← Product Code [NPL](/submissions/DE/subpart-d%E2%80%94prosthetic-devices/NPL) · K192612

# Kerecis Gingiva Graft (K192612)

_Kerecis Limited · NPL · Nov 13, 2020 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/DE/subpart-d%E2%80%94prosthetic-devices/NPL/K192612

## Device Facts

- **Applicant:** Kerecis Limited
- **Product Code:** [NPL](/submissions/DE/subpart-d%E2%80%94prosthetic-devices/NPL.md)
- **Decision Date:** Nov 13, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.3930
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Therapeutic

## Indications for Use

Kerecis Gingiva Graft is indicated for: - Localized gingival augmentation to increase keratinized tissue (KT) around teeth or implants.

## Device Story

Kerecis Gingiva Graft is a sterile, non-crosslinked, bioresorbable collagen membrane derived from North Atlantic cod fish skin. Used in oral surgical settings by dental clinicians, the graft is placed at the site of soft tissue defects to facilitate guided tissue regeneration. The device acts as a scaffold, allowing host cellular ingrowth and gradual replacement by new host tissue. It is supplied in various rectangular sizes, is pliable, and supports fixation via sutures. By increasing keratinized tissue around teeth or implants, the device aids in maintaining long-term periodontal health. The device is integrated into the surrounding tissue over time, with the piscine collagen structure promoting tissue adherence and healing.

## Clinical Evidence

Clinical evidence includes an open-label, non-comparative study of 6 individuals. All subjects completed the study with no adverse effects or complications. Results showed an increase in keratinized tissue (KT) width at all treated sites, with stability observed over a 12-month follow-up as the graft was replaced by host tissue. Results were consistent with published data for the predicate device.

## Technological Characteristics

Material: North Atlantic cod fish skin (collagen-rich). Form: Single-layer, porous, non-crosslinked, bioresorbable sheet. Dimensions: 15x20mm, 20x30mm, 30x40mm. Sterilization: Ethylene Oxide (SAL 10^-6). Biocompatibility: ISO 10993 compliant. Shelf life: 3 years (ASTM F1980). Packaging: ISO 11607 compliant.

## Regulatory Identification

Bone grafting material is a material such as hydroxyapatite, tricalcium phosphate, polylactic and polyglycolic acids, or collagen, that is intended to fill, augment, or reconstruct periodontal or bony defects of the oral and maxillofacial region.

## Special Controls

*Classification.* (1) Class II (special controls) for bone grafting materials that do not contain a drug that is a therapeutic biologic. The special control is FDA's “Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices.” (See § 872.1(e) for the availability of this guidance document.)(2) Class III (premarket approval) for bone grafting materials that contain a drug that is a therapeutic biologic. Bone grafting materials that contain a drug that is a therapeutic biologic, such as biological response modifiers, require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

## Predicate Devices

- MUCOGRAFT® Collagen Matrix ([K102531](/device/K102531.md))

## Reference Devices

- Kerecis MariGen Wound Extra ([K190528](/device/K190528.md))
- Kerecis SecureMesh ([K153364](/device/K153364.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 13, 2020

Kerecis Limited Gudmundur Sigurjonsson CEO Eyrargata 2 400 Isafjordur ICELAND

Re: K192612

Trade/Device Name: Kerecis Gingiva Graft Regulation Number: 21 CFR 872.3930 Regulation Name: Bone Grafting Material Regulatory Class: Class II Product Code: NPL Dated: October 9, 2020 Received: October 13, 2020

Dear Gudmundur Sigurjonsson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

for

Srinivas Nandkumar, Ph.D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K192612

Device Name Kerecis Gingiva Graft

Indications for Use (Describe)

Kerecis Gingiva Graft is indicated for:

- Localized gingival augmentation to increase keratinized tissue (KT) around teeth or implants.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size: 20px;">☑</span> |
|----------------------------------------------|-----------------------------------------|
| Over-The-Counter Use (21 CFR 801 Subpart C)  | ☐                                       |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

# 1 Submitter/510(k) Holder

| Address:        | Kerecis Limited<br>Eyrargata 2<br>400 Isafjordur<br>Iceland |
|-----------------|-------------------------------------------------------------|
| Contact Person: | Gudmundur Fertram Sigurjonsson<br>President and CEO         |
| Telephone:      | 011 354 562 2601                                            |
| Date Prepared:  | November 12, 2020                                           |

## 2 Device Name

| Proprietary Name:            | Kerecis Gingiva Graft             |
|------------------------------|-----------------------------------|
| 510(k) Number:               | K192612                           |
| Classification Name:         | Barrier, animal source, intraoral |
| Classification Product Code: | NPL                               |
| Regulatory Number:           | 872.3930                          |
| Class:                       | II                                |

# 3 Predicate Devices

Primary Predicate: MUCOGRAFT® Collagen Matrix (K102531) Reference Devices: Kerecis MariGen Wound Extra (K190528), Kerecis SecureMesh (K153364).

