Bioceramic Root Canal Sealer and Repair Materials (i-MTA BP)

K242934 · Longly Biotechnology (Wuhan) Co., Ltd. · KIF · Dec 12, 2024 · Dental

Device Facts

Record IDK242934
Device NameBioceramic Root Canal Sealer and Repair Materials (i-MTA BP)
ApplicantLongly Biotechnology (Wuhan) Co., Ltd.
Product CodeKIF · Dental
Decision DateDec 12, 2024
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.3820
Device ClassClass 2
AttributesTherapeutic

Indications for Use

i-MTA BP Bioceramic Root Canal Sealer and Repair Material is developed for permanent root canal sealer and repair and surgical applications. - Repair of Root Perforation - Repair of Root Resorption - Root End Filling - Apexification - Pulp Capping

Device Story

i-MTA BP is a ready-to-use, premixed bioceramic paste for permanent root canal sealing and surgical endodontic repair. Composed of calcium silicate, zirconium oxide, tantalum pentoxide, and calcium sulfate; material is insoluble, radiopaque, and aluminum-free. Device is supplied in a preloaded syringe for direct application by dentists/endodontists in a clinical setting. Upon contact with moisture in dentin, the paste sets and hardens without shrinkage. The material acts as a physical barrier and sealing agent within the root canal system or surgical site. Benefits include biocompatibility, effective sealing of root structures, and promotion of healing in endodontic applications.

Clinical Evidence

No clinical data. Bench testing only. Performance testing per ISO 6876:2012 demonstrated: setting time 2.5h (req: 2-6h), solubility 2.80% (req: <3%), radiopacity >3mm Al, size change -0.50% (req: <1.0%), and pH 12.29 (req: 11.5-13.5). Biocompatibility testing included cytotoxicity, sensitization, irritation, acute/subchronic systemic toxicity, genotoxicity (TK gene mutation, bacterial reverse mutation), implantation, pulp capping, and endodontic usage tests, all meeting ISO 10993/7405 requirements.

Technological Characteristics

Premixed bioceramic paste in preloaded syringe. Composition: calcium silicate, zirconium oxide, tantalum pentoxide, calcium sulfate. Setting principle: moisture-activated hardening. Standards: ISO 6876:2012 (performance), ISO 10993-1, ISO 7405 (biocompatibility). No electronic components or software.

Indications for Use

Indicated for permanent root canal sealing, root perforation repair, root resorption repair, root end filling, apexification, and pulp capping in patients requiring endodontic treatment.

