← Product Code [DRY](/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/DRY) · K090147

# BLOOD MONITORING UNIT, MODEL BMU 40 (K090147)

_Maquet Cardiopulmonary, AG · DRY · Apr 22, 2009 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/DRY/K090147

## Device Facts

- **Applicant:** Maquet Cardiopulmonary, AG
- **Product Code:** [DRY](/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/DRY.md)
- **Decision Date:** Apr 22, 2009
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.4330
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The intended purpose of the MAQUET Blood Monitoring Unit BMU 40 is to monitor blood parameters during cardiopulmonary bypass (CPB) or similar procedures with extracorporeal circulation, which require continuous monitoring of the arterial and/or venous blood parameters: partial pressure of oxygen (pO2), temperature (Ta and Tv), oxygen saturation (SO2), hemoglobin (Hb) and hematocrit (Hct). Oxygen consumption (VO2) can also be calculated. Blood flow (QBlood) can be entered manually or values can be received from a connected heart-lung machine. The duration of application of the disposable products (arterial BMU Sensor and venous BMU Cell) is limited to six hours. The BMU 40 is designed for continuous operation.

## Device Story

System monitors blood parameters during cardiopulmonary bypass; consists of BMU 40 control unit and sterile single-use disposable connectors (BMU Sensor/Cell) clamped to arterial/venous lines. Inputs include real-time sensor data from extracorporeal circuit and manual/automated blood flow (QBlood) values from heart-lung machine. Device processes inputs to display measured values, time courses, and calculated oxygen consumption (VO2). Used in clinical settings (e.g., OR) by perfusionists/clinicians. Output provides continuous physiological status to guide clinical decision-making during bypass; benefits patient through real-time monitoring of oxygenation and perfusion status.

## Clinical Evidence

Bench testing only. Performance characteristics were exhaustively tested and compared with the predicate device (CDI 500). No clinical data presented.

## Technological Characteristics

System includes monitor (BMU 40) and disposable sensors/cells. Features continuous monitoring of pO2, temperature, SO2, Hb, Hct. Connectivity includes manual input or interface with heart-lung machine for blood flow data. Disposable components are sterile, single-use, limited to 6-hour duration.

## Regulatory Identification

A cardiopulmonary bypass on-line blood gas monitor is a device used in conjunction with a blood gas sensor to measure the level of gases in the blood.

## Predicate Devices

- CDI Blood Parameter Monitoring System 500 ([K972962](/device/K972962.md))
- M3 Monitor ([K072131](/device/K072131.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

KC90147)

## MAQUET

### 510 (K) Summary [as required by 21 CFR 807.92(c) ]

| Submitter:            | Maquet Cardiopulmonary AG<br>Hechinger Strasse 38<br>72145 Hirrlingen<br>Germany                                 | APR 22 2009 |
|-----------------------|------------------------------------------------------------------------------------------------------------------|-------------|
| Contact Person:       | Anastasia Mueller<br>Phone: +49 7478 921-244<br>Fax: +49 7478 921-400<br>E-mail: anastasia.mueller@maquet-cp.com |             |
| Date Prepared:        | January 14, 2009                                                                                                 |             |
| Device Trade Name:    | Blood Monitoring Unit BMU 40                                                                                     |             |
| Common/Usual name:    | Blood Parameter Monitor,<br>Sensor / Cell                                                                        |             |
| Classification names: | Cardiopulmonary bypass online gas monitor<br>Cardiopulmonary bypass in-line blood gas sensor                     |             |
| Predicate Devices:    | CDI Blood Parameter Monitoring System 500,<br>K972962<br>M3 Monitor, K072131                                     |             |
| Device Description:   |                                                                                                                  |             |

The Blood Monitoring Unit BMU 40 monitors blood parameters during cardiopulmonary bypass or similar procedures with extracorporeal circulation, which require continuous monitoring of the arterial and/or venous blood parameters

The Blood Monitoring Unit BMU 40 is blood monitoring system consisting of the following componens:

- the control unit (monitor, called BMU 40) which comprises a display i showing the actual measured sensor values and time course.
- Sterile single use connectors (BMU Sensor/ BMU Cell) to be clamped י on the probes, one in the venous line and one in the arterial line. The

{1}------------------------------------------------

# MAQUET

connectors are available in different sizes. BMU Sensor is connected to arterial probe and BMU Cell is connected to venous probe.