# 4 Device Description

The subject device is fish skin medical device indicated gingiva augmentation.

The subject device is obtained from cod fish skin by a standardized controlled manufacturing process and supplied in a peel-pouch terminally sterile packaging in the following rectangular sizes:

{4}------------------------------------------------

- 15mm x 20mm ●
- 20mm x 30mm
- . 30mm x 40mm

The subject device becomes completely integrated into the surrounding tissue over time, with corresponding new host tissue deposition. The physical properties of the subject device allow cellular ingrowth for augmentation of keratinized tissue.

The subject device is biocompatible, non-crosslinked, bioresorbable, strong, pliable and supports fixation by sutures.

# 5 Intended Use

Kerecis Gingiva Graft is indicated for:

- Localized gingival augmentation to increase keratinized tissue (KT) around teeth or implants.
# 6 Technological Characteristics and Substantial Equivalence

Comparisons of the subject device with the predicate device (K102531), and the reference devices (K153364 and K 190528), demonstrate that it is substantially equivalent with regards to: intended use, materials, design, and operational principles.

See Table 6.1. Kerecis Gingiva Graft in comparison with predicate and references device

{5}------------------------------------------------

#### Table 6-1 Kerecis Gingiva Graft in comparison with predicate and references devices

|                              | Subject Device                                                                                                                                                             | Predicate Device                                                                                                                                                                                                                                             | Discussion                                                                                                                                                                                                                                                  |  |
|------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Device Name                  | Kerecis Gingiva Graft                                                                                                                                                      | Geislich MUCOGRAFT®<br>Collagen Matrix                                                                                                                                                                                                                       | No discussion<br>required                                                                                                                                                                                                                                   |  |
| 510(k)                       | Subject Device                                                                                                                                                             | K102531                                                                                                                                                                                                                                                      | Predicate device<br>and reference<br>devices are cleared.                                                                                                                                                                                                   |  |
| Regulation                   | 21 CFR 872.3930                                                                                                                                                            | 21 CFR 872.3930                                                                                                                                                                                                                                              | Same as predicate                                                                                                                                                                                                                                           |  |
| Product Code                 | NPL                                                                                                                                                                        | NPL                                                                                                                                                                                                                                                          | Same as predicate                                                                                                                                                                                                                                           |  |
| Device Classification        | Barrier, Animal Source,<br>Intraoral                                                                                                                                       | Barrier, Animal Source,<br>Intraoral                                                                                                                                                                                                                         | Same as predicate                                                                                                                                                                                                                                           |  |
| Intended Use                 | Kerecis Gingiva Graft is a<br>biocompatible, sterile<br>collagen membrane<br>intended for augmentation<br>and regeneration of soft<br>tissue in oral surgical<br>settings. | MUCOGRAFT® Collagen<br>Matrix is sterile resorbable<br>bilayer extra cellular<br>collagen membrane matrix<br>for soft tissue<br>augmentation, guided<br>tissue regeneration and<br>multiple oral tissue defect<br>regeneration in oral<br>surgical settings. | Subset of the<br>intended use of the<br>predicate device.                                                                                                                                                                                                   |  |
| Indications                  | Localized gingival<br>augmentation to increase<br>Keratinized tissue (KT)<br>around teeth and implants.                                                                    | • Covering of implants<br>placed in immediate<br>extraction sockets;<br>• Localized gingival<br>augmentation to<br>increase keratinized<br>tissue (KT) around<br>teeth and implants;                                                                         | Subset of the<br>indications of the<br>predicate device                                                                                                                                                                                                     |  |
|                              |                                                                                                                                                                            |                                                                                                                                                                                                                                                              | • Alveolar ridge<br>reconstruction for<br>prosthetic treatment<br>• Guided tissue<br>regeneration<br>procedures in<br>recession defects for<br>root coverage<br>• Guided tissue<br>regeneration<br>procedures in<br>recession defects for<br>root coverage. |  |
| Animal Origin<br>Material    | North Atlantic Cod fish: skin<br>tissue, single layer sheet                                                                                                                | Porcine skin and<br>connective tissue, double<br>layer sheet                                                                                                                                                                                                 | Different animal<br>source, same<br>anatomical tissue                                                                                                                                                                                                       |  |
| Biocompatibility             | Yes                                                                                                                                                                        | Yes                                                                                                                                                                                                                                                          | Same as predicate                                                                                                                                                                                                                                           |  |
| NON-Pyrogenic                | Yes                                                                                                                                                                        | Yes                                                                                                                                                                                                                                                          | Same as predicate                                                                                                                                                                                                                                           |  |
| Resorbable                   | Yes                                                                                                                                                                        | Yes                                                                                                                                                                                                                                                          | Same as predicate                                                                                                                                                                                                                                           |  |
| Sizes                        | 15mm x 20mm<br>20mm x 30mm<br>30mm x 40mm                                                                                                                                  | 15mm x 20mm<br>20mm x 30mm<br>30mm x 40mm                                                                                                                                                                                                                    | Same as predicate                                                                                                                                                                                                                                           |  |
| Sterilization                | Ethylene Oxide                                                                                                                                                             | Gamma Irradiation                                                                                                                                                                                                                                            | Traditional<br>Sterilization<br>Method                                                                                                                                                                                                                      |  |
| Sterility Assurance<br>Level | 10-6                                                                                                                                                                       | 10-6                                                                                                                                                                                                                                                         | Same as predicate                                                                                                                                                                                                                                           |  |
| Shelf life                   | 3 years                                                                                                                                                                    | 3 years                                                                                                                                                                                                                                                      | The shelf life of the<br>product has been<br>validated                                                                                                                                                                                                      |  |

