DEN210006 · Alung Technologies, Inc. · QOH · Nov 13, 2021 · Cardiovascular
Device Facts
Record ID
DEN210006
Device Name
Hemolung Respiratory Assist System
Applicant
Alung Technologies, Inc.
Product Code
QOH · Cardiovascular
Decision Date
Nov 13, 2021
Decision
DENG
Submission Type
Direct
Regulation
21 CFR 870.4150
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Hemolung Respiratory Assist System is indicated for respiratory support that provides extracorporeal carbon dioxide (CO2) removal from the patient's blood for up to 5 days in adults with acute, reversible respiratory failure for whom ventilation of CO2 cannot be adequately or safely achieved using other available treatment options and continued clinical deterioration is expected.
Device Story
Hemolung Respiratory Assist System provides veno-venous extracorporeal carbon dioxide removal (ECCO2R) for patients with acute respiratory failure. System inputs: venous blood via 15.5 Fr dual-lumen catheter. Components: Hemolung Controller (hardware/software), Hemolung Cartridge (integrated gas exchanger/blood pump), and Hemolung Catheter. Operation: blood circulates through cartridge hollow fiber membranes; room air sweep gas flows through membrane interior; CO2 diffuses out of blood into sweep gas; oxygen diffuses into blood. Used in clinical settings by trained staff. Controller manages blood flow rate and sweep gas flow rate. Output: corrected hypercapnia and respiratory acidosis; de-escalation of mechanical ventilation. Benefits: avoids invasive mechanical ventilation in NIV patients; corrects respiratory acidosis. Risks: bleeding, thrombosis, hemolysis, hemodynamic instability. System requires systemic anticoagulation.
Clinical Evidence
Special controls require in vivo evaluation, which may include animal testing and clinical data, to demonstrate performance over the intended duration of use. No specific clinical trial results or metrics (e.g., sensitivity, specificity) are provided in the summary document.
Technological Characteristics
System includes Hemolung Controller, Cartridge (hollow fiber membranes), and 15.5 Fr dual-lumen catheter. Patient-contacting components are heparin-coated. Biocompatibility per ISO 10993-1. Sterilization via ethylene oxide (ISO 11135). Controller manages blood/sweep gas flow. Electrical safety per IEC 60601-1; EMC per IEC 60601-1-2. Software validated per ANSI/AAMI/IEC 62304.
Indications for Use
Indicated for adults with acute, reversible respiratory failure requiring extracorporeal CO2 removal for up to 5 days when ventilation is inadequate/unsafe and clinical deterioration is expected.
Regulatory Classification
Identification
An extracorporeal system for carbon dioxide removal is a system of devices and accessories that provides assisted extracorporeal carbon dioxide removal from the patient's blood in patients with acute respiratory failure, where other available treatment options have failed, and continued clinical deterioration is expected or the risk of death is imminent. The main devices and accessories of the system include, but are not limited to, the console (hardware), software, and disposables, including, but not limited to, a gas exchanger, blood pump, cannulae, tubing, filters, and other accessories (e.g., monitors, detectors, sensors, connectors).
Special Controls
In combination with the general controls of the FD&C Act, the extracorporeal system for carbon dioxide removal is subject to the following special controls:
*Classification.* Class II (special controls). The special controls for this device are:(1) In vivo evaluation, which may include animal testing and clinical data, of the devices and accessories in the circuit must demonstrate their performance over the intended duration of use, including a detailed summary of the in vivo evaluation pertinent to the use of the devices and accessories to demonstrate their effectiveness.
(2) The technological characteristics of the device must ensure that the geometry and design parameters are consistent with the intended use, and that the devices and accessories in the circuit are compatible.
(3) Non-clinical performance testing of the devices and accessories in the circuit must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
(i) Mechanical integrity;
(ii) Durability; and
(iii) Reliability.
(4) All patient contacting components of the device must be demonstrated to be biocompatible.
(5) Performance testing must demonstrate the electrical safety and electromagnetic compatibility (EMC) of any electrical components.
(6) Software validation, verification, and hazard analysis must be performed.
(7) Performance testing must demonstrate the sterility of all patient-contacting components.
(8) Performance testing must support the shelf life of the device by demonstrating continued sterility and device functionality over the identified shelf life.
(9) Labeling must include the following:
(i) A detailed summary of the non-clinical and in vivo evaluations pertinent to use of the device and accessories in the circuit;
(ii) Adequate instructions with respect to circuit setup, performance characteristics with respect to compatibility among different devices and accessories in the circuit, and maintenance during a procedure; and
(iii) A shelf life.
