← Product Code [DTY](/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/DTY) · K093650

# BIOTREND OXYGEN SATURATION AND HEMATOCRIT SYSTEM (K093650)

_Medtronic, Inc. · DTY · Feb 17, 2010 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/DTY/K093650

## Device Facts

- **Applicant:** Medtronic, Inc.
- **Product Code:** [DTY](/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/DTY.md)
- **Decision Date:** Feb 17, 2010
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.4410
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The BioTrend Oxygen Saturation and Hematocrit System measures percent oxygen saturation and hematocrit of the blood in the extracorporeal circuit. The extracorporeal circuit is used for, but is not limited to, cardiopulmonary bypass, closed-chest support, and limb perfusion.

## Device Story

BioTrend Oxygen Saturation and Hematocrit System is an on-line monitoring instrument for extracorporeal circuits. It utilizes fiber optic technology to continuously measure venous (SvO2) and arterial (SaO2) oxygen saturation and hematocrit. The system comprises the BioTrend instrument, sensor cables, and disposable Tri-optic Measurement Cells (TMC) placed in the blood pathway. The instrument processes optical signals transmitted via sensor cables, which provide patient electrical isolation. Results are displayed on color-coded LEDs; the system includes a built-in self-check for equipment failure and error reporting. It operates on AC or internal rechargeable battery power. Used in clinical settings like cardiopulmonary bypass, the device provides real-time data to clinicians to assist in monitoring patient blood gas status during procedures.

## Clinical Evidence

No clinical data. Substantial equivalence established via bench testing, including blood testing for measurement accuracy, environmental testing, packaging testing (ASTM D 4169), system-level requirements verification, software requirements verification, hardware PCB assembly performance, and UL/TUV safety/emissions testing.

## Technological Characteristics

Fiber optic sensing principle; LED display; rechargeable battery/AC power. Components include instrument, sensor cables, and disposable Tri-optic Measurement Cells. Software-based signal processing. Packaging meets ASTM D 4169. Device provides patient electrical isolation.

## Regulatory Identification

A cardiopulmonary bypass in-line blood gas sensor is a transducer that measures the level of gases in the blood.

## Predicate Devices

- BioTrend® Oxygen Saturation and Hematocrit System ([K954501](/device/K954501.md))

## Reference Devices

- Tri-optic Measurement Cell ([K910421](/device/K910421.md))
- Tri-optic Measurement Cell ([K012743](/device/K012743.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K093652

Image /page/0/Picture/1 description: The image shows the Medtronic logo. The logo consists of a circular graphic on the left and the word "Medtronic" in bold, black letters on the right. The circular graphic appears to have a stylized representation of a human figure within it.

FEB 1 7 2010

# 510(k) Summary

| Date Prepared  | November 24, 2009                           |
|----------------|---------------------------------------------|
| Submitter      | Medtronic, Inc.<br>710 Medtronic Parkway NE |
|                | Minneapolis, MN 55432-5604                  |
|                | Establish Registration Number: 2184009      |
| Contact Person | Jeffrey L. Koll                             |
|                | Regulatory Affairs Specialist               |
|                | Phone: (763) 514-9842                       |
|                | Fax: (763) 367-8360                         |
|                | Email: jeffrey.l.koll@medtronic.com         |

## Device Name and Classification

Trade Name: Common Name: Regulation Number: Product Code: Classification:

BioTrend® Oxygen Saturation and Hematocrit System Cardiopulmonary bypass in-line blood gas sensor. 21 CFR 870.4410 DTY Class II

## Predicate Device

BioTrend® Oxygen Saturation and Hematocrit System (K954501)

#### Device Description

The modified BioTrend Oxygen Saturation and Hematocrit System is an on-line monitoring instrument that combines both venous (SvO2) and arterial (SaO2) oxygen saturation measurement with hematocrit (Hct) measurement. The BioTrend System consists of the BioTrend Instrument, two Sensor Cables, and a power cord. The BioTrend system is designed to be used with the Tri-optic Measurement Cells.

The Tri-optic Measurement Cell is a disposable device in the extracorporeal circuit that provides a sealed interface between the blood pathway and the BioTrend sensor cable. No change is being made to the Tri-optic Measurement Cell, which was previously cleared under K910421 and K012743.

Using fiber optic technology, the BioTrend System continuously measures the percentage of oxygen saturation and hematocrit and displays the results on large, easy-to-read, color-coded

Alleviating Pain · Restoring Health · Extending Life

11

{1}------------------------------------------------

Light Emitting Diodes (LEDs). The display panel indicates operating status and error messages, and provides a means for system calibration.