{6}------------------------------------------------

{7}------------------------------------------------

## 7 Performance Testing - Bench

The following performance studies were conducted on representative products to verify that material properties remain unchanged and support the substantial equivalence determination:

### 7.1 Morphology Observation

The subject device and the predicate device are based on the collagen rich animal tissue, piscine and porcine, respectively. Based on H&E staining, both materials are rich in collagen and porous, therefore favoring cellular infiltration. Scanning Electron Microscope (SEM) shows equivalent preserved collagen structure of the animal origin tissues used for both devices. Cross section of both devices showed that the porous surfaces in the skin derived collagen structure of both materials allows tissue adherence and promotes tissue regeneration by favoring cellular ingrowth when applied to soft tissue defect areas.

#### 7.2. Cellular ingrowth comparison

Both materials were tested for cellular ingrowth capability by fibroblast seeding onto the materials in vitro cellular modes. Both materials showed favorable cellular infiltration of fibroblasts after 14 days which is a key component for tissue augmentation and re-epithelization of defected keratinized tissue in the oral cavity.

#### 7.3. Tensile Strength

The tensile strength of the subject device was determined to be comparable to the predicate device measured by ultimate tensile strength.

#### 7.4. Heavy Metal Analysis

A heavy metal analysis was evaluated to show that the limits of cadmium (Cd), lead (Pb), arsenic (As) and mercury (Hg) contained within the subject device were acceptable under the ICH guidelines: Q3D Elemental Impurities-Guidance for Industry.

#### 7.5. Stability of Kerecis Gingiva Graft in a simulated physiological environment

A stability test was done in a simulated physiological oral environment (artificial saliva buffer) to investigate the dissolution of both material over time and to compare the effects that the products have on the pH levels and conductivity of the buffer over 24 hours. The buffer incubated with the two products was stable over time. Incubation of the subject device to the buffer raises the pH slightly, while incubation of the predicate device decreases it slightly. The subject device is structurally more stable than the predicate device since it dissolved slower than the predicate device at neutral pH 7.

#### 7.6. Suture Pull-Out Strength

The suture pull-out strength of the subject device meets or exceeds that of the predicate with a confidence of greater than 95%. For oral surgery and gingival augmentation, the products are equivalent.

#### 7.7. Pin Pull-Out Strength

The pin pull-out strength of the subject device exceeds that of the predicate at a confidence level of greater than 95%. For oral surgery and gingival augmentation, the products are equivalent.

#### Kerecis Limited Traditional 510k for Kerecis Gingiva Graft

{8}------------------------------------------------

#### 7.8. Compression

The compressive Peak-Load, Load-at-Break, Probe Penetration-at-Break, and Energy-to-Break of the subject device meet or exceed those of the predicate device, with a confidence of greater than 95%. For oral surgery and gingival augmentation, the products are equivalent.

