The Lucas Lifecare's LifeRay™ handheld is indicated for use only by a trained and qualified dentist or dental technicians for both adult and pediatric subjects for producing diagnostic dental images using various intraoral image sensors/receptors.
Device Story
LifeRay™ is a battery-operated, handheld dental X-ray system; utilizes 100 kHz high-frequency DC X-ray generator; oil-cooled and shielded tube head; powered by rechargeable 21.6 V Li-ion battery. Device includes beam-limiting cone (60 mm diameter) to focus exposure and reduce leakage; transparent circular backscatter shield mounted at cone exit for clinician safety. Operated by dentists/technicians in clinical settings; includes wrist lanyard for secure grip and inclination display for repeatability. Clinician selects tube voltage (60, 65, 70 kVp) and exposure time (10 ms to 1 s) via keypad/display. Output is diagnostic dental radiograph captured on intraoral sensor. Benefits include portability, ease of use in proximity to patient, and adjustable parameters for optimized image quality.
Clinical Evidence
Clinical performance evaluated via comparative review of radiographic images produced by LifeRay™ and the predicate device. Both systems generated adequate anatomical detail, edge definition, and grayscale for bony and soft tissue structures using intraoral sensors. No significant differences in diagnostic quality were observed.
Technological Characteristics
Handheld dental X-ray system; 100 kHz high-frequency DC generator; 21.6 V 3200 mAh Li-ion battery (IEC 62133-2 compliant). Features 60 mm diameter beam-limiting cone and Pb-filled acrylic backscatter shield. Adjustable kVp (60, 65, 70) and exposure time (0.01-1.0s). Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-2-28, IEC 60601-2-65. No network connectivity or cybersecurity risks.
Indications for Use
Indicated for use by trained dentists or dental technicians for adult and pediatric patients to produce diagnostic dental X-ray images using intraoral sensors/receptors.
Regulatory Classification
Identification
An extraoral source x-ray system is an AC-powered device that produces x-rays and is intended for dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures. The x-ray source (a tube) is located outside the mouth. This generic type of device may include patient and equipment supports and component parts.
K253864 — NOMAD Pro 3 · Dental Imaging Technologies Corporation · Mar 20, 2026
K213282 — ERI Handheld Dental X-ray System (Model AG100) · Energy Resources International Co., Ltd. Hsinchu Branch · Apr 4, 2022
K172928 — Portable X-ray System · Rolence Enterprise, Inc. · Mar 16, 2018
K200182 — EzRay Air Portable (Model: VEX-P300) · VATECH Co., Ltd. · Feb 11, 2020
Submission Summary (Full Text)
{0}------------------------------------------------
July 5, 2024
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Lucas Lifecare % Dave Kim Regulatory Affairs Consultant Mtech Group LLC 7505 Fannin St. Suite 610 HOUSTON, TX 77054
Re: K241305
Trade/Device Name: LifeRay™ Intraoral Handheld X-ray Regulation Number: 21 CFR 872.1800 Regulation Name: Extraoral Source X-Ray System Regulatory Class: Class II Product Code: EHD Dated: April 30, 2024 Received: May 9, 2024
Dear Mr. Kim:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Lu Jiang
Lu Jiang, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K241305
Device Name LifeRay™ Intraoral Handheld X-ray
### Indications for Use (Describe)
The Lucas Lifecare's Life Ray handheld is indicated for use only by a trained and qualified dentist or dental technicians for both adult and pediatric subjects for producing diagnostic dental images using various intraoral image sensors/receptors.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size:20px">☒</span> |
|----------------------------------------------|---------------------------------------|
| Over-The-Counter Use (21 CFR 801 Subpart C) | <span>☐</span> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image is a logo for Lucas Lifecare. The logo features a blue circle with a white "L" inside of it. To the right of the circle is the word "LUCAS" in blue, with the word "Lifecare" in gray underneath it. There is a light blue line under the word "LUCAS".
# 510(k) Summary (K241305)
This summary of 510(k) is being submitted in accordance with requirements of 21 CFR 872.1800
Date 510(k)submitted: 05/01/2024.
