TI2200 TRANSILLUMINATION CABLE

K071429 · Kerr Corporation · NTK · Sep 10, 2007 · Dental

Device Facts

Record IDK071429
Device NameTI2200 TRANSILLUMINATION CABLE
ApplicantKerr Corporation
Product CodeNTK · Dental
Decision DateSep 10, 2007
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.1745
Device ClassClass 2

Indications for Use

The Ti2200 Transillumination Cable is a diagnostic aid used to locate decay, calculus, fracture lines, endodontic orifices, cracks and fissures underneath the tooth surface utilizing a fiber optic cable and handle attached to a light source.

Device Story

The Ti2200 Transillumination Cable is a fiber-optic diagnostic tool used by dental professionals. It consists of a 2.1mm fiber-optic bundle with a 45-degree stainless steel end-tip, which connects to an external light source. The clinician places the end-tip against the tooth surface to illuminate the internal structure. Because decay, calculus, fractures, and other dental anomalies reduce light transmission, they appear as dark areas against the illuminated tooth structure. This visual contrast aids the clinician in identifying and diagnosing these conditions. The device is a passive optical instrument requiring no complex processing or software.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Fiber-optic transilluminator; 2.1mm fiber-optic bundle; 45-degree stainless steel end-tip; passive optical device; connects to external light source; no electronic or software components.

Indications for Use

Indicated for use as a diagnostic aid to locate decay, calculus, fracture lines, endodontic orifices, cracks, and fissures beneath the tooth surface in dental patients.

Regulatory Classification

Identification

A laser fluorescence caries detection device is a laser, a fluorescence detector housed in a dental handpiece, and a control console that performs device calibration, as well as variable tone emitting and fluorescence measurement functions. The intended use of the device is to aid in the detection of tooth decay by measuring increased laser induced fluorescence.

Special Controls

*Classification.* Class II, subject to the following special controls:(1) Sale, distribution, and use of this device are restricted to prescription use in accordance with § 801.109 of this chapter; (2) Premarket notifications must include clinical studies, or other relevant information, that demonstrates that the device aids in the detection of tooth decay by measuring increased laser induced fluorescence; and (3) The labeling must include detailed use instructions with precautions that urge users to: (i) Read and understand all directions before using the device, (ii) Store probe tips under proper conditions, (iii) Properly sterilize the emitter-detector handpick before each use, and (iv) Properly maintain and handle the instrument in the specified manner and condition.

1. restrictions on the sale, distribution, and use of the device to prescription use in accordance with 21 CFR 801.109; 2. clinical studies, or other relevant information, that demonstrates that the device aids in the detection of tooth decay by measuring increased laser induced fluorescence; and 3. labeling that includes detailed use instructions with precautions that urge users to (i) read and understand all directions before using the device, (ii) store probe tips under proper conditions, (iii) properly sterilize the emitter-detector handpiece before each use, and (iv) properly maintain and handle the instrument in the specified manner and condition.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the text "K071429" written in a handwritten style at the top. Below this, there is a stylized logo that appears to read "sds" in a bold, sans-serif font, with curved lines accentuating the letters. The logo is positioned above the words "SURGICAL DENTAL SPECIALTIES" written in smaller, uppercase letters. SYBRON DENTAL SPECIALTIES Section III - 510(k) Summary of Safety and Effectiveness Submitter: Sybron Dental Specialties, Inc. 1717 West Collins Avenue Orange, California 92867 (714) 516-7484 - Phone (714) 516-7488 - Facsimile Colleen Boswell - Contact Person Date Summary Prepared: May 2007 ## Device Name: - Trade Name Ti2200 Transillumination Cable . - Common Name Fiber optic transilluminator ● - . Classification Name - Laser fluorescence caries detection device, per 21 CFR & 872.1745 ### Devices for Which Substantial Equivalence is Claimed: - . Addent Inc., Microlux Transilluminator ### Device Description: The Ti2200 Transillumination Cable is a fiber optic transilluminator connected directly to a light source. The Ti2200 Transillumination Cable uses a 2.1mm fiber-optic bundle with a 45 degree stainless steel end-tip. The end-tip, when placed on the tooth surface (and when other sources of light are minimized), creates an illuminated tooth structure to help locate decay, calculus, fracture lines, endodontic orifices, cracks and fissures. These clinical entities reduce the ability to transmit light and therefore show up as dark areas in an otherwise bright structure thus aiding in diagnosis. ### Intended Use of the Device: The intended use of Ti2200 Transillumination Cable is to locate decay, calculus, fracture lines, endodontic orifices, cracks and fissures underneath the tooth surface. ### Substantial Equivalence: Ti2200 Transillumination Cable is substantially equivalent to other legally marketed devices in the United States. Ti2200 Transillumination Cable functions in a manner similar to and is intended for the same use as the Microlux Transilluminator designed by Addent Inc. SEP 1 0 2007 {1}------------------------------------------------ Image /page/1/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its wing and tail. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle. The logo is simple and recognizable, representing the department's role in providing health and human services to the United States. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 SEP 1 0 2007 Kerr Corporation C/O Ms. Colleen Boswell Vice President, Regulatory Affairs Sybron Dental Specialties, Incorporated 1717 West Collins Avenue Orange, California 92867 Re: K071429 Trade/Device Name: Ti2200 Transillumination Cable Regulation Number: 21 CFR 872.1745 Regulation Name: Laser Fluorescence Caries Detection Device. Regulatory Class: II Product Code: NTK Dated: August 28, 2007 Received: September 4, 2007 Dear Ms. Boswell: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting vour device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {2}------------------------------------------------ Page 2 - Ms. Boswell Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital. Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Indications for Use 510(k) Number (if known): Device Name: Ti2200 Transillumination Cable Indications for Use: The Ti2200 Transillumination Cable is a diagnostic aid used to locate decay, calculus, fracture lines, endodontic orifices, cracks and fissures underneath the tooth surface utilizing a fiber optic cable and handle attached to a light source. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) 10-1 (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) dian Karrar (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices 510(k) Number: k071429 and so on. I am unable to provide any meaningful assistance to you.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...