← Product Code [NPL](/submissions/DE/subpart-d%E2%80%94prosthetic-devices/NPL) · K100695

# COLLADENTAL BARRIER (K100695)

_Collamatrix, Inc. · NPL · Jan 3, 2013 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/DE/subpart-d%E2%80%94prosthetic-devices/NPL/K100695

## Device Facts

- **Applicant:** Collamatrix, Inc.
- **Product Code:** [NPL](/submissions/DE/subpart-d%E2%80%94prosthetic-devices/NPL.md)
- **Decision Date:** Jan 3, 2013
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.3930
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Therapeutic

## Indications for Use

CollaDental Barrier is intended for use in oral surgical procedures including use in augmentation around implants placed in immediate extraction sockets, delayed extraction sockets; filling of bone defects after roots resection, cystectomy, removal of retained teeth; guided bone regeneration in dental implant associated bony dehiscence defects and guided tissue regeneration procedures in bony dehiscence defects around teeth.

## Device Story

CollaDental Barrier is a resorbable, nonfriable membrane derived from purified porcine type I collagen. Used in oral surgery to facilitate guided bone and tissue regeneration. Device is flexible and conforms to defect sites when moistened with saline, blood, serum, or water. Applied by dental surgeons during procedures such as implant placement, root resection, or cystectomy. Acts as a barrier to support healing in bony dehiscence defects. Single-use only; sterilized via gamma irradiation.

## Clinical Evidence

No clinical data provided. Safety supported by bench testing and biocompatibility evaluation conducted in accordance with ISO 10993 Part 1.

## Technological Characteristics

Resorbable membrane; purified porcine type I collagen; flexible form factor; gamma irradiation sterilization; single-use. Biocompatibility tested per ISO 10993-1.

## Regulatory Identification

Bone grafting material is a material such as hydroxyapatite, tricalcium phosphate, polylactic and polyglycolic acids, or collagen, that is intended to fill, augment, or reconstruct periodontal or bony defects of the oral and maxillofacial region.

## Special Controls

*Classification.* (1) Class II (special controls) for bone grafting materials that do not contain a drug that is a therapeutic biologic. The special control is FDA's “Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices.” (See § 872.1(e) for the availability of this guidance document.)(2) Class III (premarket approval) for bone grafting materials that contain a drug that is a therapeutic biologic. Bone grafting materials that contain a drug that is a therapeutic biologic, such as biological response modifiers, require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

## Predicate Devices

- BioMend Extend absorbable collagen membrane ([K992216](/device/K992216.md))
- BIO-GIDE ([K042197](/device/K042197.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K100695

## COLLAMATRIX Co. Ltd.

## 510(k) summarv

## JAN 0 3 2013

Summary information

#### 1. Date Prepared

March 3, 2010

#### 2. Submitter name and address

Collamatrix Inc.

1F, No.50-1, Keyan Road, Jhunan Science Park Miaoli County, 350, Taiwan

#### 3. Contact person

| Name: | Dennis J. N. Seah |
|-------|-------------------|
| Tel:  | + 886 2 7711 3299 |
| Fax:  | + 886 2 7711 3599 |

#### 4. Device names

Propriety name: CollaDental Barrier Common name: Collagen dental matrix Classification name: Dressing, Wound

#### ર. Device classification

| Regulatory class: | Barrier, Animal Source, Intraoral, Class II |
|-------------------|---------------------------------------------|
| Product code:     | NPL                                         |

#### Device description 6.

"

CollaDental Barrier is a nonfriable, resorbable membrane made of purified type I collagen derived from pig skin using standardized controlled manufacturing process. The collagen is obtained from veterinary certified pigs and purified to avoid its antigenicity. The manufacturing process complies with the standards for virus inactivation. The CollaDental

> Page 1 of 3 I F, No. 50-1, Keyan Road, Jhunan Science Park, Miaoli Country, 350, Taiwan Tel: +886 2 7711 3299 Fax: +886 2 7711 3599

{1}------------------------------------------------

K100695
2 of 3

# COLLAMATRIX Co. Ltd.

Barrier-has-been-tested for-purity-using standard-purity-testing procedures, sterlized-by gamma irradiation and for single use only. It is flexible and conforms to the contours of the defect site. When moistened with water, saline, serum or blood, the device is flexible and conforms to the contours of the defect site. CollaDental Barrier has not been tested on persons less than 18 years of age.

