Indications: The Circulator Boot System alone - or in combination with other drug or device therapies - may be prescribed by the physician to treat: Poor arterial flow in extremities associated with: - Ischemic ulcers - Rest pain or claudication (pain with walking) - Threatened gangrene - Insufficient blood supply at an amputation site - Persisting ischemia after embolectomy or bypass surgery - Pre- and post-arterial reconstruction to improve runoff Diabetes complicated by the above or other conditions possibly related to arterial insufficiency including: - Nocturnal leg cramps - Necrobiosis diabeticorum Venous diseases (once risk of emboli minimized): - Prophylaxis of deep vein thrombophlebitis - Edema and induration associated with chronic venous stasis - Venous stasis ulcers Athletic injuries: "Charlie horses", pulled muscles, and edematous muscles
Device Story
Multicrus Circulator Boot System provides pneumatic compression therapy to extremities. System inputs: ECG signal (QRS complex) via heart monitor. Operation: Computer detects QRS complex; triggers pneumatic compression in end-diastole to maximize blood inflow; releases pressure 0.04 seconds before next QRS to reduce afterload/heart work. Device consists of telescoping, adjustable rigid plastic boot (eight overlapping pieces) applied to leg/thigh; secured with Velcro/straps. Technician adjusts pressure via valve assembly. Used in clinics or home settings by trained nurses, therapists, or patients under physician supervision. Output: Controlled pneumatic compression. Clinical benefit: Improved arterial flow, reduced heart work, and management of venous/ischemic conditions.
Clinical Evidence
No clinical trial data provided. Evidence based on bench testing and clinical usage observations. Device performance verified in clinical settings; desired pressures achieved in all patients during five weeks of clinical usage.
Technological Characteristics
Rigid plastic boot composed of eight overlapping, telescoping pieces. Adjustable in length, height, and width. Pneumatic compression principle. Energy source: Compressed air. Connectivity: Wired connection to heart monitor. Software: Automated timing based on ECG QRS complex detection.
Indications for Use
Indicated for patients with poor arterial flow in extremities (ischemic ulcers, claudication, threatened gangrene, post-surgical ischemia), diabetes-related arterial insufficiency (nocturnal cramps, necrobiosis diabeticorum), venous diseases (DVT prophylaxis, edema, venous stasis ulcers), and athletic injuries (muscle strains/edema).
Regulatory Classification
Identification
An external counter-pulsating device is a noninvasive, prescription device used to assist the heart by applying positive or negative pressure to one or more of the body's limbs in synchrony with the heart cycle.
Special Controls
*Classification.* (1) Class II (special controls) when the device is intended for the treatment of chronic stable angina that is refractory to optimal anti-anginal medical therapy and without options for revascularization. The special controls for this device are:(i) Nonclinical performance evaluation of the device must demonstrate a reasonable assurance of safety and effectiveness for applied pressure, synchronization of therapy with the appropriate phase of the cardiac cycle, and functionality of alarms during a device malfunction or an abnormal patient condition;
(ii) Reliabilities of the mechanical and electrical systems must be established through bench testing under simulated use conditions and matched by appropriate maintenance schedules;
(iii) Software design and verification and validation must be appropriately documented;
(iv) The skin-contacting components of the device must be demonstrated to be biocompatible;
(v) Appropriate analysis and testing must be conducted to verify electrical safety and electromagnetic compatibility of the device; and
(vi) Labeling must include a detailed summary of the device-related and procedure-related complications pertinent to use of the device.
(2) Class III (premarket approval) for the following intended uses: Unstable angina pectoris; acute myocardial infarction; cardiogenic shock; congestive heart failure; postoperative treatment of patients who have undergone coronary artery bypass surgery; peripheral arterial disease associated with ischemic ulcers rest pain or claudication, threatened gangrene, insufficient blood supply at an amputation site, persisting ischemia after embolectomy or bypass surgery, and/or pre- and post-arterial reconstruction to improve runoff; diabetes complicated by peripheral arterial disease or other conditions possibly related to arterial insufficiency including nocturnal leg cramps and/or necrobiosis diabeticorum; venous diseases, including prophylaxis of deep vein thrombophlebitis, edema (e.g., chronic lymphedema) and/or induration (e.g., stasis dermatitis) associated with chronic venous stasis, venous stasis ulcers, and/or thrombophlebitis; athletic injuries, including Charley horses, pulled muscles and/or edematous muscles; necrotizing cellulitis.
