K091362 · Ez-Pedo, Inc. · ELZ · Jun 29, 2009 · Dental
Device Facts
Record ID
K091362
Device Name
CERAMIC PEDATRIC PREFORMED CROWN
Applicant
Ez-Pedo, Inc.
Product Code
ELZ · Dental
Decision Date
Jun 29, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3330
Device Class
Class 1
Attributes
Therapeutic, Pediatric
Indications for Use
The EZ-Pedo Inc. Ceramic Pediatric Preformed Crown is intended to be used as a functional restoration for primary teeth, including for a badly decayed deciduous (baby) tooth, until the adult tooth erupts. It is not intended for permanent restoration. The EZ-Pedo Inc. Ceramic Pediatric Preformed Crown is indicated for use in infants, children, and adolescents.
Device Story
EPI Ceramic Pediatric Preformed Crown is a pre-manufactured, tooth-shaped zirconia cap; used as temporary restoration for decayed primary teeth. Device is precision-milled from densely sintered zirconia blocks; distributed in finished form requiring no further manipulation by dentist. Placed by dentists in clinical settings using standard dental adhesion techniques; fully encases visible portion of tooth at/above gum line. Restores tooth shape, size, and strength; improves appearance. Benefits patient by providing functional restoration until adult tooth eruption.
Clinical Evidence
No clinical data. Bench testing only, confirming material adequacy for intended use.
Technological Characteristics
Material: Biologically inert Zirconia (dental porcelain). Form factor: Preformed, tooth-shaped crown. Manufacturing: Precision milled from densely sintered zirconia blocks. Connectivity: None. Sterilization: Not specified.
Indications for Use
Indicated for infants, children, and adolescents requiring functional restoration of primary (deciduous) teeth, including badly decayed teeth, until permanent tooth eruption.
Regulatory Classification
Identification
A preformed crown is a prefabricated device made of plastic or austenitic alloys or alloys containing 75 percent or greater gold and metals of the platinum group intended to be affixed temporarily to a tooth after removal of, or breakage of, the natural crown (that portion of the tooth that normally protrudes above the gums). It is intended for use as a functional restoration until a permanent crown is constructed. The device also may be intended for use as a functional restoration for a badly decayed deciduous (baby) tooth until the adult tooth erupts.
Predicate Devices
King Dental Corporation Artificial Teeth (K931162)
NuSmile Preformed Crowns Primary Crowns (Class I, 510(k) exempt)
K252165 — ZIRCONOMER P · Shofu Dental Corporation · Oct 28, 2025
K121980 — ZIRPRIME · Kuraray Noritake Dental, Inc. · Oct 19, 2012
K101814 — DIRECT CROWN NATURAL 1 · Candulor USA, Inc. · Sep 27, 2010
Submission Summary (Full Text)
{0}------------------------------------------------
K091362
00018
40
JUN 2 9 2009
# 5. 510(K) SUMMARY
Applicant Name and Address:
EZ-Pedo Inc. 2350 Professional Drive Suite 200 Roseville, CA 95661
Establishment Registration Number: None at this time Device Name:
Classification: Classification Name: Product Code:
Ceramic Pediatric Preformed Crown 872.3330 Preformed Crown ELZ
# Date Prepared: May 8, 2009
### 510(k) Contact Person and Phone Number:
Company: Dr. Jeffery P. Fisher Chief Executive Officer, EZ-Pedo Inc. (916) 768 -7592 (Phone) (916) 774-9437 (Fax)
#### Name and Address of Manufacturing Site
Company: EZ-Pedo Inc. . 2350 Professional Drive Suite 200 Roseville, CA 95661
# Manufacturing Site Contact Person and Phone Number:
Elizabeth Weaver (916) 768 -7592
#### Predicate Devices:
The EPI Ceramic Pediatric Preformed Crown is claimed to be substantially equivalent to the following legally marketed predicate devices:
- 1. King Dental Corporation Artificial Teeth (Class I, K931162)
- NuSmile Preformed Crowns Primary Crowns (Class I, 510(k) exempt) 2.
