EMBOSPHERE MICROSPHERES AND EMBOGOLD MICROSPHERES FOR USE IN UTERINE FIBROID EMBOLIZATION
K021397 · Biosphere Medical, Inc. · NAJ · Nov 22, 2002 · Cardiovascular
Device Facts
Record ID
K021397
Device Name
EMBOSPHERE MICROSPHERES AND EMBOGOLD MICROSPHERES FOR USE IN UTERINE FIBROID EMBOLIZATION
Applicant
Biosphere Medical, Inc.
Product Code
NAJ · Cardiovascular
Decision Date
Nov 22, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Embosphere® and EmboGold™ Microspheres are intended for embolization of arteriovenous malformations and hypervascular tumors, including and uterine fibroids.
Device Story
Embosphere® and EmboGold™ Microspheres are small, flexible, hydrophilic, biocompatible spheres composed of acrylic polymer and porcine-derived gelatin. Used for embolization of hypervascular tumors, arteriovenous malformations, and symptomatic uterine fibroids. Delivered to target vessel via catheter under fluoroscopic guidance by physicians. Microspheres act as mechanical embolic agents to occlude blood flow to target tissue. Available in six size ranges (40-1200 microns); 500+ micron sizes required for uterine fibroid embolization. EmboGold™ variant includes colorization for improved visibility during clinical handling. Device provides therapeutic benefit by reducing blood supply to target lesions.
Clinical Evidence
Prospective clinical study conducted to support safety and effectiveness of Embosphere® Microspheres for uterine fibroid embolization.
Technological Characteristics
Materials: Acrylic polymer and porcine-derived gelatin. Form factor: Flexible, hydrophilic microspheres. Sizes: 40-120, 100-300, 300-500, 500-700, 700-900, 900-1200 microns. Delivery: Catheter-based under fluoroscopy. Sterilization: Sterile, nonpyrogenic.
Indications for Use
Indicated for embolization of arteriovenous malformations, hypervascular tumors, and symptomatic uterine fibroids. For uterine fibroids, only microspheres 500 microns or greater are indicated.
Regulatory Classification
Identification
A vascular embolization device is an intravascular implant intended to control hemorrhaging due to aneurysms, certain types of tumors (e.g., nephroma, hepatoma, uterine fibroids), and arteriovenous malformations. This does not include cyanoacrylates and other embolic agents, which act by polymerization or precipitation. Embolization devices used in neurovascular applications are also not included in this classification, see § 882.5950 of this chapter.
Special Controls
*Classification.* Class II (special controls.) The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Vascular and Neurovascular Embolization Devices.” For availability of this guidance document, see § 870.1(e).
{0}------------------------------------------------
K021397
## 510(k) Summary
### for
## Embosphere® Microspheres and EmboGold™ Microspheres for use in Uterine Fibroid Embolization
- 1. SPONSOR
Biosphere Medial, Inc. 1050 Hingham Street Rockland, MA 02370
Contact Person: Jon McGrath, Vice President of Research Telephone: 781-681-7950
Date Prepared: April 30, 2002
#### 2. DEVICE NAME
| Proprietary Name: | Embosphere® Microspheres and EmboGold™<br>Microspheres for use in Uterine Fibroid Embolization |
|----------------------|------------------------------------------------------------------------------------------------|
| Common/Usual Name: | Artificial emboli |
| Classification Name: | Artificial Embolization Device (21 CFR 882.5950 and<br>21 CFR 870 3300) |
#### 3. PREDICATE DEVICES
Embosphere® Microspheres as reviewed under K991549 and EmboGold™ Microspheres as reviewed under K010026.
#### 4. INTENDED USE
Embosphere® and EmboGold™ Microspheres are intended for embolization of arteriovenous malformations and hypervascular tumors, including and uterine fibroids.
{1}------------------------------------------------
#### 5. DEVICE DESCRIPTION
Embosphere® and EmboGold™ Microspheres are small, flexible, hydrophilic, biocompatible spheres made of acrylic polymer and porcine-derived gelatin. The microspheres are packaged in 0.9% saline and are provided sterile and nonpyrogenic. They are delivered to the target site by a catheter under fluoroscopic control.
Both products are provided in six size ranges in order to allow physicians to choose the calibration necessary for the vessel being embolized. The size ranges available are:
- 40-120 microns 100-300 microns 300-500 microns 500-700 microns 700-900 microns 900-1200 microns
Only microspheres of 500 microns or greater should be used for the embolization of uterine fibroids.
The only difference between Embosphere® and EmboGold™ Microspheres is the colorization of the EmboGold™ Microspheres. Embosphere® Microspheres are translucent, although they are visible to the naked eye when in suspension. EmboGold™ Microspheres are purple/red in color for improved visibility during preparation and handling by the physician.
#### 6. BASIS FOR SUBSTANTIAL EQUIVALENCE
Embosphere® and EmboGold™ Microspheres for uterine fibroid embolization are substantially equivalent to the same products when indicated for the more general intended use of embolization of hypervascular tumors. Clinical data was collected in a prospective clinical study to support the safety and effectiveness of Embosphere® Microspheres for uterine fibroid embolization. EmboGold™ Microspheres are the same product as Embosphere® Microspheres, except for the colorization. Data to support the substantial equivalence of EmboGold™ Microspheres to Embosphere® Microspheres was previously provided in K010026.
{2}------------------------------------------------
od and Drug Administration 00 Corporate Boulevard Rockville MD 20850
NOV 22 2002
BioSphere Medical, Inc. % Sheila Hemeon-Heyer, J.D., R.A.C. Senior Staff Consultant Medical Device Consultants 49 Plain Street NORTH ATTLEBORO MA 02760
Re: K021397 Trade/Device Name: Embosphere® Microspheres for UFE. Regulation Number: 21 CFR 882.5950 Regulation Name: Artificial Embolization device Regulatory Class: III Product Code: 85 NAJ Dated: August 27, 2002 Received: August 28, 2002
Dear Ms. Hemeon-Heyer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 8xx.1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Henry C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive. Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# 510(k) Number (if known):
Embosphere® Microspheres for use in Device Name: Uterine Fibroid Embolization
Indications for Use:
Embosphere Microspheres are indicated for use in embolization of arteriovenous malformations, hypervascular turnors, and symptomatic uterine fibroids.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Nancy C. Beardon
Prescription Use (Per 21 CFR 801.109)
OR
Over-The-Counter Use __________
(Optional Format 1-2-96
Embosphere Microspheres & EmboGold Additional Information - K021397
August 27, 2002 CONFIDENTIAL Appendix E · Page 1
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.