← Product Code [LBH](/submissions/DE/subpart-d%E2%80%94prosthetic-devices/LBH) · K200614

# FiteBac Antimicrobial Cavity Cleanser (K200614)

_Largent Health, LLC · LBH · Jun 25, 2020 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/DE/subpart-d%E2%80%94prosthetic-devices/LBH/K200614

## Device Facts

- **Applicant:** Largent Health, LLC
- **Product Code:** [LBH](/submissions/DE/subpart-d%E2%80%94prosthetic-devices/LBH.md)
- **Decision Date:** Jun 25, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.3260
- **Device Class:** Class 2
- **Review Panel:** Dental

## Indications for Use

The FiteBac® Antimicrobial Cavity Cleanser with 2% K21 QAS is an antimicrobial aqueous ethanolic solution intended for cleansing and moistening/re-wetting of cavity preparations.

## Device Story

FiteBac® Antimicrobial Cavity Cleanser is a 2% K21 Quaternary Ammonium Silane-functionalized (QAS) aqueous ethanolic solution. Used by dental professionals upon completion of tooth preparation or etching, prior to sealing dentinal tubules. The solution removes debris from carious lesion preparations and penetrates exposed dentin tubules, facilitating the binding of restorative adhesives to the dentin surface. The antimicrobial properties reduce microorganisms such as Streptococcus mutans, Actinomyces naeslundii, Lactobacillus acidophilus, and Candida albicans. The device is supplied in 8 mL low-density polyethylene bottles.

## Clinical Evidence

No clinical data. Evidence consists of in vitro antimicrobial effectiveness testing against Streptococcus mutans, Actinomyces naeslundii, Lactobacillus acidophilus, and Candida albicans, as documented in referenced literature (Gou et al. 2018; Daod et al. 2020) and internal data.

## Technological Characteristics

Antimicrobial aqueous ethanolic solution containing 2% K21 Quaternary Ammonium Silane (QAS). Form factor: liquid in 8 mL low-density polyethylene (LDPE) bottles. Principle of operation: chemical cleansing and re-wetting of prepared dental surfaces.

## Regulatory Identification

Cavity varnish is a device that consists of a compound intended to coat a prepared cavity of a tooth before insertion of restorative materials. The device is intended to prevent penetration of restorative materials, such as amalgam, into the dentinal tissue.

## Special Controls

*Classification.* Class II (special controls). The device, when it is an external cleaning solution, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 872.9.

## Predicate Devices

- FiteBac® Cavity Cleanser ([K190271](/device/K190271.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and "ADMINISTRATION" in a smaller font size below.

June 25, 2020

Largent Health, LLC % Elizabeth O'keeffe Director of Regulatory Affairs Secure BioMed Evalutations 7828 Hickory Flat Highway Suite 120 Woodstock. Georgia 30188

Re: K200614

Trade/Device Name: FiteBac Antimicrobial Cavity Cleanser Regulation Number: 21 CFR 872.3260 Regulation Name: Cavity Varnish Regulatory Class: Class II Product Code: LBH Dated: March 28, 2020 Received: May 28, 2020

Dear Elizabeth O'keeffe:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Srinivas "Nandu" Nandkumar, Ph.D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

# Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement on last page.

510(k) Number (if known) K200614

Device Name

FiteBac® Antimicrobial Cavity Cleanser

Indications for Use (Describe)

The FiteBac® Antimicrobial Cavity Cleanser with 2% K21 QAS is an antimicrobial aqueous ethanolic solution intended for cleansing and moistening/re-wetting of cavity preparations.

Type of Use (Select one or both, as applicable) 図 Prescription Use (Part 21 CFR 801 Subpart D)

□ Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of

information unless it displays a currently valid OMB number."

FORM FDA 3881 (7/17)

PSC Publishing Services(301)443-6740 EF

{3}------------------------------------------------

#### K200614

#### 6 510(k) Summary

In accordance with 21 CFR 807.87 (h) and 21 CRF 807.92, the 510(k) summary for the FiteBac® Antimicrobial Cavity Cleanser is provided below.

| Date                 | March 6, 2020                                                                                                                                                             |
|----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitted by         | Largent Health, LLC<br>3698 Largent Way NW, Suite 101<br>Marietta, GA 30064<br>Phone: 770-218-6221                                                                        |
| 510(k) Contact       | Secure BioMed Evaluations<br>Elizabeth O'Keeffe, Ph.D.<br>7828 Hickory Flat Highway Suite 120<br>Woodstock, GA 30188<br>770-837-2681 (direct)<br>Regulatory@SecureBME.com |
| Trade Name           | FiteBac® Antimicrobial Cavity Cleanser                                                                                                                                    |
| Common Name          | Varnish, Cavity                                                                                                                                                           |
| Code -Classification | LBH: Class II                                                                                                                                                             |
| Predicate Device     | K190271 FiteBac® Cavity Cleanser                                                                                                                                          |

#### Device Description

The FiteBac® Antimicrobial Cavity Cleanser is an antimicrobial 2% K21 Quaternary Ammonium Silanefunctionalized (QAS) aqueous ethanolic solution intended for cleansing and moistening/re-wetting of prepared dental surfaces. It is recommended for use upon completion of tooth preparation or etching, prior to sealing dentinal tubules. FiteBac® Antimicrobial Cavity Cleanser acts on the microorganisms in the table below, and not only removes debris in carious lesion preparations but can penetrate exposed dentin tubules allowing restorative adhesives to tightly bind to the prepared dentin surface. In vitro studies, referenced below, demonstrated a reduction in the following microorganisms.

| Microorganisms Susceptible to FiteBac® Antimicrobial Cavity Cleanser |
|----------------------------------------------------------------------|
| Streptococcus mutans1                                                |
| Actinomyces naeslundii1                                              |
| Lactobacillus acidophilus2                                           |
| Candida albicans3                                                    |

<sup>1</sup> Gou YP, Li JY, Meghil MM, et al. Quatemary ammonium silane-based antibacterial and anti-proteolytic cavity cleanser. Dent Mater. 2018;34(12):1814-1827.