(c)
*Date premarket approval application (PMA) or notice of completion of product development protocol (PDP) is required.* A PMA or notice of completion of a PDP is required to be filed with FDA on or before March 31, 2014, for any external counter-pulsating device, with an intended use described in paragraph (b)(2) of this section, that was in commercial distribution before May 28, 1976, or that has, on or before March 31, 2014, been found to be substantially equivalent to any external counter-pulsating device, with an intended use described in paragraph (b)(2) of this section, that was in commercial distribution before May 28, 1976. Any other external counter-pulsating device with an intended use described in paragraph (b)(2) of this section shall have an approved PMA or declared completed PDP in effect before being placed in commercial distribution.
K971026 — THE CIRCULATOR BOOT · Circulator Boot Corp. · Aug 14, 1997
K240499 — ARTAIRA Arterial Compression Device (AACD01) · Airos Medical, Inc. · Oct 11, 2024
K102582 — ARTERIOFLOW MODEL 7500 SEQUENTIAL COMPRESSION DEVICE · Devon Medical, Inc. · Oct 7, 2010
K050172 — CARDIASSIST COUNTER PULSATION SYSTEM SERIES 4000/5000 · Cardiomedics, Inc. · Mar 31, 2005
K181409 — Pressure Therapy System PT1003 · Xiamen Senyang Co., Ltd. · Feb 19, 2019
Submission Summary (Full Text)
{0}------------------------------------------------
# Premarket Notification 510(k) Summary As required by section 807.92
This summary was submitted by:
Richard S. Dillon, M.D. MAY - 7 2009 Circulator Boot Corporation 72 Pennsylvania Avenue Malvern, PA 19355 FDA Registration number: 2518635 Telephone number: 610-240-9980 (Circulator Boot Corp Office) Fax: 610-240-9982 Dr. Dillon @ home office: 610-527-0874
Contact person: Richard S. Dillon, M.D. 1260 Round Hill Rd. Bryn Mawr PA 19010 Telephone: 610-527-0874
Date this summary prepared: 11/30/2008
Device name:
Proprietary name: Multicrus Circulator Boot Systems Common name: End-diastolic boot Current Classification: Listing Number Listing Status Product Code Device Name Enhanced External Counter Pulsation D033631 Active DRN Q012536 Active IPM Cover. Limb Instrument, Compression Q013236 Actice HWN
Name of device component originating this submission: Multicrus Boot.
NAMES OF LEGALLY MARKETED DEVICES FOR WHICH A CLAIM OF SUBSTANTIAL EQUIVALENCE IS MADE as required by 807.92(a)(3):
Circulator Boot - K792354 (1980) - End-diastolic therapy using Long rigid boots for treatment of leg vascular diseases and congestive heart failure
Circulator Boot - K833627/A (1984) - Modification of end-diastolic therapy to limit area of treatment to specific area needing the therapy: the lower leg. The Miniboot.
Circulator Boot - K971026 (1997) - Addition of computer in heart monitor to automate the function of the attending boot technician in continually adjusting the delay time after the detected
{1}------------------------------------------------
ORS complex to allow the preset compression time to occur in end-diastole and to end 0.04 seconds before the next QRS complex. Thus, maximal inflow of blood into the leg is allowed before boot compressions and a maximum reduction in afterload is produced to assist the heart. A nomogram allows therapy in cases of atrial fibrillation.
Vasomedical and CardioMedics: K020857and K22107. Indications approved for the latter include ischemic heart disease, congestive heart failure, angina pectoris, myocardial infarction and cardiogenic shock.