Related Devices
K191407 — Novalung System · Fresenius Medical Care Renal Therapies Group, LLC · Feb 21, 2020
K232926 — Novalung ultimate kit (US) · Fresenius Medical Care Renal Therapies Group, LLC · Apr 15, 2024
K232767 — Nautilus VF ECMO Oxygenator · Mc3, Inc. · Oct 4, 2023
{0}------------------------------------------------
### DE NOVO CLASSIFICATION REQUEST FOR HEMOLUNG RESPIRATORY ASSIST SYSTEM
#### REGULATORY INFORMATION
FDA identifies this generic type of device as:
Extracorporeal system for carbon dioxide removal. An extracorporeal system for carbon dioxide removal is a system of devices and accessories that provides assisted extracorporeal carbon dioxide removal from the patient's blood in patients with acute respiratory failure, where other available treatment options have failed, and continued clinical deterioration is expected or the risk of death is imminent. The main devices and accessories of the system include, but are not limited to, the console (hardware), software, and disposables, including, but not limited to, a gas exchanger, blood pump, cannulae, tubing, filters, and other accessories (e.g., monitors, detectors, sensors, connectors).
NEW REGULATION NUMBER: 21 CFR 870.4150
CLASSIFICATION: Class II
PRODUCT CODE: QOH
### BACKGROUND
DEVICE NAME: Hemolung Respiratory Assist System
SUBMISSION NUMBER: DEN210006
DATE DE NOVO RECEIVED: March 4, 2021
#### SPONSOR INFORMATION:
ALung Technologies, Inc. 2500 Jane Street. Suite 1 Pittsburgh, Pennsylvania 15203
### INDICATIONS FOR USE
The Hemolung Respiratory Assist System is indicated for respiratory support that provides extracorporeal carbon dioxide (CO2) removal from the patient's blood for up to 5 days in adults
with acute, reversible respiratory failure for whom ventilation of CO2 cannot be adequately or safely achieved using other available treatment options and continued clinical deterioration is expected.
{1}------------------------------------------------
### LIMITATIONS
The sale, distribution, and use of the Hemolung RAS are restricted to prescription use in accordance with 21 CFR 801.109.
### Contraindications:
The Hemolung RAS is contraindicated for patients with known sensitivity to heparin (e.g., history of heparin-induced thrombocytopenia). The Hemolung Cartridge membranes are coated with heparin and systemic anticoagulation is required when using the device.
Use of the Hemolung 15.5 Fr Femoral Catheter is contraindicated for patients with an inferior vena cava filter.
PLEASE REFER TO THE LABELING FOR A COMPLETE LIST OF WARNINGS, PRECAUTIONS AND CONTRAINDICATIONS.
# DEVICE DESCRIPTION
The Hemolung RAS provides low blood flow, veno-venous extracorporeal carbon dioxide removal (ECCO2R) using a single, 15.5 French dual lumen catheter inserted percutaneously in the femoral or jugular vein. The Hemolung RAS is not intended to provide therapeutic levels of oxygenation. During Hemolung therapy, blood passing through the circuit is oxygenated via room air sweep gas; however, at ultra-low extracorporeal blood flows, the limited oxygen carrying capacity of blood precludes meaningful oxygenation of mixed venous blood.
The Hemolung RAS consists of three main components:
Image /page/1/Figure/9 description: The image shows three different medical devices. Image A shows a cylindrical container with a clear top and bottom and a white filter inside. Image B shows two catheters with red and blue connectors. Image C shows a medical device on wheels with a screen displaying the word "HEMOLLUNG".
C. Hemolung Controller is the mechanism for operating the Hemolung Respiratory Assist System. It controls the extracorporeal blood flow rate and the sweep gas flow rate.
- A. Hemolung Cartridge is an integrated extracorporeal gas exchanger and blood pump. Blood is circulated around the outside of the Cartridge's hollow fiber membranes while a sweep gas flows through the inside of the membranes. Carbon dioxide diffuses out of the blood and is swept away by the sweep gas while oxygen diffuses from the sweep gas into the blood. Blood tubing and other accompanying disposable products are included in the Hemolung Cartridge Kit.
- B. Hemolung Catheter is a dual lumen venous catheter designed specifically for use with the Hemolung RAS. It exhibits low resistance to flow while also resisting kinks. Individual femoral and jugular Hemolung Catheter Kits are available for use. Each kit includes a Catheter Insertion Kit.
