PLA · Esophageal Thermal Regulation And Gastric Suctioning Device

Cardiovascular · 21 CFR 870.5910 · Class 2

Overview

Product CodePLA
Device NameEsophageal Thermal Regulation And Gastric Suctioning Device
Regulation21 CFR 870.5910
Device ClassClass 2
Review PanelCardiovascular
3rd-Party ReviewableYes

Identification

An esophageal thermal regulation device is a prescription device used to apply a specified temperature to the endoluminal surface of the esophagus via an external controller. This device may incorporate a mechanism for gastric decompression and suctioning. The device is used to regulate patient temperature.

Classification Rationale

Class II (special controls). The special controls for this device are:

Special Controls

In combination with the general controls of the FD&C Act, the Esophageal Cooling Device is subject to the following special controls:

*Classification.* Class II (special controls). The special controls for this device are:(1) The patient contacting materials must be demonstrated to be biocompatible. (2) Non-clinical performance evaluation must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested: (i) Mechanical integrity testing. (ii) Testing to determine temperature change rate(s). (iii) Testing to demonstrate compatibility with the indicated external controller. (iv) Shelf life testing. (3) Animal testing must demonstrate that the device does not cause esophageal injury and that body temperature remains within appropriate boundaries under anticipated conditions of use. (4) Labeling must include the following: (i) Detailed insertion instructions. (ii) Warning against attaching the device to unintended connections, such as external controllers for which the device is not indicated, or pressurized air outlets instead of vacuum outlets for those devices, including gastric suction. (iii) The operating parameters, name, and model number of the indicated external controller. (iv) The intended duration of use.

Recent Cleared Devices (9 of 9)

RecordDevice NameApplicantDecision DateDecision
K250632Arctx Cool Catheter SetArctx MedicalOct 30, 2025SESE
K233357EsoCool Thermal Regulation CatheterNuvaira, Inc.Jun 27, 2024SESE
K180742EnsoETMAdvanced Cooling Therapy, Inc. D/B/A Attune MedicalMay 17, 2018SESE
K180244EnsoETMAdvanced Cooling Therapy, Inc. D/B/A Attune MedicalFeb 28, 2018SESE
K172029EnsoETMAttune MedicalJan 5, 2018SESE
K172493EnsoETMAttune MedicalOct 17, 2017SESE
K170009Esophageal Cooling DeviceAdvanced Cooling Therapy, LLCMar 30, 2017SESE
K152450Esophageal Cooling DeviceAdvanced Cooling Therapy, LLCJan 15, 2016SESE
DEN140018ESOPHAGEAL COOLING DEVICEAdvanced Cooling Therapy, LLCJun 23, 2015DENG

Top Applicants

Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...