← Product Code [LOX](/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices/LOX) · K242969

# Pantera Pro; Pantera LEO (K242969)

_Biotronik, Inc. · LOX · Oct 25, 2024 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices/LOX/K242969

## Device Facts

- **Applicant:** Biotronik, Inc.
- **Product Code:** [LOX](/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices/LOX.md)
- **Decision Date:** Oct 25, 2024
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 870.5100
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

The Pantera Pro is indicated for balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis for the purpose of improving myocardial perfusion. The Pantera Pro (balloon diameter 2.0 - 4.0 mm) is also indicated for post-delivery expansion of balloon expandable stents. The Pantera LEO is indicated for balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis for the purpose of improving myocardial perfusion and for post dilatation of coronary stents.

## Device Story

Pantera Pro and Pantera LEO are rapid exchange PTCA catheters; distal balloon inflated via proximal Luer port; guide wire lumen provided for positioning; radiopaque markers facilitate fluoroscopic placement. Used by physicians in clinical settings for coronary artery/bypass graft stenosis dilatation and stent expansion. Operates via hydraulic pressurization of balloon. Output is mechanical dilatation of vessel/stent. Benefits include improved myocardial perfusion and optimized stent apposition.

## Clinical Evidence

Bench testing only. Testing included coating integrity, particulate analysis, biocompatibility, bioburden, and shelf-life. Results demonstrate conformance to design specifications and substantial equivalence to predicates.

## Technological Characteristics

PTCA rapid exchange catheter; Polyamide 12 balloon; hydrophilic/hydrophobic coatings; radiopaque markers; 0.014 inch guide wire compatibility; 5F guiding catheter compatibility. Sterilization via EO gas. Single-use.

## Regulatory Identification

Standard PTCA Catheter: A PTCA catheter is a device that operates on the principle of hydraulic pressurization applied through an inflatable balloon attached to the distal end. A PTCA balloon catheter has a single or double lumen shaft. The catheter features a balloon of appropriate compliance for the clinical application, constructed from a polymer. The balloon is designed to uniformly expand to a specified diameter and length at a specific pressure as labeled, with well characterized rates of inflation and deflation and a defined burst pressure. The device generally features a type of radiographic marker to facilitate fluoroscopic visualization of the balloon during use. A PTCA catheter is intended for balloon dilatation of a hemodynamically significant coronary artery or bypass graft stenosis in patients evidencing coronary ischemia for the purpose of improving myocardial perfusion. A PTCA catheter may also be intended for the treatment of acute myocardial infarction; treatment of in-stent restenosis (ISR) and/or post-deployment stent expansion. Cutting/scoring PTCA Catheter: A cutting/scoring PTCA catheter is a balloon-tipped catheter with cutting/scoring elements attached, which is used in those circumstances where a high pressure balloon resistant lesion is encountered. A cutting/scoring PTCA catheter is intended for the treatment of hemodynamically significant coronary artery stenosis for the purpose of improving myocardial perfusion. A cutting/scoring PTCA catheter may also be indicated for use in complex type C lesions or for the treatment of in-stent restenosis.

## Special Controls

*Classification.* Class II (special controls). The special control for this device is “Class II Special Controls Guidance Document for Certain Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheters.” See § 870.1(e) for the availability of this guidance document.(b)
*Cutting/scoring PTCA Catheter* —(1)*Identification.* A cutting/scoring PTCA catheter is a balloon-tipped catheter with cutting/scoring elements attached, which is used in those circumstances where a high pressure balloon resistant lesion is encountered. A cutting/scoring PTCA catheter is intended for the treatment of hemodynamically significant coronary artery stenosis for the purpose of improving myocardial perfusion. A cutting/scoring PTCA catheter may also be indicated for use in complex type C lesions or for the treatment of in-stent restenosis.(2)
*Classification.* Class III (premarket approval). As of May 28, 1976, an approval under section 515 of the act is required before this device may be commercially distributed. See § 870.3.

## Predicate Devices

- Pantera Pro ([K160985](/device/K160985.md))
- Pantera LEO ([K163660](/device/K163660.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. Next to that is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

October 25, 2024

Biotronik, Inc. Jon Brumbaugh Vice President, Regulatory Affairs and New Product Development 6024 Jean Road Lake Oswego, Oregon 97035

# Re: K242969

Trade/Device Name: Pantera Pro: Pantera LEO Regulation Number: 21 CFR 870.5100 Regulation Name: Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheter Regulatory Class: Class II Product Code: LOX Dated: September 26, 2024 Received: September 26, 2024

# Dear Jon Brumbaugh:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Image /page/1/Picture/7 description: The image shows a digital signature. The signature is for Lydia S. Glaw -S. The date of the signature is 2024.10.25, and the time is 15:22:32 -04'00'.

{2}------------------------------------------------

 Lydia Glaw Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

# Indications for Use

Submission Number (if known)

K242969

Device Name

Pantera Pro; Pantera LEO

Indications for Use (Describe)

The Pantera Pro is indicated for balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis for the purpose of improving myocardial perfusion. The Pantera Pro (balloon diameter 2.0 - 4.0 mm) is also indicated for post-delivery expansion of balloon expandable stents.

