Dual Stage Venous Cannulae

K253671 · Sorin Group Italia S.R.L. · DWF · Jan 30, 2026 · Cardiovascular

Device Facts

Record IDK253671
Device NameDual Stage Venous Cannulae
ApplicantSorin Group Italia S.R.L.
Product CodeDWF · Cardiovascular
Decision DateJan 30, 2026
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.4210
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Dual Stage Venous Return Cannula is indicated for single tube venous drainage from the right atrium and vena cava during cardiopulmonary bypass surgery for up to six hours.

Device Story

Dual Stage Venous Return Cannulae are single-use, sterile, non-pyrogenic PVC polymer tubes used during cardiopulmonary bypass surgery. The device features a lighthouse-shaped distal tip and multiple perforations at various stages to facilitate venous blood drainage from the right atrium and vena cava. Some models include a malleable obturator to assist in vessel placement, which is removed once positioned. The proximal end is clear and non-reinforced to allow for clamping and may include a pre-mounted 1/2" barbed connector for integration with standard bypass tubing. The device is operated by a surgeon in a clinical setting. By providing a pathway for venous return, the cannula supports the extracorporeal circulation process, benefiting patients by enabling necessary cardiac surgical procedures.

Clinical Evidence

No clinical data was provided. Substantial equivalence is supported by non-clinical bench testing, including visual inspection, connector pull tests, clamp tests, blood pathway integrity, flow rate/pressure drop, kink resistance, collapse resistance, pull strength, blood trauma characterization, and biocompatibility, all performed in accordance with ISO 18193:2021.

Technological Characteristics

Materials: PVC polymer, wire reinforcement. Energy: None (passive). Dimensions: French sizes 40-50 Fr (proximal) and 32-36 Fr (distal). Connectivity: Standalone, compatible with 1/2" bypass tubing. Sterilization: Ethylene oxide. Single-use, non-pyrogenic fluid path.

Indications for Use

Indicated for patients undergoing cardiopulmonary bypass surgery requiring single tube venous drainage from the right atrium and vena cava for up to six hours.