#### DEVICE DESCRIPTION as required by 807.92(a)(4):
The Circulator Boot Systems are designed to allow pneumatic compression therapy to any extremity (its whole or a part) as prescribed by a physician. A computer within a heart monitor delays compressions after the detection of the QRS complex to allow maximum inflow of blood into the extremities and signals release of the extremity 0.04 seconds before the time of the next QRS complex to maximally reduce afterload and heart work. The part to be treated is first placed into a double-walled plastic disposable bag (legs or arms) or enclosed by a small cuff (such as an elbow). The bagged extremity is then placed within a rigid plastic boot which is adjusted to limit the dead space around the leg. In the case of K792354, adjustment for differences in the length of the extremity is made by applying one of three different length Long Boots ("A". "B" or "C") while adjustments in the width of the extremity is made by moving inner walls snuggly against the part; adjustments for the A-P (anterior-posterior) dimension are accomplished by arcing the thick plastic aprons attached to the inner side of the lateral movable walls tightly down over the extremity.
The new "Multicrus" boot K082134 is likewise a rigid plastic boot but telescoping and adjustable in all three dimensions: length, height and width to minimize the space around the bagged leg. It is composed of eight overlapping right-angled plastic pieces which are applied firmly against the extremity. Thus, a piece might be placed under the thigh and knee and against the inner part of the leg (as high up the leg as desired). A second piece may be placed against the lateral aspect of the thigh and knee with its bottom section under the first part. These two parts are pushed together to snuggly hug the inner and lateral parts of the extremity. A third part may then be placed against the inner calf and under the heel and distal part of the first part. A fourth part is placed against the lateral calf and foot and under the heel and the third part and under the second part. The third and fourth parts are again pushed together to snuggly hug the calf and foot. The vertical wall of the fifth part is placed between the vertical walls of the first and third parts while its horizontal wall is placed over the upper leg, again as far up the leg as desired. The vertical wall of the sixth part is placed between the vertical walls of the second and fourth parts and its horizontal part is placed to overlap the thigh and the fifth part. Finally, a seventh and eighth part are placed over the top of the lower leg and foot and over the distal horizontal aspects of the fifth and sixth parts and again are pushed together to minimize the dead space around the
{2}------------------------------------------------
lower leg. All are held in place by Velcro patches between the overlapping parts and by straps surrounding the assembled boot.
The valve assembly is attached to a plate included on the top of the eighth part and connected then to the compressed air line and the heart monitor. The technician adjusts the pressure knob of the valve assembly as needed to achieve the desired pressure in all of the boot systems. Desired pressures were gained in every patient treated with the Multicrus boot during its five weeks of clinical usage. With all of our boot systems, the physician is encouraged to improvise as necessary to treat patients with special needs (e.g. deformed legs that cannot be straightened out sufficiently to enter our rigid boots). A small cuff is one such improvisation that has allowed treatment of portions of any extremity in concert with treatment of the legs in the Long Boots. This small "Cuff Boot"consists of a juvenile blood pressure cuff to surround the extremity (perhaps a tennis elbow) with a bladder from an adult blood pressure cuff connected to both it and a air-bulb. The bladder is placed over the thigh and beneath the plastic airbag of the Long Boot and, hence, is compressed simultaneously with the leg. The small cuff applies pressure in end-diastole with the same pressure introduced into the Long Boot.
# INTENDED USE as required by 807.92(a)5) Indications for Use:
2008 Indications and Contraindications for the Circulator Boot Systems
"Indications: The Circulator Boot System alone - or in combination with other drug or device therapies - may be prescribed by the physician to treat:
# Poor arterial flow in extremities associated with:
- . Ischemic ulcers
- Rest pain or claudication (pain with walking) ◆
- t Threatened gangrene
- # Insufficient blood supply at an amputation site
- Persisting ischemia after embolectomy or bypass surgery
- . Pre- and post-arterial reconstruction to improve runoff
# Diabetes complicated by the above or other conditions possibly related to arterial insufficiency including:
- Nocturnal leg cramps .