The Pantera LEO is indicated for balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis for the purpose of improving myocardial perfusion and for post dilatation of coronary stents.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

#### Special 510(k) Premarket Notification

| Date Prepared:           | October 23, 2024                                                                                                                                                                                                                    |
|--------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact:                 | Jon Brumbaugh<br>Vice President, Regulatory Affairs & New Product Development<br>BIOTRONIK, Inc.<br>6024 Jean Road<br>Lake Oswego, OR 97035 USA<br>Phone (888) 345-0374<br>Fax (800) 913-6993<br>Email: jon.brumbaugh@biotronik.com |
| Trade Names:             | Pantera Pro and Pantera LEO                                                                                                                                                                                                         |
| Generic/Common<br>Names: | Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheter                                                                                                                                                                      |
| Classification Name:     | Catheter, transluminal coronary angioplasty, percutaneous                                                                                                                                                                           |
| Product Code:            | LOX                                                                                                                                                                                                                                 |
| Predicate Device:        | Pantera Pro (K160985) and Pantera LEO (K163660).                                                                                                                                                                                    |

#### 510(K) SUMMARY: Pantera Pro and Pantera LEO Catheters

#### Device Description

Pantera Pro is a PTCA rapid exchange system with a balloon at the distal end of the catheter. A Luer port at the proximal end enables the attachment of an inflation device for the inflation of the balloon. The catheter provides a lumen, which enables the use of a guide wire to position the catheter. Radiopaque balloon markers aid in the placement of the catheter's balloon segment under fluoroscopy.

Pantera LEO is a PTCA rapid exchange system with a balloon at the distal end of the catheter. A Luer port at the proximal end enables the attachment of an inflation device for the inflation of the balloon. The catheter provides a lumen, which enables the use of a guide wire to position the catheter. Radiopaque balloon markers aid in the placement of the catheter's balloon segment under fluoroscopy.

#### Indications for Use

The Pantera Pro is indicated for balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis for the purpose of improving myocardial perfusion. The Pantera Pro (balloon diameter 2.0 - 4.0 mm) is also indicated for post-delivery expansion of balloon expandable stents.

{5}------------------------------------------------

The Pantera LEO is indicated for balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis for the purpose of improving myocardial perfusion and for post dilatation of coronary stents.