Regulatory Classification

Identification

A cardiopulmonary bypass vascular catheter, cannula, or tubing is a device used in cardiopulmonary surgery to cannulate the vessels, perfuse the coronary arteries, and to interconnect the catheters and cannulas with an oxygenator. The device includes accessory bypass equipment.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD &amp; DRUG ADMINISTRATION January 30, 2026 Sorin Group Italia S.R.L. Luigi Vecchi Director RA Via Statale 12 Nord, 86 Mirandola, MO 41037 Italy Re: K253671 Trade/Device Name: Dual Stage Venous Cannulae Regulation Number: 21 CFR 870.4210 Regulation Name: Cardiopulmonary bypass vascular catheter, cannula, or tubing Regulatory Class: Class II Product Code: DWF Dated: October 19, 2025 Received: November 21, 2025 Dear Luigi Vecchi: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1} K253671 - Luigi Vecchi Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {2} K253671 - Luigi Vecchi Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Meaghan Erlewein -S For Nicole Gillette Assistant Director DHT2B: Division of Circulatory Support, Structural, and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} | DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration **Indications for Use** | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. | | --- | --- | 510(k) Number (if known)<br/>K253671 Device Name<br/>Dual Stage Venous Return Cannulae Indications for Use (Describe)<br/>The Dual Stage Venous Return Cannula is indicated for single tube venous drainage from the right atrium and vena cava during cardiopulmonary bypass surgery for up to six hours. Type of Use (Select one or both, as applicable)<br/>☑ Prescription Use (Part 21 CFR 801 Subpart D)<br/>☐ Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. **“DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.”** The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services<br/>Food and Drug Administration<br/>Office of Chief Information Officer<br/>Paperwork Reduction Act (PRA) Staff<br/>PRAStaff@fda.hhs.gov “An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.” FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 613-6740 {4} LivaNova Health innovation that matters # 510(k) Summary (in accordance with 21 CFR 807.92) 510(k) Number: K253671 ## I. Applicant Information Applicant: SORIN GROUP ITALIA S.R.L. Via Statale 12 Nord, 86 Mirandola, MO 41047 Italy Contact Person: Luigi Vecchi Regulatory Affairs Director Tel: +39 0535 29957 e-mail: luigi.vecchi@livanova.com Application Correspondent: SORIN GROUP ITALIA S.R.L. Via Statale 12 Nord, 86 Mirandola, MO 41047 Italy Contact Person: Luigi Vecchi Regulatory Affairs Director Tel: +39 0535 29957 e-mail: luigi.vecchi@livanova.com Date Prepared: January 23th 2026 ## II. Subject Device Identification Device Trade Name: Dual Stage Venous Return Cannulae Classification Name: Cardiopulmonary bypass vascular catheter, cannula, or tubing Regulation Number: 21 CFR 870.4210 Product Code: DWF Classification: Class II Classification Panel: Cardiovascular ## III. Predicate Device The Dual Stage Venous Return Cannulae are substantially equivalent to the following cleared predicate device. Both modified and unmodified models have the same fundamental scientific technology and intended use: | Venous cannulae | 510(k) Number: | K943934 | | --- | --- | --- | | | Device Trade Name: | Venous Cannulae | | | Classification Name: | Cardiopulmonary bypass vascular catheter, cannula, or tubing | | | Regulation Number: | 21 CFR 870.4210 | | | Product Code: | DWF | | | Classification: | Class II | | | Classification Panel: | Cardiovascular | Traditional 510(k) – Dual Stage Venous Return Cannulae {5} LivaNova Health innovation that matters ## IV. Device Description Dual Stage Venous Return Cannulae are single-use, non-toxic, non-pyrogenic fluid path devices and supplied sterile and individually packaged. The devices are composed of the cannula: an open lumen PVC polymer tube incorporating wire reinforcement in distal section. In some models is present a malleable obturator inserted into the cannula to allow the placement of the cannula along the vein. Once the cannula is positioned onto the vessel the surgeon removes the obturator from the cannula tubing and connects it to the venous line. The distal end of the cannula has a lighthouse-shaped tip, allowing the venous blood to flow from the vessel into the cannula, in addition the distal sections of the cannula are perforated with multiple holes at multiple stages to better allow the fluid flows from outside into the cannula body. The clear proximal section is not reinforced to allow clamping; the proximal end typically does not have any pre-mounted connector and can accommodate a barbed connector for standard cardiopulmonary bypass tubing (½" diameter). In some models a pre-mounted 1/2" barbed connector is present. The following cannula models are available: | Device trade name | Model | French size (Outside diameter) | Characteristics | | --- | --- | --- | --- | | Dual Stage Venous Cannulae | RDS-61040 | Proximal: 40 Fr Distal: 32 Fr. | Without connector | | Dual Stage Venous Cannulae | RDS-61034 | Proximal: 46 Fr Distal: 34 Fr. | Without connector | | Dual Stage Venous Cannulae | RDS-61046 | Proximal: 46 Fr Distal: 36 Fr. | Without connector | | Dual Stage Venous Cannulae | RDS-61050 | Proximal: 50 Fr Distal: 36 Fr. | Without connector | | Dual Stage Venous Cannulae | RDS-61140 | Proximal: 40 Fr Distal: 32 Fr. | With connector | | Dual Stage Venous Cannulae | RDS-61134 | Proximal: 46 Fr Distal: 34 Fr. | With connector | | Dual Stage Venous Cannulae | RDS-61146 | Proximal: 46 Fr Distal: 36 Fr. | With connector | | Dual Stage Venous Cannulae | RDS-61150 | Proximal: 50 Fr Distal: 36 Fr. | With connector | The Dual Stage Venous Return Cannulae are the modified version of the disposables currently marketed under K943934. Both modified and unmodified cannulae models are recommended for use as Venous cannula during cardiopulmonary bypass up to six hours. ## V. Indications for Use The Dual Stage Venous Return Cannula is indicated for single tube venous drainage from the right atrium and vena cava during cardiopulmonary bypass surgery for up to six hours. Traditional 510(k) – Dual Stage Venous Return Cannulae {6} LivaNova Health innovation that matters ## VI. Summary of Technical Characteristics The Dual Stage Venous Return Cannulae have the same fundamental technological characteristics, principles of operation and control mechanisms as the unmodified devices. The devices are ethylene oxide sterilized and have a non-pyrogenic fluid path. They are for single use only. ## VII. Non-Clinical Performance Data Sorin Group Italia S.r.l. has conducted extensive verification and validation testing of the Dual Stage Venous Return Cannulae; specifically, the following tests were carried out according to standard ISO 18193 ed. 2021: - Visual inspection - Connector testing 180° pull test - Cannula clamp test - Blood pathway integrity - Flow rate and pressure drop through cannulae - Cannula kink test - Label legibility - Drainage cannula collapse resistance - Cannula integrity - Pull strength - Blood trauma characterization test - Biocompatibility The devices passed all the testing in accordance with standards. ## VIII. Clinical Performance Data No clinical testing was conducted in support of the Dual Stage Venous Return Cannulae, the indications for use and technical characteristics are equivalent to those of the predicate devices, which have been on the market for several years with proven safety and efficacy of use. The non-clinical testing summarized in this submission supports the substantial equivalence of the subject devices with the predicate devices when used according to their intended use. ## IX. Statement of Substantial Equivalence Based on equivalent intended use and technological the Dual Stage Venous Return Cannulae can be deemed to be substantially equivalent to their predicate devices: - the unmodified Venous cannulae, cleared under K943934; The Dual Stage Venous Return Cannulae, as designed and manufactured, do not raise new questions regarding safety and effectiveness as compared to their predicate devices and are determined to be substantially equivalent to their predicate devices listed above. Traditional 510(k) – Dual Stage Venous Return Cannulae
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...