Impression Material

K201483 · Zhengzhou Huaer Electro Optics Technology Co., Ltd. · ELW · Oct 8, 2020 · Dental

Device Facts

Record IDK201483
Device NameImpression Material
ApplicantZhengzhou Huaer Electro Optics Technology Co., Ltd.
Product CodeELW · Dental
Decision DateOct 8, 2020
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.3660
Device ClassClass 2

Indications for Use

Used for all crown, bridge, and orthodontic impression techniques.

Device Story

Impression Material is an addition-cure rubber dental impression material composed of vinyl polysiloxane and fillers. It consists of a base and a catalyst component. The device is used by dental professionals to create accurate dental impressions for crown, bridge, and orthodontic procedures. The material is hand-kneaded at a 1:1 ratio, applied to the patient's dental anatomy, and allowed to set. The resulting impression provides a negative mold used to fabricate dental restorations or appliances. The device is provided non-sterile.

Clinical Evidence

No clinical data. Bench testing only. Compliance with ISO 10993-5 (cytotoxicity), ISO 10993-10 (irritation/sensitization), ISO 10993-11 (systemic toxicity), and ISO 4823 (elastomeric impression materials) was demonstrated. Performance parameters (consistency, compatibility, dimensional change, elastic recovery, strain in compression) met all acceptance criteria for Type 0 materials.

Technological Characteristics

Addition-cure vinyl polysiloxane impression material. Components: base (vinyl polysiloxane, methylhydrogensiloxane dimethylsiloxane, dimeticone, white oil, silicon) and catalyst (vinyl polysiloxane, platinum catalyst, dimeticone, white oil, silicon). Form factor: putty. Mixing ratio: 1:1. Manual manipulation. Non-sterile. Conforms to ISO 4823 and ISO 10993 standards.

Indications for Use

Indicated for patients requiring dental impressions for crown, bridge, and orthodontic procedures.

Regulatory Classification

Identification

Impression material is a device composed of materials such as alginate or polysulfide intended to be placed on a preformed impression tray and used to reproduce the structure of a patient's teeth and gums. The device is intended to provide models for study and for production of restorative prosthetic devices, such as gold inlays and dentures.

Special Controls

*Classification.* Class II (Special Controls).