#### Indications for Use:

The intended purpose of the MAQUET Blood Monitoring Unit BMU 40 is to monitor blood parameters during cardiopulmonary bypass (CPB) or similar procedures with extracorporeal circulation, which require continuous monitoring of the arterial and/or venous blood parameters: partial pressure of oxygen (pO2), temperature (Ta and Tv), oxygen saturation (SO2), hemoglobin (Hb) and hematocrit (Hct). Oxygen consumption (VO2) can also be calculated. Blood flow (QBlood) can be entered manually or values can be received from a connected heart-lung machine.

The duration of application of the disposable products (arterial BMU Sensor and venous BMU Cell) is limited to six hours.

The BMU 40 is designed for continuous operation.

### Statement of Technical Comparison:

The Blood Monitoring Unit BMU 40 is comparable to the CDI Blood Parameter Monitoring System 500 as well as to the M3 Monitor regarding the intended use, design and performance.

### Non-clinical Testing and Performance:

The Blood Monitoring Unit BMU 40 performs as intended according to its performance specifications.

The performance characteristics of the Blood Monitoring Unit BMU 40 were exhaustively tested and compared with the predicate device (CDI 500).

1587

{2}------------------------------------------------

## MAQUET

### Determination of Substantial Equivalence

Testing and evaluation on safety and effectiveness was executed to demonstrate that the Blood Monitoring Unit BMU 40 described in this submission is substantially equivalent to the CDI Blood Parameter Monitoring System 500 and to the M3 Monitor.

The following areas have been tested:

- Performance
- Electrical and mechanical safety on the monitor- BMU 40
- Software Validation on the monitor- BMU 40
- Biocompatibility on the BMU Sensor and BMU Cell
- Sterility on the BMU Sensor and BMU Cell
- Integrity on the BMU Sensor and BMU Cell

### Conclusion

The data given demonstrate that the Blood Monitoring Unit BMU 40 is substantially equivalent to the named predicate devices.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of an eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

APR 2 2 2009

Maquet Cardiopulmonary AG c/o Ms. Anastasia Mueller Hechinger Strasse 38 72145 Hiirrlingen Germany

Re: K090147

Blood Monitoring Unit BMU40 Regulation Number: 21 CFR 870.4410 Regulation Name: Cardiopulmonary bypass online gas monitor Regulatory Class: Class II Product Code: DRY Dated: January 14, 2009 Received: January 21, 2009

Dear Ms. Mueller:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

#### Page 2 - Ms. Anastasia Mueller

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

i

Bram D. Zuckerman, M.D. Division Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

### Indications for Use

510(k) Number (if known): KO90H7

Blood Monitoring Unit BMU 40 _________________________________________________________________________________________________________________________________________________

Indications for Use:

The intended purpose of the MAQUET Blood Monitoring Unit BMU 40 is to monitor blood parameters during cardiopulmonary bypass (CPB) or similar procedures with extracorporeal circulation, which require continuous monitoring of the arterial and/or venous.blood parameters: partial pressure of oxygen (p02), temperature (Ta and Tv), oxygen saturation (SO2), hemoglobin (Hb) and hematocrit (Hct). Oxygen consumption (VO2) can also be calculated. Blood flow (QBlood) can be entered manually or values can be received from a connected heart lung machine.

The duration of application of the disposable products (arterial BMU Sensor and venous BMU Cell) is limited to six hours.

The BMU 40 is designed for continuous operation.

| Prescription Use            | X |
|-----------------------------|---|
| (Part 21 CFR 801 Subpart D) |   |

AND/OR

| Over-The-Counter Use   |  |
|------------------------|--|
| (21 CFR 801 Subpart C) |  |

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE
OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation.(ODE)

| (Division Sign-Off)                |         |
|------------------------------------|---------|
| Division of Cardiovascular Devices |         |
| 510(k) Number                      | K090147 |

| Page | 1 of 1 |
|------|--------|
|------|--------|

(Posted November 13, 2003)

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/DRY/K090147](https://fda.innolitics.com/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/DRY/K090147)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
