DIDECO D920, LILLIPUT 1 TWIN RESERVOIR

K984322 · Dideco S.P.A. · DTP · Mar 2, 1999 · Cardiovascular

Device Facts

Record IDK984322
Device NameDIDECO D920, LILLIPUT 1 TWIN RESERVOIR
ApplicantDideco S.P.A.
Product CodeDTP · Cardiovascular
Decision DateMar 2, 1999
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.4230
Device ClassClass 2
AttributesTherapeutic, Pediatric

Indications for Use

The Dideco D920, Lilliput I twin reservoir is intended for use in infants who undergo cardiopulmonary bypass surgery requiring extracorporeal circulation for whom a low circuit priming volume is required. The venous reservoir with cardiotomy filter is intended for use as a storage reservoir (gravity or vacuumassisted) for venous return blood and as a filtered reservoir for cardiac suction during ECC. The cardiotomy reservoir has the same intended use as the venous reservoir, as a filtered reservoir for cardiac suction blood in a bypass extracorporeal circuit, but does not have a venous return. Following intraoperative use, the reservoirs are used for the collection and autotransfusion of shed blood.

Device Story

D 920 Lilliput 1 twin reservoir; rigid hardshell polycarbonate venous/cardiotomy reservoir. Inputs: venous return blood, intracardiac sucker/vent blood. Operation: gravity or vacuum-assisted storage of venous blood; filtration of cardiotomy suction blood via internal defoamer sponge and 30-micron polyester screen filter. Output: processed blood for oxygenator; autotransfusion of shed blood post-operatively. Used in surgical theater during cardiopulmonary bypass; operated by perfusionists/surgeons. Benefits: low priming volume for infants; removal of microaggregates/microemboli.

Clinical Evidence

No clinical data. Bench testing only. Biocompatibility testing performed per ISO 10993-1:1992. In vitro performance testing conducted per ISO 7199 international standard.

Technological Characteristics

Rigid hardshell polycarbonate housing; polyester cardiotomy filter (30 micron); polyurethane reservoir filter; polyester screen filter (120 micron). Connections: 3/16"-1/4" venous return/outlet. Sterilization: Ethylene oxide. Max volume: 700 ml. Max blood flow: 1.5 LPM. Bayonet/YSI 400 compatible temperature probe.

Indications for Use

Indicated for infants undergoing cardiopulmonary bypass surgery requiring extracorporeal circulation where low circuit priming volume is required.

Regulatory Classification

Identification

A cardiopulmonary bypass defoamer is a device used in conjunction with an oxygenator during cardiopulmonary bypass surgery to remove gas bubbles from the blood.

Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance for Extracorporeal Blood Circuit Defoamer 510(k) Submissions.”

