DR300 HOLTER MONITOR

K142424 · North East Monitoring, Inc. · MWJ · Nov 25, 2014 · Cardiovascular

Device Facts

Record IDK142424
Device NameDR300 HOLTER MONITOR
ApplicantNorth East Monitoring, Inc.
Product CodeMWJ · Cardiovascular
Decision DateNov 25, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.2800
Device ClassClass 2
AttributesPediatric

Indications for Use

Holter Mode: Detection of Arrhythmias, Efficacy of Pharmacological Treatment and Pacemaker Evaluation. Event Recorder: The event recorder module is a patient activated device designed to record and for diagnostic evaluation of transient symptoms (such as dizziness, palpitations, syncope, and chest pain). Once data is recorded, the data is transmitted for evaluation.

Device Story

Ambulatory ECG recorder (Holter/Event monitor) for cardiac rhythm monitoring; records ECG signals via patient-worn electrodes and cables; stores data on removable SD card. Features Bluetooth transmitter for wireless data transmission to NorthEast Monitoring gateway or PC USB adapter. Operated by physician/technician for hookup; patient-activated for event recording. Data downloaded to TelePro analysis software for physician review. Assists in diagnosing arrhythmias and evaluating treatment efficacy. Not for real-time telemetry of life-threatening arrhythmias.

Clinical Evidence

Bench testing only. Verification included 11 Holter tests (24-hour to 14-day duration) to confirm functionality with added Bluetooth and shielding. Wireless communication of encrypted data was validated through simulated bench testing (24-hour to 48-hour duration). Compliance with IEC 60601-1, 60601-1-6, 60601-1-11, 60601-2-47, and 60601-1-2 standards demonstrated.

Technological Characteristics

Ambulatory ECG recorder; 2 or 3-channel recording; uses standard Ag/AgCl electrodes. Powered by 1.5V AA alkaline, NiMH, or Lithium L91 battery. Connectivity via Bluetooth wireless transmitter (encrypted data). Includes LCD for status/lead quality. Standards: IEC 60601-1, IEC 60601-1-6, IEC 60601-1-11, IEC 60601-2-47, IEC 60601-1-2.

Indications for Use

Indicated for patients, including infants <10 kg, requiring ambulatory cardiac monitoring for arrhythmia detection, pharmacological treatment efficacy, pacemaker evaluation, or diagnostic evaluation of transient symptoms like dizziness, palpitations, syncope, and chest pain.

