-pyrogenic | Non-pyrogenic<br>(USP <85> and ISO 10993-<br>11) | {6}------------------------------------------------ Image /page/6/Picture/0 description: The image contains the logo for DSM, a global science-based company. The logo consists of a colorful, abstract swirl design on the left, with the letters "DSM" in bold, blue font on the right. The swirl design incorporates various colors, including blue, green, orange, and purple, creating a dynamic and visually appealing element. DSM Biomedical Dental Bone Graft Plus and the predicate device have identical indications for use and are for use in the same target population. They are provided in the same sizes. Both are sterilized by gamma irradiation and supplied in a double tray configuration. The materials are shown to be an osteoconductive hydroxyapatite bone mineral with a collagen component to aid in bone formation. The DSM Biomedical Dental Bone Graft Plus is derived from porcine bone and bovine collagen while the predicate device is derived from bovine bone and porcine collagen. This does not raise questions of substantial equivalence, as chemical and physical characteristics have been shown to be substantially equivalent in benchtop and animal testing. ### VII. NON-CLINICAL TESTING DATA: The following performance data was provided in support of the substantial equivalence determination. {7}------------------------------------------------ Image /page/7/Picture/0 description: The image contains the logo for DSM, a global science-based company. The logo consists of a colorful abstract graphic on the left and the letters "DSM" in bold, blue font on the right. The graphic is composed of overlapping, curved shapes in shades of blue, green, orange, and purple, creating a dynamic and modern design. ### Mechanical and Physical Testing Material characterization of the DSM Biomedical Dental Bone Graft Plus was completed in accordance with the FDA Guidance Document, Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices. Bench testing was completed to verify that the device can be hydrated and trimmed prior to implantation. The packaging of DSM Dental Bone Graft Plus was designed to meet the requirements of ISO 11607-1. The product is gamma sterilized and has been validated per ISO 11137-1 and ISO 11137-2. Stability testing on the device included evaluations of appearance, mass, collagen characterization, and simulated use. The stability assessments verify that the passage of time did not affect the device performance, collagen characterization or package integrity. # Biocompatibility Testing Biocompatibility testing was completed in accordance with the requirements of ISO 10993-1 for a permanent implant device with tissue/bone contact. Biocompatibility testing included the following biological effects: - . Cytotoxicity per ISO 10993-5 - Sensitization per ISO 10993-10 ● - Irritation per ISO 10993-10 - Acute Systemic Toxicity per ISO 10993-11 ● - Genotoxicity per ISO 10993-3 ● - Subacute Systemic Toxicity per ISO 10993-11 - Chronic Systemic Toxicity per ISO 10993-6 . - . Implantation per ISO 10993-6 Pyrogenicity testing per USP <85> and Rabbit Material Mediated Pyrogenicity testing per ISO 10993-11 have been completed to verify that the device is non-pyrogenic. The porcine bone and bovine collagen are sourced in accordance with ISO 22442-2 and per the FDA Guidance Document, Medical Devices Containing Materials Derived from Animal Sources (Except for In Vitro Diagnostic Devices). Viral inactivation studies per ISO 22442-3 and a residual chemical assessment per ISO 10993-17 were conducted. Results indicate that the device's biocompatibility profile is acceptable. {8}------------------------------------------------ Image /page/8/Picture/0 description: The image shows the logo for DSM, a global science-based company. The logo consists of a colorful abstract graphic on the left and the letters "DSM" in bold blue font on the right. The graphic is composed of overlapping shapes in various colors, including blue, green, orange, and purple, creating a sense of dynamism and innovation. The overall design is modern and professional, reflecting the company's focus on science and technology. # Animal Testing A canine intrabony defect animal study was performed to evaluate bone healing following treatment with the DSM Biomedical Dental Bone Graft Plus compared to the predicate device. # VIII. CONCLUSIONS Pursuant to section 510(k), DSM Biomedical Dental Bone Graft Plus is substantially equivalent to the predicate device Bio-OSS® Collagen regarding indication for use, material, technological characteristics, including principles of operation, and performance characteristics as shown in an anatomically relevant animal model.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...