BioTrend Sensor Cables connect to the BioTrend instrument and the in-line Tri-optic Measurement Cells (TMC) to transmit optical measurement signals. The BioTrend sensor cables isolate the patient from the instrument electronics, providing patient protection. The BioTrend instrument contains a built-in, rechargeable battery pack to provide the battery power. Consequently, the BioTrend instrument can be operated on AC or battery power. A continuous, built-in self-check immediately alerts the operator of equipment failure and displays a corresponding error code.

#### Indications for Use

5

The BioTrend Oxygen Saturation and Hematocrit System measures percent oxygen saturation and hematocrit of the blood in the extracorporeal circuit. The extracorporeal circuit is used for, but is not limited to, cardiopulmonary bypass, closed-chest support, and limb perfusion.

## Comparison to Predicate Device

A comparison of the modified product and the currently marketed oxygen saturation and hematocrit system has the following similarities to the system which received 510(k) clearance:

- . Same intended use.
- . Same operating principle.
- Same performance claims. .
- Same Tri-Optic Measurement Cells (K910421 and K012743) for transmission of . optical measurement signals.
- Same Algorithm used to calculate oxygen saturation and hematocrit. .

#### Intended Use

The intended use is unchanged.

#### Labeling

A summary of the labeling changes is as follows:

- Update IFU and labeling to address hardware and software changes to the device. .
- Add model number to device label. .

12

{2}------------------------------------------------

#### Summary of Performance Data

Preclinical testing data were used to establish the performance characteristics of the modifications to this device. Clinical testing was not required to establish substantial equivalence. The following bench testing was conducted:

- Blood testing: to verify oxygen saturation and hemacrit measurement accuracy .
- Environmental testing: to verify operational temperature, storage temperature, . thermal shock, storage humidity, vibration, mechanical shock, cautery, defibrillation, spill resistance, and chemical resistance
- Packaging testing: to verify packaging requirements per ASTM D 4169 performance . testing of shipping Containers and Systems, Distribution Cycle 13, Assurance Level 1. :
- System testing: to verify the device meets the system level requirements called out in . the system requirements document
- Software testing: to verify the device meets the software requirements called out in . the software requirements document
- Hardware testing: to verify the PCB assemblies meet the performance requirements . called out in the design specifications
- UL/TUV testing: to verify the device meets the emissions, immunity/susceptibility, . and safety requirements

#### Conclusion

The modifications to the BioTrend® Oxygen Saturation and Hematocrit System described in this submission result in a substantially equivalent device because the fundamental scientific technology and the intended use are unchanged.

13

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, facing to the left.

Food and Druq Administration 10903 New Hampshire Avenue Document Control Room W-066-0609 Silver Spring, MD 20993-0002

FEB 1 7 2010

Medtronic Cardiovascular MVS83 c/o Mr. Jeffrey L. Koll Regulatory Affairs Specialist 8200 Coral Sea Street NE Mounds View, MN 55112

Re: K093650

> BioTrend® Oxygen Saturation and Hematocrit System Regulation Number: 21 CFR 870.4410 Regulation Name: Sensor, Blood-Gas, In-Line, Cardiopulmonary Bypass Regulatory Class: Class II Product Code: DTY Dated: January 13, 2010 Received: January 14, 2010

Dear Mr. Koll:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Mr. Jeffrey L. Koll

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ycm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

ie R. -Vachner

Image /page/4/Picture/8 description: The image shows a free-form sketch of a shape that resembles a mountain range or a stylized crown. The lines are dark and somewhat jagged, giving the impression of a quick, hand-drawn illustration. The shape is simple and abstract, with no shading or additional details.

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

## Indications for Use

510(k) Number (if known): K د و 3 65 O

Device Name: BioTrend Oxygen Saturation and Hematocrit System

Indications for Use:

The BioTrend Oxygen Saturation and Hematocrit System measures percent oxygen saturation and hematocrit of the blood in the extracorporeal circuit. The extracorporeal circuit is used for, but is not limited to, cardiopulmonary bypass, closed-chest support, and limb perfusion.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

una R. viclunen

(Division Sign-Off) Division of Cardiovascular Devices

K093650 510(k) Number_

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/DTY/K093650](https://fda.innolitics.com/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/DTY/K093650)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