- Necrobiosis diabeticorum
# Venous diseases (once risk of emboli minimized):
- . Prophylaxis of deep vein thrombophlebitis
- Edema and induration associated with chronic venous stasis .
- Venous stasis ulcers .
#### Athletic injuries: "Charlie horses", pulled muscles, and edematous muscles
{3}------------------------------------------------
#### SUMMARY OF TECHNOLOGICAL CHARACTERITICS OF MULTICRUS BOOT DEVICE COMPARED TO THE PREDICATE LONG BOOT DEVICES K 792354 (as required by 807.92(a)(6)
The Multicrus Boot is substantially equivalent in safety and effectiveness to the legally marketed (predicate) "A", "B" and "C" adjustable boots (K792354). They function similarly, share the same scientific principles, may be used for the same indications, share the same construction materials. They differ in that the predicate boots have adjustable interiors in two dimensions: the width and anterior-posterior height of the leg; the Multicrus boot in addition has an adjustable length. Improvisations like the small Cuff Boot are in keeping with the overall medical goals of the Circulator Boot Systems. In the case of the Cuff Boot, a rigid boot is not necessary because of the small volume of the compressed air required to produce adequate pressures. Past and new labels were included in the notebook submitted October, 2008.
#### SUMMARY OF NONCLINICAL TESTING FOR THE DEVICE and CONCLUSIONS as required by 807 .92(b)( 1 Y)3)
The Circulator Boot Systems are normally applied and operated by a trained nurse, physical therapist, technician or doctor. Lay personnel have been successfully trained in its usage to treat patients under the supervision of a physician. Again, trained lay personnel under the supervision of their doctor have successfully treated family members in the home. Necessary precautions and warnings are stated in the Manual. The Verification and Validation data was accumulated in the clinic.
CONCLUSION: The summary above shows that there are no new questions of safety and effectiveness for the Multicrus Boot as compared to the predicate device.
#### END OF SUMMARY
{4}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
DEPARTMENT OF HEALTH & HUMAN SERVICES. USA
MAY - 7 2009
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Circulator Boot Corporation c/o Richard S. Dillon, M.D. 72 Pennsylvania Avenue Malvern, PA 19355
Re: K082134
Circulator Multicrus Boot System Regulation Number: 21 CFR 870.5225 Regulation Name: Counter-Pulsating, External Device Regulatory Class: Class III Product Code: DRN Dated: March 7, 2009 Received: March 13, 2009
Dear Dr. Dillon:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Richard S. Dillon, M.D.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Bram D. Zuckerman, M.
Bram D. Zuckerman, M.D. Division Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# STATEMENT: INDICATION FOR USE
510 (k) Number K082134
Device Name: Circulator Multicrus Boot
#### Indications for Use:
# 2008 Indications and Contraindications for the Circulator Boot Systems
"Indications: The Circulator Boot System alone - or in combination with other drug or device therapies - may be prescribed by the physician to treat:
Poor arterial flow in extremities associated with:
- Ischemic ulcers .
- Rest pain or claudication (pain with walking) .
- Threatened gangrene .
- Insufficient blood supply at an amputation site .
- Persisting ischemia after embolectomy or bypass surgery
- Pre- and post-arterial reconstruction to improve runoff ●
Diabetes complicated by the above or other conditions possibly related to arterial insufficiency including:
- Nocturnal leg cramps .
- Necrobiosis diabeticorum .
Venous diseases (once risk of emboli minimized):
- Prophylaxis of deep vein thrombophlebitis .
- Edema and induration associated with chronic venous stasis .
- Venous stasis ulcers .
Athletic injuries: "Charlie horses", pulled muscles, and edematous muscles
Prescription Use
Over the Counter Use
(Part 21 CFR 801 Subpart D) and/or (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGES IF NEEDED)
> Concurrence of of Device Evaluation (ODE) (Division Sign-Off) Division of Cardiovascular Devices K082134 510(k) Number
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.