Sterilizable Internal Defibrillation Paddles for use with LIFEPAK defibrillators/monitors

K182503 · Physio-Control, Inc. · LDD · Jun 3, 2019 · Cardiovascular

Device Facts

Record IDK182503
Device NameSterilizable Internal Defibrillation Paddles for use with LIFEPAK defibrillators/monitors
ApplicantPhysio-Control, Inc.
Product CodeLDD · Cardiovascular
Decision DateJun 3, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.5300
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Sterilizable Internal Defibrillation Paddles are intended for use with LIFEPAK defibrillators to internally detect ECG rhythm and provide defibrillation or synchronized cardioversion directly to the surgically exposed heart within a sterile use environment. Defibrillation is indicated for the termination of certain potentially fatal arrhythmias, such as ventricular fibrillation and symptomatic ventricular tachycardia. Synchronized cardioversion is indicated for the treatment of atrial fibrillation, atrial flutter, paroxysmal supraventricular tachycardia, supraventricular tachycardia, and in relatively stable patients, ventricular tachycardia.

Device Story

Accessory for LIFEPAK defibrillators/monitors used during open-heart surgery. Consists of metal paddle electrodes and molded plastic handles with integrated discharge control. Operated by highly-trained medical professionals in sterile environments. Device detects ECG rhythm directly from surgically exposed heart; delivers defibrillation or synchronized cardioversion therapy. Requires sterilization before and after each use. Design improvements include integrated handle-electrode assembly and material changes for increased durability against repeated sterilization cycles.

Clinical Evidence

No human clinical studies submitted. Evidence consists of bench testing, including design verification, biocompatibility, electrical safety, electromagnetic compatibility, animal testing, and simulated use testing.

Technological Characteristics

Metal paddle electrodes and molded plastic handles. Integrated handle-electrode assembly. Two connector configurations. Designed for repeated sterilization. Electrical energy delivery via LIFEPAK defibrillator. Class II device.

Indications for Use

Indicated for patients requiring internal defibrillation or synchronized cardioversion for arrhythmias including ventricular fibrillation, symptomatic ventricular tachycardia, atrial fibrillation, atrial flutter, and paroxysmal supraventricular tachycardia. Contraindicated for Pulseless Electrical Activity (PEA), idioventricular/ventricular escape rhythms, asystole, and ventricular fibrillation (for cardioversion).