{2}------------------------------------------------
# SUMMARY OF NONCLINICAL/BENCH STUDIES
The following bench studies were performed to demonstrate that the technological characteristics of the Hemolung system are consistent with the device's indication for use:
| Pump characterization | The Hemolung system was tested in a recirculation loop using a blood analogue at 37°C.<br>Pressure-flow characteristics were measured for the system at various pump speeds when<br>connected to Hemolung 15.5 Fr Femoral and Jugular Catheters | |
|------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Gas exchange | Gas exchange testing was performed to characterize gas exchange performance of the<br>Hemolung system over the full range of blood and sweep gas flow rates. The system was<br>tested in a recirculation loop using heparinized bovine blood at 37°C. At each blood flow<br>rate (350 mL/min, 450 mL/min, 550 mL/min), gas exchange was measured at 1 L/min, 5.5<br>L/min and 10 L/min room air sweep gas flow rates. All results are normalized for an inlet<br>pCO2 of 45mmHg. | |
| Heat loss/gain | Heat loss/gain testing was performed to simulate worst case conditions for heat loss or gain<br>during normal operation of the Hemolung. The system was tested in a recirculation loop<br>using heparinized bovine blood at 37°C and room air sweep gas. Temperatures were<br>measured at the inlet and outlet of the 6' blood lines to account for heat loss/gain across the<br>entire extracorporeal circuit. | |
| Reliability and<br>Physical Integrity | Reliability testing evaluated physical integrity of the Hemolung system under worst case<br>operating conditions. The test was performed in a recirculation loop of 40:60 glycerol:<br>water, 0.9% NaCl solution at 37°C for a duration of 14 days. The Cartridges were operated<br>at maximum pump speed (1400 RPM), maximum sweep gas flow (10 LPM), and 1.5X the<br>maximum operating pressure. Following the 14-day simulated use, physical integrity of the<br>blood and gas pathways were evaluated, and the Cartridge and Catheter were examined for<br>any wear or corrosion | |
| Hemolysis | Hemolysis testing was performed to characterize blood damage within the system by<br>measuring plasma free hemoglobin in a recirculation loop. | |
| Heparin coating<br>leachability | Phosphate buffered saline (PBS) was recirculated through each Cartridge for a period of 7<br>days under worst case conditions (37°C, 1400 RPM, 550 mL/min). Eluted heparin was<br>quantified in the PBS solution throughout the test, and fiber mat heparin activity was<br>measured at the conclusion of the 7-day study. | |
| Catheter Performance<br>Testing | Flow Test | Characterize flows and pressures through the catheter in<br>both straight and bent configurations over a range of pump<br>speeds in a recirculation loop using a blood<br>analog at 37°C |
| | Physical Requirements | Ensure markings are present on catheters following a<br>simulated soak for 2X the duration of use |
| | Kink Resistance | Distal top of the catheter was fixed to a 5cm radium wheel<br>and bent around the wheel 25 times. |
| | Antiseptic Resistance | Antiseptics were placed on the catheter materials including<br>glue joints and soaked for 24 hours at 37 °C. The antiseptics<br>used were:<br>• 10% povidone-iodine solution in water (i.e.Betadine®),<br>• 4% Chlorhexidine gluconate,<br>• 2% chlorhexidine gluconate in 70% isopropyl alcohol (i.e.<br>Chloraprep®),<br>• Bacitracin zinc ointment (i.e. Polysporin®) |
| | Leak Test (Pressure and<br>vacuum) | Catheters placed under vacuum and pressure representing<br>2X the maximum vacuum and pressure of the system for 10<br>min using DI water following simulated use. |
| | Force to Break Test | Measuring the force to break each distinct section of the<br>Catheter using a tensile strain rate of 20mm/min/mm of<br>gauge length. |