#### 7. Intended use

CollaDental Barrier is intended for use in oral surgical procedures including use in augmentation around implants placed in immediate extraction sockets, delayed extraction sockets; filling of bone defects after roots resection, cystectorny, removal of retained teeth; guided bone regeneration in dental implant associated bony dehiscence defects and guided tissue regeneration procedures in bony dehiscence defects around teeth.

#### 8. Statement of Substantial equivalence

CollaDental Barrier is a device similar to predicate devices that are previously approved by the agency. CollaDental Barrier is substantially equivalent in indications and design principles to predicate devices, BioMend Extend absorbable collagen membrane (K992216) and BIO-GIDE® (K042197), each of which has been determined by FDA to be substantially equivalent to preamendment devices. CollaDental Barrier has the following similarities to the predicate devices in terms of indication for use, technological characteristics, material use and the process for sterilization. In summary, CollaDental Barrier is substantially equivalent to the predicate devices under the 510(k) regulations.

#### 9. Biocompatibility

. '

CollaDental Barrier has been demonstrated to be safe. To support the biocompatibility of this product, safety tests were conducted in accordance with ISO 10993 Part 1 Biological Evaluation of Medical Devices.

All test results from tests conducted on CollaDental Barrier are taken together as a whole, CollaDental Barrier have been demonstrated to be a safe device in accordance with ISO 10993-1.

> IF, No. 50-1, Keyan Road, Jhunan Science Park, Miaoli Country, 350, Taiwan Tel: +886 2 7711 3299 Fax: +886 2 7711 3599

Page 2 of 3

{2}------------------------------------------------

K100695
3073

# COLLAMATRIX Co. Ltd.

### 10 - Conclusion

. ·

CollaDental Barrier is essentially equivalent in indication for use, technological characteristics and material to the commercially available predicate device, and therefore meets the requirements as defined in 21 CFR § 807.

> 1F, No. 50-1, Keyan Road, Jhunan Science Park, Miaoli Country, 350, Taiwan Page 3 of 3 Tel: +886 2 7711 3299 Fax: +886 2 7711 3599

{3}------------------------------------------------

### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the department's name around the perimeter. In the center of the seal is a stylized caduceus, a symbol often associated with healthcare. The caduceus features a staff with a snake winding around it, representing healing and medicine.

-Public Health Service -

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring MD 20993-0002

## January 3, 2013

Mr. Dennis J.N. Seah Collamatrix, Incorporated 26F No. 105, Section 2 Dunhua South Road, DA-AN Distric Taipei, China 106

Re: K100695

Trade/Device Name: CollaDental Barrier Regulation Number: 21 CFR 872.3930 Regulation Name: Bone Grafting Material Regulatory Class: II Product Code: NPL Dated: January 5, 2011 Received: December 14, 2012

Dear Mr. Seah:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

2013.01.03 Susan Runner DDS, MA 08:32:49 -05'00'

Anthony D. Watson, B.S., M.S., M.B.A. Director

Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation

Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

## Statement of indications for use

510(K) Number (if known): K100695

Device Name: CollaDental Barrier

Indications for Use:

CollaDental barrier is intended for use in oral surgical procedures including use in augmentation around implants placed in immediate extraction sockets, delayed extraction sockets; filling of bone defects after roots resection, cystectomy, removal of retained teeth; guided bone regeneration in dental implant associated bony dehiscence defects and guided tissue regeneration procedures in bony dehiscence defects around teeth.

Prescription Use ನ್ (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

## Concurrence of CDRH, Office of Device Evaluation (ODE)

2012.12.31 Susan Runner DDS, MA 10:48:45 -05'00' ·

(Division Sign-Off) (Division of Anesthesiology, General Hospital Infection Control, Dental Devices

**510(k) Number:** K100695

---

**Source:** [https://fda.innolitics.com/submissions/DE/subpart-d%E2%80%94prosthetic-devices/NPL/K100695](https://fda.innolitics.com/submissions/DE/subpart-d%E2%80%94prosthetic-devices/NPL/K100695)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
