Single-Loop Snare Retrieval Kit and Triple-Loop Snare Retrieval Kit are intended for the percutaneous removal of retrievable inferior vena cava (IVC) filters that are no longer medically required, via jugular approach.
Device Story
Single-use percutaneous retrieval system; includes 9F inner sheath, 11F outer sheath, 8F dilator, hemostasis valve, stopcock, and snare catheter with either single-loop or triple-loop snare. Snares feature radiopaque loops; catheter components include radiopaque marker bands for fluoroscopic visualization. Operated by physicians in clinical settings; device inserted percutaneously via jugular approach to capture and remove IVC filters. Fluoroscopy guides placement and retrieval. Modification involves change to inner layer material of 11F outer sheath.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical bench testing, including supplemental verification of the modified outer sheath and biocompatibility testing (cytotoxicity, sensitization, irritation, pyrogenicity, acute systemic toxicity, and hemocompatibility).
Technological Characteristics
Disposable percutaneous retrieval kit. Materials include radiopaque snare loops and marker bands. Biocompatibility per ISO 10993-4, 5, 10, 11, 23. Testing includes ASTM hemolysis, complement activation, PTT, and platelet/leukocyte counts. Mechanical testing includes tensile strength, liquid leakage, and simulated use. No energy source; manual operation.
Indications for Use
Indicated for percutaneous removal of retrievable IVC filters no longer medically required via jugular approach.
Regulatory Classification
Identification
An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.
K191758 — Single-Loop Snare Retrieval Kit, Triple-Loop Snare Retrieval Kit · Argon Medical Devices, Inc. · Dec 17, 2019
K133681 — CRUX SNARE FILTER RETRIEVAL SET · Volcano Corporation · Jan 30, 2014
K152560 — Bard Snare Retrieval Kit · Neometrics, Inc. · Oct 23, 2015
K152136 — Recovery Cone Removal System · C.R. Bard, Inc. · Feb 18, 2016
K181757 — Günther Tulip® Vena Cava Filter Retrieval Set · Cook Incorporated · Nov 6, 2018
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
September 7, 2023
Argon Medical Devices, Inc. Ana Jimenez-Hughes Senior Regulatory Affairs Specialist 1445 Flat Creek Road Athens, Texas 75751
Re: K232443
Trade/Device Name: Single-Loop Snare Retrieval Kit, Triple-Loop Snare Retrieval Kit Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy Catheter Regulatory Class: Class II Product Code: MMX Dated: August 9, 2023 Received: August 14, 2023
Dear Ana Jimenez-Hughes:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Gregory W. Digitally signed by Gregory W. O'connell -S O'connell -S Date: 2023.09.07
11:13:13-04'00'
Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K232443
Device Name Single-Loop Snare Retrieval Kit Triple-Loop Snare Retrieval Kit
Indications for Use (Describe)
Single-Loop Snare Retrieval Kit and Triple-Loop Snare Retrieval Kit are intended for the percutaneous removal of retrievable inferior vena cava (IVC) filters that are no longer medically required, via jugular approach.