| | Corrosion Resistance | Catheters soaked in a sodium chloride solution for five<br>hours, immersed in boiling distilled water for 30 min and<br>then placed in an oven at 37°C for 48 hours and then<br>inspected for corrosion. |
| | Ambulation-Dislodgement | Ensure suture ring can secure the catheter during simulated<br>patient ambulation of axial, transverse, and upward forces<br>of 15N and torque each applied for 15 seconds |
| | Ambulation-Weight on<br>Catheter | Catheters placed in a water bath at 37°C and the extension<br>tubes were bent in all four directions 25 times to simulate<br>patient movement. |
| | Recirculation | Recirculation properties of the Catheter were tested in a<br>vena cava simulated flow circuit using both SVC and IVC<br>flow conditions at the minimum and maximum catheter<br>flow rates, as compared to an FDA-approved dialysis<br>catheter. |
| Biocompatibility<br>Testing | Biocompatibility testing was performed on all patient contacting components as specified in<br>ISO 10993-1 and Use of International Standard ISO 10993-1, "Biological evaluation of<br>medical devices - Part 1: Evaluation and testing within a risk management process" -<br>Guidance for Industry and Food and Drug Administration Staff. All components of the<br>system were tested in their final, finished, and sterilized (2X EtO) form. | |
| | Cytotoxicity (ISO-10993-5) | MEM Extraction Assay with L929 mammalian fibroblasts |
| | Sensitization (ISO 10993-10) | Guinea pig maximization test |
| | Irritation (ISO 109993-10) | Intracutaneous Reactivity |
| | Systemic Toxicity (ISO<br>10993-11) | Acute Systemic Injection Test<br>Material mediated rabbit pyrogen test<br>Sub Chronic Systemic Injection Test<br>Leverage from functional large animal study |
| | Genotoxicity (ISO 10993-3) | Bacterial Mutagenicity Test - Ames Assay<br>In Vitro Mouse Lymphoma Assay<br>In Vivo Mouse Bone Marrow Micronucleus Assay |
| | Hemocompatibility (ISO<br>10993-4/A1 & ASTM F756<br>& ASTM F2382-18 &<br>ASTM F2888-19) | Hemolysis; Saline Extract<br>Hemolysis; Direct Contact<br>Complement Activation - SC5b-9 Assay<br>Partial Thromboplastin Time (PTT) with Comparison<br>Article<br>Heparinized Blood Platelet and Leukocyte Count with<br>Comparison Article<br>Surface Morphology Assessment<br>In Vivo thrombogenicity; In Vivo animal study |
| | Implantation (ISO 10993-6) | In Vivo animal study |
| Shelf-Life Testing | Real-Time Aging | Device performance testing was completed using product<br>following 2-year real time aging to characterize any impacts<br>from aging on device performance or safety. The aged<br>products were used to confirm that the device met all of the<br>following product requirements: gas exchange, pump<br>characterization, reliability, heparin stability, catheter<br>performance. |
| | Accelerated Aging | Device performance testing was completed using 2-year<br>accelerated aged product to support shelf-life where real<br>time aging is not expected to impact device performance.<br>The aged products were used to confirm that the device met<br>all of the following product requirements: hemolysis, heat<br>loss/gain, heparin/uniformity and leachability, verification<br>by analysis, catheter recirculation, in vivo animal study. |
| Sterilization<br>Validation | Sterilization validation was performed in accordance with ANSI/AAMI/ISO 11135-1:2014/<br>AMD1:2018 including two sublethal exposures, four half-cycle exposures, cold<br>conditioning, and one full cycle exposure. An ethylene oxide (EO) sterilization validation<br>study was performed for the Hemolung family of sterile products to confirm a 10-6 sterility<br>assurance level (SAL) for the designated product load and acceptable levels of EO residuals.<br><br>(ISO 11135:2014, ISO 11138-1:2017, ISO 11138-2:2017, ISO 11737-1:2018, ISO 117737-<br>2: 2019, AAMI TIR No. 14 (2016), 15(2016), 16 (2017)) | |
