K253232 · ABBOTT MEDICAL · KRH · Oct 29, 2025 · Cardiovascular
Device Facts
Record ID
K253232
Device Name
Seguin Annuloplasty Ring
Applicant
ABBOTT MEDICAL
Product Code
KRH · Cardiovascular
Decision Date
Oct 29, 2025
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.3800
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Séguin™ Annuloplasty Ring is indicated for use in the repair of mitral valves that are diseased or damaged due to acquired or congenital processes.
Device Story
Seguin Annuloplasty Ring is a semi-rigid or semi-flexible implant designed to support mitral valve repair. Device consists of a polyethylene core covered by a knitted polyester sewing cuff. It is provided in sizes 24-40mm (2mm increments). Used by surgeons in clinical settings to provide structural support to the mitral valve annulus during repair procedures, potentially improving valve function and patient outcomes. Device is supplied sterile via ethylene oxide (EO) sterilization.
Clinical Evidence
No clinical testing was conducted. Substantial equivalence is supported by non-clinical bench testing, including design validation and real-time aging studies for packaging, confirming a 5-year shelf life.
Indicated for repair of mitral valves diseased or damaged due to acquired or congenital processes.
Regulatory Classification
Identification
An annuloplasty ring is a rigid or flexible ring implanted around the mitral or tricuspid heart valve for reconstructive treatment of valvular insufficiency.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance for Annuloplasty Rings 510(k) Submissions.”
K092310 — SEGUIN ANNULOPLASTY RING · St Jude Medical · Aug 27, 2009
K250859 — TransForm McCarthy Mitral Annuloplasty Ring (TF) · Genesee Biomedical, Inc. · Jul 25, 2025
K190506 — NeoForm Annuloplasty Ring · Genesee Biomedical, Inc. · Apr 18, 2019
K961246 — SJM SEGUIN ANNULOPLASTY RING MODEL SAR-M · St. Jude Medical, Inc. · Feb 14, 1997
K072655 — ATS SIMULUS SEMI-RIGID MITRAL ANNULOPLASTY RING, MODEL 800SR · Genesee Biomedical, Inc. · Oct 23, 2007
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
October 29, 2025
Abbott Medical
Bryan Soto
Senior Regulatory Affairs Specialist
177 County Road B East
St. Paul, Minnesota 55117
Re: K253232
Trade/Device Name: Seguin Annuloplasty Ring
Regulation Number: 21 CFR 870.3800
Regulation Name: Annuloplasty Ring
Regulatory Class: Class II
Product Code: KRH
Dated: September 26, 2025
Received: September 29, 2025
Dear Bryan Soto:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253232 - Bryan Soto
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K253232 - Bryan Soto
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
JULIE B. MACKEL -S
Samuel Raben
Acting Assistant Director
DHT2B: Division of Circulatory Support, Structural, and Vascular Devices
OHT2: Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K253232 | |
| Device Name Seguin Annuloplasty Ring | |
| Indications for Use (Describe) The Seguin Annuloplasty Ring is indicated for use in the repair of mitral valves that are diseased or damaged due to acquired or congenital processes. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
K253232 - Page 1 of 2
| 510(k) #: K253232 | 510(k) Summary | Prepared on: 2025-10-28 |
| --- | --- | --- |
| Contact Details | | 21 CFR 807.92(a)(1) |
| Applicant Name | Abbott Medical | |
| Applicant Address | 177 County Road B East St. Paul MN 55117 United States | |
| Applicant Contact Telephone | 1(651)269-1805 | |
| Applicant Contact | Mr. Bryan Soto | |
| Applicant Contact Email | bryan.soto1@abbott.com | |
| Device Name | | 21 CFR 807.92(a)(2) |
| Device Trade Name | Seguin Annuloplasty Ring | |
| Common Name | Seguin Annuloplasty Ring | |
| Classification Name | Ring, Annuloplasty | |
| Regulation Number | 870.8300 | |
| Product Code(s) | KRH | |
| Legally Marketed Predicate Devices | | 21 CFR 807.92(a)(3) |
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
| K092310 | SJM Seguin Annuloplasty Ring | KRH |
| Device Description Summary | | 21 CFR 807.92(a)(4) |
| The Seguin Annuloplasty Ring (SARP) is a semi-rigid or semi-flexible annuloplasty ring designed to support mitral valve repair. The ring is constructed from a polyethylene core covered by a knitted polyester sewing cuff. The Seguin Ring is sterilized by EO sterilization and supplied sterile. The Seguin ring is offered in sizes 24-40mm in increments of 2 mm. | | |
| Intended Use/Indications for Use | | 21 CFR 807.92(a)(5) |
| The Séguin™ Annuloplasty Ring is indicated for use in the repair of mitral valves that are diseased or damaged due to acquired or congenital processes. | | |
| Indications for Use Comparison | | 21 CFR 807.92(a)(5) |
| The indications for use of the device currently in scope of this submission are identical to those of the predicate device. | | |
| Technological Comparison | | 21 CFR 807.92(a)(6) |
| There is no difference in the technological characteristics between the device in scope of this submission to the predicate device. | | |
| Non-Clinical and/or Clinical Tests Summary & Conclusions | | 21 CFR 807.92(b) |
| There was no clinical testing needed for this determination. The following non-clinical testing was conducted in support of a substantial | | |
{5}
K253232 - Page 2 of 2
equivalence determination:
- Design Validation: Design validation testing was successfully conducted and the test results support a 5 year shelf life.
- Packaging: A real time aging testing was conducted and packaging system successfully met all the applicable stability requirements at 5-year timepoint.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.