MITRA-LIFT

K103812 · Sciencity Co., Ltd. · KRH · May 18, 2011 · Cardiovascular

Device Facts

Record IDK103812
Device NameMITRA-LIFT
ApplicantSciencity Co., Ltd.
Product CodeKRH · Cardiovascular
Decision DateMay 18, 2011
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.3800
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Mitra-Lift Set is indicated as reinforcement for repair of the human mitral valves damaged by acquired or congenital disease, or as a replacement for a previously implanted annuloplasty ring. The annuloplasty ring should be used only in cases where visual inspection confirms that the valve is repairable and does not require replacement.

Device Story

Mitra-Lift Set is a surgical implant system for mitral annuloplasty; consists of an annuloplasty ring, sizers, and a holder handle. Used by surgeons in an operating room setting to reinforce damaged human mitral valves. The ring is constructed of polyester textile materials. The device is supplied sterile for single use. During surgery, the clinician uses the sizers and holder handle to facilitate the implantation of the ring into the mitral valve annulus. The ring provides structural support to the valve, potentially improving valve function and patient outcomes by avoiding the need for full valve replacement when repair is feasible.

Clinical Evidence

Bench testing only. Biocompatibility testing confirmed materials are non-toxic, non-hemolytic, and non-pyrogenic per 21 CFR Part 58. Mechanical integrity testing, specifically suture retention, demonstrated performance equivalent to the predicate device.

Technological Characteristics

Constructed of polyester textile materials. Annuloplasty ring form factor with associated sizers and holder handle. Sterile, single-use device. No electronic components or software.

Indications for Use

Indicated for patients requiring mitral valve repair due to acquired or congenital disease, or replacement of a previously implanted annuloplasty ring. Contraindicated if visual inspection indicates the valve is not repairable and requires replacement.

Regulatory Classification

Identification

An annuloplasty ring is a rigid or flexible ring implanted around the mitral or tricuspid heart valve for reconstructive treatment of valvular insufficiency.

Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance for Annuloplasty Rings 510(k) Submissions.”

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## 510(k) Summary #### Submitter: ScienCity Co., Ltd. Hur Eun #101,102,107,108, 3-dong, Woniu Medical Instrument Industry Park. 1720-26, Taejang-dong, Wonju-shi, Kangwon-do, 220-120, Republic of Korea Phone: 82-2-3437-0954 Fax: 82-2-3437-6888 # MAY 18 2011 #### Official Correspondent: Kodent Inc. Joyce Bang 325 N. Puente St. Unit B Brea, CA 92821 Email: kodentinc@gmail.com Phone: 714-525-0114 Fax: 714-525-0116 #### Device Information Trade Name: Mitra-Lift Set Device: Ring, Annuloplasty Regulation Description: Annuloplasty Ring Product Code: KRH Regulation Number: 870.3800 Device Class: Class II Submission Types: 510(K) Submission #### General Description Mitra-Lift Set is set of devices for mitral annuloplasty. This set consists of Mitra-Lift (implant device for mitral annuloplasty), and its accessories (Sizers and holders). The main device (Mitra-Lift series) is supplied sterile and intended for single use only. Mitra-Lift series is constructed of polyester textile materials and provided as the various sizes. It is intended to be used with its accessories, Mitral Sizer and Holder Handle. #### Indication for Use The Mitra-Lift is indicated as reinforcement for repair of the human cardiac mitral and tricuspid valves damaged by acquired or congenital disease, or as a replacement for a previously implanted annuloplasty ring. The annuloplasty ring should be used only in cases where visual inspection confirms that the valve is repairable and does not require replacement. #### Predicate Devices: {1}------------------------------------------------ The subject device is substantially equivalent to the following predicate devices: - . Simplici-T Annuloplasty System (K052970) manufactured by Medtronic Heart Valves (Medtronic). #### Performance Data Non-clinical laboratory testing was performed demonstration that the device complied with the USP Monographs and with the EP Monographs for Absorbable surgical sutures. The results of biocompatibility testing support that the materials used in the manufacture of the Mitra-Lift are non-toxic, non-hemolytic, and non-pyrogenic. All testing was conducted under Good Laboratory Practices per 21 CFR Part 58. Mechanical Integrity testing for the COMVAR Set (including Mitra-Lift Set) includes suture retention testing which demonstrated that the design provided for a more than adequate retention force as compared to the predicate device. Testing demonstrated that the Mitra-Lift Set is substantially equivalent to the predicate device for repair of the mitral or tricuspid valve. #### Comparison to Predicate Devices Testing and other comparisons have established that the subject of Mitra-Lift Set is substantially equivalent in design, materials, indications and intended use, packaging, and performance to other predicate devices of the type currently marketed in the U.S. #### Conclusion The Mitra-Lift, subject of this submission, constitutes a safe, reliable and effective medical device, meeting all the declared requirements of its intended use. Device presents no adverse health effects or safety risks to patients when used as intended. The Mitra-Lift has the same intended use and fundamental scientific technology as its predicate devices Simplici-T Annuloplasty System (K052970) manufactured by Medtronic Heart Valves (Medtronic). Therefore, Mitra-Lift and its predicate devices are believed to be substantially equivalent. {2}------------------------------------------------ Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 MAY 1 8 2011 Sciencity Co.,Ltd. c/o Ms. Joyce Bang Kodent, Inc. 325 N. Puente St. Unit B Brea, California 92821 Re: K103812 Mitra-Lift Set Regulation Number: 21 CFR 870.3800 Regulation Name: Annuloplasty Ring Regulatory Class: Class II (two) Product Code: KRH Dated: May 9, 2011 Received: May 12, 2011 Dear Ms. Bang: This letter corrects our substantially equivalent letter of May 18, 2011. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Page 2 - Ms. Joyce Bang Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Bram D. Zuckerman, M.D. Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Sciencity Co., Ltd. ## Indication for Use | 510(K) Number (if known): | K103812 | |---------------------------|---------| |---------------------------|---------| Device Name: Mitra-Lift Set ## Indication for Use: : The Mitra-Lift Set is indicated as reinforcement for repair of the human mitral valves damaged by acquired or congenital disease, or as a replacement for a previously implanted annuloplasty ring. The annuloplasty ring should be used only in cases where visual inspection confirms that the valve is repairable and does not require replacement. Prescription Use _____________________________________________________________________________________________________________________________________________________________ AND/OR Over-The-Counter (Part 21 CFR 801 Subpart D) (Per 21 CFR 801 Subpart C) ## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Page 1 of 1 | | <i>(Division Sign-Off)</i> | |---------------|------------------------------------| | | Division of Cardiovascular Devices | | 510(k) Number | K103812 |
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...