| Packaging Validation | A packaging validation study was performed to evaluate the ability of each Hemolung<br>product sterile packaging configuration to maintain strength and integrity of the sterile<br>barriers' seals following simulated shipping and/or aging. Testing was conducted in<br>accordance with ISO 11607-1 Packaging for Terminally Sterilized Medical Devices - Part<br>1: Requirements for materials, sterile barrier systems, and packaging systems.<br><br>(ISTA 2A-2011, ISTA 3E:2017, ASTM D 4169-16, ASTM D 6653/D653M-13, ASTM<br>F1980-16. ASTM F2096-11 (2019), ISO 11607-1: 2019, ASTM F 88/F 88M-15) | |
| Hardware and<br>Software Verification | The Hemolung CR4 Controller's hardware and software was tested to verify functionality<br>and safety before clinical testing. The following functional areas were tested: interface with<br>disposable components, power system, generation, control, and monitoring of sweep gas<br>flow, CO2 monitoring, blood pump RPM control and monitoring, and blood flow<br>monitoring.<br><br>(ANSI/AMI/IEC 62304:2006 + AMDI: 2015) | |
| Electrical Safety and<br>Compliance Testing<br>and other safety-<br>related requirements | Electrical Safety and Compliance Testing was conducted for the Hemolung CR4 Controller<br>to verify functionality and safety before clinical testing. This testing verified compliance<br>with IEC 60601-1 electrical safety standard and IEC 60601-1-8 alarm system standard. This<br>testing consisted of verifying short circuit protection of all off-board connections and<br>operation of AC Power indicator. In addition, testing was conducted to verify detection of<br>air in blood lines.<br>In addition, the testing was used to verify the following functional areas: ventilation to<br>prevent battery damage, measurement of enclosure temperature, and label resistance to<br>bleach, soap and water.<br>(IEC 60601-1:2015+ AMI: 2012, IEC 60601-1-2:2014, IEC 60601-1-6:2010+A1: 2013, IEC<br>60601-1-8: 2006 + AMD1:2012) | |
| Electromagnetic<br>Compatibility | Electromagnetic Compatibility Testing was conducted for the Hemolung CR4 Controller to<br>verify conformance with IEC 60601-1-2 and collateral standards and conformance with<br>AIM 7351731.<br>(IEC 60601-1-2:2014, AIM 7351731, EN 55011:2009+A1:2010, IEC 61000-3-2:2014, IEC<br>61000-3-3: 2013. IEC 61000-4-2: 2008. IEC 61000-4-3:2010, IEC 61000-4-4:2012, IEC<br>61000-4-5:2014, IEC 61000-4-6:2013, IEC 61000-4-8:2019. IEC 61000-4-11:2010) | |
| Environmental<br>Verification | Environmental Verification Testing was conducted for the Hemolung CR4 Controller to<br>verify fluid ingress rating of IPX1, test shipping to ISTA 3E, and that the Corner drop /<br>Edge drop are in compliance with 60068-2-31. In addition, the following environmental<br>conditions were verified in the testing: 1) operating temperature and humidity from 10C to<br>35C at 20% to 90% non-condensing humidity, 2) storage temperature and humidity from -<br>20C to 50C at 15% to 95% non-condensing humidity, and 3) altitude requirements: maintain<br>sweep gas for of 10.0 SLPM ±0.3 SLPM at simulated altitudes up to 8000 ft (2500 m). | |
| Usability | Usability testing was conducted in accordance with IEC 62366-1:2015 to verify user<br>interface components. In the testing, critical care nurses were trained in the use of the<br>system and then after a decay period completed a summative test to validate all disposables-<br>related tasks (unpacking disposables, system priming, recirculation, catheter unpackaging,<br>patient connection, cartridge replacement, rinse back, and vacuum canister replacement). | |
{3}------------------------------------------------
{4}------------------------------------------------
# PERFORMANCE TESTING - ANIMAL
{5}------------------------------------------------
The objective of the in vivo Animal Study was to perform a safety and performance evaluation of the Hemolung system under simulated clinical use.
