← Product Code [LDD](/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices/LDD) · K962551

# CARDIOSERV S (K962551)

_Marquette Electronics, Inc. · LDD · Nov 19, 1996 · Cardiovascular · ST_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices/LDD/K962551

## Device Facts

- **Applicant:** Marquette Electronics, Inc.
- **Product Code:** [LDD](/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices/LDD.md)
- **Decision Date:** Nov 19, 1996
- **Decision:** ST
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.5300
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

CardioServ is intended to be used for the emergency resuscitation of cardiac arrest victims and clinical cardiac dysrhythmia. ◆ CardioServ is intended to be used by trained operators ◆ CardioServ is designed for external and internal defibrillation (including cardioversion) ◆ CardioServ is capable of monitoring the heart rate with adjustable alarm limits. ◆ CardioServ is capable of monitoring the saturation of O₂ with adjustable alarm limits. The intended use of CardioServ is identical to the intended use of the predicate devices.

## Device Story

Portable defibrillator with integrated ECG monitor, recorder, and pulse oximeter; used by trained operators for emergency resuscitation and cardiac dysrhythmia management. Device performs external and internal defibrillation/cardioversion; monitors heart rate and O2 saturation with adjustable alarm limits. Output provided via integrated recorder and visual/audible alarms to assist clinicians in immediate cardiac intervention. Benefits include rapid rhythm assessment and therapeutic shock delivery.

## Clinical Evidence

Bench testing only. Compliance demonstrated with ANSI/AAMI DF2-1989, ANSI/AAMI ES1-1993, IEC 606-1, IEC 601-1-1, IEC 601-2-4, and prEN865. Quality assurance included requirements reviews, code inspections, hardware/software testing, safety/environmental testing, and independent final validation.

## Technological Characteristics

Portable defibrillator/monitor/oximeter. Standards: ANSI/AAMI DF2-1989, ANSI/AAMI ES1-1993, IEC 606-1, IEC 601-1-1, IEC 601-2-4, prEN865. Features integrated recorder, ECG monitoring, and O2 saturation monitoring. Technology identical to predicate devices.

## Regulatory Identification

Low-energy DC-defibrillator: A device that delivers into a 50 ohm test load an electrical shock of a maximum of 360 joules of energy used for defibrillating (restoring normal heart rhythm) the atria or ventricles of the heart or to terminate other cardiac arrhythmias. This generic type of device includes low energy defibrillators with a maximum electrical output of less than 360 joules of energy that are used in pediatric defibrillation or in cardiac surgery. The device may either synchronize the shock with the proper phase of the electrocardiogram or may operate asynchronously. The device delivers the electrical shock through paddles placed either directly across the heart or on the surface of the body. High-energy DC-defibrillator: A device that delivers into a 50 ohm test load an electrical shock of greater than 360 joules of energy used for defibrillating the atria or ventricles of the heart or to terminate other cardiac arrhythmias. The device may either synchronize the shock with the proper phase of the electrocardiogram or may operate asynchronously. The device delivers the electrical shock through paddles placed either directly across the heart or on the surface of the body.

## Predicate Devices

- HELLIGE CardioServ SCP 910
- NELLCOR Escort 300

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

K962551

NOV 19 1996

# SECTION 2 - 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

|  Submitter: | Marquette Electronics, Inc.
8200 W. Tower Avenue
Milwaukee, WI 53223
Telephone: (414) 355-5000
FAX: (414) 362-3553  |
| --- | --- |
|  Contact Person: | Kristin Pabst  |
|  Device: | Trade Name: CardioServ S
Classification Name: DC-Defibrillator, low energy (including paddles)
Oximeter  |
|  Predicate Devices: | HELLIGE CardioServ SCP 910
NELLCOR Escort 300  |
|  Device Description: | CardioServ S is a portable defibrillator with ECG monitor,
built-in recorder, and the capability of monitoring the saturation of O₂ with
adjustable alarm limits.  |
|  Intended Use: | CardioServ is intended to be used for the emergency
resuscitation of cardiac arrest victims and clinical cardiac
dysrhythmia.

◆ CardioServ is intended to be used by trained operators
◆ CardioServ is designed for external and internal defibrillation
(including cardioversion)
◆ CardioServ is capable of monitoring the heart rate with
adjustable alarm limits.
◆ CardioServ is capable of monitoring the saturation of O₂ with
adjustable alarm limits.

The intended use of CardioServ is identical to the intended use of the
predicate devices.  |
|  Technology: | CardioServ employs the same technology as the predicate devices.  |

510(k) Notification "CardioServ S" - June 26, 1996

{1}

6

Performance:

CardioServ complies with the voluntary standards ANSI/AAMI DF2-1989, ANSI/AAMI ES1-1993, IEC 606-1, IEC 601-1-1, IEC 601-2-4, and prEN865.

The following quality assurance measures were applied to the development of CardioServ:

Requirements specification reviews, code inspections, software and hardware testing, safety testing, environmental testing, final validation testing by an independent test group, field tests.

The results of these measurements demonstrated that CardioServ is as safe, as effective, and performs as well as the predicate devices CardioServ SCP 910 and NELLCOR Escort 300 Series

510(k) Notification "CardioServ S" - June 26, 1996

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices/LDD/K962551](https://fda.innolitics.com/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices/LDD/K962551)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