- 3. Procera Bridge Zirconia (K041283)
- 4. C5 Medical Werks ZirDent CAD/CAM Blocks (K081253)
#### General Description:
The EPI Ceramic Pediatric Preformed Crown is a temporary tooth restoration. The dental crown is a tooth-shaped "cap" that is placed over a tooth - covering the tooth to restore its shape and size, strength, and/or to improve its appearance. The preformed crowns are manufatured in specific sizes for a proper fit, which are contained within the range of sizes commonly available for existing preformed crowns. Sizing is exclusively determined by the dentist for their particular patient and the EPI crowns are not meant to undergo further manipulation by the dentist. The EPI Ceramic Pediatric Preformed Crown is secured within the patient's mouth using standard denal adhesion techniques. The crowns, when cemented into place, fully encase the entre visible portion of a tooth that lies at and above the gum line.
The EPI Ceramic Pediatric Preformed Crown is made of Zirconia. Zirconia is a biologically inert, high-tech ceramic material (dental porcelain) intended for use in the construction of irentium oxide ceramic prosthetics. Because the EPI crowns are distributed in their final, finished form, there are no pellets, casting alloys, powders or auto mix systems that require firther manipulation by the dentist.
{1}------------------------------------------------
# Indications for Use:
The EZ-Pedo Inc. Ceramic Pediatric Preformed Crown is intended to be used as a functional restoration, including for a badly decayed deciduous (baby) tooth, until the adult tooth erupts. The EZ-Pedo Inc. Ceramic Pediatric Preformed Crown is indicated for use in infants, children and adolescents.
#### Summary of Technical Characteristics
The technical characteristics of the EPI Ceramic Pediatric Preformed Crown are substantially equivalent to those of the King Dental Corporation Artifical Teeth (K931162) and the NuSmile Preformed Primary Crowns. Both devices are preformed crowns, and the sizes of the EPI crowns are contained within the sizes available for existing preformed crowns (i.e. NuSmile Preformed Crown - Primary Crowns, Class I devices, 510(k) exempt).
The technical characteristics of the EPI Ceramic Pediatric Preformed Crown are also substantially equivalent to those of the Nobel Biocare Procera Bridge Zirconia device (K041283) and the C5 Medical Werks CAD/CAM Blocks (K081253). All three devices are dental products intended for long term placement in the mouth, specifically contacting the gums (mucpus membrenee). All three devices are made of biologically inert Zirconia material. In addition, both the EPI crown and the Nobel Biocare Procera Bridge are precision milled from a single densely sintered Zirconia block. The EPI crown is precision milled directly from the C5 CAD/CAM blocks.
# Summary of Non-Clinical Testing
Non-chincel product testing was conducted on the preformed crown, including material adequate for its intended use.
#### Conclusion
The EPI Ceramic Pediatric Preformed Crown is substantially equivalent with respect to the indications for use, technological characteristics, sizing, performance characteristics and processing to the following the legally marketed predicate devices:
- I. King Dental Corporation Artificial Teeth (K931162)
- NuSmile Preformed Crowns Primary Crowns (Class I, 510(k) exempt) 11.
- ============================================================================================================================================================================== Nobel Biocare Procera Bridge Zirconia (K041283)
- IV. C5 Medical Werks CAD/CAM Blocks (K081253)
Section 5 -2 of 2
{2}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. The eagle is black, and the text is also black. The logo is simple and recognizable.
Public Health Service
JUN 29 2009
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Dr. Jeffery Fisher CEO EZ-Pedo, Incorporated 2350 Professional Drive, Suite 200 Roseville, California 95661
Re: K091362
Trade/Device Name: Ceramic Pediatric Preformed Crown Regulation Number: 21 CFR 872.3330 Regulation Name: Preformed Crown Regulatory Class: I Product Code: ELZ Dated: May 8, 2009 Received: May 8, 2009
Dear Dr. Fisher:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
# Page 2 - Dr. Fisher
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). . please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please contact the CDRH/Office of Surveillance and Biometrics/Division of Postmarket Surveillance at 240-276-3464. For more information regarding the reporting of adverse events, please go to http://www.fda.gov/cdrh/mdr/.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Susan Keenan
Susan Runner, D.D.S., M.A. Acting Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
#### Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number:
Device Name: EZ-Pedo Inc. Ceramic Pediatric Preformed Crown
Indications for Use:
The EZ-Pedo Inc. Ceramic Pediatric Preformed Crown is intended to be used as a functional restoration for primary teeth, including for a badly decayed deciduous (baby) tooth, until the adult tooth erupts. It is not intended for permanent restoration. The EZ-Pedo Inc. Ceramic Pediatric Preformed Crown is indicated for use in infants, children, and adolescents.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-the Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Robert Betz DDS for Dr. Kevin Mulry
(Division Sign-Off)
Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K091362
Page 1 of
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.