#### 7.9. User Evaluation of Device for Cutting and Shaping

The subject device was evaluated in comparison to the predicate by four dental clinicians for use in the oral environment using a questionnaire to assess: ease of placement, stability over the site, robustness of the device, and satisfaction with handling the device. The questionnaire results showed a favorable usability that was substantially equivalent to the predicate for cutting and shaping the device for use as a dental barrier membrane.

#### 7.10.Biocompatibility, Sterilization, Shelf-life and Animal origin.

Testing from the applicant's own predicate device (K190528 and K153364) were leveraged in support of substantial equivalence.

- . Biocompatibility per ISO 10993 series
	- . Cytotoxicity
	- Sensitization
	- . Irritation or Intracutaneous reactivity
	- Acute systemic toxicity
	- . Subacute/sub-chronic toxicity
	- Genotoxicity
	- Implantation
	- Materials-Mediated Pyrogenicity
	- Chronic Toxicity
	- Carcinogenicity
- Sterilization validation per ISO 11737-1, Ethylene Oxide residual test following ISO 10993-7
- . Endotoxin validation (<20 EU/device) of sterilization method per LAL turbidimetric kinetic method following ISO 10993-11
- Shelf life per ASTM F1980 and Q5C (R2)[ICH] using accelerated and real-time aged samples
- Packaging per ISO 11607-series, ASTM F88 and ASTM F1886
- . Animal Origin and Viral inactivation per ISO 22422 series

Based upon our assessment of the design and applicable performance data, the subject device has been determined to be substantially equivalent to the identified predicate device.

## 8 Performance Testing - Animal

Kerecis performed preclinical testing in canines to demonstrate the capacity of the device to facilitate guided tissue repair (GTR) in mucogingival defects in the oral environment. Each animal received four oral defects: two smaller soft tissue defects in the mandibular canines and two larger soft tissue+GTR defects on the maxillary defects. In both cases, the oral membranes (subject or predicate devices, respectively) were inserted between the exposed tooth and the soft tissue and secured by suturing the defect. Three animals were terminated at 30 days after surgery, three animals at 60 days after surgery

{9}------------------------------------------------

and four animals at 90 days after surgery. The study endpoints were as follows:

Primary analyses outcome variable:

- . Histomorphometric measurements (thickness and length of keratinized tissue (KT))
- Histological analysis (inflammation, healing, remodeling)

Secondary analyses outcome variable(s):

- . Thickness of keratinized tissue measured by Periodontal Probe (in vivo and postmortem).
Overall, the findings of the study support the conclusion that all animals remained in good general health throughout the duration of the study and gross pathological findings suggest an acceptable safety profile for the subject device. For all the parameters analyzed, including KT thickness, KT length, root coverage, inflammation and membrane degradation, no statistically significant difference was detected between the subject and predicate devices after 90 days of healing. In vivo measurements indicate that both the subject and predicate device were able to increase keratinized tissue around teeth for both the smaller (mandibular) and larger (maxillary) defects.

The overall conclusion is that regarding the thickness and length of keratinized tissue 90 days after the creation of a soft tissue defect, the subject device is equivalent to the predicate device.

# 9 Performance Testing - Clinical

The clinical study (and the respective publications) provided safety and clinical evidence supporting the device use for gingival augmentation. The device was tested in six individuals in an open label, noncomparative study. All the subjects completed the clinical investigation. There were no adverse effects or complications during the duration of the clinical study. Three patients were treated for unilateral deformities and the other three patients were treated for bilateral deformities. An increased in the in the width of the KT was noticed for all treated sites. This average gain has been reported as adequate to maintain long-term periodontal health. During the 12-month follow up, the device showed stability as the KT slowly replaced the piscine xenograft. In addition, the results with the subject device are similar to published results reported for the predicate device.

# 10 Conclusion

Based on the data provided within this 510(k) submission, as summarized above, it can be concluded that the subject device is substantially equivalent to the predicate device with regard to intended use, indications for use, and technological characteristics, including principles of operation and performance characteristics as shown in a series of biocompatibility, bench, and clinical testing. Thus, the subject device is substantially equivalent.

---

**Source:** [https://fda.innolitics.com/submissions/DE/subpart-d%E2%80%94prosthetic-devices/NPL/K192612](https://fda.innolitics.com/submissions/DE/subpart-d%E2%80%94prosthetic-devices/NPL/K192612)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
