K953901 · 3M Health Care, Sarns · DWB · Apr 19, 1996 · Cardiovascular
Device Facts
Record ID
K953901
Device Name
SARNS 800 ROLLER PUMP
Applicant
3M Health Care, Sarns
Product Code
DWB · Cardiovascular
Decision Date
Apr 19, 1996
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.4370
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Sarns 8000 Roller Pump is indicated for use in extracorporeal circulation of blood for arterial regional perfusion, and cardiopulmonary bypass procedures only, when used by a qualified perfusionist who is experienced in the operation of Sarns or similar equipment.
Device Story
Sarns 8000 Roller Pump is a microprocessor-controlled, 2-roller peristaltic pump; provides extracorporeal blood circulation for cardiopulmonary bypass and arterial regional perfusion. Operates at speeds up to 250 RPM; achieves flows up to 9.99 L/min depending on tubing. Used as stand-alone or integrated with Sarns 8000 Modular Perfusion System. Operated by qualified perfusionists in clinical settings. Device design incorporates SMT circuit boards and updated software for I/O address remapping. Provides mechanical pumping action to maintain blood flow during procedures; assists clinical decision-making by enabling precise control of perfusion parameters.
Clinical Evidence
Bench testing only. No clinical data provided. Electromagnetic Compatibility (EMC) testing performed to ensure equivalent performance to predicate.
Technological Characteristics
Microprocessor-controlled 2-roller peristaltic pump. Components utilize Surface Mount Technology (SMT) on circuit boards. Power supply included. Stand-alone or modular integration. Flow capacity up to 9.99 L/min; speed up to 250 RPM.
Indications for Use
Indicated for patients requiring extracorporeal circulation of blood for arterial regional perfusion and cardiopulmonary bypass procedures. Intended for use by qualified, experienced perfusionists.
Regulatory Classification
Identification
A roller-type cardiopulmonary bypass blood pump is a device that uses a revolving roller mechanism to pump the blood through the cardiopulmonary bypass circuit during bypass surgery.
K953904 — SARNS 9000 UNIVERSAL ROLLER PUMP · 3M Health Care, Ltd. · Apr 19, 1996
K131618 — LARGE (6) ROLLER PUMP FOR TERUMO ADVANCED PERFUSION SYSTEM 1; LARGE (6) ROLLER PUMP FOR TERUMO ADVANCED PERFUSION SYST · Terumo Cardiovascular Systems Corporation · Sep 27, 2013
K162843 — Small (4) Roller Pump for the Terumo Advanced Perfusion System 1, Large (6) Roller Pump for the Terumo Advanced Perfusion System 1 · Terumo Cardiovascular Systems Corporation · Nov 17, 2016
K112587 — ADVANCED PERFUSION SYSTEM 1 (APS1) TUBE CLAMP ASSEMBLY, ROLLER PUMP 6 INCH CLASSIC COLOR, · Terumo Cardiovascular Systems Corp. · Dec 19, 2011
K960974 — COBE CENTURY PERFUSION PUMP · Cobe Cardiovascular, Inc. · Jul 14, 1997
Submission Summary (Full Text)
{0}
# Section 2 - Summary & Certification
## 2.1 510(k) Summary
A 510(k) Summary of Safety and Effectiveness is on the next page.
K 953901
APR 19 1996
3M Confidential
Page 5 of 668
{1}
3M Health Care
Sarns™ and CDI™ Products
6200 Jackson Road
PO Box 1247
Ann Arbor, MI 48106-1247
USA
800 521 2818 Tel
313 663 4145 Tel
313 663 7981 Fax
3M
August 9, 1995
# 510(k) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. The assigned 510(k) number is K95391.
This 510(k) is for the Sarns 8000 Roller Pump, a "Roller-type cardiopulmonary bypass blood pump" and "Cardiopulmonary bypass pump speed control". The new Sarns 8000 Roller Pump is substantially equivalent to the existing Sarns 8000 Roller Pump (K915183).
The Sarns 8000 Roller Pump consists of a Universal Roller Pump, which is a microprocessor controlled 2-roller peristaltic pump with adjustable occlusion, and a power supply. The pump is capable of flows up to 9.99 L/min (depending on pump head tubing used) at pump speeds up to 250 RPM. The Sarns 8000 Roller Pump can be used as a stand-alone device or it is compatible with the Sarns 8000 Modular Perfusion System.
The Sarns 8000 Roller Pump is indicated for use in extracorporeal circulation of blood for arterial regional perfusion, and cardiopulmonary bypass procedures only, when used by a qualified perfusionist who is experienced in the operation of Sarns or similar equipment.
The Sarns 8000 Roller Pump has different technological characteristics from the predicate device. The design and the manufacturing process have been changed. The modified roller pump is manufactured using some Surface Mount Technology (SMT) components and processes to produce the circuit boards which contain the pump electronics. The electronic packaging of some of the components on these circuit boards is SMT instead of Through-Hole Technology (THT). Also, the software has undergone minor changes to remap the I/O addresses.