| Methods: | Eight (8) male calves were implanted with the Hemolung system. Calves were to be recovered<br>and survived on therapy for 7 days. Blood chemistry, hematology, fibrinogen, plasma free<br>hemoglobin, activated clotting time and hematocrit were regularly tested throughout the study.<br>Calves were routinely observed by veterinary staff and Subjective Objective Assessment Plan<br>examinations were performed by veterinarians. Clinical abnormalities were scored using an<br>abnormality grading scale adapted from the Common Terminology Criteria for Adverse<br>Events.<br>At the end of the therapy, the Hemolung system was shut down and the circuit was thoroughly<br>examined for clots/thrombi. The calves were humanely euthanized, and a detailed necropsy<br>was performed to examine implant site and organs for gross abnormalities, excising<br>representative samples for histopathology. Histopathology was performed on naïve and test<br>catheter implanted sections of the right jugular vein, lungs, heart, lymph nodes, adrenal glands,<br>liver, kidney, spleen and any other tissue with abnormal observations. |
|--------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Results: | There were no major clinical events. Therapy was discontinued early on two (2) of the eight<br>calves: one calf was taken off therapy on Day 1.7 post implant and terminated at Day 2 post<br>implant, and one calf was taken off therapy on Day 6.2 post implant but terminated as planned<br>on Day 7. |
| Conclusions: | There were no procedural complications directly related to the test article that put the calf's<br>health at risk or created any significant clinical health concerns. There were adjustments<br>required for the anticoagulation therapy as well as for routine maintenance and observation of<br>the catheter. There were no significant health concerns to the calves resulting from these<br>adjustments. Cartridge replacements were not required and cessation of therapy in two calves<br>was due to decreases in blood flow through the test article. There was no clinical evidence of<br>severe coagulation insufficiency in any of the calves. Based on the abnormality grading<br>system, none of the calves were found to have any severe abnormalities that required a grade 3<br>or higher. The two calves that were taken off therapy were removed due the low blood flow<br>alarms and were not removed for health reasons. Based on the daily PCV values none of the<br>calves were clinically anemic. There was no clinical evidence of hemolysis. Plasma free<br>hemoglobin (PFH) was increased in one calf but was found to be clinically insignificant as the<br>calf did not have evidence of anemia, low hematocrit or hyperbilirubinemia. No obvious<br>indications of excessive bleeding, inflammation, or infection were noted. No evidence of<br>infection was observed at the test Catheter insertion site, and no gross thrombi were noted at<br>the site of the insertion. Overall, the calves remained stable throughout the 7-day time point<br>and did not have any clinical emergencies that threatened the life of the calf. |
# SUMMARY OF CLINICAL INFORMATION
### Hemolung Clinical Performance
The clinical performance data provided to support the de novo request included data from 234 patients treated with Hemolung therapies from prospective clinical trials, real-world use, and the Hemolung Post-market Registry. Patients were systematically evaluated for inclusion in the effectiveness analysis if all 4 of the following criteria were met: 1) Original patient level efficacy data was available, 2) Pre-Hemolung pH and PaCO2 data was available, 3) Data coinciding with at least one additional time point during the first 35 hours after commencement of Hemolung therapy was available, and 4) the patient received Hemolung therapy for at least 6 hours. The two primary clinical outcomes used to evaluate effectiveness of Hemolung therapy were:
- 1) Correction of refractory hypercapnia and respiratory acidosis
{6}------------------------------------------------
#### OR
- 2) De-escalation of mechanical ventilatory support while preventing respiratory acidosis
Data was stratified based on baseline (Pre-Hemolung) pH as follows: Acidotic patients (pH < 7.35) were analyzed for correction of refractory hypercapnia and respiratory acidosis; Nonacidotic patients (pH ≥ 7.35) were analyzed for de-escalation of mechanical ventilatory support while preventing respiratory acidosis. An acute timepoint (first day of Hemolung therapy) was chosen to analyze Hemolung clinical effectiveness as the physiologic response from Hemolung therapy is observed within hours of initiating treatment. The acute time frame also eliminates longer-term confounding effects from underlying illness, additional critical care support technologies and patient-specific titration of Hemolung therapy.
The benefits of Hemolung therapy from this clinical data analysis are summarized as follows:
- . Extra-corporeal carbon dioxide removal using the Hemolung demonstrated clinically and statistically significant correction of respiratory acidosis and de-escalation of mechanical ventilatory support in patients who have not been on ventilatory support and primarily received standard of care therapy until the Hemolung therapy was initiated
- 91% of Hemolung patients demonstrated a clinically beneficial response, where there was . a correction in respiratory acidosis after one day of Hemolung therapy.
- . 92% of non-invasively ventilated (NIV) patients avoided intubation and invasive mechanical ventilation (IMV) with Hemolung therapy. For acidotic patients failing NIV (n=43). Hemolung therapy resulted in correction of respiratory acidosis (pH =7.24 -> pH=7.36)
# Hemolung Clinical Safety
Assessment of Hemolung safety was supported through analysis of clinical complication data from 1,034 patients that received Hemolung therapy Safety data was derived from four primary sources: 1) monitored and independently adjudicated adverse event data from prospective clinical trials, 2) US Expanded Access or Emergency Use Authorization data collection and surveillance, 3) the OUS post-market Hemolung Registry, and 4) OUS post-market surveillance. The Hemolung Registry Form was provided to treating physicians for all commercial therapies outside the US. For commercial therapies completed without a returned Registry Form, ALung post-market surveillance procedures were used to collect safety data directly from the treating physician where possible.