<sup>2</sup> Daod U. Burrow MF, Yu CKY. Effect of a novel quaternary ammonium silane cavity disinfectant on cariogenic biofilm formation. Clin Oral Investig. 2020;24(2):649-661.

<sup>3</sup> Data on File

{4}------------------------------------------------

#### Indications for Use

The FiteBac® Antimicrobial Cavity Cleanser with 2% K21 QAS is an antimicrobial aqueous ethanolic solution intended for cleansing and moistening/re-wetting of cavity preparations.

## Technological Characteristics

As the subject device is identical in formulation and manufacture, the subject device has substantially equivalent technological characteristics to the predicate device in terms of principles of operation, intended use, material performance, and biocompatibility.

## Non-clinical Testing – Bench Study Comparison

As the subject device has the same formulation and manufacturing as the predicate device, the subject device has mechanical and physical properties identical to the predicate device. No additional biocompatibility testing was conducted for this submission.

In the referenced literature, the subject device underwent antimicrobial effectiveness testing against Streptococcus mutans, Actinomyces naeslundii, Lactobacillus, and Candida albicans. Briefly, dentin blocks impregnated with each microorganism were treated with the subject device to determine the effectiveness of the subject device at reducing the microorganism from the dentin, in the reduction of each microorganism.

#### Substantial Equivalence Summary (Conclusion)

FiteBac® Antimicrobial Cavity Cleanser subject device has equivalent intended use, principles of operation and the exact same technological characteristics as the predicate device FiteBac® Cavity Cleanser (K190271). Although the subject device includes and expanded antimicrobial claim, the subject device and the predicate device are the same formulation with the same manufacturing process. FiteBac® Antimicrobial Cavity Cleanser subject device is as safe and effective as the predicate device currently cleared for marketing in the United States and does not raise additional questions of safety and effectiveness.

| Trait                     | FiteBac® Antimicrobial Cavity Cleanser™<br>(Subject Device) | FiteBac® Cavity Cleanser™<br>(Predicate Device) | Comparison/<br>Equivalent to |
|---------------------------|-------------------------------------------------------------|-------------------------------------------------|------------------------------|
| 510(k) number             | TBD                                                         | K190271                                         | N/A                          |
| FDA Regulation            | 872.3260                                                    | 872.3260                                        | Equivalent                   |
| Product Code              | LBH                                                         | LBH                                             | Equivalent                   |
| Product<br>Classification | Class II                                                    | Class II                                        | Equivalent                   |
| Use                       | Prescription Use<br>Part 21 CFR 801 Subpart D               | Prescription Use<br>Part 21 CFR 801 Subpart D   | Equivalent                   |

A comparison of the subject device to the predicate device is shown in the following table.

{5}------------------------------------------------

| Trait                       | FiteBac® Antimicrobial Cavity Cleanser™<br>(Subject Device)                                                                                                                                      | FiteBac® Cavity Cleanser™<br>(Predicate Device)                                                                                                             | Comparison/<br>Equivalent to              |
|-----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------|
| Indications for<br>Use      | The FiteBac® Antimicrobial Cavity<br>Cleanser with 2% K21 QAS is an<br>antimicrobial aqueous ethanolic solution<br>intended for cleansing and<br>moistening/re-wetting of cavity<br>preparations | The FiteBac® Cavity Cleanser is a 2% K21<br>QAS aqueous ethanolic solution<br>intended for cleansing and<br>moistening/re-wetting of cavity<br>preparations | Equivalent plus<br>Antimicrobial<br>Claim |
| Principle of<br>Operation   | Cleansing and re-wetting of carious<br>preparations                                                                                                                                              | Cleansing and re-wetting of carious<br>preparations                                                                                                         | Equivalent                                |
| Composition                 | 2% K21 QAS                                                                                                                                                                                       | 2% K21 QAS                                                                                                                                                  | Equivalent                                |
| Available<br>Configurations | Liquid                                                                                                                                                                                           | Liquid                                                                                                                                                      | Equivalent                                |
| Size                        | 8 mL                                                                                                                                                                                             | 8 mL                                                                                                                                                        | Equivalent                                |
| Package                     | Low Density Polyethylene Bottles                                                                                                                                                                 | Low Density Polyethylene Bottles                                                                                                                            | Equivalent                                |
| Biocompatibility            | Yes                                                                                                                                                                                              | Yes                                                                                                                                                         | Equivalent                                |

---

**Source:** [https://fda.innolitics.com/submissions/DE/subpart-d%E2%80%94prosthetic-devices/LBH/K200614](https://fda.innolitics.com/submissions/DE/subpart-d%E2%80%94prosthetic-devices/LBH/K200614)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
