K240735 · Bredent GmbH & Co. KG · EBD · Dec 12, 2024 · Dental
Device Facts
Record ID
K240735
Device Name
visio.lign color, visio.lign shield
Applicant
Bredent GmbH & Co. KG
Product Code
EBD · Dental
Decision Date
Dec 12, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3310
Device Class
Class 2
Indications for Use
visio.lign color: Characterization with colour effects on the surface of composite restorations, denture base materials and artificial denture teeth. visio.lign shield: Surface coating and wear resistance of composite restorations, denture base materials and artificial denture teeth.
Device Story
visio.lign color and visio.lign shield are light-curing, acrylate-based dental coating materials. visio.lign color provides surface characterization via 17 available shades; visio.lign shield provides surface coating and wear resistance, available in low-viscosity (LV) and high-viscosity (HV) versions. Used in dental laboratories, surgeries, and clinics by dentists and dental technicians. Applied to composite restorations, denture bases, and artificial teeth to improve aesthetics and surface properties. Benefits include enhanced color matching and increased surface durability of dental prosthetics.
Clinical Evidence
Bench testing only. Performance evaluated via bond strength, color stability, surface roughness, and viscosity testing. Biocompatibility assessed per ISO 10993-1:2009. No clinical data provided.
Technological Characteristics
Light-curing, acrylate-based resin coating. Available in multiple shades (color) and viscosities (shield). Biocompatibility per ISO 10993-1:2009. Curing mechanism identical to predicate.
Indications for Use
Indicated for characterization (color effects) or surface coating and wear resistance of composite restorations, denture base materials, and artificial denture teeth. For use by dentists, dental technicians, and trained dental staff in dental laboratories, surgeries, and clinics.
Regulatory Classification
Identification
A coating material for resin fillings is a device intended to be applied to the surface of a restorative resin dental filling to attain a smooth, glaze-like finish on the surface of the filling.
K123761 — VITA ENAMIC STAINS · Innovation Meditech GmbH · Mar 1, 2013
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a sans-serif font.
December 12, 2024
Bredent GmbH & Co. KG Christine Pakebusch Regulatory Affairs Manager Weissenhorner Str. 2 Senden, 89250 GERMANY
Re: K240735
Trade/Device Name: visio.lign color, visio.lign shield Regulation Number: 21 CFR 872.3310 Regulation Name: Coating Material For Resin Fillings Regulatory Class: Class II Product Code: EBD, EBI, ELM Dated: November 15, 2024 Received: November 15, 2024
Dear Christine Pakebusch:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# MICHAEL E. ADJODHA -S
Michael E. Adjodha, MChE, RAC, CQIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
#### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.
Submission Number (if known)
#### K240735
Device Name
visio.lign color,
visio.lign shield
Indications for Use (Describe)
visio.lign color: Characterization with colour effects on the surface of composite restorations, denture base materials and artificial denture teeth.
visio.lign shield: Surface coating and wear resistance of composite restorations, denture base materials and artificial denture teeth.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
| 510(k) #: | K240735 | 510(k) Summary | Prepared on: | 2024-11-15 |
|-----------|---------|----------------|--------------|------------|
|-----------|---------|----------------|--------------|------------|
| Contact Details | | 21 CFR 807.92(a)(1) |
|-----------------------------|-------------------------------------------|---------------------|
| Applicant Name | bredent GmbH & Co. KG | |
| Applicant Address | Weissenhorner Str. 2 Senden 89250 Germany | |
| Applicant Contact Telephone | +49 7309 872-53 | |
| Applicant Contact | Christine Pakebusch | |
| Applicant Contact Email | christine.pakebusch@bredent.com | |
| Device Name | | 21 CFR 807.92(a)(2) |
|---------------------|--------------------------------------|---------------------|
| Device Trade Name | visio.lign color ; visio.lign shield | |
| Common Name | Coating material for resin fillings | |
| Classification Name | Coating, Filling Material, Resin | |
| Regulation Number | 872.3310 | |
| Product Code(s) | EBD | |
| Legally Marketed Predicate Devices | | 21 CFR 807.92(a)(3) |
|------------------------------------|--|---------------------|
|------------------------------------|--|---------------------|
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
|-------------|----------------------------------------------------------|--------------|
| K133836 | Optiglaze Color | EBD |
| Device Description Summary | | 21 CFR 807.92(a)(4) |
|----------------------------|--|---------------------|
|----------------------------|--|---------------------|
Description
visio.lign shield & color is a light-curing, transparent or coloured, acrylate-based glossy coating.
visio.lign color achieves colour effects on the surface of composite restorations, denture base materials and artificial denture teeth.
visio.lign color is available in 17 shades.
visio.lign shield provides a surface coating on composite restorations, denture base materials and artificial denture teeth. visio.lign shield is available in the thin-bodied version visio.lign shield LV and in the higher-viscosity version visio.lign shield HV.
Indication of Use
visio.lign color: Characterization with colour effects on the surface of composite restorations, denture base materials and artificial denture teeth.
visio.lign shield: Surface coating and wear resistance of composite restorations, denture base materials and artificial denture teeth.
User profile:
The product must only be used by doctors, dentists, dental technicians and suitably trained dental staff.
Use environment:
Dental laboratories, dental surgeries, and clinics.
{5}------------------------------------------------
## Intended Use/Indications for Use
visio.lign color: Characterization with colour effects on the surface of composite restorations, denture base materials and artificial denture teeth.
visio.lign shield: Surface coating and wear restorations, denture base materials and artificial denture teeth.
### Indications for Use Comparison
### 21 CFR 807.92(a)(5)
Indication for Use
The predicate and the subject device are both intended for the use on restorations.
The predicate and the subject device achieve both characterization via colour effects on restorations.
Both products obtain surface smoothness via their surface coating.
The predicate device can be used for composite restorations, acrylic teeth. This can be compared with the indicated materials for the subject device.
## Technological Comparison
### 21 CFR 807 92(a)(6)
Composition and technological characteristics:
The subject and the predicate device share the same curing mechanism with similar composition. Biocompatibility:
The biocompatibility assessment for both devices was made in accordance to ISO 10993 1:2009.
Summary:
Based on available 510(k) information provided herein, the products of visio.lign color & shield are considered substantial equivalent to the predicate device in terms of indications for use, material, technology, design and performance specifications. There are no differences between the devices which would raise new concerns regarding safety or effectiveness.
#### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)
Performance Bench testing:
Substantial equivalence was demonstrated by the following performance tests: Bond strength, Colour stability, Surface roughness and Viscosity.
#### Not Applicable
The performance of the subject device was tested with the same kind of Bench testings as the predicate device. The design specifications of the subject device were met in the bench testings carried out. This demonstrates that the product fulfills its intended purpose and device description and shows substantial equivalence to the predicate device.
### 21 CFR 807.92(a)(5)
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.