Gunther Tulip Vena Cava Filter Set for Femoral Vein Approach, Gunther Tulip Vena Cava Filter Set for Jugular Vein Approach, Gunther Tulip Vena Cava Filter Set for Femoral and Jugular Vein Approach
K172557 · William Cook Europe Aps · DTK · Nov 20, 2017 · Cardiovascular
Device Facts
Record ID
K172557
Device Name
Gunther Tulip Vena Cava Filter Set for Femoral Vein Approach, Gunther Tulip Vena Cava Filter Set for Jugular Vein Approach, Gunther Tulip Vena Cava Filter Set for Femoral and Jugular Vein Approach
Applicant
William Cook Europe Aps
Product Code
DTK · Cardiovascular
Decision Date
Nov 20, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.3375
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Günther Tulip Filter implant is intended for the prevention of recurrent pulmonary embolism (PE) via placement in the vena cava in the following situations: - Pulmonary thromboembolism when anticoagulant therapy is contraindicated; - Failure of anticoagulant therapy in thromboembolic diseases; - Emergency treatment following massive PE where anticipated benefits of conventional therapy are reduced; and - Chronic, recurrent PE where anticoagulant therapy has failed or is contraindicated. The Günther Tulip Filter implant may be retrieved if clinically indicated; please refer to the "Optional Filter Retrieval" section of the Instructions for Use for more information. The product is intended for percutaneous placement via a femoral or jugular vein for filtration of inferior vena cava (IVC) blood to prevent PE.
Device Story
The Günther Tulip Vena Cava Filter is a permanent/retrievable implant designed to prevent pulmonary embolism by trapping emboli in the inferior vena cava (IVC). The device consists of a cobalt-chromium alloy filter with a hooked leg design for vessel wall securement and webbed wires to maintain shape and increase trapping surface area. It is delivered percutaneously via a 7 French coaxial introducer system (femoral or jugular approach). The filter is placed by a physician in a clinical setting. The device functions mechanically; it does not utilize software or algorithms. The filter remains in the IVC to filter blood flow; it may be retrieved if clinically indicated. The device benefits patients by providing mechanical prophylaxis against PE when anticoagulation is not an option or has failed.
Clinical Evidence
No clinical data; bench testing only. The manufacturer states that because no changes were made to the design, manufacturing, sterilization, or principles of operation, no new performance testing was required.
Technological Characteristics
Material: Cobalt chromium alloy. Form factor: Filter length ~50 mm, compressed diameter 30 mm. Delivery: 7 French coaxial introducer system with Check-Flo valve, hydrophilic coated pre-dilator, and three-way stopcock. Features: Hooked leg design for wall securement, webbed wires for shape maintenance and emboli trapping. Energy source: None (mechanical). Connectivity: None. Sterilization: Not specified.
Indications for Use
Indicated for patients requiring prevention of recurrent pulmonary embolism (PE) where anticoagulant therapy is contraindicated, has failed, or in emergency cases of massive PE. Intended for percutaneous placement in the inferior vena cava via femoral or jugular vein.
Regulatory Classification
Identification
A cardiovascular intravascular filter is an implant that is placed in the inferior vena cava for the purpose of preventing pulmonary thromboemboli (blood clots generated in the lower limbs and broken loose into the blood stream) from flowing into the right side of the heart and the pulmonary circulation.
Special Controls
*Classification.* Class II. The special controls for this device are:(1) “Use of International Standards Organization's ISO 10993 ‘Biological Evaluation of Medical Devices Part I: Evaluation and Testing,’ ” and
(2) FDA's:
(i) “510(k) Sterility Review Guidance and Revision of 2/12/90 (K90-1)” and
(ii) “Guidance for Cardiovascular Intravascular Filter 510(k) Submissions.”
K043509 — GUNTHER TULIP VENA CAVA FILTER SET · Cook, Inc. · May 5, 2005
K211874 — Günther Tulip Vena Cava Filter Set for Femoral Vein Approach, Günther Tulip Vena Cava Filter Set for Jugular Vein Approach, Günther Tulip Vena Cava Filter Set for Femoral and Jugular Vein Approach · William Cook Europe Aps · Jul 28, 2021
K172584 — Günther Tulip Vena Cava Filter Set for Femoral Vein Approach, Günther Tulip Vena Cava Filter Set for Femoral and Jugular Vein Approach · William Cook Europe Aps · Nov 20, 2017
K211875 — Cook Celect Platinum Vena Cava Filter Set for Femoral Vein Approach, Cook Celect Platinum Vena Cava Filte Set for Jugular Vein Approach, Cook Celect Platinum Vena Cava Filter Set for Femoral and Jugular Vein Approach · William Cook Europe Aps · Jul 28, 2021
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
November 20, 2017
William Cook Europe ApS Henriette S. Christiansen Director of Regulatory Affairs Sandet 6, 4632 Bjaeverskov Denmark
Re: K172557
Trade/Device Name: Gunther Tulip Vena Cava Filter Set for Femoral Vein Approach, Gunther Tulip Vena Cava Filter Set for Jugular Vein Approach, Gunther Tulip Vena Cava Filter Set for Femoral and Jugular Vein Approach Regulation Number: 21 CFR 870.3375 Regulation Name: Cardiovascular Intravascular Filter Regulatory Class: Class II Product Code: DTK Dated: August 23, 2017 Received: August 24, 2017
Dear Henriette S. Christiansen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Nicole G. Ibrahim -S
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
#### 510(k) Number (if known) K172557
#### Device Name
Günther Tulip Vena Cava Filter Set for Femoral Vein Approach Günther Tulip Vena Cava Filter Set for Jugular Vein Approach Günther Tulip Vena Cava Filter Set for Femoral and Jugular Vein Approach
#### Indications for Use (Describe)
The Günther Tulip Filter implant is intended for the prevention of recurrent pulmonary embolism (PE) via placement in the vena cava in the following situations:
- · Pulmonary thromboembolism when anticoagulant therapy is contraindicated;
- · Failure of anticoagulant therapy in thromboembolic diseases;
- · Emergency treatment following massive PE where anticipated benefits of conventional therapy are reduced; and
- · Chronic, recurrent PE where anticoagulant therapy has failed or is contraindicated.