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
Date Prepared: August 9, 2023
| Company: | Argon Medical Devices, Inc.<br>1445 Flat Creek Road<br>Athens, Texas 75751 USA<br>Facility Registration number: 1625425 |
|----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Ana Jimenez-Hughes<br>Senior Regulatory Affairs Specialist<br>Phone: 903-676-4276<br>Fax: 903-677-9396<br>Email: <a href="mailto:ana.hughes@argonmedical.com">ana.hughes@argonmedical.com</a> |
| Device Trade Name: | Single-Loop Snare Retrieval Kit<br>Triple-Loop Snare Retrieval Kit |
| Device Common Name: | Device, Percutaneous Retrieval |
| Device Classification: | Device, Percutaneous Retrieval<br>Product code, MMX<br>21 CFR 870.5150<br>Class II<br>Review Panel: Cardiovascular Devices |
| Predicate Device(s): | Primary: K191758 Single-Loop Snare Retrieval Kit, Triple-Loop Snare Retrieval Kit |
| Description of the Device: | The Single-Loop Snare Retrieval Kit and Triple-Loop Snare Retrieval Kit are single use devices. |
| | The disposable system consists of: [A] 9F (ID) Inner Sheath, [B] 11F (ID) Outer Sheath,<br>[C] 8F (OD) Dilator, [D] Hemostasis Valve with Sideport, [E] High Pressure Stopcock, [F]<br>7F [2.4mm] (OD) x 76cm Snare Catheter with Tuohy-Borst Y-Port Adapter, 20mm x<br>93cm Single-Loop Snare (fully expanded) or 30mm x 93cm Triple-Loop Snare (fully<br>expanded) with Torque Handle. |
| | The snares have radiopaque loops and are preloaded in the snare catheter. The snare<br>catheter, inner sheath, and outer sheath have a radiopaque marker band at the distal<br>tip for enhanced fluoroscopic visualization. |
{4}------------------------------------------------
| Indication for<br>Use: | Single-Loop Snare Retrieval Kit and Triple-Loop Snare Retrieval Kit are intended for<br>the percutaneous removal of retrievable inferior vena cava (IVC) filters that are no<br>longer medically required, via jugular approach. |
|-------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device<br>Modification: | The device modification included in this submission is limited to the change of inner<br>layer material of the 11F (ID) Outer Sheath. |
| Substantial<br>Equivalence: | There is no change of intended use or fundamental scientific technology between<br>the proposed modified and predicate device. The proposed modified device has<br>the same indication for use as the predicate, K191758.<br>Non-Clinical Testing<br>In accordance with the Design Failure Modes and Effects Analysis, supplemental<br>verification testing was identified to support the substantial equivalence of the<br>modified Single-Loop Snare Retrieval Kit and Triple-Loop Snare Retrieval Kit. Testing<br>included:<br>Outer Sheath - Visual Outer Sheath - Dimensional Outer Sheath - Tensile Strength Outer Sheath - Liquid Leakage Simulated Use Outer Sheath - Delamination Biocompatibility: Cytotoxicity (ISO 10993-5) Sensitization (ISO 10993-10) Irritation or Intracutaneous Reactivity (ISO 10993-23) Material Mediated Pyrogen (ISO 10993-11) Acute Systemic Toxicity (ISO 10993-11) Hemocompatibility (ISO10993-4) ASTM Hemolysis, Direct and Extract Complement Activation, SC5b-9 Partial Thromboplastin (PTT) Platelet and Leukocyte Count <i>In vitro</i> Blood Loop |
| | The following testing was leveraged from K191758:<br>Visual/Dimensional (Inner Sheath, Dilator, Delivery Catheter Single-Loop Snare & Triple Loop Snare) Leak Test (Inner Sheath, Dilator and Delivery Catheter) |
| • | Burst Test |
| • | Pull Test/Tensile Strength (Inner Sheath, Dilator, Delivery Catheter Single-Loop Snare & Triple Loop Snare) |
| • | Corrosion Resistance (Single-Loop Snare & Triple Loop Snare) |
| • | Torque Response (Snare Assembly) |
| • | Radiopacity |
| • | Luer Testing |
| • | Contrast Medium Injection |
| • | High Pressure Stopcock Testing |
| • | Flexural Modulus and Tip Flexibility Testing |
| • | Radial Force Testing |
| • | Design Validation Testing and Summative Usability Testing |
| Animal testing was not required for the determination of substantial equivalence. | |
| Clinical testing was not required for the determination of substantial equivalence. | |
| Test results demonstrate that all acceptance criteria were met; therefore, the device meets the established product specifications. | |
| <b>Conclusion:</b> | The proposed device modifications to the Single-Loop Snare Retrieval Kit and Triple-Loop Snare Retrieval kit do not change its intended use or principles of operation. Based on the Indication for Use, design, and performance testing, the Single-Loop Snare Retrieval Kit and Triple-Loop Snare Retrieval Kit meet the requirements for its intended use and is substantially equivalent to the predicate device. |
{5}------------------------------------------------
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.