V. Mueller Cosgrove Flex Clamps, Model Number-CV1033, CV1061, CV1133, CV1161, CV1186

K210324 · Carefusion 2200, Inc. · DXC · Mar 4, 2021 · Cardiovascular

Device Facts

Record IDK210324
Device NameV. Mueller Cosgrove Flex Clamps, Model Number-CV1033, CV1061, CV1133, CV1161, CV1186
ApplicantCarefusion 2200, Inc.
Product CodeDXC · Cardiovascular
Decision DateMar 4, 2021
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.4450
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The V. Mueller Cosgrove™ Flex Clamps are indicated to occlude a blood vessel temporarily. Used in pulmonary and gastrointestinal procedures and can be used to clamp over indwelling catheters. Also used in minimally invasive and standard open cardiovascular procedures for temporary occlusion of a blood vessel.

Device Story

V. Mueller Cosgrove Flex Clamps are reusable surgical instruments designed for temporary vascular occlusion. The device features a ring handle, a flexible shaft, and a jaw. The flexible shaft allows the surgeon to bend the instrument after clamping, improving visualization of the surgical site. Used by surgeons in cardiovascular, pulmonary, and gastrointestinal procedures. The device is supplied non-sterile and requires cleaning and sterilization by the user. The current design incorporates a modified wire configuration in the cable compared to predicate versions. The device functions as a mechanical clamp; it does not involve electronic, software, or algorithmic components.

Clinical Evidence

No clinical tests were conducted for this submission. Substantial equivalence is supported by non-clinical bench testing.

Technological Characteristics

Reusable vascular clamp constructed of stainless steel. Features a ring handle, flexible shaft, and jaw (available in 33mm, 61mm, 86mm lengths). Mechanical operation. Sterilization via pre-vacuum steam or ethylene oxide. Complies with ISO 10993 for biocompatibility and various AAMI/ISO standards for cleaning and sterilization.

Indications for Use

Indicated for temporary occlusion of blood vessels in patients undergoing pulmonary, gastrointestinal, or cardiovascular procedures (minimally invasive or open); also indicated for clamping over indwelling catheters.

Regulatory Classification

Identification

A vascular clamp is a surgical instrument used to occlude a blood vessel temporarily.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION". March 4, 2021 CareFusion 2200, Inc. Jane Weber Regulatory Affairs Sr. Manager 75 North Fairway Drive Vernon Hills, Illinois 60061 Re: K210324 Trade/Device Name: V. Mueller Cosgrove Flex Clamps, Model Number-CV1033, CV1061, CV1133, CV1161, CV1186 Regulation Number: 21 CFR 870.4450 Regulation Name: Vascular Clamp Regulatory Class: Class II Product Code: DXC Dated: February 2, 2021 Received: February 4, 2021 Dear Jane Weber: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. Rachel Neubrander Assistant Director DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K210324 Device Name V. Mueller Cosgrove Flex Clamps Indications for Use (Describe) The V. Mueller Cosgrove™ Flex Clamps are indicated to occlude a blood vessel temporarily. Used in pulmonary and gastrointestinal procedures and can be used to clamp over indwelling catheters. Also used in minimally invasive and standard open cardiovascular procedures for temporary occlusion of a blood vessel. | Type of Use (Select one or both, as applicable) | | |--------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------| | <span style="font-size: 10pt;"> </span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size: 10pt;"> </span> Over-The-Counter Use (21 CFR 801 Subpart C) | ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the CareFusion logo. The logo consists of an orange circle with a white shield inside, and the text "CareFusion" in gray. Below the text, in smaller gray font, it says "has joined BD". # 510(k) SUMMARY – K210324 A summary of 510(k) safety and effectiveness information in accordance with 21 CFR 807.92. | SUBMITTER INFORMATION | | |---------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Name | CareFusion 2200 Inc. | | Address | 75 North Fairway Drive<br>Vernon Hills, IL 60061 | | Phone number | (847) 362-8094 | | Establishment Registration<br>Number | 1423507 | | Name of contact person | Jane Weber | | Date prepared | March 4, 2021 | | DEVICE INFORMATION | | | Trade or proprietary name | V. Mueller Cosgrove™ Flex Clamp | | Common or usual name | Vascular Clamp | | Classification name | Vascular Clamp | | Classification panel | 74 Cardiovascular | | Regulation | Class II per 21CFR 870.4450, Product code DXC | | Product Code(s) | CV1033, CV1061, CV1133, CV1161, CV1186 | | Legally marketed device(s) to<br>which equivalence is claimed | Cosgrove Flex Clamp - K974769<br>Cosgrove Flex Clamp Quick-Bend™ - K991589 | | Reason for 510(k) submission | Design and labeling modifications | | Device description | Each V. Mueller Cosgrove™ Flex Clamp device consists of a ring<br>handle, flexible shaft and jaw. The design of the flexible shaft<br>allows the end user to bend the shaft following clamping of the<br>blood vessel providing the user a clear field of visualization to the<br>surgical site.<br><br>The proposed device and the predicate device have a similar design, the same materials and the same manufacturing and fabrication methods/processes. The main difference between the subject device and the predicate device is a change to the wire configuration of the cable. In addition, the cleaning and sterilization instructions in the instructions for use were updated in accordance with current guidance. These changes do not raise different questions of safety or effectiveness. | | Intended use of the device | To be used to occlude a blood vessel temporarily | {4}------------------------------------------------ ## SUMMARY OF THE TECHNOLOGICAL CHARACTERISTICS OF THE DEVICE COMPARED TO THE PREDICATE DEVICE | Characteristic | Predicate Devices | New Device | |--------------------------|-----------------------------------------------------|------------------------------------| | Type of Device | Reusable | Reusable | | How Supplied | Non-Sterile | Non-Sterile | | Sterilization Modalities | Pre-vacuum Steam<br>Gravity Steam<br>Ethylene Oxide | Pre-vacuum Steam<br>Ethylene Oxide | | Insert Jaw Lengths | 33mm, 61mm, 86mm | 33mm, 61mm, 86mm | | Materials | Stainless Steel | Stainless Steel | ### PERFORMANCE DATA ### SUMMARY OF NON-CLINICAL TESTS CONDUCTED FOR DETERMINATION OF SUBSTANTIAL EQUIVALENCE | Characteristic | Standard / Test / FDA Guidance | Results Summary | |------------------------------------------------|----------------------------------------------------------------------------|-----------------| | Instrument strength | Strength Tests, Life Cycle Test | PASS | | Device must be reusable | Cleaning and Sterilization Tests | PASS | | Device is able to be cleaned<br>and sterilized | AAMI TIR12, AAMI TIR30, ANSI AAMI ST79,<br>ISO 11138, ISO 17664, ISO 17665 | PASS | | Device materials are<br>biocompatible | ISO 10993 | PASS | SUMMARY OF CLINICAL TESTS CONDUCTED FOR DETERMINATION OF SUBSTANTIAL EQUIVALENCE AND/OR OF CLINICAL INFORMATION N/A - No clinical tests were conducted for this submission. CONCLUSIONS DRAWN FROM NON-CLINICAL AND CLINICAL DATA The results of the non-clinical tests demonstrate the V. Mueller Cosgrove™ Flex Clamps meet all performance requirements and are substantially equivalent to the predicate devices.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...