Regulatory Classification

Identification

Low-energy DC-defibrillator: A device that delivers into a 50 ohm test load an electrical shock of a maximum of 360 joules of energy used for defibrillating (restoring normal heart rhythm) the atria or ventricles of the heart or to terminate other cardiac arrhythmias. This generic type of device includes low energy defibrillators with a maximum electrical output of less than 360 joules of energy that are used in pediatric defibrillation or in cardiac surgery. The device may either synchronize the shock with the proper phase of the electrocardiogram or may operate asynchronously. The device delivers the electrical shock through paddles placed either directly across the heart or on the surface of the body. High-energy DC-defibrillator: A device that delivers into a 50 ohm test load an electrical shock of greater than 360 joules of energy used for defibrillating the atria or ventricles of the heart or to terminate other cardiac arrhythmias. The device may either synchronize the shock with the proper phase of the electrocardiogram or may operate asynchronously. The device delivers the electrical shock through paddles placed either directly across the heart or on the surface of the body.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo for the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. Physio-Control, Inc. Chelsea Cullen Sr. Regulatory Affairs Specialist 11811 Willow Road NE Redmond, Washington 98052 Re: K182503 Trade/Device Name: Sterilizable Internal Defibrillation Paddles for use with LIFEPAK defibrillators/monitors Regulation Number: 21 CFR 870.5300 Regulation Name: DC-Defibrillator (Including Paddles) Regulatory Class: Class II Product Code: LDD Dated: April 30, 2019 Received: May 1, 2019 Dear Chelsea Cullen: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, For Matthew Hillebrenner Deputy Director Division of Cardiac Electrophysiology, Diagnostics, and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K182503 Device Name Sterilizable Internal Defibrillation Paddles Indications for Use (Describe) The Sterilizable Internal Defibrillation Paddles are intended for use with LIFEPAK defibrillators to internally detect ECG rhythm and provide defibrillation or synchronized cardioversion directly to the surgically exposed heart within a sterile use environment. | Type of Use (Select one or both, as applicable) | | |----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------| | <span style="font-family: DejaVu Sans, sans-serif">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: DejaVu Sans, sans-serif">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # SECTION 5: 510(k) SUMMARY ### Submitter: Physio-Control, Inc. 11811 Willows Road Northeast Redmond, Washington 98073-9706 Registration Number: 3015876, 3006703820 # Contact: Chelsea Cullen Staff Regulatory Affairs Specialist 425 867 4138 (o) (425) 867-4154 (f) chelsea.cullen(@stryker.com April 23, 2019 Date of Preparation: ## Device Trade/Proprietary Name: Sterilizable Internal Defibrillation Paddles for use with LIFEPAK® defibrillators/monitors ### Device Common Name: Internal Defibrillation Paddles #### Device Classification: | Device Classification and<br>CFR Reference | Classification Panel | Product Code | |----------------------------------------------------------------------------------|------------------------|--------------| | Low Energy DC-Defibrillator<br>(Including Paddles), Class II<br>(21CFR 870.5300) | Cardiovascular Devices | LDD | | | | | #### Predicate Devices: The features and functions of the proposed Sterilizable Internal Defibrillation Paddles accessory are equivalent to the following previously cleared Internal Paddles: | Predicate Device | 510(k) Number(s)/ Clearance Date | |------------------------------------------------------------------|----------------------------------| | Internal Paddles and Paddle<br>Handles with Discharge<br>Control | K895379 / cleared in 1990 | {4}------------------------------------------------ ## Device Description: The Sterilizable Internal Defibrillation Paddles is an accessory designed to be used with biphasic LIFEPAK defibrillator/monitors during open heart cardiac surgery. The Sterilizable Internal Defibrillation Paddles are intended for use by highly-trained medical professionals to internally detect electrocardiogram (ECG) rhythm and deliver internal defibrillation and synchronized cardioversion therapy directly to the surgically exposed heart. This accessory consists of metal paddle electrodes, and molded plastic handles with discharge control (located on the right handle) which connects directly to the defibrillator. The Sterilizable Internal Defibrillation Paddles accessory requires sterilization before initial use and after each use, per the Sterilizable Internal Defibrillation Paddles Instructions for Use. ## Indications for Use: The Sterilizable Internal Defibrillation Paddles are intended for use with LIFEPAK defibrillators to internally detect ECG rhythm and provide defibrillation or synchronized cardioversion directly to the surgically exposed heart within a sterile use environment. ## Intended Use: Defibrillation is indicated for the termination of certain potentially fatal arrhythmias, such as ventricular fibrillation and symptomatic ventricular tachycardia. Synchronized cardioversion is indicated for the treatment of atrial fibrillation, atrial flutter, paroxysmal supraventricular tachycardia, supraventricular tachycardia, and in relatively stable patients, ventricular tachycardia. ## Contraindications: Defibrillation is contraindicated in the treatment of Pulseless Electrical Activity (PEA), such as idioventricular or ventricular escape rhythms, and in the treatment of asystole. Synchronized cardioversion is contraindicated in the treatment of Pulseless Electrical Activity (PEA) such as idioventricular or ventricular escape rhythms, asystole, and ventricular fibrillation. ## Summary of Technological Characteristics: The intended use, features, and functional characteristics of the proposed Sterilizable Internal Defibrillation Paddles are equivalent to the predicate device. The main differences between the previously cleared predicate Internal paddle handles with discharge control and the proposed Sterilizable Internal Defibrillation Paddles are 1) an integrated handle-electrode assembly which is available in two (2) connector configurations, and 2) material change to the device to increase durability and resistance of the paddles to better withstand the harshness of cleaning and repeated sterilization methods. The proposed changes have not raised any new issues when compared to the existing predicate devices. {5}------------------------------------------------ # Performance Data: Performance testing has been completed to demonstrate that the proposed Sterilizable Internal Defibrillation Paddles meet the safety and performance requirements established in the design specifications. Comprehensive testing included the following: - o Design Verification Testing - Biocompatibility Testing - Electrical Safety and Electromagnetic Compatibility Testing o - o Design Validation Testing - O Animal Testing, and Simulated Use Testing No human clinical studies were submitted as part of this 510(k) Premarket Notification. ## Conclusion: The information in this 510(k) notification demonstrates that the Sterilizable Internal Defibrillation Paddles are substantially equivalent to the predicate Internal Deffbrillation Handles and Electrodes with respect to performance.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...