# Therapy-Related Complications
A total of 172 therapy-related complications were reported as being definitely, probably, or possibly related to use of the Hemolung, or with unknown determination of causality (Table 1). The total number of patients who experienced therapy-related complications was 121 of the 1,034 treatments (i.e., 172 complications occurred in 121 patients). Of the 172-total therapyrelated complications, 66 had no patient impact, 96 required medical intervention and 10 resulted in death. There were no reported unanticipated adverse device events.
{7}------------------------------------------------
| Therapy-Related Complications | TOTAL<br>(n=1034 pts) |
|------------------------------------------------|-----------------------|
| Bleeding (n, incidence) | 72 (7.0%) |
| Hemolysis (n, incidence) | 27 (2.6%) |
| Thrombocytopenia (n, incidence) | 22 (2.1%) |
| Thrombosis/coagulation disorder (n, incidence) | 13 (1.3%) |
| Hemodynamic instability (n, incidence) | 16 (1.6%) |
| Other (n, incidence) | 22 (2.1%) |
| TOTAL # OF THERAPY-RELATED COMPLICATIONS | 172 |
| Required medical intervention | 96 |
| Resulted in death | 10 |
| No patient impact | 66 |
Table 1. Summary of Hemolung therapy-related complications. Bleeding includes complications categorized as bleeding or anemia. Thrombosis/ coagulation disorder includes complications categorized as coagulation disorder, disseminated intravascular coagulopathy, pulmonary embolism, and vein thrombosis. Hemodynamic instability includes complications categorized as cardiac arrythmia, cardiac arrest, hemodynamic instability, hypovolemia, septic shock, and shock.
# Procedural, Operational and Component-Related Complications
A total of 128 procedural, operational, and component-related complications were reported (Table 2). Of the 128 total complications, 116 had no patient impact, 8 required medical intervention and 4 resulted in death.
| | TOTAL<br>(n=1,034 pts) |
|--------------------------------------------------|------------------------|
| Procedural-related complications (n, incidence) | 62 (6.0%) |
| Required medical intervention | 7 |
| Resulted in death | 4 |
| No patient impact | 51 |
| Operational-related complications (n, incidence) | 33 (3.2%) |
| Required medical intervention | 0 |
| Resulted in death | 0 |
| No patient impact | 33 |
| Component-related complications (n, incidence) | 33 (3.2%) |
| Required medical intervention | 1 |
| Resulted in death | 0 |
| No patient impact | 32 |
{8}------------------------------------------------
Table 2. Summary of Hemolung procedural-related, operational-related and component- related complications
### Pediatric Extrapolation
In this De Novo request, existing clinical data were not leveraged to support the use of the device in a pediatric patient population.
### LABELING
The Hemolung Respiratory Assist System labeling consists of Instructions for Use that includes a detailed summary of the non-clinical and in vivo evaluations pertinent to use of the device and accessories in the circuit. The Instructions for Use also includes adequate instructions with respect to circuit setup, performance characteristics with respect to compatibility among different devices and accessories in the circuit, and maintenance during a procedure.
### RISKS TO HEALTH
The table below identifies the risks to health that may be associated with use of an extracorporeal system for carbon dioxide removal and the measures necessary to mitigate these risks.
| Identified Risks to Health | Mitigation Measures |
|---------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|
| Bleeding, thrombocytopenia, hemolysis, thrombosis | In vivo evaluation<br>Non-clinical performance testing<br>Labeling |
| Infection | In vivo evaluation<br>Sterility<br>Shelf-life testing<br>Labeling |
| Adverse tissue and/or<br>hematologic reaction | In vivo evaluation<br>Biocompatibility<br>Labeling |
| Mechanical failure | In vivo evaluation<br>Non-clinical performance testing<br>Labeling<br>Software validation, verification, and hazard<br>analysis |
| Hemodynamic instability | In vivo evaluation<br>Non-clinical performance testing<br>Labeling |
| Hypothermia | In vivo evaluation<br>Non-clinical performance testing<br>Labeling |
| Mechanical injury to access<br>vessels | In vivo evaluation<br>Non-clinical performance testing<br>Labeling |
{9}------------------------------------------------
| Inadequate gas exchange | In vivo evaluation<br>Non-clinical performance testing<br>Labeling |
|-------------------------|--------------------------------------------------------------------|
| Hemodilution | In vivo evaluation<br>Non-clinical performance testing<br>Labeling |
| Gas embolism | In vivo evaluation<br>Non-clinical performance testing<br>Labeling |
### SPECIAL CONTROLS
In combination with the general controls of the FD&C Act, the extracorporeal system for carbon dioxide removal is subject to the following special controls:
- 1) In vivo evaluation, which may include animal testing and clinical data, of the devices and accessories in the circuit must demonstrate their performance over the intended duration of use, including a detailed summary of the in vivo evaluation pertinent to the use of the devices and accessories to demonstrate their effectiveness
- 2) The technological characteristics of the device must ensure that the geometry and design parameters are consistent with the intended use, and that the devices and accessories in the circuit are compatible
- 3) Non-clinical performance testing of the devices and accessories in the circuit must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
- a. Mechanical integrity:
- b. Durability; and
- c. Reliability.