The Sarns 8000 Roller Pump has the same performance as the predicate device. Nonclinical tests were performed on the device to determine substantial equivalence. Electromagnetic Compatibility (EMC) tests were also performed to ensure equivalent or better results.
Cheryl Rosenberg
Cheryl Rosenberg
(313) 663-4145 Phone
(313) 663-5062 Fax
Page 6 of 668
{2}
3M Confidential
Page 7 of 668
## 2.2 Safety Literature Search
I certify that I have conducted a reasonable search of all information known or otherwise available to me about the types and causes of safety and/or effectiveness problems that have been reported for blood roller pumps. I further certify that I am aware of the types of problems to which blood roller pumps are susceptible and that the following list of safety and/or effectiveness problems about blood roller pumps is complete and accurate:
1. Air Emboli
2. Hemolysis
3. Loss of speed control
4. Random pump stops
Below is a bibliography of the materials upon which the above summary is based:
1. Customer Field Input Log, various dates, Sams, 3M Health Care
2. Intermittent Failure of a Stockert/Shiley Multiflow Roller Pump Module, The Journal of Extra-Corporeal Technology, Volume 25, Number 3, 1993
3. Medical Device Report M566956, M186992, M186044, M138277, various dates, Sams, 3M Health Care
4. Medical Device Report M494536, filed 4/21/94, Sams, 3M Health Care
5. Medical Device Report M444060, M123912, M111730, M101274, various dates, Sams, 3M Health Care
6. Medical Device Report M435457, M242746, M225052, M184140, M138914, M174191, various dates, Sams, 3M Health Care
7. Medical Device Report M377761, filed 2/26/93, Sams, 3M Health Care
8. Medical Device Report M125141, received 6/24/86, Cobe Labs, Inc.
9. Medical Device Report M269021, M244886, M244883, M225527, M210562, M210551, M210545, M129763, various dates, Shiley, Inc.
10. Medical Device Report M266724, M245762, M243742, M221182, various dates, Shiley, Inc.
11. Medical Device Report M253241, M250682, M225194, M221859, various dates, Shiley, Inc.
12. Medical Device Report M235614, M233235, various dates, Shiley, Inc.
13. Medical Device Report M220822, M212362, M212346, M210217, M166630, M164988, M137248, various dates, Shiley, Inc.
14. Medical Device Report M210544, received 8/17/90, Shiley, Inc.
15. Medical Device Report M218408, received 12/10/90, Shiley, Inc.
16. Medical Device Report M218946, received 12/17/90, Shiley, Inc.
17. Medical Device Report M162834, received 1/5/89, Shiley, Inc.
18. Medical Device Report M162828, M162829, M162830, M162831, M162832, M162833, M162836, M162838, M162839, M162840, M162841, M164234, M164858, M164860, M164850, M164934, M169675, various dates, Shiley, Inc.
19. Medical Device Report M151200, received 2/26/88, Shiley, Inc.
20. Medical Device Report M113582, received 11/5/92, Shiley, Inc.
{3}
# Bibliography (continued)
21.) *Medical Device Report* M149951, M154495, M158908, M162835, M162837, M166744, M169035, M181659, M183581, M183582, M201061, M203578, M204035, M204038, M204528, M206003, M206119, M206314, M206208, M209511, M214496, M218406, M218555, M218949, M218952, M218954, M220245, M220248, M220824, M221189, M221190, M222866, M225528, M225533, M232442, M231336, M231337, M231338, M235611, M235947, M238231, M248275, M243327, M244887, M245205, M245936, M245940, M247740, M247741, M249816, M249823, M252531, M262466, M268638, M269022, M269713, M272031, various dates, Shiley, Inc.
22.) *Medical Device Report* M364517, M311533, M310489, M308281, M306710, M306708, M306205, M283826, M282507, M273690, various dates, Sorin Biomedical, Inc.
23.) *Medical Device Report* M345060, M317126, various dates, Sorin Biomedical, Inc.
24.) *Medical Device Report* M344176, received 9/2/92, Sorin Biomedical, Inc.
25.) *Medical Device Report* M324153, received 9/30/92, Sorin Biomedical, Inc.
26.) *Medical Device Report* M321079, received 8/13/92, Sorin Biomedical, Inc.
27.) *Medical Device Report* M285491, received 4/29/92, Sorin Biomedical, Inc.
28.) *Medical Device Report* M283413, received 4/15/92, Sorin Biomedical, Inc.
Printed name of person required to submit 510(k): Cheryl Rosenberg
Signature of person required to submit 510(k): Cheryl Rosenberg
Title of person submitting 510(k): Advanced Regulatory Affairs Coordinator
Name of Company: 3M Health Care
Date: August 9, 1995
3M Confidential
Page 8 of 668
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.