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # 510(k) Safety and Effectiveness Summary #### 1. Submitted by: Barry Sall Senior Regulatory Consultant PAREXEL International Corporation Waltham, MA, USA #### 2. Contact Person: Luigi Vecchi Regulatory Affairs Manager Dideco, S.p.A. via Statale 12 Nord, 86 P.O. Box 87 41037 Mirandola (MODENA) Italy Telephone number: 01139 0535 29811 #### 3 Date of Preparation: December 2, 1998 #### Name and Address of Owner/Operator and Manufacturer 4 Dideco, S.p.A. via Statale 12 Nord, 86 P.O. Box 87 41037 Mirandola (MODENA) Italy #### 5 Product Name Trade Name: D 920, Lilliput 1 twin reservoir Common Name: Infant-newborn Venous Cardiotomy Reservoir #### 6. Predicate Devices #### 1) VENOMIDICARD/MIDICARD D752/D762 manufactured by: Dideco S.p.A.. Via Statale 12 Nord, 86 41037 Mirandola (MO) ITALY K941215, cleared on May 5, 1995 {1}------------------------------------------------ 2) MICRO SAFE manufactured by: Polystan AJS Walgerholm 8 3500 Vaerlose DENMARK K953976, cleared on February 4, 1997 #### 7. Claim of Substantial Equivalence The D 920 is an infant/newborn hard-shell venous reservoir, like the predicate devices. The basic function of all hardshell reservoirs is the same. They store the blood coming from the surgical theater, draw it from the venous reservoir and push it through an oxygenator via an external pump. The operating principles and control mechanisms are exactly the same for the D 920 and the predicate devices. As both the Venomidicard and the D 920 are manufactured by Dideco, they share the same incoming raw materials inspections, manufacturing in process controls; in addition the D 920 utilizes the same material type and a similar blood flow path as the Venomidicard and the filtration sequence is the same for D 920 and Venomidicard. The indications for use are the same for both D 920, Micro safe and Venomidicard: all devices are intended for the infant population who undergo cardiopulmonary bypass surgery requiring extracorporeal circulation. All predicate devices have a bayonet/YSI 400 compatible temperature probe. | Parameters | VENOMIDICARD/<br>MIDICARD | MICRO-SAFE | D 920 | |------------------------------------------|---------------------------|----------------|----------------| | Manufacturer | Dideco S.p.A. | Polystan A/V. | Dideco S.p.A. | | Method of sterilization | Ethylene Oxyde | Ethylene Oxyde | Ethylene oxide | | Maximum blood flow rate | 4 LPM | 0.8 LPM | 1.5 LPM | | Recommended operating<br>blood flow rate | 4000 ml/min | 800 ml/min | 800 ml/min | | Filtering pore size | 20 micron | 40 micron | 30 micron | ## Substantial Equivalence Comparison Chart: Performance Characteristics {2}------------------------------------------------ ### Substantial Equivalence Comparison Chart: General features | Parameters | VENOMIDICARD/<br>MIDICARD | MICRO-SAFE | D 920 | |----------------------------|----------------------------|---------------------------|----------------------------| | Connections: | | | | | Venous return | 3/16" | 3/16" - 1/4" | 3/16" - 1/4" | | Venous outlet | 3/8" | 3/16" - 1/4" | 3/16" - 1/4" | | Cardiotomy inlet | 3/16" - 1/4" | 3/16" - 1/4" | 3/16" - 1/4" | | Reservoir venous inlet | 3/8" | 1/4" | 1/4" | | Filter materials: | | | | | Cardiotomy filter | polyester 20 microns | polyester 40 microns | polyester 30 microns | | Reservoir filter | polyurethane | polyurethane | polyurethane | | Screen filter | polyester 120 microns | polyester 175 microns | polyester 120 microns | | Available<br>configuration | Card. /venous<br>reservoir | Venous reservoir | Card. /venous<br>reservoir | | Housing: | | | | | type | rigid hardshell | rigid hardshell | rigid hardshell | | material | polycarbonate | polycarbonate | polycarbonate | | max. volume | 2000 ml | 400 ml | 700 ml | | Packaging: | | | | | type | Single/multi unit box | single/multi unit box | single/multi unit box | | material | Tyvek pouch/carton<br>box | Tyvek pouch/carton<br>box | Tyvek pouch/carton<br>box | | Patient population | Children | Infants | Infants | #### 8. Device Description The D 920 serves as a collection reservoir for venous blood recovered by intracardiac sucker and ventricular vent devices. The D 920 device can also be attached to an oxygenator, previously cleared by the Agency for the same patient population, in order to collect blood during normal operations, assuring the proper oxygenation capability of the device is currently being developed in an open system configuration which includes the Dideco D901 Lilliput oxygenator, cleared by the Agency on March 15th 1996 (K953835). The hard-shell cardiotomy/venous reservoirs are comprised of a rigid polycarbonate housing with an internal support; in which a filtering system is placed