The Günther Tulip Filter implant may be retrieved if clinically indicated; please refer to the "Optional Filter Retrieval" section of the Instructions for Use for more information.
The product is intended for percutaneous placement via a femoral or jugular vein for filtration of inferior vena cava (IVC) blood to prevent PE.
Type of Use (Select one or both, as applicable)
| <div> <span style="font-size: 10px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </div> |
|---------------------------------------------------------------------------------------------------------------|
| <div> <span style="font-size: 10px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| 510(k) Summary | K172557 |
|----------------------|------------------------------------------------------------------------------|
| Date Prepared: | 23 August 2017 |
| Submitted By: | William Cook Europe ApS |
| | Sandet 6 |
| | 4632 Bjaeverskov, Denmark |
| Contact: | Henriette Stordal Christiansen |
| Phone: | +45 56 86 87 56 |
| Fax: | +45 56 86 86 96 |
| Email: | Henriette.Christiansen@CookMedical.com |
| Device: | |
| Trade Name: | Günther Tulip® Vena Cava Filter Set for Femoral Vein Approach |
| | Günther Tulip® Vena Cava Filter Set for Jugular Vein Approach |
| | Günther Tulip® Vena Cava Filter Set for Femoral and Jugular<br>Vein Approach |
| Common Name: | Inferior Vena Cava Filter |
| Review Panel: | Cardiovascular |
| Classification Name: | Intravascular Filter (21 CFR 870.3375, Product Code DTK) |
| Classification: | Class II |
{4}------------------------------------------------
### Indications for Use:
The Günther Tulip Filter implant is intended for the prevention of recurrent pulmonary embolism (PE) via placement in the vena cava in the following situations:
- Pulmonary thromboembolism when anticoagulant therapy is contraindicated; .
- Failure of anticoagulant therapy in thromboembolic diseases; .
- Emergency treatment following massive PE where anticipated benefits of conventional . therapy are reduced; and
- . Chronic, recurrent PE where anticoagulant therapy has failed or is contraindicated.
The Günther Tulip Filter implant may be retrieved if clinically indicated; please refer to the "Optional Filter Retrieval" section of the Instructions for Use for more information.
The product is intended for percutaneous placement via a femoral or jugular vein for filtration of inferior vena cava (IVC) blood to prevent PE.
### Predicate Device:
The Günther Tulip Vena Cava Filter Sets are substantially equivalent to the predicate devices:
- . Günther Tulip Vena Cava Filter Sets cleared under K090140 on 03 April 2009.
## Comparison to Predicate Device:
It has been demonstrated that the subject Günther Tulip Vena Cava Filter Sets are substantially equivalent to the predicate Günther Tulip Vena Cava Filter Sets (K090140). No changes have been made to the Günther Tulip Vena Cava Filter or its delivery systems; the Günther Tulip Vena Cava Filter Sets are identical to the predicate devices in terms of indications for use, principles of operation, design, materials of construction, manufacturing processes, sterilization process, and basic technological characteristics. The purpose of this submission is to update device labeling based on safety and performance information gathered from post-market experience and to align with current international and regulatory standards.
{5}------------------------------------------------
### Device Description:
The Günther Tulip Vena Cava Filter is an IVC filter intended for use in prevention of recurrent PE. The filter is intended for percutaneous placement via either the jugular vein or femoral vein for filtration of blood within the IVC. The Günther Tulip Vena Cava Filter Set is available in a Femoral Set and a Jugular Set. Each set consists of a preloaded filter, a coaxial introducer system, a hydrophilic coated pre-dilator, and a three-way stopcock. The filter is introduced and placed via a 7 French coaxial introducer system with a Check-Flo® valve. The introducer dilator has 8 sideports and two radiopaque markers 30 mm apart. The Günther Tulip Vena Cava Filter Set is also available in a universal vein access version consisting of components for both femoral and jugular vein approaches; which includes a preloaded filter in the femoral approach. The Günther Tulip Vena Cava Filter is constructed from a cobalt chromium alloy. The end of each leg is hooked outward; this hook design enables securement of the filter to the wall of the inferior vena cava. "Webbed" wires (like tulip petals) between the legs are bent secondary legs which maintain the shape of the filter by pressing outward the vein walls. These webs also increase the area into which emboli can be trapped. The filter is approximately 50 mm long along its main axis when compressed to a diameter of 30 mm.
### Performance Testing:
No changes to the design, manufacturing, sterilization, or principles of operation have been introduced with the subject devices; therefore, no performance testing was considered necessary.
#### Conclusion:
The Günther Tulip Vena Cava Filter Sets are substantially equivalent to the predicate devices. No changes have been made to the filter implant or the delivery systems. The only changes being made are to the device labeling.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.