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text. October 8, 2020 Zhengzhou Huaer Electro Optics Technology Co., Ltd % Ivy Wang Technical Manager Shanghai Sungo Management Consulting Company Limited 13th Floor, 1500# Central Avenue Shanghai, Shanghai 200122 CHINA Re: K201483 Trade/Device Name: Impression Material Regulation Number: 21 CFR 872.3660 Regulation Name: Impression Material Regulatory Class: Class II Product Code: ELW Dated: September 8, 2020 Received: September 11, 2020 Dear Ivy Wang: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. for Srinivas "Nandu" Nandkumar, Ph.D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K201483 Device Name Impression Material | Indications for Use (Describe) | |--------------------------------------------------------------------| | Used for all crown, bridge, and orthodontic impression techniques. | | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--| |-------------------------------------------------|--| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## K201483 # 510(K) Summary Date of Summary prepared: 2020-10-02 ### A. Applicant: Zhengzhou Huaer Electro Optics Technology Co., Ltd Address: Floor 11, B of Building 18, the National University Science Park of Henan Province, Zhengzhou city, China, 450000 Contact Person: Li XiaoJun Tel: +86 13592524674 Mail: admin@cnhuaer.com Submission Correspondent: Primary contact: Ms. Ivy Wang Shanghai SUNGO Management Consulting Co., Ltd. Room 1309, Dongfang Building, 1500# Century Ave., Shanghai 200122, China Tel: +86-21-58817802 Email: haiyu.wang@sungoglobal.com Secondary contact: Mr. Raymond Luo Room 1309, Dongfang Building, 1500# Century Ave., Shanghai 200122, China Tel: +86-21-68828050 Email: fda.sungo@gmail.com ### B. Device: Trade Name: Impression Material Common Name: Dental Impression Material Regulatory Information Classification Name: Material, Impression Classification: Class II. Product code: ELW Regulation Number: 872.3660 Review Panel: Dental ### C. Predicate device: K152518 Vonflex S™ Putty VERICOM Co., Ltd. ### D. Intended use of the device: Used for all crown, bridge, and orthodontic impression techniques. {4}------------------------------------------------ ## K201483 ### E. Device Description: The Impression Material is a kind of addition-cure rubber impression material composed of vinyl polysiloxane and various fillers, with neutral smell and applicable to impression in dentistry. It consists of base and catalyst, in which putty base contains Vinyl polysiloxane, Methylhydrogensiloxane dimethylsiloxane, dimeticone, white oil and silicon. Catalyst mainly contains vinyl polysiloxane, platinum catalyst, dimeticone, white oil and silicon. The product is provided non-sterile. ### F. Non-Clinical Test Conclusion Non-clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards: - A ISO 10993-5: 2009 Biological Evaluation of Medical Devices -- Part 5: Tests For In Vitro Cytotoxicity - A ISO 10993-10: 2010 Biological Evaluation of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization - A ISO 10993-11 Biological Evaluation of Medical Devices - Part 11: Tests for Systemic Toxicity - A ISO 4823 Dentistry - Elastomeric impression materials ### G. Clinical Test Conclusion No clinical study is included in this submission. ### H. Comparison with predicate device Table 1 Comparison Table | Device | Proposed Device | Predicate Device | Result | |---------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------| | Manufacturer | Zhengzhou Huaer Electro<br>Optics Technology Co., Ltd | VERICOM Co., Ltd. | - | | 510K number | K201483 | K152518 | - | | Model Name | Impression Material | Vonflex S™ Putty | - | | Classification | Class II Device, ELW (21 CFR<br>872.3660) | Class II Device, ELW (21 CFR<br>872.3660) | Same | | Intend use | Used for all crown, bridge,<br>and orthodontic<br>impression techniques. | It is used for all crown, bridge<br>and orthodontic impression<br>techniques. | Same | | Standard conformed | ISO4823 | ISO4823 | Same | | Physical properties | - Classification according to<br>ISO4823: Type 0<br>- Dimensional accuracy:<br>Max.1.5%<br>- Consistency: Max. 35mm<br>- Compatibility with the die and<br>cast materials: 75 μm | - Classification according to<br>ISO4823: Type 0<br>- Dimensional accuracy:<br>Max.1.5%<br>- Consistency: Max. 35mm<br>- Compatibility with the die and<br>cast materials: 75 μm | Same | {5}------------------------------------------------ ## K201483 | | reproduction<br>- Curve of the shrinkage (Strain<br>in compression): Min.0.8 ~<br>Max.20% | reproduction<br>- Curve of the shrinkage (Strain<br>in compression): Min.0.8 ~<br>Max.20% | | |------------------------|-------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------|------| | Raw Material | Vinyl polysiloxane | Vinyl polysiloxane | Same | | Mixing Ratio | 1:1 | 1:1 | Same | | Sterility | Non-sterile | Non-sterile | Same | | Method of Manipulation | Hand-kneaded mixes | Hand-kneaded mixes | Same | | Biocompatibility | ISO 10993 | ISO 10993 | Same | ### Table 2 Performance parameter | Item | Proposed device | Acceptance Criteria (Type 0) | Result | |-----------------------------------------------------|-----------------|------------------------------|--------| | Consistency | 32 mm | < 35mm | PASS | | Compatibility with<br>the die and cast<br>materials | Complied | < 75 um reproduction | PASS | | Linear dimensional<br>change % | 0.69 | < 1.5 | PASS | | Elastic recovery | 97.73% | $\ge 96.5%$ | PASS | | Strain in<br>compression % | 4.52 | 0.8 ~ 20% | PASS | ### I. Conclusion The conclusions drawn from the nonclinical tests demonstrate that the subject device is substantial equivalence with the legally marketed predicate device, Vonflex S™ Putty under K152518.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...