#### Submitter Information: 1.
| Sumitter's Name | : Lucas Lifecare |
|------------------------|----------------------------------------------------|
| Address | : 2302 Beloit Avenue, Janesville, Wisconsin 53546. |
| Contact Person | : Joseph Paul, CTO |
| Telephone Number | : 608-373-4301, 044 7549 996549, +91 95919-74913 |
| Email | : Joseph@lucaslifecare.com |
| Official Correspondent | : Dave Kim, MBA, MTech Group LLC |
| Address | : 7505 Fannin St. Suite 610, Houston, TX 77054 |
| Tel | : 713-467-2607 |
| Email | : davekim@mtechgroupllc.com |
#### 2. Trade/ Proprietary Name:
| Trade/ Device Name | : LifeRay™ Intraoral Handheld X-ray |
|--------------------------|-------------------------------------|
| System Manufacturer Name | : Lucas Lifecare |
| Common Name | : Portable X-ray System |
| Regulation Name | : Extraoral Source X-ray System |
| Regulation Number | : 21 CFR 872.1800 |
| Primary Product Code | : EHD |
| Classification Name | : Unit, X-ray, Extraoral with Timer |
| Regulatory Class | : II |
| 510K Review Panel | : Dental |
#### 3. Legally Marketed Predicate Device Information: (K173319)
| Trade/ Device Name | : KaVo NOMAD Pro 2 Handheld X-ray System |
|----------------------|------------------------------------------|
| Manufacturer Name | : Aribex |
| Common Name | : Portable X-ray System |
| Regulation Name | : Extraoral Source X-ray System |
| Regulation Number | : 21 CFR 872.1800 |
| Primary Product Code | : EHD |
| Classification Name | : Unit, X-ray, Extraoral with Timer |
| Regulatory Class | : II |
| 510K Review Panel | : Dental |
LUCAS LIFECARE MANUFACTURING SITE| 2302 BELOIT AVENUE, JANESVILLE, WISCONSIN 53546 EMAIL ID | info@lucaslifecare.com PHONE NO | +1 443 742 5556
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Lucas Lifecare. The logo consists of a blue circle with a white "L" inside of it, followed by the word "LUCAS" in blue. Below the word "LUCAS" is the word "Lifecare" in a smaller, gray font.
| | LUCAS LifeRay™ Intra oral<br>Handheld X-ray System | KaVo NOMAD Pro2<br>Handheld X-ray System<br>(K173319) | Remarks |
|----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|
| INDICATIONS<br>FOR USE: | The Lucas Lifecare's LifeRay™<br>handheld is indicated for use only<br>by a trained and qualified dentist or<br>dental technicians for both adult<br>and pediatric subjects for producing<br>diagnostic dental images using<br>various intraoral image<br>sensors/receptors. | The KaVo NOMAD Pro 2<br>Handheld X-ray System is<br>indicated for use only by a<br>trained and qualified dentist or<br>dental technician for both adult<br>and pediatric subjects as an<br>extraoral diagnostic dental X-<br>ray source to produce X-ray<br>images using intraoralimage<br>receptors. | Equivalent |
| MECHANICAL: | | | |
| Size: Body<br>(mm/inch) | 12.8" L x 5.0" W x 11.3" H | 11"L x 10.5"H x 5.5"W | Similar |
| Weight | 5.3 lbs. | 6.0 lbs. | Better |
| Source to skin<br>distance | 20 cm | 21 cm | Equivalent |
| Cone diameter | 6 cm | 6 cm | Same |
| User Interface | Similar, plus several user-selectable preset times. | | Equivalent |
| Backscatter<br>radiation<br>protection | 6.75" dia. Pb-filled acrylic plastic<br>scatter shield | 6.75" dia. Pb-filled acrylic plastic<br>scatter shield | Same |
| Exposure switch | Trigger located on handset | Trigger located on handset | Same |
| Tube head<br>mounting | Handheld | | Same |
| ELECTRICAL: | | | |
| Tube head | RADII™ KL11-0.4-70 Series | Cyprus PSOC CY8C29866 | Equivalent |
| Energy Source | Rechargeable 21.6 V 3200 mAh Li-<br>ion battery | Rechargeable 21.6 V DC Li-ion<br>battery core pack | Equivalent/Better |
| Capacity | 3.2 A-hr. | 1.7 A-hr. | Equivalent/Better |
| Recharge<br>capability | After 300 cycles or full drained | 70% remaining capacity after<br>300 cycles | Equivalent |
| Exposure Time | 0.01s -1.0s in 0.01s increments | 0.02s - 1.0s in 0.01s increments | Equivalent/Better |
| mA | 2.6 mA fixed | 2.5 mA fixed | Equivalent/Better |
| kVp | Adjustable kVp 60,65,70 | 60 kVp fixed | Better |
| Electrical Safety<br>Standards | IEC 60601-1:2006+A1:2013+<br>AC:2014 +A12:2014 +A2:2020 | AAMI ES60601-:2005/(R)2012<br>And A1:2012 | Same |
| EMI Standards | IEC60601-1-2 Ed. 4 | IEC60601-1-2 Ed. 4 | Same |
LUCAS LIFECARE MANUFACTURING SITE| 2302 BELOIT AVENUE, JANESVILLE, WISCONSIN 53546 EMAIL ID | info@lucaslifecare.com PHONE NO | +1 443 742 5556
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logo for Lucas Lifecare. The logo features a blue circle with a white "L" inside of it. To the right of the circle is the word "LUCAS" in blue, with the word "Lifecare" in gray underneath.