| Description                       | Pantera Pro<br>(K160985 -<br>Predicate)                                                                                                                                                                                                                                                                                                                                                                                                                                                | Pantera LEO<br>(K163660 -<br>Predicate)                                                                                                                                                                                                                                                                                                                                                                                                                                                | Subject Devices<br>Rationale for<br>Substantial<br>Equivalence                     |
|-----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|
| 510(k) Number                     | K160985                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | K163660                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | K242969                                                                            |
| Classification                    | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Identical                                                                          |
| Product Code                      | LOX                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | LOX                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Identical                                                                          |
| Regulation                        | 21 CFR 870.5100                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | 21 CFR 870.5100                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Identical                                                                          |
| Device Description                | Pantera Pro is a<br>PTCA rapid exchange<br>system with a<br>balloon at the distal<br>end of the catheter.<br>A Luer port at the<br>proximal end<br>enables the<br>attachment of an<br>inflation device for<br>the inflation of the<br>balloon. The<br>catheter provides a<br>lumen, which<br>enables the use of a<br>guide wire to<br>position the<br>catheter.<br>Radiopaque balloon<br>markers aid in the<br>placement of the<br>catheter's balloon<br>segment under<br>fluoroscopy. | Pantera LEO is a<br>PTCA rapid exchange<br>system with a<br>balloon at the distal<br>end of the catheter.<br>A Luer port at the<br>proximal end<br>enables the<br>attachment of an<br>inflation device for<br>the inflation of the<br>balloon. The<br>catheter provides a<br>lumen, which<br>enables the use of a<br>guide wire to<br>position the<br>catheter.<br>Radiopaque balloon<br>markers aid in the<br>placement of the<br>catheter's balloon<br>segment under<br>fluoroscopy. | Identical                                                                          |
| Indications for Use               | The Pantera Pro is<br>indicated for balloon<br>dilatation of the<br>stenotic portion of a<br>coronary artery or<br>bypass graft<br>stenosis for the<br>purpose of<br>improving<br>myocardial<br>perfusion. The<br>Pantera Pro (balloon<br>diameter 2.0 – 4.0<br>mm) is also<br>indicated for post-                                                                                                                                                                                     | The Pantera LEO is<br>indicated for balloon<br>dilatation of the<br>stenotic portion of a<br>coronary artery or<br>bypass graft<br>stenosis for the<br>purpose of<br>improving<br>myocardial perfusion<br>and for post<br>dilatation of<br>coronary stents.                                                                                                                                                                                                                            | Identical                                                                          |
|                                   |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |                                                                                    |
|                                   | delivery expansion<br>of balloon<br>expandable stents.                                                                                                                                                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |                                                                                    |
| Principle of<br>Operation         | Device operates on<br>the principle of<br>hydraulic<br>pressurization<br>applied through an<br>inflatable balloon<br>attached to the<br>distal end                                                                                                                                                                                                                                                                                                                                     | Device operates on<br>the principle of<br>hydraulic<br>pressurization<br>applied through an<br>inflatable balloon<br>attached to the<br>distal end                                                                                                                                                                                                                                                                                                                                     | Identical                                                                          |
| Catheter Type                     | Percutaneous<br>Transluminal<br>Coronary<br>Angioplasty (PTCA)<br>Catheter                                                                                                                                                                                                                                                                                                                                                                                                             | Percutaneous<br>Transluminal<br>Coronary<br>Angioplasty (PTCA)<br>Catheter                                                                                                                                                                                                                                                                                                                                                                                                             | Identical                                                                          |
| Guiding catheter<br>compatibility | Single Device: 5F<br>(0.056 inch/1.42<br>mm)                                                                                                                                                                                                                                                                                                                                                                                                                                           | Single Device: 5F<br>(0.056 inch/1.42<br>mm)                                                                                                                                                                                                                                                                                                                                                                                                                                           | Identical                                                                          |
| Balloon Material                  | Polyamide 12                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Polyamide 12                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Identical                                                                          |
| Balloon Diameter                  | 1.25, 1.5, 2.0, 2.5,<br>3.0, 3.5, 4.0                                                                                                                                                                                                                                                                                                                                                                                                                                                  | 2.0, 2.25, 2.5, 2.75,<br>3.0, 3.25, 3.5, 3.75,<br>4.0, 4.5, 5.0                                                                                                                                                                                                                                                                                                                                                                                                                        | Identical                                                                          |
| Balloon Length                    | 6, 10, 15, 20, 25, 30                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | 8, 12, 15, 20, 30                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Identical                                                                          |
| Radiopaque Markers                | Ø1.25 - 1.5 mm: One<br>marker band<br>Ø >1.5 mm: Two<br>marker bands                                                                                                                                                                                                                                                                                                                                                                                                                   | Ø all sizes: Two<br>marker bands                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Identical                                                                          |
| Usable Length                     | 140                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | 145                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Identical                                                                          |
| Guide wire<br>compatibility       | 0.014 inch (0.36<br>mm)                                                                                                                                                                                                                                                                                                                                                                                                                                                                | 0.014 inch (0.36<br>mm)                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Identical                                                                          |
| Balloon Nominal<br>pressure [atm] | all sizes: 7 atm                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | all sizes: 14 atm                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Identical                                                                          |
| Balloon RBP [atm]                 | all sizes: 14 atm                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Ø: 2.0-4.0 mm: 20 atm<br>Ø: 4.5-5.0 mm: 18<br>atm                                                                                                                                                                                                                                                                                                                                                                                                                                      | Identical                                                                          |
| Distal outer shaft<br>coating     | Hydrophilic Coating<br>Slip Coat Base<br>Coat Type 90174 B<br>Slip Coat Top Coat<br>Type 90172 A                                                                                                                                                                                                                                                                                                                                                                                       | Hydrophilic Coating<br>Slip Coat Base<br>Coat Type 90174 B<br>Slip Coat Top Coat<br>Type 90172 A                                                                                                                                                                                                                                                                                                                                                                                       | Similar – alternative<br>basecoat<br>formulation.<br>Top coat remains<br>unchanged |
| Balloon coating                   | Ø 1.25 – 2.0 mm:<br>Hydrophilic coating<br>Ø 2.5 - 4.0 mm:<br>Hydrophobic coating                                                                                                                                                                                                                                                                                                                                                                                                      | all sizes:<br>Hydrophobic coating                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Identical                                                                          |
| Sterilization Method              | EO gas                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | EO Gas                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Identical                                                                          |
| Single Use                        | Single-Use                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Single-Use                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Identical                                                                          |

## Comparison of Technological Characteristics with Predicate Devices

{6}------------------------------------------------

{7}------------------------------------------------

#### Performance Data

All necessary performance testing was conducted on the Pantera Pro and Pantera LEO Catheters to ensure that the devices conforms to the design specification and to support a determination of substantial equivalence to the predicate devices.

- . Product performance - coating integrity and particulates
- . Biocompatibility
- . Microbiological - Bioburden
- . Shelf-Life - coating integrity and particulates

The collective results of the performed testing demonstrated that the materials chosen, the manufacturing processes, and design of the components meet the established specifications necessary for consistent performance durinq its intended use. In addition, the collective bench testing demonstrates that the proposed device does not introduce new issues of safety or effectiveness when compared to the predicate device.

## Conclusions

Based on the performance testing and the technological characteristics, it can be concluded that the Pantera Pro and Pantera LEO catheters meet their established performance for their intended use and are substantially equivalent to the predicate devices.

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices/LOX/K242969](https://fda.innolitics.com/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices/LOX/K242969)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