Regulatory Classification

Identification

A root canal filling resin is a device composed of material, such as methylmethacrylate, intended for use during endodontic therapy to fill the root canal of a tooth.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square and the words "U.S. FOOD & DRUG ADMINISTRATION". December 12, 2024 Longly Biotechnology (Wuhan) Co., Ltd. % Shirely Kong RA Beijing Xinranyicheng Medicine & Technology Co., Ltd. A-15B01, Langqin International Building, No.168, Guang'anmen Outer Street, Xicheng District Beijing, Beijing 100053 China Re: K242934 Trade/Device Name: Bioceramic Root Canal Sealer and Repair Materials (i-MTA BP) Regulation Number: 21 CFR 872.3820 Regulation Name: Root Canal Filling Resin Regulatory Class: Class II Product Code: KIF. Dated: September 14, 2024 Received: September 25, 2024 Dear Shirely Kong: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory {2}------------------------------------------------ assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Image /page/2/Picture/3 description: The image contains the name "Bobak Shirmohammadi" stacked on top of each other. Below the name is the character string "-S". To the left of the name is a faded image of the letters "FDA". For Michael E. Adjodha, MChE, RAC, CQIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Indications for Use 510(k) Number (if known) K242934 Device Name Bioceramic Root Canal Sealer and Repair Materials Indications for Use (Describe) i-MTA BP Bioceramic Root Canal Sealer and Repair Material is developed for permanent root canal sealer and repair and surgical applications. - Repair of Root Perforation - Repair of Root Resorption - Root End Filling - Apexification - Pulp Capping | Type of Use (Select one or both, as applicable) | | | |-------------------------------------------------|--|--| | | | | |X Prescription Use (Part 21 CFR 801 Subpart D) | | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ # 510(k) Summary K242934 This summary of 510(k) substantial equivalence is being submitted in accordance with requirements of 21 CFR 807.92. ## 1.0 Submitter's Information Name: Longly Biotechnology (wuhan) Co., Ltd Address: No.02 of 2nd Floor, Building A2-2, Optics Valley Biomedical Park, Gaoxin Avenue No.858, East Lake High-tech Development Zone, Wuhan, China Tel: +86 - 13907166169 Contact: Yan Liang (Registration manager) Email: longlyyanliang466@163.com Date of Preparation: Sep.8,2024 ## 2.0 Device Information Trade Name: Bioceramic Root Canal Sealer and Repair Materials Model: i-MTA BP Common Name: Root canal filling resin Classification Name: Resin, Root Canal Filling Device Panel: Dental Classification Regulation: 21 CFR 872.3820 Device Classification: Class II Product Code: KIF ### 3.0 Identification of Predicate Device and Reference Device | Information | Predicate Device | Reference Device | |---------------------------|----------------------------------------------------|-----------------------------| | 510(k) Number | K240365 | K092715 | | Device Name | Bioceramic Root Repair Material<br>(C-Root BP) | IROOT BP PLUS | | Common Name | Bioceramic Root Repair Material | Root Repair Material | | Classification Name | Resin, Root Canal Filling | Resin, Root Canal Filling | | Device Panel | Dental | Dental | | Classification Regulation | 21 CFR 872.3820 | 21 CFR 872.3820 | | Device Classification | Class II | Class II | | Product Code | KIF | KIF | | Manufacturer | Beijing C-Root Dental Medical<br>Devices Co., LTD. | Innovative BioCeramix, Inc. | ### 4.0 Device Description {5}------------------------------------------------ i-MTA BP Bioceramic Root Canal Sealer and Repair Materials is a ready-to-use premixed bioceramic paste developed for permanent root canal sealer and repair and surgical applications. i-MTA BP is an insoluble, radiopaque and aluminum-free material based on a calcium silicate composition, which requires the presence of water to set and harden. i-MTA BP does not shrink during setting and demonstrates excellent physical properties. i-MTA BP is packaged in a preloaded syringe. ### 5.0 Indication for Use i-MTA BP Bioceramic Root Canal Sealer and Repair Material is developed for permanent root canal sealer and repair and surgical applications. - Repair of Root Perforation - Repair of Root Resorption - Root End Filling - Apexification - . Pulp Capping #### 6.0 Comparison of Product Properties with the Predicate Device The subject device Bioceramic Root Canal Sealer and Repair Materials(i-MTA BP) met all design specifications as was Substantially Equivalent (SE) to the predicated device. {6}------------------------------------------------ 510(k) Summary | Comparison<br>Items | Subject Device | Predicate Device | Reference Device | Comparison | |----------------------------------------|------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------|---------------------------------| | K Number | K242934 | K240365 | K092715 | -- | | Trade name | Bioceramic Root Canal Sealer and Repair<br>Materials | Bioceramic Root Repair Material | IROOT BP PLUS | -- | | Model | i-MTA BP | C-Root BP | iRoot BP Plus | -- | | Classification<br>name | Resin, Root Canal Filling | Resin, Root Canal Filling | Resin, Root Canal Filling | Same | | Product code | KIF | KIF | KIF | Same | | Class | II | II | II | Same | | Intended<br>use/Indications<br>for Use | • Root End Filling<br>• Repair of Root Perforation<br>• Repair of Root Resorption<br>• Apexification<br>• Pulp Capping | • Repair of Root Perforation<br>• Repair of Root Resorption<br>• Root End Filling<br>• Apexification<br>• Pulp Capping | • Repair of Root Perforation<br>• Repair of Root Resorption<br>• Root End Filling<br>• Apexification<br>• Pulp Capping | Same | | Targeted Site | Root canal | Root canal | Root canal | Same | | Prescription<br>Use | Yes | Yes | Yes | Same | | Chemical<br>Composition | Calcium silicate, zirconium oxide, tantalum<br>pentoxide, calcium sulfate and filler agents. | Zirconium oxide, strontium<br>silicate, calcium phosphates, calcium<br>hydroxide, tantalum oxide and filler<br>agents. | Tricalcium silicate, dicalcium silicate,<br>zirconium oxide, tantalum pentoxide,<br>calcium sulfate and filler agents. | Analysis 1 | | Performance<br>Standard<br>Conformance | Standards or other<br>guidelines/ referential<br>used | Standards or other<br>guidelines/ referential<br>used | Standards or other<br>guidelines/ referential<br>used | Same as<br>predicated<br>device | | | ISO 6876:2012 | ISO 6876:2012 | ISO 6876:2012 | | | | ISO 