#### Description of Device: 4.
LifeRay™ is a battery-operated, handheld, dental X-ray device. LifeRay™ employs a 100 kHz highfrequency DC X-ray generator. This oil-cooled and shielded tube head is energized by a rechargeable lithium-ion battery pack. The battery is housed under to enable portability. The battery gets charged when the device is docked into its cradle, which is powered through an AC - DC power supply. To focus exposure on the specified zone of the subject, and limit radiation leaking elsewhere out of the device, a shielded beam-limiting cone is engaged to the tube head. This restricts the X-ray field to 60 mm diameter, compatible to dental image receptors. An additional transparent, circular backscatter shield is provided which shall be permanently mounted around the exit of the beam limiting cone to arrest radiation scattering back from the subject, so clinicians can perform in proximity. A wrist lanyard facilitates secure grip, and an inclination display aids in consistent repeatability of radiographs. LifeRay™ employs a fixed tube current of 2.6 mA, but has three selectable tube voltages: 60, 65, or 70 kVp. Exposure duration can be set between 10 ms to 1 s in steps of 10 ms. All these presets can be selected from the home screen of the display through a practical keypad to achieve the best radiograph results.
#### ട്. Indications for Use:
The Lucas Lifecare's LifeRay™ handheld is indicated for use only by a trained and qualified dentist or dental technicians for both adult and pediatric subjects for producing diagnostic dental images using various intraoral image sensors/receptors.
#### 6. Description of Substantial Equivalence: Technological Characteristics:
# Technological Characteristics:
LifeRay™ has similar design components and operating features as the predicate device, KaVo NOMAD Pro 2 Handheld X-ray System (K173319).
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image contains the logo for Lucas Lifecare. The logo features a blue circle with a white "L" inside of it. To the right of the circle is the word "LUCAS" in blue, with the word "Lifecare" in gray underneath.
The handheld device features a main unit (tube head), rechargeable battery and charger. main components of the tube head including X-ray tube, internal shielding and The external backscatter shielding are like the predicate device. The functionality of the user interface is also like the predicate device.
The power is supplied by a rechargeable Lithium-Ion battery core pack built into a handset. The design for Liferay is equipped with a 21.6 VDC, 3.2 Ahr battery core pack compared to the predicate device (KaVo NOMAD Pro 2 Handheld X-ray System - K173319) design of 21.6 VDC, 1.7 Ahr. The LifeRay™ battery core pack is compliant with IEC 62133-2.
LifeRay™ is not capable of network connection and it does not contain other connected capabilities. LifeRay™ is not a cyber device and not subject to cybersecurity risks and threats per Section 524B.
### Non-Clinical Test Data:
Testing was performed in accordance with the following international standards:
IEC 60601-1 Edition 3.2 2020-08 CONSOLIDATED VERSION (Recognition No19-49)
IEC 60601-1-2 Edition 4.1 2020 Medical Electrical Equipment - Part 1-2: GeneralRequirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Disturbances -Requirements And Tests
IEC 60601-4-2 Edition 1.0 2016-05 Medical electrical equipment - Part 4-2: Guidance and interpretation - Electromagnetic immunity: performance of medical equipment and medical electrical systems (Recognition No: 19-19)
IEC 60601-1-3 Edition 2.1 2013-04 Medical Electrical Equipment - Part 1-3: General Requirements for Basic Safety and Essential Performance – Collateral Standard: RadiationProtection in Diagnostic Xray Equipment (Recognition No: 12-269)
IEC 60601-2-28 Edition 3.0 2017 Medical electrical equipment - Part 2-28: Particular requirements for the basic safety and essential performance of X-ray tube assemblies for medical diagnosis (Recognition No: 12-204)
IEC 60601-2-65 Edition 1.2 2021 Medical electrical equipment - Part 2-65: Particular requirements for the basic safety and essential performance of dental intra-oral-X-ray equipment (Recognition No: 12-340)
IEC 62133-2: Edition 1.0 2017: Lithium System (Recognition No: 19-33)
IEC 62304 Edition 1.1 2015-06 Medical device software life cycle processes (Recognition No: 13-79)
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image contains the logo for Lucas Lifecare. The logo consists of a blue circle with a white "L" inside, followed by the text "LUCAS" in blue and "Lifecare" in gray. A light blue line is located under the "LUCAS" text.
# FDA Guidance Documents Utilized:
Radiation Safety Considerations for X-ray Equipment Designed for Hand-Held Use, issued.
December 24, 2008
Guidance for the Content of Premarket Submissions for Software Contained in MedicalDevices, issued June 14, 2023
Pediatric Information for X-ray Imaging Device Premarket Notifications, issued November 28,2017
# Clinical Performance Data:
Clinical radiographic images using of LifeRay™ and the predicate device have been reviewed and compared to determine the substantial equivalence. Both intraoral X-ray systems generate adequate anatomical details using an intra oral sensor. Difference in edge definition and grayscale of bony structures for both images were negligible. Both intraoral X-ray systems generate sufficient X-ray levels to obtain acceptable edge definition and grayscale of bony and soft tissue images.
# Conclusion as to Substantial Equivalence:
Based on a comparison of intended use, indications, technological characteristics, principle of operation, features and performance data, the LifeRay™ Handheld Dental X-Ray is deemed to be substantially equivalent to the predicate device.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.