- 4) All patient contacting components of the device must be demonstrated to be biocompatible.
- 5) Performance testing must demonstrate the electrical safety and electromagnetic compatibility (EMC) of any electrical components.
- 6) Software validation, verification, and hazard analysis must be performed.
- 7) Performance testing must demonstrate the sterility of all patient-contacting components.
- 8) Performance testing must support the shelf life of the device by demonstrating continued sterility and device functionality over the identified shelf life.
- 9) Labeling must include the following:
- a. A detailed summary of the non-clinical and in vivo evaluations pertinent to use of the device and accessories in the circuit; and
- b. adequate instructions with respect to circuit setup, performance characteristics with respect to compatibility among different devices and accessories in the circuit, and maintenance during a procedure; and
- c. A shelf life.
# BENEFIT-RISK DETERMINATION
{10}------------------------------------------------
The benefits and risks of the device are based on nonclinical bench and animal studies as well as clinical data collected in clinical studies described above and real-world evidence.
Hemolung benefits were consistent across primary diagnoses and data sources, demonstrating generalizability of the clinical benefits for numerous acute respiratory failure etiologies. The two primary clinical benefits that were observed were 1) Correction of refractory hypercapnia and respiratory acidosis and 2) De-escalation of mechanical ventilatory support while preventing respiratory acidosis. 91% of Hemolung patients demonstrated a clinically beneficial response to Hemolung therapy. 92% of non-invasively ventilated (NIV) patients avoided intubation and invasive mechanical ventilation (IMV) with Hemolung therapy. The observed benefits are clinically meaningful enough to outweigh the potential complications related to device therapy.
Complications that were Hemolung therapy-related, procedural-related, operational-related, and component-related, occurred with low incidences, most often did not result in any patient harm, and rarely resulted in permanent patient harm or death. Complications from Hemolung therapy included bleeding, thrombosis/ coagulation disorder, and hemodynamic instability. Complications from procedures referred to insertion of the Hemolung Catheter, which is performed using a percutaneous technique. Complications during normal operational use of the Hemolung included circuit thrombosis, low blood flow due to patient/catheter positioning or hypovolemia, air in circuit, catheter dislodgement, and incorrect system setup. These complications are all expected risks of extracorporeal blood circulation. Complications related to Hemolung component failure or malfunction were primarily due to hardware/software issues with the Hemolung Controller, the majority of which have since been resolved through software bug fixes and hardware improvements. All risks from the Hemolung therapy are characterized and mitigated through non-clinical testing, including biocompatibility, in vivo animal studies, sterilization and packaging validation, mechanical reliability, and functional performance and safety testing. In addition, all risks from Hemolung therapy are characterized in ALune's Risk Management System and are appropriately mitigated through special controls. These risks are clinically acceptable and consistent with other technologically similar devices. Overall, the clinical benefits of the Hemolung Respiratory Assist System outweigh the probable risks.
# PATIENT PERSPECTIVES
This submission did not include specific information on patient perspectives for this device.
### BENEFIT/RISK CONCLUSION
In conclusion, given the available information above, for the following indication statement:
The Hemolung Respiratory Assist System is indicated for respiratory support that provides extracorporeal carbon dioxide (CO2) removal from the patient's blood for up to 5 days in adults with acute, reversible respiratory failure for whom ventilation of CO2 cannot be adequately or safely achieved using other available treatment options and continued clinical deterioration is expected.
{11}------------------------------------------------
The probable benefits outweigh the probable risks for the Hemolung Respiratory Assist System. The device provides benefits, and the risks can be mitigated by the use of general controls and the identified special controls.
### CONCLUSION
The De Novo request for the Hemolung Respiratory Assist System is granted, and the device is classified as follows:
Product Code: QOH Device Type: Cardiovascular Surgical Devices Regulation Number: 21 CFR 870.4150 Class: II
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.