6876:2012 | ISO 6876:2012 | ISO 6876:2012 | | | | ISO 6876:2012 | ISO 6876:2012 | ISO 6876:2012 | | | | Technical Specifications | Technical Requirements | Technical Requirements | | | | Setting time | Setting time | Setting time | | | | Solubility | Solubility | Solubility | | | | Radio-opacity | Radio-opacity | Radio-opacity | | | | Composition | Appearance | Appearance | | | Comparison<br>Items | Subject Device | Predicate Device | Reference Device | Comparison | | | Technical Specifications | | | | | | Technical<br>Specifications=ISO6876:2<br>001 | Technical Requirements | Technical Requirements | | | | Technical Specifications | | | | | | Technical Specifications | Technical Requirements | Technical Requirements | | | | Technical Specifications | | | | | | | Dimension<br>changes after<br>setting | Dimension<br>changes after<br>setting | | | | Content | | | | | | Size change after<br>curing | | | | | | pH | pH | pH | | | | Acid-soluble<br>arsenic and lead<br>contents | | | | | | Heavy metal | | | | | Duration of<br>Contact | Permanent contact - contact exceeds 30d | Permanent contact - contact exceeds 30d | Permanent contact - contact exceeds 30d | Same | | Biocompatibility | Comply with ISO 10993-1, ISO 7405 and<br>FDA Guidance | Comply with ISO 10993-1, ISO 7405 and<br>FDA Guidance | Comply with ISO 10993-1, ISO 7405 and<br>FDA Guidance | Same | {7}------------------------------------------------ #### 510(k) Summary {8}------------------------------------------------ # Analysis - 1. The subject device Bioceramic Root Canal Sealer and Repair Materials(i-MTA BP) has the same intended use and comparable materials as the predicated device C-Root BP (K240365). - 2. The chemical composition of the subject device Bioceramic Root Canal Sealer and Repair Materials is different with the predicated device C-Root BP (K240365). The subject device uses calcium silicates whereas the predicated device uses strontium silicate. Calcium silicates and strontium silicates are all belong to silicate. The subject device uses calcium sulfate whereas the predicated device uses calcium phosphates and calcium hydroxide. They are all belong to calcium salt. They all can be set and harden upon the moisture in the dentin. They all play the same role in each's formula. Also, Calcium silicates have better biocompatibility. The subject device Bioceramic Root Canal Sealer and Repair Materials(i-MTA BP) has the same formulation/ processing/ non-sterilization/ principal of setting/ final form after setting as the reference device iRoot BP Plus (K092715), and no other chemical composition has been added. All the raw materials of the subject device are all used in the reference device (K092715). Biocompatibility testing of the subject device and the reference device (K092715) performed meet ISO 10993 biocompatibility and FDA guidance requirements. - 3. Also, the performance comparison testing including composition/appearance, setting time, size changes after curing, solubility, radiopacity, pH of 'Bioceramic Root Canal Sealer and Repair Materials' and C-Root BP provide the evidence that chemical and physical properties are substantially equivalent. # 7.0 Summary of Testing # Performance Testing - Bench Performance testing has been carried out to demonstrate that this device meets the performance specifications for its intended use. The following tests were performed on the device. | Standards or other<br>guidelines/ referential<br>used | Performances | Requirements | Compliance<br>result | |-------------------------------------------------------|----------------------------------------------|---------------------------------------------------------------------------------------------------|--------------------------------| | ISO 6876:2012 | Setting time | 2h-6h | 2.5h | | ISO 6876:2012 | Solubility | < 3% | 2.80% | | ISO 6876:2012 | Radio-opacity | > 3mm Al | > 3mm Al | | Technical Specifications | Composition | The product composition should<br>not contain any visible foreign<br>objects to the naked eye. | Compliant with<br>requirements | | Technical Specifications | Content | Not less than 93% of the indicated<br>loading capacity. | Compliant with<br>requirements | | Technical<br>Specifications=ISO6876<br>:2001 | Size change<br>after curing | < 1.0% | -0.50% | | Technical Specifications | pH | 11.5-13.5 | 12.29 | | Technical Specifications | Acid-soluble<br>arsenic and<br>lead contents | The maximum content of acid-<br>soluble arsenic is 2mg/kg, the<br>maximum content of acid-soluble | Compliant with<br>requirements | {9}------------------------------------------------ | | | lead is 100mg/kg | | |--------------------------|-------------|------------------|------------| | Technical Specifications | Heavy metal | < 100 µg/g | < 100 µg/g | Clinical testing was not required for this submission. ## Biocompatibility Testing Biocompatibility testing was performed in accordance with the ISO 10993-1, ISO 7405 and FDA guidance document Use of International Standard ISO 10993-1, "Biological Evaluation of Medical Devices - Part 1: Evaluation and testing within a risk management process." Testing included the following: - Cytotoxicity (ISO 7405:2018) - · Sensitization (ISO 10993-10:2021) - Irritation (ISO 10993-10:2021) - · Acute Systemic Toxicity (ISO 10993-11: 2017) - · TK Gene Mutation Test (ISO 10993-3:2014) - · Bacterial Reverse Mutation Test (ISO 10993-3:2014) - · Subchronic Systemic Toxicity (ISO 10993-11: 2017) - Implantation (ISO10993-6:2016) - · Pulp Capping Testing (ISO7405:2018) - Endodontic Usage Test (ISO 7405:2018) All the results of biocompatibility testing demonstrate that the subject devices are complied with the biocompatibility requirements. ### Summary Bioceramic Root Canal Sealer and Repair Materials has the same Indications for Use and the principle of operations as the predicate device and reference device. It is intended purpose as they are placed into the root canal as a root filling material which met the requirement according to ISO 6876. It has similar physical and biocompatible properties and demonstrates comparable performance specifications to the predicate devices. # 8.0 Conclusion The conclusions drawn from the comparison and analysis above demonstrate that the differences between the subject device and the predicated device C-Root BP are insignificant in terms of substantial equivalence. The subject device is substantially equivalent to the predicate device C-Root BP.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...