Regulatory Classification

Identification

A medical magnetic tape recorder is a device used to record and play back signals from, for example, physiological amplifiers, signal conditioners, or computers.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, composed of three stylized human profiles facing to the right. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 November 25, 2014 North East Monitoring, Inc. c/o Mr. Joe Azary Official Correspondent 2 Clock Tower Place, Suite 555 Maynard, Massachusetts 01754 Re: K142424 > Trade/Device Name: DR300 Holter Monitor Regulation Number: 21 CFR 870.2800 Regulation Name: Medical Magnetic Tape Recorder Regulatory Class: Class II Product Code: MWJ Dated: October 3, 2014 Received: October 6, 2014 Dear Mr. Joe Azary, We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in {1}------------------------------------------------ the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Melissa A. Torres -S For Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ K142424 Page __1__ of__1_ 510(k) Number (if known): Device Name: __ DR300 Holter Monitor__________________________________________________________________________________________________________________________________________ Indications For Use: Holter Mode: Detection of Arrhythmias, Efficacy of Pharmacological Treatment and Pacemaker Evaluation. Event Recorder: The event recorder module is a patient activated device designed to record and for diagnostic evaluation of transient symptoms (such as dizziness, palpitations, syncope, and chest pain). Once data is recorded, the data is transmitted for evaluation. (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use X (Per 21 CFR 801.109) Over-The-Counter Use OR {3}------------------------------------------------ # K142424 Page 1 of 5 # 510(k) Summary for North East Monitoring DR300 ### 1. SUBMITTER/510(K) HOLDER North East Monitoring Two Clock Tower Place Suite 555 Maynard, MA 01754 Contact Name: Joseph Azary Email: jazary@erols.com Telephone: (203) 242-6670 Date Revised: November 4, 2014 ### 2. DEVICE NAME | Proprietary Name: | DR300 Holter Monitor | |----------------------|-------------------------------------------------------------------------------------------| | Common/Usual Name: | Ambulatory ECG recorder, Ambulatory Electrocardiograph (without analysis), Holter Monitor | | Classification: | Class 2, 21 CFR 870.2800 | | Classification Name: | Ambulatory ECG recorder | | Product Code: | MWJ | #### 3. PREDICATE DEVICES - K061293 Telaheart Digital Recorder / DR200 . #### 4. DEVICE DESCRIPTION The DR300 is an ambulatory monitor, sometimes called a Holter, which is a painless method to monitor the heart beat for a period of time (such as 24 hours, 48 hours, or 72 hours). The Holter is a small recording device that records the heart beat while being worn by the patient. The physician or technician places electrodes and wires on the patient. The wires are connected to the Holter or digital recorder. {4}------------------------------------------------ The DR300 Recorder has two modes that allow it to be used either as a standard Holter monitor or a looping Event recorder. The device is designed to facilitate ambulatory cardiac monitoring on the order of a physician, of those patients (including infants weighing less than 10 kg) who may benefit from such monitoring. The data obtained during monitoring is not analyzed at the time of recording. After the recording is complete, the data must later be downloaded to a compatible NorthEast Monitoring Holter or Event analysis system (TelePro Software) to be analyzed. The Holter Analysis Software was cleared by FDA under K930564. The device is not intended to replace real time telemetry monitoring for patient suspected of having life threatening arrhythmias. The DR300 Digital Recorder package includes: - . DR300 Digital Recorder - Operation Manual - SD Card ● - Patient Cable The DR300 digital recorder is powered by one 1.5 volt AA alkaline battery (MN1500 or the equivalent), one AA rechargeable NiMH (nickel metal hydride) battery, or one AA Eveready Lithium L91 battery. Batteries should not be re-used for a second patient. The batteries are not included; users are instructed to purchase 2 AA batteries. The device is compatible with standard silver / silver chloride ECG electrodes. Electrodes are not provided with the subject device. The user is instructed to purchase standard silver / silver chloride ECG electrodes. The DR300 digital recorder uses NorthEast Monitoring NorthEast Monitoring shielded patient cables with either 5 or 7 patient leads for a 3 channel recording or three leads for 2 channel Holter Recording. The DR300 has an LCD that is used to display the time of day (during the recording), prompts and error messages (during the hookup procedure or during recording), or lead quality (during the hookup procedure). The data collected by the DR300 digital recorder is stored on a removable SD Card. The DR300 is packaged in a plastic bag in a cardboard shipping carton. The shipping carton will also include a patient cable and an SD Card. {5}------------------------------------------------ The difference between the DR300 and the DR200/Telaheart is the wireless capability. The DR300 digital recorder is equipped with wireless Bluetooth transmitter. A NorthEast Monitoring Bluetooth Gateway or USB Dongle is able to receive the encrypted Holter and Event data. In order for the wireless transmission to occur, the DR300 will need to come in range of a North East Monitoring transceiver - either a gateway or a PC with a USB Adapter. If there is sufficient data to send, the recorder will attempt to locate a transceiver every 20 minutes. When the transceiver is found the data will be sent. Transmission may be started manually by pressing the enter button. ### 5. INTENDED USE The DR300 Digital Recorder can be used in Holter mode and Event Recorder mode. ### Holter Mode Detection of Arrhythmias, Efficacy of Pharmacological Treatment, and Pacemaker Evaluation. ### Event Recorder The DR300 event recorder module is a patient activated device designed to record and for diagnostic evaluation of transient symptoms (such as dizziness, palpitations, syncope, and chest pain). Once data is recorded, the data is transmitted wirelessly for evaluation. #### 6. TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE The DR300 substantially equivalent to the Telaheart / DR 200 cleared under 510(k) K061293. The DR300 is equipped with a wireless Bluetooth transmitter and has additional shielding to be immune to radiofrequency interference and in order to pass FCC and EMC test requirements. When the wireless function is enabled by the physician/technician, once every preset interval (normally set at 20 minutes) the device will check for the availability of any new data since the last successful transmission. If there is any, the Bluetooth module will be enabled and an attempt made to establish a link. If a link is established, then all available new ECG data will be transmitted. After each "block" of data is transmitted (normally 16 K bytes or less) a verification will be made for a successful transmission by the detection of a reply from the receiving end. The data will be re-transmitted if needed. {6}------------------------------------------------ ### 7. PERFORMANCE TESTING The subject device conforms to the following standards: - . IEC 60601-1-6: 2010 Medical Electrical Equipment Part 1-6: General Requirements for Safety - Collateral Standard: Usability including IEC 62366: Application of usability engineering to medical devices. - IEC 60601-1-11: 2010 Medical Electrical Equipment Part 1-11: General . Requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment. - IEC 60601-1: 2012 Medical Electrical Equipment Part 1: General Requirements . for Safety - IEC 60601-1-2-47 Medical Electrical Equipment Part 2 47: Particular ● Requirements for the Safety including Essential Performance of Ambulatory Electrocardiographic Systems. - IEC 60601-1-2 Medical Electrical Equipment Part 1-2: General Requirements . for Basic Safety and Essential Performance Collateral Standard: Electromagnetic Compatibility. - FDA Guidance Radio Frequency Wireless in Medical Devices (August 14, 2013) ● ## 8. SAFETY AND EFFICACY The subject device was subjected to testing including: - · IEC 60601-1-6: 2010 Medical Electrical Equipment Part 1-6: General Requirements for Safety - Collateral Standard: Usability including IEC 62366: Application of usability engineering to medical devices. · IEC 60601-1-11: 2010 Medical Electrical Equipment Part 1-11: General Requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment. · IEC 60601-1: 2012 Medical Electrical Equipment Part 1: General Requirements for Safety · IEC 60601-1-2-47 Medical Electrical Equipment Part 2 – 47: Particular Requirements for the Safety including Essential Performance of Ambulatory Electrocardiographic Systems. · IEC 60601-1-2 Medical Electrical Equipment Part 1-2: General Requirements for Basic Safety and Essential Performance Collateral Standard: Electromagnetic Compatibility. - Software Verification and Validation ● - Simulated bench and wireless testing (see summary below) {7}------------------------------------------------ The testing included 11 Holter Tests for the purposes of verifying that the DR300 functions as intended. Testing was conducted related to the functionality of the Holter and Event functions of the DR300 without the wireless function being utilized. These tests verified that the device functions as intended after the inclusion of the Bluetooth transmitter and additional shielding. The testing ranged from monitoring over a 24 hour to 14 day period. Additional testing was performed to evaluate the wireless Bluetooth technology and communication of encrypted data. The wireless testing ranged from monitoring over a 24 hour to 48 hour period. All testing confirmed that the device using the wireless technology functioned as intended and found no issues of safety or effectiveness. ## 9. CONCLUSION North East Monitoring believes that based on the indications for use, technological characteristics, and comparison to predicate devices the DR300 has been shown to be substantially equivalent to the predicate.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...