K992657 · International Biophysics Corp. · DWD · Feb 18, 2000 · Cardiovascular
Device Facts
Record ID
K992657
Device Name
IBC VRV, MODEL 1350
Applicant
International Biophysics Corp.
Product Code
DWD · Cardiovascular
Decision Date
Feb 18, 2000
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.4430
Device Class
Class 2
Indications for Use
The IBC VRV is indicated for use as an intracardiac suction control valve for left ventricular venting during cardiopulmonary bypass surgery (up to six hours).
Device Story
IBC VRV is a single-use, disposable vacuum and pressure relief check valve used in suction lines during open-heart surgery for left ventricular venting. Operated by perfusionists within a perfusion circuit, the device manages blood recovery line parameters. It features a polycarbonate body and a curved duckbill check valve design to reduce flow resistance and hemolysis compared to predicate devices. The device provides vacuum relief and over-pressure venting to prevent complications during pump failure. It is used in clinical settings; output is observed by the perfusionist to ensure proper venting and suction control. Benefits include improved visibility of the vacuum relief point and reduced hemolysis, enhancing patient safety during bypass procedures.
Clinical Evidence
Bench testing only. Comparative dynamic circuit testing performed on the IBC VRV and Omni RLV-2100"B" using fresh bovine blood (45% hematocrit) at 2 L/min flow. Metrics included vacuum relief rate, pressure drop, and hemolysis over 6 hours. Reverse-flow and over-pressure venting were tested with air to simulate pump failure. Results demonstrated substantial equivalence in performance and safety profiles.
Technological Characteristics
Single-use, disposable check valve. Materials: Polycarbonate body. Sensing/Actuation: Mechanical vacuum and pressure relief via curved duckbill check valve. Dimensions/Form Factor: Inline suction line component. Sterilization: Standard IBC sterilization process. Biocompatibility: Compliant with tripartite ISO standards and FDA 1995 modified matrix. No software or electronic components.
Indications for Use
Indicated for patients undergoing cardiopulmonary bypass surgery requiring left ventricular venting via an intracardiac suction control valve for up to six hours.
Regulatory Classification
Identification
A cardiopulmonary bypass intracardiac suction control is a device which provides the vacuum and control for a cardiotomy return sucker.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
Predicate Devices
Omni Model RLV-2100"B" VRV valve
Related Devices
K992964 — IBC VENT VALVE, MODEL 1400 · International Biophysics Corp. · Nov 19, 1999
K992785 — IBC FLOGUARD, MODEL 6050 · International Biophysics Corp. · Jan 14, 2000
K982406 — COBE VACUUM RELIEF CHECK VALVE · Cobe Cardiovascular, Inc. · Oct 7, 1998
K960937 — SUCTION SAFETY DEVICE · R D Intl. · Jun 5, 1996
K964391 — CHASE LEFT VENT SAFETY VALVE · Chase Medical, Inc. · May 13, 1997
Submission Summary (Full Text)
{0}------------------------------------------------
K992657
## EXECUTIVE SUMMARY
The IBC VRV Valve is a single use, disposable, vacuum and pressure relief, check valve that is intended for use during open-heart surgery for left ventricular venting. It is sometimes commonly referred to as a suction control valve for left ventricular venting. The device is designed for use in a suction line and must be used in accordance with the operating parameters of that blood recovery line. The fully assembled IBC VRV Valve is geometrically similar and clinically the same to the Omni Model RLV-2100"B" VRV valve, which is the predicate device for purposes of this 510(k) submission. The operating principles of the two devices are identical.
The differences between the IBC VRV and the predicate device are found in the material selection and specific dimensions of the component parts neither of which effects safety or effectiveness of the product. The first of these differences is the choice of polymer for molding the plastic components. The main flow through body of the IBC VRV is made of polycarbonate and the Omni RLV-2100"B" is made of ABS. The polycarbonate is clear, has higher impact resistance, greater tensile and compression strength and superior chemical. Upon review of the predicate device, we found the inability to see the interior of the device made it difficult for the user to identify the vacuum relief point. The use of polycarbonate eliminated that problem. Secondly, the straight duck bill check valve in the predicate device represents a substantial resistance to flow which contributes significantly to hemolysis. The unique curved design of the duckbill check valve in the IBC VRV substantially reduces flow resistance and hemolysis.
The IBC VRV (sterile, 6 month aged, environmentally conditioned) was subjected to a series of dynamic tests in a side by side comparison with the Omni RLV-2100"B". The vacuum relief rate for both products was measured in a dynamic circuit flowing blood at 2 liters per minute. The pressure drop of the check valves for both products were measured also at 2 L.P.M. flow with blood. Hemolysis was measured in a simulated clinical circuit at 2 L.P.M. for 6 hours using fresh bovine blood at 45% hematocrit. The reverse-flow and over-pressure vent was tested with air for both products to simulate reverse flow pump failure mode. Comparison of the IBC VRV to the Omni RLV-2100"B" demonstrates that the two products are substantially equivalent when used clinically.
The IBC VRV is manufactured in a Class 100,000 clean room. The Bioburden prior to sterilization is extremely low and comparable to other 510(k) listed products manufactured by IBC. The device will be packaged and sterilized for single use using the same packaging and sterilization as other IBC products. Additionally, the IBC VRV will be packaged in bulk form and non-sterile for the Custom Perfusion Pack market. The materials used to manufacture the IBC VRV are non-toxic using the tripartite biocompatible ISO standards and the FDA modified matrix of 1995.
A perfusionist substituting an IBC VRV for an Omni RLV-2100"B" in his or her perfusion circuit will be unable to distinguish between the two valves functionally.
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the logo is an abstract symbol that resembles an eagle or bird in flight, composed of three curved lines.
## Public Health Service
## FEB 1 8 2000
Mr. H. David Shockley,Jr. President International Biophysics Corporation 4020 South Industrial Dr. Suite 160 Austin, Tx 78744
Re: K992657/S1 IBC VRV, Model 1350 Regulatory Class: II Product Code: DWD Dated: November 19, 1999 Received: November 22, 1999
Dear Mr. Shockley:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent [(for the indications for use stated in the enclosure)] to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of The general controls provisions of the Act include the Act. requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval) it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General (QS) regulation (21 CFR Part 820) and that, through periodic QS inspections, FDA will verify such assumptions. ' Failure to comply with the GMP regulation may result in regulatory action. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to continue marketing your device as described in your 510(k) premarket notification. The FDA finding
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
{2}------------------------------------------------
Page 2 - Mr. H. David Shockley, Jr.
of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at their toll free number (800) 638-2041 or at (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Celia M. Witten, Ph.D., M.D.
Acting Director
Acting Director Division of Cardiovascular Respiratory and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
{3}------------------------------------------------
## INDICATIONS FOR USE
K942657 510(k) Number:
Device Name: IBC VRV
Indications for use:
The IBC VRV is indicated for use as an intracardiac suction control valve for left ventricular venting during cardiopulmonary bypass surgery (up to six hours).
(PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)
Bose R. Lamperle
vision Sign-Off) Division of Cardiovascular, Respiratory, and Neurological Devices
510(k) Number K992657
or
Prescription Use (Per 21 CFR 801.109) Over-The-Counter-Use_
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.