← Product Code [DTL](/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/DTL) · K101113

# GUARDIAN II NC HEMOSTASIS VALVE (K101113)

_Zerusa Limited · DTL · May 21, 2010 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/DTL/K101113

## Device Facts

- **Applicant:** Zerusa Limited
- **Product Code:** [DTL](/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/DTL.md)
- **Decision Date:** May 21, 2010
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 870.4290
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The Guardian® II NC Hemostasis Valve is intended to maintain hemostasis during the introduction, withdrawal and use of diagnostic/interventional devices during vascular procedures.

## Device Story

Guardian® II NC Hemostasis Valve functions as a conduit for interventional devices (up to 8.0F) during vascular procedures. Device features dual-seal system: low-pressure wiper seal and high-pressure seal. Low-pressure seal uses spring-loaded mechanism; opens by depressing cap, closes automatically upon release. High-pressure seal operated by rotating nut clockwise; secures device in position and enables pressure injections up to 150 psi (10 ATM). Includes guidewire introducer (austenitic stainless steel tube/polycarbonate hub) and guidewire torquer for steering. Used in clinical settings by physicians to maintain hemostasis, preventing blood loss while allowing device manipulation within vasculature.

## Clinical Evidence

Bench testing only. No clinical data provided.

## Technological Characteristics

Materials include polycarbonate and austenitic stainless steel. Features dual-seal design: spring-loaded low-pressure wiper seal and rotating nut high-pressure seal. Supports diagnostic/interventional devices up to 8.0F. Rated for pressure injections up to 150 psi (10 ATM). Mechanical operation; no software or electronic components.

## Regulatory Identification

A cardiopulmonary bypass adaptor, stopcock, manifold, or fitting is a device used in cardiovascular diagnostic, surgical, and therapeutic applications to interconnect tubing, catheters, or other devices.

## Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.

## Predicate Devices

- Guardian® II Hemostasis Valve ([K092711](/device/K092711.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

### 510(K) SUMMARY

#### 1. SUBMITTER:

MAY 2 2010

Zerusa Limited 219-220 Business Innovation Centre, NUIG Galway, Ireland 011-353-91-861611 Telephone: Establishment Registration Number: "3005395947

Official contact: Mr. Liam Mulloy, CEO 011-353-91-863060 Phone: Date Prepared: April 22, 2010

MAY 2 ] 2010

#### 2. DEVICE:

| Tradename:           | Guardian® II NC Hemostasis Valve                   |
|----------------------|----------------------------------------------------|
| Classification Name: | Cardiopulmonary Bypass Adaptor, Stopcock, Manifold |
|                      | or Fitting                                         |
| Classification:      | Class II                                           |
| Common Name:         | Hemostatic Valve                                   |
| Product Code:        | 74 DTL                                             |
| Regulation Number:   | 870.4290                                           |

#### 3. PREDICATE DEVICE:

The only change being requested in this Special 510(k) Premarket Notification is to modify the manner in which the device's low pressure seal is activated. The low pressure seal on Zerusa's currently marketed Guardian® II Hemostasis Valve is clicked "open" and clicked "closed" in two separate actions by the user (i.e., push / push mechanism). The proposed Guardian® II NC Hemostasis Valve incorporates a low pressure seal which is held open by depressing the cap. The seal will automatically close once the user releases the cap (i.e., spring loaded action).

The predicate device used to determine substantial equivalence for this device was the Zerusa Limited's currently marketed Guardian® II Hemostasis Valve (#K092711).

#### DEVICE DESCRIPTION: 4.

The Zerusa Guardian® II NC Hemostasis Valve is designed to be used as a conduit when interventional devices with diameters up to 8.0F (2.67mm or 0.105") are inserted into the human vascular system.

{1}------------------------------------------------

The device has two seals: the low-pressure seal (or wiper seal) and the high-pressure seal. Depressing the cap opens the low-pressure seal, releasing the cap closes the seal. The high-pressure seal is operated by rotating the nut clockwise. Closure of the highpressure seal, which is achieved when the nut can no longer rotate, secures the diagnostic/interventional device in position within the vasculature and also allows for pressure injections up to 150 psi (10 ATM).

Included with the Guardian® II NC Hemostasis Valve is a Guidewire Introducer, which is used to facilitate entry of the guidewire into the Guardian® II NC Hemostasis Valve. It consists of an austenitic stainless steel tube connected to a hub constructed of polycarbonate.

Also included with the Guardian® II NC HV is a simple Guidewire Torquer which is used to manipulate the steering of a guidewire within the vascular regions.

#### INTENDED USE: ડ.

The Guardian® II NC Hemostasis Valve is intended to maintain hemostasis during the introduction, withdrawal and use of diagnostic/interventional devices during vascular procedures.

#### INDICATIONS FOR USE: 6.

The Guardian® II NC HV is intended to maintain hemostasis during the use of diagnostic/interventional devices. The device is indicated for maintaining a seal around diagnostic/interventional devices with outside diameters up to 8.0F (2.67 mm or 0.105") during diagnostic/interventional procedures.

The guidewire introducer is included to facilitate the guidewire's passage through the Guardian® II NC HV.

The Torque Device is intended to manipulate the steering of the guidewire within the vascular regions.

#### COMPARISON OF CHARACTERISTICS: 7.

Comparisons of the proposed and predicate devices show that the technological characteristics such as materials, performance characteristics, sterilization and packaging are identical or substantially equivalent to the currently marketed predicate devices.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows handwritten text on a white background. The text at the top reads "K101113". Below that, the text reads "pg 3 of 3". The handwriting appears to be somewhat messy and the text is slightly slanted.

#### 8. PERFORMANCE DATA:

The Guardian® II NC Hemostasis Valve was subjected to a full battery of performance testing. The results of the performance testing demonstrated the safety and effectiveness of the device.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002

MAY 2 ] 2010

Zerusa Limited c/o Mr. Stephen M. Page Regulatory Consultant 219-220 Business Innovation Centre, NUIG Galway, Ireland

Re: K101113

Guardian® II NC Hemostasis Valve Regulation Number: 21 CFR 870.4290 Regulation Name: Cardiopulmonary Bypass Adaptor, Stopcock, Manifold or Fitting Regulatory Class: Class II Product Code: DTL Dated: April 22, 2010 Received: April 21, 2010

Dear Mr. Page:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Mr. Stephen M. Page

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

ff you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

R., Viihnen

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

# KI01113

## Indications for Use

510(k) Number (if known): K101113

Device Name: Guardian® II NC Hemostasis Valve

Indications for Use: The Guardian® II NC HV is intended to maintain hemostasis during the use of diagnostic/interventional devices. The device is indicated for maintaining a seal around diagnostic/interventional devices with outside diameters up to 8.0F (2.67 mm or 0.105") during diagnostic/interventional procedures.

The guidewire introducer is included to facilitate the guidewire's passage through the Guardian® II NC HV.

The Torque Device is included to manipulate the steering of the guidewire within the vascular regions.

Prescription Use XXX (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

printer R. balmes

(Division Sign-Off) Division of Cardiovascular Devices

510(k) Number_ ¥ 10 113 ---

Page 1 of 1

(Posted November 13, 2003)

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/DTL/K101113](https://fda.innolitics.com/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/DTL/K101113)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