around the support. The D 920 has two distinct (upper and lower) sections connected by an overflow system. The upper section contains a defoamer sponge and screen filter which are designed to remove microaggregates and microemboli from cardiotomy suction and vent blood. The devices are available with a 30 micron filter. The venous return blood enters the lower section and passes through a defoamer sponge. Only the lower section is contained within a polyester outer screen. The D 920 has four cardiotomy suction/vent inlet ports which accommodate both 1/4 and 3/16" inner diameter tubing. Three priming ports are positioned on the top of the reservoir: 3 x luer lock (filtered), 1 x luer lock (unfiltered) and 1 x 1/4-3/16". Fluid added through the one outer priming port bypasses the integral cardiotomy section. Fluids added through the three inner luer and 1/4-3/16" ports flow through the defoamer and filter elements in the cardiotomy section. Venous blood enters through the 1/4-3/16" inlet port at the bottom of the unit. The D 920 has a 1/4-3/16" blood outlet port at the bottom of the unit. The 1/4" gas outlet port is located at the top of the unit. Both devices are supported by a holder. {3}------------------------------------------------ ## 9. Intended Use The Dideco D 920 is intended for use in infants who undergo cardiopulmonary bypass surgery requiring extracorporeal circulation (ECC) for whom a low circuit priming volume is required. The venous reservoir with cardiotomy filter is intended for use as a storage reservoir (gravity or vacuumassisted) for venous return blood and as a filtered reservoir for cardiac suction during ECC. The cardiotomy reservoir has the same intended use as the venous reservoir above mentioned as a filtered reservoir for cardiac suction blood in a bypass extracorporeal circuit, but does not have a venous return. Following intraoperative use, the reservoirs are used for the collection and autotransfusion of shed blood. #### 10. Summary of Nonclinical Data The Dideco D 920 has successfully completed the full range of biocompatability testing as specified in ISO 10993-1:1992 and FDA May 1, 1995 Memorandum on the use of the ISO 10993 standard for Biocompatibility Testing. Dideco has also successfully performed in vitro performance testing on the D 920 based on the ISO 7199 international standard. {4}------------------------------------------------ ### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/4/Picture/2 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the top half of the circle. Inside the circle is a stylized image of an eagle with three heads. MAR - 2 1999 Dideco, S.P.A. c/o Mr. Barry Sall Senior Regulatory Consultant Parexel International Corporation 1601 Trapelo Road Waltham, MA 02154 Re : K984322 Dideco D920, Lilliput 1 Twin Reservoir Regulatory Class: II (Two) and III (Three) Product Code: DTN and DTP Dated: December 2, 1998 Received: December 3, 1998 Dear Mr. Sall: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to A substantially equivalent determination assumes compliance with 895. the Current Good Manufacturing Practice requirements, as set forth in the Quality System Requlation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 {5}------------------------------------------------ Page 2 - Mr. Barry Sall This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in_vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html". Sincerely yours, Thomas J. Callahon Thomas J. Callaman, Ph.D. Director Division of Cardiovascular, Respiratory and Neuroloqical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ k 984322 510(k) Number (if known): Device Name: Dideco D920, Lilliput 1 twin reservoir Indications for Use: The Dideco D920, Lilliput I twin reservoir is intended for use in infants who undergo cardiopulmonary bypass surgery requiring extracorporeal circulation for whom a low circuit priming volume is required. The venous reservoir with cardiotomy filter is intended for use as a storage reservoir (gravity or vacuumassisted) for venous return blood and as a filtered reservoir for cardiac suction during ECC. The cardiotomy reservoir has the same intended use as the venous reservoir, as a filtered reservoir for cardiac suction blood in a bypass extracorporeal circuit, but does not have a venous return. Following intraoperative use, the reservoirs are used for the collection and autotransfusion of shed blood. #### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Satie & Compule (Division Sign-Off) Division of Cardiovascular, Respiratory, and Neurological Devices 510(k) Number K984322 Prescription Use (Per 21 CFR 801.109) OR Over-The-Counter Use (Optional format 1-2-96)
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...